Ten- to 13-Year Results of Mobile Bearing Posterior-Stabilized Rotating-Platform Knee Implants, Reported by Nondesigner Surgeon.

Rajesh N Maniar1, Tushar Singhi2, Parul R Maniar3, Vipan Kumar4.   

Abstract

BACKGROUND: Posterior-stabilized rotating-platform (PSRP) knee was designed with the purpose of improving postoperative flexion and stability. Its long-term performance has limited reports, which could show whether this purpose has been realized without deleterious effect. We report its long-term results at 10-13 years, of a previously studied cohort, evaluated longitudinally.
METHODS: A total of 133 consecutive PSRP implants, selected for 118 patients were studied. Twenty-one patients (24 knees) were deceased and 11 patients (12 knees) were lost to follow-up after 10 years. Eighty patients (97 knees) completed clinical evaluation, of whom 78 patients (88 knees) also completed radiological evaluation.
RESULTS: Mean flexion improved from 106.8° (30°-150°) to 127° (90°-155°). Forty-three percent patients could sit crosslegged, 32.5% could sit on floor, and 3.4% could squat. Mean knee subscore of Knee Society Score improved from 28 (1-59) to 96 (67-100). Mean function subscore improved from 53 (5-81) to 78 (-10 to 100). After 10 years, 5 patients had zero function score because of developing other debilitating medical illnesses. Twenty-one dead patients (24 knees), at their last follow-up (7.7 years), had satisfactory scores. No patient had spinout or revision. Radiologically, alignment was satisfactory and there was no osteolysis. Kaplan-Meier analysis showed 100% implant survival.
CONCLUSION: PSRP design, evaluated after 10 years of implantation in selected patients, had 100% survival with good flexion. Addition of post and cam to the original rotating platform design has provided good stability without untoward effects of wear or osteolysis at 10-13 years.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  knee arthroplasty; mobile bearing; posterior stabilized; results; rotating platform

Mesh:

Year:  2016        PMID: 27789096     DOI: 10.1016/j.arth.2016.09.001

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  2 in total

1.  Fifteen-year survival of the Cedior™ total knee prosthesis.

Authors:  Roger Erivan; Edouard Fadlallah; Guillaume Villatte; Aurélien Mulliez; Stéphane Descamps; Stéphane Boisgard
Journal:  Eur J Orthop Surg Traumatol       Date:  2019-07-06

2.  Ten- to 14-Year Results of a High-Flex Rotating Platform Knee Implant: A Follow-Up Report of a Prospective Cohort.

Authors:  Adit R Maniar; Arpit Gajjar; Nishit Bhatnagar; Abhinav Mishra; Rajesh N Maniar
Journal:  Indian J Orthop       Date:  2021-07-30       Impact factor: 1.033