Monte Carlo validation of the TrueBeam 10XFFF phase-space files for applications in lung SABR.

Tony Teke1, Cheryl Duzenli2, Alanah Bergman2, Francis Viel2, Parmveer Atwal2, Ermias Gete2.   

Abstract

PURPOSE: To establish the clinical acceptability of universal Monte Carlo phase-space data for the 10XFFF (flattening filter free) photon beam on the Varian TrueBeam Linac, including previously unreported data for small fields, output factors, and inhomogeneous media. The study was particularly aimed at confirming the suitability for use in simulations of lung stereotactic ablative radiotherapy treatment plans.
METHODS: Monte Carlo calculated percent depth doses (PDDs), transverse profiles, and output factors for the TrueBeam 10 MV FFF beam using generic phase-space data that have been released by the Varian MC research team were compared with in-house measurements and published data from multiple institutions (ten Linacs from eight different institutions). BEAMnrc was used to create field size specific phase-spaces located underneath the jaws. Doses were calculated with DOSXYZnrc in a water phantom for fields ranging from 1 × 1 to 40 × 40 cm(2). Particular attention was paid to small fields (down to 1 × 1 cm(2)) and dose per pulse effects on dosimeter response for high dose rate 10XFFF beams. Ion chamber measurements were corrected for changes in ion collection efficiency (P(ion)) with increasing dose per pulse. MC and ECLIPSE ANISOTROPIC ANALYTICAL ALGORITHM (AAA) calculated PDDs were compared to Gafchromic film measurement in inhomogeneous media (water, bone, lung).
RESULTS: Measured data from all machines agreed with Monte Carlo simulations within 1.0% and 1.5% for PDDs and in-field transverse profiles, respectively, for field sizes >1 × 1 cm(2) in a homogeneous water phantom. Agreements in the 80%-20% penumbra widths were better than 2 mm for all the fields that were compared. For all the field sizes considered, the agreement between their measured and calculated output factors was within 1.1%. Monte Carlo results for dose to water at water/bone, bone/lung, and lung/water interfaces as well as within lung agree with film measurements to within 2.8% for 10 × 10 and 3 × 3 cm(2) field sizes. This represents a significant improvement over the performance of the ECLIPSE AAA.
CONCLUSIONS: The 10XFFF phase-space data offered by the Varian Monte Carlo research team have been validated for clinical use using measured, interinstitutional beam data in water and with film dosimetry in inhomogeneous media.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26632043     DOI: 10.1118/1.4935144

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Monte carlo model and output factors of elekta infinity™ 6 and 10 MV photon beam.

Authors:  Sitti Yani; Indra Budiansah; Mohamad Fahdillah Rhani; Freddy Haryanto
Journal:  Rep Pract Oncol Radiother       Date:  2020-04-28

2.  Comparison of Air-Gaps Effect in a Small Cavity on Dose Calculation for 6 MV Linac.

Authors:  A Azzi; D Ryangga; S A Pawiro
Journal:  J Biomed Phys Eng       Date:  2021-02-01

3.  Benchmarking of electron beam parameters based on Monte Carlo linear accelerator simulation.

Authors:  Fan Zhang; Mi Zhou; Jing Liu; Lu Yue; Lihua Deng; Zhijian Xu; Gang Wang
Journal:  Transl Cancer Res       Date:  2020-02       Impact factor: 1.241

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.