Isabella D Baur1, Gerd U Auffarth1, Grzegorz Łabuz1, Christian S Mayer1, Ramin Khoramnia1.
Abstract
PURPOSE: We present the case of implantation of two different Extended depth of focus intraocular lenses (EDoF IOLs) in a patient with a history of unilateral intrastromal femtosecond laser treatment for presbyopia correction (Intracor). OBSERVATIONS: The patient reported decreasing visual acuity at near distance and increasing spectacle dependence. Ten years earlier, he had Intracor treatment for presbyopia correction in his left eye. Corrected distance visual acuity (CDVA) was 0.08 logMAR for the right eye and 0.16 logMAR for the left eye. Apart from dysfunctional lens syndrome, the examination results were unremarkable. Phacoemulsification and subsequent IOL implantation was performed in both eyes. The left eye was implanted with an IC-8 (AcuFocus, Irvine, CA, USA), whereas the fellow eye was implanted with an AcrySof IQ Vivity IOL (Alcon, Fort Worth, TX, USA). Postoperatively, CDVA improved to 0.02 and 0.04 logMAR for the right and left eye. Uncorrected intermediate visual acuity (UIVA) was 0.24 logMAR for the right eye and -0.04 logMAR for the left eye, binocular UIVA was -0.04 logMAR. The patient reported a low level of photic phenomena and spectacle independence for far and intermediate distance. CONCLUSIONS AND IMPORTANCE: Combined implantation of a non-diffractive and a small-aperture EDoF lens after previous unilateral Intracor treatment could successfully improve visual acuity at far and intermediate distance.Entities:
Keywords: Cataract surgery; Extended depth of focus IOL; Refractive surgery; Small-aperture IOL; Spectacle independence
Year: 2022 PMID: 35198820 PMCID: PMC8844772 DOI: 10.1016/j.ajoc.2022.101398
Source DB: PubMed Journal: Am J Ophthalmol Case Rep ISSN: 2451-9936
Results of preoperative visual acuity testing in logMAR. OD: right eye, OS: left eye, OU: both eyes, UDVA: uncorrected distance visual acuity, CDVA: corrected distance visual acuity, UNVA: uncorrected near visual acuity, DCNVA: distance corrected near visual acuity.
| Subjective refraction | UDVA | CDVA | UNVA | DCNVA | |
|---|---|---|---|---|---|
| OD | +1.75/-0.50/80° | 0.70 | 0.08 | 0.80 | |
| OS | +1.75/−/− | 0.30 | 0.16 | 0.56 | |
| OU | 0.20 | 0.04 | 0.62 |
Fig. 1Corneal tomography of the left eye acquired with the Oculus Pentacam (OCULUS Optikgerate GmbH, Wetzlar, Germany). The image shows the central corneal steepening that is typically found after Intracor treatment.
Fig. 2Intraoperative image with displayed digital marking system of the right eye (left part of the image) and the left eye (right part of the image).
Results of visual acuity testing at 3 months postoperatively in logMAR. OD: right eye, OS: left eye, OU: both eyes, UDVA: uncorrected distance visual acuity, CDVA: corrected distance visual acuity, UNVA: uncorrected near visual acuity, DCNVA: distance corrected near visual acuity.
| Subjective refraction | UDVA | CDVA | UNVA | DCNVA | UIVA | DCIVA | |
|---|---|---|---|---|---|---|---|
| OD | +0.25/-0.25/100° | 0.08 | 0.02 | 0.54 | 0.60 | 0.24 | 0.14 |
| OS | −0.50/-0.25/70° | 0.16 | 0.04 | 0.32 | 0.40 | −0.04 | 0.20 |
| OU | 0.02 | −0.02 | 0.24 | 0.34 | −0.04 | 0.12 |
Fig. 3Monocular and binocular defocus curves. Defocus curve testing was performed with distance correction. For binocular testing, visual acuity was 0.2 logMAR or better from +1.25 to −2.0 diopters.
Fig. 4Result of the Halo & Glare simulation (Eyeland-Design Network GmbH, Vreden, Germany). The patient reported a starburst type halo (size 34 and intensity 37 on a scale from 0 to 100). The patient did not perceive glare.