| Literature DB >> 35177029 |
Xianjun Han1, Nan Luo1, Lixue Xu1, Jiaxin Cao1, Ning Guo2, Yi He1, Min Hong3, Xibin Jia4, Zhenchang Wang1, Zhenghan Yang5.
Abstract
BACKGROUND: To investigate the influence of artificial intelligence (AI) based on deep learning on the diagnostic performance and consistency of inexperienced cardiovascular radiologists.Entities:
Keywords: Artificial intelligence (AI); CCTA; Coronary artery disease; Coronary stenosis; Inexperience readers
Mesh:
Year: 2022 PMID: 35177029 PMCID: PMC8851787 DOI: 10.1186/s12880-022-00756-y
Source DB: PubMed Journal: BMC Med Imaging ISSN: 1471-2342 Impact factor: 1.930
Fig. 1Flow diagram of the patient selection process. CCTA coronary computed tomography angiography, ICA invasive coronary angiography
Clinical features of the 196 patients
| Parameter | Men | Women |
|---|---|---|
| Total no. of patients | 124 (63.3%) | 72 (36.7%) |
| Age (y) | 61.72 ± 8.50 | 67.63 ± 8.22 |
| Median body mass index (kg/m2) | 25.58 ± 3.13 | 25.93 ± 4.23 |
| No. of patients who smoked | 93 (75.0%) | 7 (9.7%) |
| No. of patients who drank | 59 (47.6%) | 1 (1.4%) |
| No. of patients with hypertension | 89 (72.4%) | 54 (75.0%) |
| No. of patients with hyperlipidaemia | 67 (59.3%) | 33 (50.8%) |
| No. of patients with diabetes | 52 (43.3%) | 27 (37.5%) |
| No. of patients with known CHD | 11 (8.9%) | 10 (13.9%) |
| Heart rate of patients | 63.90 ± 11.59 | 65.18 ± 12.52 |
No. = number. Age and heart rate are shown as the mean ± standard deviation or the median and range. CHD = coronary heart disease. Hypertension data were available for 195 patients (123 men and 72 women), hyperlipidaemia data were available for 178 patients (113 men and 65 women), and diabetes data were available for 192 patients (120 men and 72 women)
Fig. 2Single-centre retrospective case of a noncalcified plaque lesion in the proximal LAD that caused severe stenosis (white arrow). a CPR image reconstructed manually, b ICA image of the left coronary artery, c CPR, VRT and straightened images automatically reconstructed by the AI system. ICA invasive coronary angiography, LAD left anterior descending, LCx left circumflex
Diagnostic performance of the six readers and the AI system alone for the same set of patients
| Reader | Sensitivity (%) | Specificity (%) | PPV (%) | NPV (%) | Accuracy (%) |
|---|---|---|---|---|---|
| AI | 93.5 (88.1, 97.0) | 57.9 (44.1, 70.9) | 84.4 | 78.6 | 80.0 (73.1, 84.9) |
| Reader 1 | 70.5 (62.2, 77.9) | 54.4 (40.7, 67.6) | 79.0 | 43.1 | 69.1 (62.1, 75.4) |
| Reader 2 | 78.4 (70.6, 84.9) | 57. 9 (44.1, 70.9) | 82.0 | 52.4 | 69.0 (62.0, 75.4) |
| Reader 3 | 85.6 (78.7, 91.0) | 45.6 (32.4, 59.3) | 79.3 | 56.5 | 71.2 (64.3, 77.5) |
| Reader 4 | 87.1 (80.3, 92.1) | 40.4 (27.6, 54.2) | 78.1 | 56.1 | 67.8 (60.8, 74.3) |
| Reader 5 | 65.5 (56.9, 73.3) | 68.4 (54.8, 80.1) | 83.5 | 44.8 | 77.3 (70.8, 83.0) |
| Reader 6 | 77.7 (69.9, 84.3) | 64.9 (51.1, 77.1) | 84.4 | 54.4 | 75.6 (69.0, 81.5) |
Data in parentheses are 95% confidence intervals
AI artificial intelligence, PPV positive predictive value, NPV negative predictive value
Fig. 3AUC comparison of the six readers at the patient level
Diagnostic performance of the six readers and the AI system alone for the same set of vessels
| Reader | Sensitivity (%) | Specificity (%) | PPV (%) | NPV (%) | Accuracy (%) |
|---|---|---|---|---|---|
| AI | 78.1 (72.1, 83.3) | 82.5 (79.0, 85.7) | 65.9 | 89.7 | 84.5 (81.7, 87.0) |
| Reader 1 | 53.2 (46.4, 59.9) | 89.2 (86.2, 91.7) | 67.8 | 81.7 | 82.5 (79.6, 85.2) |
| Reader 2 | 61.2 (54.6, 67.6) | 78.3 (74.6, 81.8) | 54.9 | 82.7 | 79.5 (76.4, 82.3) |
| Reader 3 | 67.1 (60.6, 73.2) | 83.2 (79.8, 86.3) | 63.5 | 85.4 | 83.3 (80.4, 85.9) |
| Reader 4 | 69.3 (62.9, 75.2) | 81.3 (77.7, 84.5) | 61.7 | 85.9 | 81.3 (78.4, 84.1) |
| Reader 5 | 50.9 (44.2, 57.6) | 93.1 (90.5, 95.1) | 76.2 | 81.3 | 84.7 (81.9, 87.2) |
| Reader 6 | 64.2 (57.5, 70.4) | 89.8 (86.9, 92.3) | 73.2 | 85.2 | 85.0 (82.2, 87.5) |
Data in parentheses are 95% confidence intervals
AI artificial intelligence, PPV positive predictive value, NPV negative predictive value
Fig. 4AUC comparison of the six readers at the vessel level