Fan Zhang1, Mi Zhou1, Jing Liu1, Lu Yue1, Lihua Deng1, Zhijian Xu1, Gang Wang1.
Abstract
BACKGROUND: The main objective in this paper is to build up a simplified benchmarking procedure to determine the incident electron beam parameters independently using a Monte Carlo linear accelerator model.Entities:
Keywords: Monte Carlo; dose distribution; electron beam parameter; photon beam
Year: 2020 PMID: 35117402 PMCID: PMC8798818 DOI: 10.21037/tcr.2019.12.02
Source DB: PubMed Journal: Transl Cancer Res ISSN: 2218-676X Impact factor: 1.241
Figure 1Percentage depth dose analysis. Measured and calculated depth doses for optimal electron energy (A); the radial intensity of electron beams (given by the standard deviation of a Gaussian distribution), σ, was 0.1 cm. The mean angular spread was 0°. Dose differences between simulations and measurements were shown in (B,C,D).
The linear regressions for rest electron energies starting from 5.4 to 6.5 MeV
| Energy/MeV | Least-square linear regression |
|---|---|
| 5.4 | y=–0.060x+0.434 |
| 5.5 | y=–0.057x+0.693 |
| 5.6 | y=–0.042x+0.582 |
| 5.7 | y=–0.017x–0.052 |
| 5.8 | y=0.0097x–0.849 |
| 6.1 | y=–0.04x+1.628 |
| 6.2 | y=0.022x–0.223 |
| 6.3 | y=0.209x+0.167 |
| 6.4 | y=0.019x+0.139 |
| 6.5 | y=–0.003X+1.149 |
Figure 2Percentage depth dose calculated with different radial intensities compared with measured doses.
Figure 3Dose differences between calculated and measured percentage depth doses for different electron energies.
Figure 4Dose profiles calculated with different energies compared with measured doses. The radial intensity, σ, was 0.1cm, and the mean angular spread was 0°.
Figure 5Dose profiles calculated with different radial intensities compared with measured doses. The electron energy was 5.95 MeV, and the mean angular spread was 0°.
Local dose differences between calculated and measured dose at a point with 5.2 cm off-CAX distance
| Radial intensity/cm | 0.0 | 0.03 | 0.05 | 0.08 | 0.10 | 0.13 | 0.15 | 0.18 | 0.20 |
|---|---|---|---|---|---|---|---|---|---|
| Dose difference/% | 9.84 | 7.85 | 7.35 | 3.26 | 0.15 | 2.01 | 3.45 | 4.18 | 8.03 |
Figure 6Dose profiles were calculated with different mean angular spreads compared with measured doses.
The electron beam parameters and effects on the dose distributions
| Electron beam parameter | Effect on | |
|---|---|---|
| Depth dose | Lateral dose profile | |
| Energy | Yesa | No |
| Radial intensity (σ) | No | Yesb |
| Mean angular spread | No | Yesc |
a, energy out of 5.5–6.4 MeV range results non-ignorable depth dose differences; b, radial intensity of σ value in the range 0.0–0.19 cm has negligible effects on the dose profiles in the main field but a notable difference with measured dose in the penumbra areas; c, 0.3° mean angular spread results an acceptable agreement for dose profiles at different depths.