Supriyanto Ardjo Pawiro1, Dwi Aprilia Mahfirotin1, Muhamad Iqbal Assegab2, Wahyu Edy Wibowo2.
Abstract
PURPOSE: The recently worldwide standard measurement of electron beam reference dosimetry include the International Atomic Energy Agency (IAEA) Technical Report Series (TRS)-398 and Association of Physicists in Medicine (AAPM) Task Group (TG)-51 protocols. Muir et al. have modified calibration methods for electron beam calibration based on AAPM TG-51. They found that the use of cylindrical chambers at low energy gave acceptable results. In this study, we propose and report a modified calibration for electron beam based on IAEA TRS-398, the standard reference dosimetry protocol worldwide.Entities:
Keywords: AAPM TG-51 protocol; IAEA TRS-398 protocol; absorbed dose to water; electron beam; modified calibration
Mesh:
Substances:
Year: 2022 PMID: 35226389 PMCID: PMC8992941 DOI: 10.1002/acm2.13573
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Ratio of absorbed dose to water of modified calibration and Technical Report Series (TRS)‐398 () obtained from parallel‐plate chamber used in this work
| Elekta Synergy Platform | Elekta Versa HD | ||||
|---|---|---|---|---|---|
| Type of chamber | Energy (MeV) |
| Discrepancy (%) |
| Discrepancy (%) |
| Exradin A11 | 6 | 1.013 | 1.3 | 1.013 | 1.3 |
| 8 | 1.013 | 1.3 | 1.013 | 1.3 | |
| 10 | 1.012 | 1.2 | 1.013 | 1.3 | |
| 12 | 1.014 | 1.4 | 1.014 | 1.4 | |
| 15 | 1.015 | 1.5 | 1.016 | 1.6 | |
Note: Discrepancy (%) = (absorbed doses for modified calibration – absorbed doses for TRS‐398)/(absorbed doses for TRS‐398 dose).
Power fitting parameters for cylindrical chamber for as a function of 12
| Power fitting parameter | |||||
|---|---|---|---|---|---|
| Manufacturer | Chambers |
|
|
| RMSD (%) |
| PTW | 30013 | 0.978 | 0.112 | 0.816 | 0.15 |
| IBA | CC13 | 0.926 | 0.129 | 0.279 | 0.10 |
| Exradin | A1Sl | 0.205 | 0.854 | 0.036 | 0.13 |
Note: Cylindrical chambers are positioned with their central axes at (no effective point of measurement).
Abbreviation: RMSD, root mean square deviation.
Exponential fitting parameters for parallel‐plate chamber for as a function of 12
| Exponential fitting parameter | |||||
|---|---|---|---|---|---|
| Manufacturer | Chamber |
|
|
| RMSD (%) |
| Exradin | A11 | 0.992 | 0.114 | 2.864 | 0.13 |
Abbreviation: RMSD, root mean square deviation.
Percentage depth dose (PDD) at ) from Elekta Synergy Platform and Elekta Versa HD linear accelerators
| Linear accelerator |
|
| PDD |
|---|---|---|---|
| Elekta Synergy Platform | 2.37 | 1.320 | 99.172 |
| 3.17 | 1.802 | 98.810 | |
| 3.84 | 2.204 | 99.854 | |
| 4.68 | 2.708 | 99.672 | |
| 5.76 | 3.356 | 98.684 | |
| Elekta Versa HD | 2.48 | 1.390 | 99.756 |
| 3.23 | 1.839 | 99.606 | |
| 4.00 | 2.301 | 99.616 | |
| 4.75 | 2.750 | 99.760 | |
| 5.96 | 3.473 | 98.752 |
The results of beam quality factor calculation for each chamber by Technical Report Series (TRS)‐398 protocol in this work
|
| ||||||
|---|---|---|---|---|---|---|
| Linear accelerator | Energy (MeV) |
| PTW 30013 | IBA CC13 | Exradin A1Sl | Exradin A11 |
| Elekta Synergy Platform | 6 | 2.37 | – | – | – | 0.932 |
| 8 | 3.17 | – | – | – | 0.921 | |
| 10 | 3.84 | 0.911 | 0.920 | 0.914 | 0.915 | |
| 12 | 4.68 | 0.908 | 0.918 | 0.913 | 0.907 | |
| 15 | 5.76 | 0.905 | 0.914 | 0.912 | 0.899 | |
| Elekta Versa HD | 6 | 2.484 | – | – | – | 0.930 |
| 8 | 3.231 | – | – | – | 0.921 | |
| 10 | 4.002 | 0.911 | 0.920 | 0.914 | 0.913 | |
| 12 | 4.750 | 0.908 | 0.918 | 0.913 | 0.906 | |
| 15 | 5.955 | 0.904 | 0.913 | 0.912 | 0.897 | |
The results of beam quality factor calculation for each chamber by modified calibration in this work
|
| ||||||
|---|---|---|---|---|---|---|
| Linear accelerator | Energy (MeV) |
| PTW 30013 | IBA CC13 | Exradin A1Sl | Exradin A11 |
| Elekta Synergy Platform | 6 | 2.37 | 1.033 | 1.027 | 1.033 | 1.042 |
| 8 | 3.17 | 1.022 | 1.019 | 1.024 | 1.030 | |
| 10 | 3.84 | 1.015 | 1.015 | 1.019 | 1.022 | |
| 12 | 4.68 | 1.010 | 1.010 | 1.013 | 1.014 | |
| 15 | 5.760 | 1.005 | 1.005 | 1.007 | 1.007 | |
| Elekta Versa HD | 6 | 2.484 | 1.031 | 1.026 | 1.031 | 1.040 |
| 8 | 3.231 | 1.021 | 1.019 | 1.024 | 1.029 | |
| 10 | 4.002 | 1.014 | 1.014 | 1.017 | 1.020 | |
| 12 | 4.750 | 1.009 | 1.010 | 1.012 | 1.014 | |
| 15 | 5.955 | 1.004 | 1.004 | 1.006 | 1.006 | |
FIGURE 1The results of the beam quality conversion factor for each ionization chamber using modified calibration and Technical Report Series (TRS)‐398
FIGURE 2The result of dose per monitor unit (MU) obtained with cylindrical and parallel‐plate chamber using a modified calibration
Ratio of absorbed dose to water of modified calibration and Technical Report Series (TRS)‐398 () obtained from three types of cylindrical chamber used in this work
| Elekta Synergy Platform | Elekta Versa HD | ||||
|---|---|---|---|---|---|
| Type of chamber | Energy (MeV) |
| Discrepancy (%) |
| Discrepancy (%) |
| PTW 30013 | 6 | – | – | – | – |
| 8 | – | – | – | – | |
| 10 | 0.999 | ‐0.1 | 0.994 | ‐0.6 | |
| 12 | 0.997 | ‐0.3 | 0.994 | ‐0.6 | |
| 15 | 0.999 | ‐0.1 | 0.996 | ‐0.4 | |
| IBA CC13 | 6 | – | – | – | – |
| 8 | – | – | – | – | |
| 10 | 0.988 | ‐1.2 | 0.987 | ‐1.3 | |
| 12 | 0.997 | ‐0.3 | 0.988 | ‐1.2 | |
| 15 | 0.996 | ‐0.4 | 0.985 | ‐1.5 | |
| Exradin A1Sl | 6 | – | – | – | – |
| 8 | – | – | – | – | |
| 10 | 1.017 | 1.7 | 1.018 | 1.8 | |
| 12 | 1.012 | 1.2 | 1.019 | 1.9 | |
| 15 | 1.012 | 1.2 | 1.013 | 1.3 | |
Note: Discrepancy (%) = (absorbed doses for modified calibration – absorbed doses for TRS‐398)/(absorbed doses for TRS‐398 dose).
FIGURE 3The result of average dose per monitor unit a from different types of chamber according to modified calibration and Technical Report Series (TRS)‐398
FIGURE 4The results of dose per monitor unit calculation from this work compared to the previous work