Jing Wu1,2, Changping Yang3, Yan Yin4, Linlin Liu5, Hui Wang5.
Abstract
AIM: To analyze the pooled incidence rate in repositioning surgery by considering different materials and designs.Entities:
Year: 2022 PMID: 35223091 PMCID: PMC8881179 DOI: 10.1155/2022/6699596
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Characteristics of toric IOLs included in the meta-analysis.
| Country | Company | Commercial name | Spherical power | Cylinder power | Design | Haptic | Material |
|---|---|---|---|---|---|---|---|
| USA | Alcon | Acrysof TIOL SN60TT | +6D∼+34D | +1.5D∼+6.0D | Single-piece | Loop | Hydrophobic acrylic |
| Acrysof IQ toric IOL SN6AT | +6D∼+34D | +1.5D∼+6.0D | Aspheric optic | Loop | Hydrophobic acrylic | ||
| USA | Abbott Medical Optics | TECNIS | +5D∼+34D | 1.00D, 1.50, 2.25, 3.00, 4.00D | Single-piece | Loop | Hydrophobic acrylic |
| USA | Staar | AA 4203TF/TL | +10D∼+28D | +2D, 3.5D | Single-piece | Plate | Silicone |
| Germany | Human Optics | Microsil 6116TU | −3D∼+30D | +2D∼+12D | 3-Piece | PMMA Z-design | Silicone |
Figure 1Flowchart demonstrating how the identified published studies were included in the meta-analysis.
Figure 2The forest plot displaying the pooled incidence rate of repositioning surgery of toric IOL.
Figure 3The forest plots displaying the pooled incidence rate of repositioning surgery of (a) loop-haptic toric IOLs and (b) plate-haptic IOLs.
The pooled incidence rate of repositioning surgery with different subgroups.
| Variable | Number of articles | Case/total | Pooled estimate [95% CI] | Heterogeneity |
| OR (95% CI) |
|
|---|---|---|---|---|---|---|---|
| Total | 19 | 77/3220 | 2 [1–3] | 53 | 0.01 | ||
|
| |||||||
|
| |||||||
| Loop | 13 | 53/2855 | 2 [1–2] | 15 | 0.29 | 0.264 (0.160–0.436) | ≤0.001 |
| Plate | 5 | 23/344 | 6 [4–9] | 0 | 0.60 | 1.00 | |
|
| |||||||
|
| |||||||
| Silicone | 6 | 24/365 | 6 [4–9] | 0 | 0.63 | 1.00 | |
| Hydrophobic acrylic | 13 | 53/2855 | 2 [1–2] | 15 | 0.29 | 0.289 (0.164–0.441) | ≤0.001 |
|
| |||||||
|
| |||||||
| Acrysof | 9 | 27/1932 | 2 [1–2] | 0 | 0.63 | 0.198 (0.112–0.349) | 0.003 |
| Staar | 5 | 23/344 | 6 [4–9] | 0 | 0.60 | 1.00 | ≤0.001 |
| TECNIS | 2 | 24/819 | 3 [2–4] | 0 | 0.58 | 0.421 (0.234–0.757) | 0.003 |
| PhysIOL SA | 1 | 1/71 | |||||
| T-flex 623T | 1 | 1/33 | |||||
| Microsil 6116TU | 1 | 1/21 | |||||
The chi-square test was used for two sample rates and list χ2 test was used for multiple sample rates. P < 0.05 was considered statistically significant.
Figure 4The forest plots displaying the pooled incidence rate of repositioning surgery of (a) hydrophobic acrylic material IOLs and (b) silicone material IOLs.
Figure 5The forest plots displaying the pooled incidence rate of repositioning surgery of (a) Acrysof toric IOLs, (b) Staar IOLs, and (c) TECNIS toric IOLs.
Pairwise comparison of multiple sample rates by the partitions of the χ2 method.
| Subgroup | Sample | No. of samples |
|
|
|---|---|---|---|---|
| Acrysof | 27 | 1905 | 38.011 | ≤0.001 |
| Staar | 23 | 321 | ||
| Total | 40 | 2226 | ||
|
| ||||
| Acrysof | 27 | 1905 | 7.428 | 0.006 |
| TECNIS | 24 | 795 | ||
| Total | 51 | 2700 | ||
|
| ||||
| Staar | 23 | 321 | 8.811 | 0.003 |
| TECNIS | 24 | 795 | ||
| Total | 47 | 1116 | ||
P < 0.0125 was considered statistically significant.
Figure 6(a) The funnel plot displaying the pooled incidence rate of repositioning surgery of toric IOLs. (b) The filled funnel plot with pseudo-95% CI (the pseudo-95% confidence interval (CI) is computed as part of the analysis that produces the funnel plot).