| Literature DB >> 35237622 |
Zhuoyi Chen1,2,3,4, Lingling Niu1,2,3,4, Jing Zhao1,2,3,4, Peijun Yao1,2,3,4, Xiaoying Wang1,2,3,4, Xingtao Zhou1,2,3,4.
Abstract
PURPOSE: To investigate the feasibility and safety of the implantable collamer lens V4c (ICL V4c) implantation without using an ophthalmic viscosurgical device (OVD-free technique).Entities:
Keywords: ICL (implantable collamer lens); myopia; new surgical and technical approach; ophthalmic viscosurgical device; refractive surgery
Year: 2022 PMID: 35237622 PMCID: PMC8882649 DOI: 10.3389/fmed.2022.790137
Source DB: PubMed Journal: Front Med (Lausanne) ISSN: 2296-858X
Figure 1The surgical steps of Implantable Collamer Lens V4c implantation with an ophthalmic viscosurgical device-free (OVD-free) technique. Case 1 (A–E): (A) Incision creation. (B–D) ICL-V4c injection and placement. (E) The status at the end of the surgery. Case 2 (F–J): (F) Incision creation. (G,H) ICL-V4c injection and placement. (I) Balanced salt solution injection. (J) The status at the end of the surgery.
Baseline characteristics of the participants.
|
|
|
|
|---|---|---|
| Age (years) | 26.09 ± 4.89 | 26.37 ± 5.31 |
| Corrected distance visual acuity (Logmar) | −0.00 ± 0.04 | −0.00 ± 0.05 |
| Spherical equivalent (D) | −10.27 ± 2.60 | −9.68 ± 2.94 |
| Anterior chamber depth (μm) | 3.21 ± 0.23 | 3.18 ± 0.25 |
| Corneal thickness (μm) | 526.23 ± 29.35 | 528.30 ± 37.48 |
| Intraocular pressure (mmHg) | 15.19 ± 2.52 | 16.15 ± 2.99 |
| Endothelial cell | ||
| Endothelial cell density (cell/mm2) | 2620.03 ± 239.50 | 2572.40 ± 211.57 |
| Coefficient of variation (%) | 41.50 ± 4.75 | 41.59 ± 5.09 |
| Percentage of hexagonal cells (%) | 39.78 ± 8.32 | 39.34 ± 7.23 |
OVD, ophthalmic viscosurgical device.
Figure 2Refractive outcomes at 12 months after the Implantable Collamer Lens V4c implantation in ophthalmic viscosurgical device-free (OVD-free)/OVD group. (A) Change in corrected distance visual acuity (CDVA). (B) Uncorrected distance visual acuity (UDVA). (C) Difference between postoperative UDVA and preoperative CDVA. (D) Spherical equivalent refractive accuracy.
Changes in endothelial cells and intraocular pressure in OVD-free/OVD group.
|
|
|
|
|
|---|---|---|---|
| ECD (cell/mm2) | |||
| 1 week | 50.54 ± 208.70 | 51.39 ± 149.82 | 0.8870 |
| 1 month | 37.95 ± 173.24 | 44.50 ± 144.84 | 0.9355 |
| 3 months | 87.26 ± 260.86 | 26.70 ± 280.39 | 0.1389 |
| 12 months | 65.34 ± 213.47 | 25.94 ± 163.66 | 0.0382 |
| CV (%) | |||
| 1 week | −3.50 ± 4.07 | −3.11 ± 3.71 | 0.8024 |
| 1 month | −1.17 ± 4.84 | −1.85 ± 3.25 | 0.3188 |
| 3 months | −1.22 ± 6.84 | −2.75 ± 6.21 | 0.1782 |
| 12 months | −2.16 ± 4.64 | −1.33 ± 3.69 | 0.5029 |
| 6A (%) | |||
| 1 week | 5.45 ± 5.58 | 6.44 ± 7.05 | 0.4996 |
| 1 month | 2.83 ± 6.96 | 2.80 ± 5.04 | 0.5034 |
| 3 months | 3.38 ± 8.35 | 5.69 ± 8.53 | 0.0380 |
| 12 months | 2.93 ± 6.22 | 1.48 ± 4.16 | 0.2503 |
| IOP (mmHg) | |||
| 1 day | 0.73 ± 2.48 | −0.05 ± 3.57 | 0.8676 |
| 1 week | 0.30 ± 3.96 | −0.57 ± 3.28 | 0.6023 |
| 1 month | −1.43 ± 2.36 | −1.90 ± 3.09 | 0.9592 |
| 3 months | −1.10 ± 2.77 | −1.60 ± 2.99 | 0.2118 |
| 12 months | −0.60 ± 1.93 | −1.64 ± 2.80 | 0.0125 |
p < 0.05.
OVD, ophthalmic viscosurgical device; ECD, endothelical cell density; CV, coefficient of variation; 6A, percentage of hexagonal cells; IOP, intraocular pressure.
Figure 3Time course of changes in endothelial cell density (ECD) after Implantable Collamer Lens V4c implantation in ophthalmic viscosurgical device-free (OVD-free)/OVD group.