Literature DB >> 35140856

Ten- to 14-Year Results of a High-Flex Rotating Platform Knee Implant: A Follow-Up Report of a Prospective Cohort.

Adit R Maniar1,2, Arpit Gajjar3, Nishit Bhatnagar4, Abhinav Mishra5,6, Rajesh N Maniar1,2.   

Abstract

BACKGROUND: High-flex posterior stabilised rotating platform (PSRP) implant was introduced to provide for deep knee flexion. Few short-term results have been reported, but there are no long-term outcomes reported.
METHODS: We prospectively followed 48 patients (53 knees) implanted with one such design. Inclusion criteria for implantation were patients with good pre-operative flexion and wishing to perform activities requiring deep knee flexion post-operatively, with femorotibial varus angle < 15° and having good flexion stability at trialling stage. Previously, we reported their outcomes at 2-6 years (FU-1). We now report their functional and radiological results at a minimum follow-up of 10 years (FU-2) in 39 patients (43 knees); 5 patients having died and 4 lost to follow-up.
RESULTS: The mean pre-operative flexion of 124° improved to 130° at FU-1 and to 134° at FU-2. Flexion of 130° or more was seen in 59.6% knees at FU-1 and 74.42% knees at FU-2. At FU-2 mean Knee score was 90.5 and Function score was 67.8. Incidence of patellofemoral symptoms increased from 7.7% at FU-1 to 11.36% at FU-2. There were no cases of bearing spin out, osteolysis or revision surgeries.
CONCLUSION: At a minimum 10-year follow-up, high-flexion PSRP design in selected patients yielded 100% survival. We recorded good knee flexion and knee society scores, with no case of spin out, implant loosening, osteolysis or revision surgery. Although deep knee flexion improved at longer follow-up, its use in ADL had reduced due to other age-related factors. There was increased incidence of patellofemoral symptoms. © Indian Orthopaedics Association 2021.

Entities:  

Keywords:  High flex; Knee arthroplasty; Outcomes; Posterior stabilised; Rotating platform; Ten-year results

Year:  2021        PMID: 35140856      PMCID: PMC8789968          DOI: 10.1007/s43465-021-00441-2

Source DB:  PubMed          Journal:  Indian J Orthop        ISSN: 0019-5413            Impact factor:   1.033


  21 in total

1.  Ten-year follow-up of a rotating-platform, posterior-stabilized total knee arthroplasty.

Authors:  Morteza Meftah; Amar S Ranawat; Chitranjan S Ranawat
Journal:  J Bone Joint Surg Am       Date:  2012-03-07       Impact factor: 5.284

2.  The P.F.C. sigma RP-F TKA designed for improved performance: a matched-pair study.

Authors:  Sanjay K Gupta; Amar S Ranawat; Vineet Shah; Bashir A Zikria; Joseph F Zikria; Chitranjan S Ranawat
Journal:  Orthopedics       Date:  2006-09       Impact factor: 1.390

3.  The natural history of anterior knee pain in 2 posterior-stabilized, modular total knee arthroplasty designs.

Authors:  Morteza Meftah; Amar S Ranawat; Chitranjan S Ranawat
Journal:  J Arthroplasty       Date:  2011-01-28       Impact factor: 4.757

4.  Prospective randomized comparison of high-flex and standard rotating platform total knee arthroplasty.

Authors:  William G Hamilton; Supatra Sritulanondha; C Anderson Engh
Journal:  J Arthroplasty       Date:  2011-07-01       Impact factor: 4.757

Review 5.  Ten-Year Follow-Up of High-Flexion Versus Conventional Total Knee Arthroplasty: A Matched-Control Study.

Authors:  Marc R Angerame; Catie L Eschen; Roseann M Johnson; Jason M Jennings; Douglas A Dennis
Journal:  J Arthroplasty       Date:  2021-03-06       Impact factor: 4.757

6.  High-flex rotating platform knee implants: two- to 6-year results of a prospective study.

Authors:  Rajesh N Maniar; Tushar Singhi
Journal:  J Arthroplasty       Date:  2011-09-23       Impact factor: 4.757

7.  Surgical technique: Lateral retinaculum release in knee arthroplasty using a stepwise, outside-in technique.

Authors:  Rajesh N Maniar; Tushar Singhi; Suyog Shashimohan Rathi; Jayesh Vinod Baviskar; Ravi Mohan Nayak
Journal:  Clin Orthop Relat Res       Date:  2012-06-16       Impact factor: 4.176

8.  Ten- to 13-Year Results of Mobile Bearing Posterior-Stabilized Rotating-Platform Knee Implants, Reported by Nondesigner Surgeon.

Authors:  Rajesh N Maniar; Tushar Singhi; Parul R Maniar; Vipan Kumar
Journal:  J Arthroplasty       Date:  2016-09-29       Impact factor: 4.757

9.  Lateral retinacular release rates in mobile- versus fixed-bearing TKA.

Authors:  Charles C Yang; Lee A McFadden; Douglas A Dennis; Raymond H Kim; Adrija Sharma
Journal:  Clin Orthop Relat Res       Date:  2008-08-16       Impact factor: 4.176

10.  Results of Total Knee Arthroplasty with NexGen LPS-flex Implant Using Navigation System (Brain Lab): Results with a 5-year Follow-up.

Authors:  Chul Hyung Kang; Kyung Jae Lee; Ki Cheor Bae; Chul Hyun Cho; Si Wook Lee; Hong Kwan Shin; Young Kook Lee; Ji Suk Bae
Journal:  Knee Surg Relat Res       Date:  2012-11-29
View more