Anterior subcapsular cataract caused by forceful irrigation during implantation of a posterior chamber phakic intraocular lens with a central hole.

Gernot Steinwender1, Kleopatra Varna-Tigka1, Mehdi Shajari1, Thomas Kohnen2.   

Abstract

PURPOSE: To report the rare entity of anterior subcapsular cataract after implantation of a posterior chamber phakic intraocular lens (pIOL) with a central hole.
SETTING: Department of Ophthalmology, Goethe-University, Frankfurt, Germany.
DESIGN: Retrospective case series.
METHODS: Medical reports were reviewed of patients with vision loss caused by anterior subcapsular cataract after implantation of the V4c Visian Implantable Collamer Lens with a central hole. Patients were identified during regular follow-up visits between January and September 2015. Possible causative factors, such as microbial contamination of surgical instruments, type of ophthalmic viscosurgical device (OVD) and irrigation substance, intracameral application of mydriatics and antibiotics, duration of surgeries, and surgical technique were analyzed.
RESULTS: Eleven eyes of 8 patients showed white opacification behind the anterior capsule of the crystalline lens 1 day postoperatively. Opacification led to a significant loss of corrected distance visual acuity in all eyes. In 9 eyes, the pIOL was explanted and in 8 eyes, phacoemulsification with IOL implantation was performed. All other analyzed possible causative factors were excluded; thus, the opacification was attributed to the surgical technique of OVD removal by irrigation with a cannula in the presence of a central hole in the pIOL. After the surgical technique was adapted, no further cases of anterior subcapsular cataract occurred.
CONCLUSIONS: Anterior subcapsular cataract was induced during implantation of a pIOL with a central hole by cannula irrigation with enforced stream onto the capsule through that hole. This entity can be avoided by a change in surgical technique.
Copyright © 2017 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2017        PMID: 28823445     DOI: 10.1016/j.jcrs.2017.03.045

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  5 in total

Review 1.  Analysis of perioperative problems related to intraocular Implantable Collamer Lens (ICL) implantation.

Authors:  Hao Zhang; Rui Gong; Xiaolan Zhang; Yingping Deng
Journal:  Int Ophthalmol       Date:  2022-06-22       Impact factor: 2.029

Review 2.  Phakic intraocular lenses: An overview.

Authors:  Soraya M R Jonker; Tos T J M Berendschot; Isabelle E Y Saelens; Noël J C Bauer; Rudy M M A Nuijts
Journal:  Indian J Ophthalmol       Date:  2020-12       Impact factor: 1.848

3.  One-year Observation of Safety of Implantable Collamer Lens V4c Implantation Without Using an Ophthalmic Viscosurgical Device.

Authors:  Zhuoyi Chen; Lingling Niu; Jing Zhao; Peijun Yao; Xiaoying Wang; Xingtao Zhou
Journal:  Front Med (Lausanne)       Date:  2022-02-14

4.  Multicenter clinical outcomes of hole implantable collamer lens implantation in middle-aged patients.

Authors:  Akihito Igarashi; Kazutaka Kamiya; Kazuo Ichikawa; Yoshihiro Kitazawa; Takashi Kojima; Tomoaki Nakamura; Kimiya Shimizu
Journal:  Sci Rep       Date:  2022-03-10       Impact factor: 4.379

5.  Safety of implantable Collamer lens implantation without ophthalmic viscosurgical device: A retrospective cohort study.

Authors:  Manqiang Peng; Qiongyan Tang; Libei Zhao; Muhanmad Ahmad Khan; Ding Lin
Journal:  Medicine (Baltimore)       Date:  2020-06-12       Impact factor: 1.817

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.