Weiyan Zhou1, Yuhang Deng1,2, Haijian Zhao1, Chuanbao Zhang1.
Abstract
BACKGROUND: Accurate measurements of serum insulin and C-peptide are needed for the therapy and classification of diabetes. This study investigated the status of serum insulin and C-peptide measurements in China by analyzing the results of five pooled serum samples measured in 94 laboratories.Entities:
Keywords: C-peptide; Diabetes mellitus; Harmonization; Insulin; Standardization
Mesh:
Substances:
Year: 2022 PMID: 35177563 PMCID: PMC8859554 DOI: 10.3343/alm.2022.42.4.428
Source DB: PubMed Journal: Ann Lab Med ISSN: 2234-3806 Impact factor: 4.941
Fig. 1Intra-laboratory %CVs for insulin and C-peptide measurements. (A) Intra-laboratory %CVs of insulin measurements, showing a decreasing trend with increasing concentration of insulin (ANOVA: P=0.0059; 202111 vs. 202113 P=0.0255; 202111 vs. 202115 P=0.0063). (B) Intra-laboratory %CVs of C-peptide measurements, with no significant trend observed (ANOVA: P=0.702). The serum samples are shown in order of insulin and C-peptide concentration.
Intra-laboratory and inter-laboratory %CVs of insulin and C-peptide measurements
| Samples | Insulin | C-peptide | ||||||
|---|---|---|---|---|---|---|---|---|
|
|
| |||||||
| Target (nmol/L) | Intra-laboratory %CV | Inter-laboratory %CV | Target (nmol/L) | Intra-laboratory %CV | Inter-laboratory %CV | |||
|
|
| |||||||
| Range | Mean | Range | Mean | |||||
| 202111 | 0.071 | 0.7–21.8 | 3.4 | 64.3 | 0.319 | 0.0–8.8 | 2.5 | 21.8 |
| 202112 | 0.123 | 0.7–18.0 | 2.8 | 8.4 | 0.937 | 0.3–10.1 | 2.3 | 17.4 |
| 202113 | 0.357 | 0.2–15.7 | 2.4 | 13.1 | 2.376 | 0.3–10.7 | 2.3 | 14.8 |
| 202114 | 0.386 | 0.5–15.6 | 2.5 | 10.0 | 3.081 | 0.3–11.1 | 2.3 | 14.4 |
| 202115 | 0.932 | 0.3–16.5 | 2.2 | 13.3 | 4.308 | 0.3–8.1 | 2.2 | 13.6 |
Abbreviations: Intra-lab, intra-laboratory; Inter-lab, inter-laboratory.
Fig. 2Insulin measurement results obtained using different measurement systems for the five pooled samples (A–E). (F) Mean results for the measurement systems. The X-axis presents the five pooled samples and the Y-axis represents the mean value of laboratories using the same measurement system. Each symbol represents a different measurement system, and the serum samples are shown in order of insulin concentration.
Fig. 3C-peptide measurement results of different measurement systems for the five pooled samples (A–E). (F) Mean results of the measurement systems. The X-axis shows the five pooled samples and the Y-axis represents the mean value of laboratories using the same measurement system. Each symbol represents a different measurement system, and the serum samples are shown in order of C-peptide concentration.
Mean results for five pooled serum samples analyzed using different measurement systems, and the intra- and inter-MS %CVs of insulin and C-peptide measurements
| Sample | Representative values before recalibration (nmol/L) | Intra-MS %CV | Inter-MS %CV | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
|
| ||||||||||||
| Roche | Abbott | Siemens | Beckman | Snibe | Mindray | Roche | Abbott | Siemens | Beckman | Snibe | Mindray | ||
| Insulin | |||||||||||||
| 202111 | 0.043 | 0.170 | 0.127 | 0.091 | 0.064 | 0.038 | 5.6 | 2.8 | 4.9 | 6.2 | 7.5 | 8.8 | 58.3 |
| 202112 | 0.126 | 0.137 | 0.124 | 0.102 | 0.113 | 0.109 | 4.8 | 2.9 | 4.3 | 5.1 | 2.4 | 5.2 | 10.7 |
| 202113 | 0.391 | 0.295 | 0.302 | 0.265 | 0.346 | 0.365 | 4.6 | 2.2 | 6.3 | 5.0 | 3.8 | 6.7 | 14.5 |
| 202114 | 0.414 | 0.345 | 0.346 | 0.302 | 0.369 | 0.395 | 4.7 | 2.9 | 6.6 | 5.2 | 2.8 | 5.7 | 11.0 |
| 202115 | 1.006 | 0.749 | 0.802 | 0.699 | 0.924 | 1.081 | 4.4 | 2.9 | 6.4 | 4.0 | 1.4 | 9.0 | 17.3 |
| C-peptide | |||||||||||||
| 202111 | 0.363 | 0.265 | 0.195 | 0.173 | 0.385 | 0.256 | 4.5 | 3.7 | 6.9 | 9.6 | 6.7 | 4.7 | 31.6 |
| 202112 | 1.031 | 0.847 | 0.634 | 0.555 | 1.034 | 0.915 | 4.8 | 3.9 | 6.5 | 9.6 | 7.6 | 5.9 | 24.1 |
| 202113 | 2.529 | 2.227 | 1.723 | 1.511 | 2.696 | 2.562 | 4.7 | 4.0 | 8.2 | 11.7 | 10.4 | 4.2 | 22.1 |
| 202114 | 3.270 | 2.813 | 2.255 | 2.161 | 3.530 | 3.359 | 4.7 | 6.5 | 7.5 | 10.5 | 9.3 | 4.3 | 20.2 |
| 202115 | 4.546 | 4.066 | 3.293 | 2.804 | 4.688 | 4.755 | 4.8 | 4.6 | 6.5 | 9.5 | 8.7 | 4.0 | 20.1 |
Abbreviation: MS, measurement systems.
Differences between measurement systems after recalibration
| Samples | Representative values after recalibration (nmol/L) | Inter-MS %CV (after recalibration) | |||||
|---|---|---|---|---|---|---|---|
|
| |||||||
| Roche | Abbott | Siemens | Beckman | Snibe | Mindray | ||
| Insulin | |||||||
| 202111 | 0.047 | 0.159 | 0.122 | 0.104 | 0.073 | 0.063 | 44.1 |
| 202113 | 0.346 | 0.315 | 0.318 | 0.325 | 0.336 | 0.319 | 3.8 |
| 202114 | 0.367 | 0.376 | 0.367 | 0.373 | 0.358 | 0.342 | 3.6 |
| C-peptide | |||||||
| 202111 | 0.230 | 0.260 | 0.310 | 0.294 | 0.269 | 0.289 | 10.4 |
| 202113 | 2.195 | 2.204 | 2.142 | 2.191 | 2.287 | 2.204 | 2.1 |
| 202114 | 2.867 | 2.785 | 2.781 | 3.113 | 3.015 | 2.867 | 4.6 |
Abbreviation: MS, measurement systems.