Microincision Hydrophobic Acrylic Aspheric Toric Intraocular Lens for Astigmatism and Cataract Correction.

Hiroko Bissen-Miyajima, Kazuno Negishi, Osamu Hieda, Shigeru Kinoshita.   

Abstract

PURPOSE: To evaluate the efficacy and safety of a new acrylic one-piece toric intraocular lens (IOL).
METHODS: This prospective multicenter clinical trial included 93 eyes of 61 patients that were implanted with a hydrophobic acrylic toric IOL from 2010 to 2012 and followed for 1 year. This IOL uses the platform of a microincision one-piece aspheric IOL, the NY-60 IOL (HOYA, Tokyo, Japan), with three increments in cylindrical power (NHT15, 1.5 diopters [D]; NHT23, 2.25 D; and NHT30, 3.0 D). The inclusion criterion was preoperative corneal astigmatism from 0.75 to 3.00 D. The primary endpoint was uncorrected distance visual acuity (UDVA) of 0.0 logMAR (20/20 Snellen) or better 6 months postoperatively. In addition to UDVA, corrected distance visual acuity (CDVA), residual astigmatism, stability of the IOL alignment, need of realignment, and the rate of Nd:YAG laser capsulotomy were evaluated up to 1 year postoperatively. Errors in astigmatic correction were assessed using Alpin's vector analysis.
RESULTS: The primary endpoint was achieved in 54.8% of eyes. One year postoperatively, the logMAR UDVAs were 0.02 ± 0.13, 0.05 ± 0.17, and 0.09 ± 0.14 with models NHT15, NHT23, and NHT30, which corresponds to 0.96 (19/20 Snellen), 0.89 (18/20 Snellen), and 0.82 (16/20 Snellen), respectively. One year postoperatively, the residual astigmatism was 0.66 ± 0.58 D. In each evaluation, the mean absolute change in the position of the axis mark was between 1.93° and 2.32°. Three eyes required repositioning of the IOL axis and 2 eyes received Nd:YAG laser capsulotomy. The correction error showed an undercorrection with against-the-rule astigmatism and overcorrection with with-the-rule astigmatism.
CONCLUSIONS: The new one-piece toric IOL provided desirable clinical outcomes and stability in eyes with corneal astigmatism. Copyright 2015, SLACK Incorporated.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26046701     DOI: 10.3928/1081597X-20150521-01

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  3 in total

1.  Rotational stability of modified toric intraocular lens.

Authors:  Ryoko Osawa; Tetsuro Oshika; Masahiko Sano; Takuma Yuguchi; Tadayoshi Kaiya
Journal:  PLoS One       Date:  2021-03-01       Impact factor: 3.240

Review 2.  Impact of Material and Lens Design on Repositioning Surgery of Toric Intraocular Lenses: A Single-Arm Meta-Analysis.

Authors:  Jing Wu; Changping Yang; Yan Yin; Linlin Liu; Hui Wang
Journal:  J Ophthalmol       Date:  2022-01-27       Impact factor: 1.909

3.  Comparison of two techniques for toric intraocular lens implantation: hydroimplantation versus ophthalmic viscosurgical devices.

Authors:  Yueqin Chen; Qian Cao; Chunyan Xue; Zhenping Huang
Journal:  BMC Ophthalmol       Date:  2018-04-24       Impact factor: 2.209

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.