Pablo Lozano1, Marta Peña1, Mariano Herrero-Climent2, Jose Vicente Rios-Santos1, Blanca Rios-Carrasco1, Aritza Brizuela3, Javier Gil4,5.
Abstract
The procedure generally used to remove bacterial biofilm adhering to the surface of titanium on dental implants is implantoplasty. This treatment is based on the machining of the titanium surface to remove bacterial plaque. In this study, we used 60 grade 4 titanium implants and performed the implantoplasty protocol. Using X-ray diffraction, we determined the stresses accumulated in each of the as-received, machined and debris implants. The resistance to corrosion in open circuit and potentiodynamically in physiological medium has been determined, and the corrosion potentials and intensities have been determined. Tests have been carried out to determine ion release by ICP-MS at different immersion times. The results show that the corrosion resistance and the release of titanium ions into the medium are related to the accumulated energy or the degree of deformation. The titanium debris exhibit compressive residual stresses of -202 MPa, the implant treated with implantoplasty -120 MPa, and as-received -77 MPa, with their corrosion behavior resulting in corrosion rates of 0.501, 0.77, and 0.444 mm/year, respectively. Debris is the material with the worst corrosion resistance and the one that releases the most titanium ions to the physiological medium (15.3 ppb after 21 days vs. 7 ppb for as-received samples). Pitting has been observed on the surface of the debris released into the physiological environment. This behavior should be taken into account by clinicians for the good long-term behavior of implants with implantoplasty.Entities:
Keywords: corrosion; debris; implantoplasty; residual stresses; titanium ion release
Year: 2022 PMID: 35208101 PMCID: PMC8875252 DOI: 10.3390/ma15041563
Source DB: PubMed Journal: Materials (Basel) ISSN: 1996-1944 Impact factor: 3.623
Figure 1(A) Dental implant with peri-implantitis, (B) Resective surgery. (C) Polishing of the titanium surface to remove plaque. Implantoplasty process.
Figure 2Instruments used for the implantoplasty.
Chemical composition of Hank’s solution.
| Chemical Product | Composition (mM) |
|---|---|
| K2HPO4 | 0.44 |
| KCl | 5.4 |
| CaCl2 | 1.3 |
| Na2HPO4 | 0.25 |
| NaCl | 137 |
| NaHCO3 | 4.2 |
| MgSO4 | 1.0 |
| C6H12O6 | 5.5 |
Figure 3Corrosion resistance equipment.
Figure 4(a) As-received titanium dental implant. (b) Passivated surface on the titanium.
Figure 5(a) Titanium dental implant after implatoplasty porcess. (b) Slippimg bands produce by the high stress applied.
Figure 6(a) Titanium debris produced by implantoplasty in titanium dental implant. (b) Widmanstatten microstucture of cp Ti with high deformation.
Surface residual stresses calculated at the different samples.
| Material | σ (MPa) |
|---|---|
| As-received | −77.2 ± 5.2 |
| Implantoplasty | −120.0 ± 10.3 |
| Debris | −202.1 ± 12.2 |
Electrochemical and corrosion parameters assessed for Ti alloy meshes with different passivation treatments.
| Samples | EOCP
| icorr
| Rp
| ECORR
| Vc
|
|---|---|---|---|---|---|
| As-received | −195 ± 9 | 0.049 ± 0.007 |
| −340 ± 32 |
|
| Implantoplasty | −273 ± 10 | 0.056 ± 0.005 |
| −368 ± 47 |
|
| Debris | −334 ± 17 | 0.063 ± 0.009 |
| −411 ± 21 |
|
Figure 7Open-circuit corrosion potential versus time of immersion.
Figure 8Potentiodynamic curves for the different samples studied.
Figure 9Pitting corrosion marks produced after completing the potentiodynamic test on a debris titanium released aby the implantoplasty process.
Ti ion release (ppb) at different incubation times in Hank’s solution.
| Mesh | 1 Day | 3 Days | 7 Days | 14 Days | 21 Days |
|---|---|---|---|---|---|
| As-receieved | 1.3 ± 0.2 | 2.7 ± 0.5 | 2,9 ± 0.5 | 4.5 ± 0.4 | 7.0 ± 0.6 |
| Implantoplasty | 1.7 ± 0.4 | 3.3 ± 0.4 | 4.1 ± 0.2 | 5.7 ± 0.3 | 9.1 ± 0.5 |
| Debris | 2.9 ± 0.8 | 4.8 ± 0.8 | 5.3 ± 0.9 | 10.4 ± 5.9 | 15.3 ± 1.9 |