William L Walter1, Neri A Baker2, Daniel Marsden-Jones2, Ameneh Sadeghpour2.
Abstract
PURPOSE: The purposes of the present research were to assess the accuracy and usability of the inertial navigation system (INS).Entities:
Keywords: THA; accurate; imageless; inertial; intraoperative; navigation
Year: 2022 PMID: 35115849 PMCID: PMC8806051 DOI: 10.2147/MDER.S339669
Source DB: PubMed Journal: Med Devices (Auckl) ISSN: 1179-1470
Figure 1INS graphical user interface.
Figure 2INS components.
Figure 3Device activation.
Figure 4Device interface with pelvis.
Figure 5Device interface with impactor.
Figure 6Patient position for registration.
Figure 7Phantom model.
Calibrated Phantom Angles for Each Position
| Phantom Position | Radiographic Angle | |||
|---|---|---|---|---|
| Base Position | Fan Position | Shaft Position | Inclination (°) | Anteversion (°) |
| 0 | 0 | N,E,S,W | 0 | 0.014 |
| 0 | 2 | N,E,S,W | 0 | 30.004 |
| 0 | 4 | N,E,S,W | 0 | 60.007 |
| 0 | 6 | N,E,S,W | 0 | 89.903 |
| 0 | 8 | N,E,S,W | 0 | 119.981 |
| 0 | 10 | N,E,S,W | 0 | 149.994 |
| 0 | 12 | N,E,S,W | 0 | 179.999 |
| 2 | 0 | N,E,S,W | 19.9965 | 0.014 |
| 2 | 2 | N,E,S,W | 19.9965 | 30.004 |
| 2 | 4 | N,E,S,W | 19.9965 | 60.007 |
| 2 | 6 | N,E,S,W | 19.9965 | 89.903 |
| 2 | 8 | N,E,S,W | 19.9965 | 119.981 |
| 2 | 10 | N,E,S,W | 19.9965 | 149.994 |
| 2 | 12 | N,E,S,W | 19.9965 | 179.999 |
| 4 | 0 | N,E,S,W | 39.9955 | 0.014 |
| 4 | 2 | N,E,S,W | 39.9955 | 30.004 |
| 4 | 4 | N,E,S,W | 39.9955 | 60.007 |
| 4 | 6 | N,E,S,W | 39.9955 | 89.903 |
| 4 | 8 | N,E,S,W | 39.9955 | 119.981 |
| 4 | 10 | N,E,S,W | 39.9955 | 149.994 |
| 4 | 12 | N,E,S,W | 39.9955 | 179.999 |
| 6 | 0 | N,E,S,W | 59.9945 | 0.014 |
| 6 | 2 | N,E,S,W | 59.9945 | 30.004 |
| 6 | 4 | N,E,S,W | 59.9945 | 60.007 |
| 6 | 6 | N,E,S,W | 59.9945 | 89.903 |
| 6 | 8 | N,E,S,W | 59.9945 | 119.981 |
| 6 | 10 | N,E,S,W | 59.9945 | 149.994 |
| 6 | 12 | N,E,S,W | 59.9945 | 179.999 |
| 8 | 0 | N,E,S,W | 79.996 | 0.014 |
| 8 | 2 | N,E,S,W | 79.996 | 30.004 |
| 8 | 4 | N,E,S,W | 79.996 | 60.007 |
| 8 | 6 | N,E,S,W | 79.996 | 89.903 |
| 8 | 8 | N,E,S,W | 79.996 | 119.981 |
| 8 | 10 | N,E,S,W | 79.996 | 149.994 |
| 8 | 12 | N,E,S,W | 79.996 | 179.999 |
| 10 | 0 | N,E,S,W | 99.9995 | 0.014 |
| 10 | 2 | N,E,S,W | 99.9995 | 30.004 |
| 10 | 4 | N,E,S,W | 99.9995 | 60.007 |
| 10 | 6 | N,E,S,W | 99.9995 | 89.903 |
| 10 | 8 | N,E,S,W | 99.9995 | 119.981 |
| 10 | 10 | N,E,S,W | 99.9995 | 149.994 |
| 10 | 12 | N,E,S,W | 99.9995 | 179.999 |
| 12 | 0 | N,E,S,W | 120.0045 | 0.014 |
| 12 | 2 | N,E,S,W | 120.0045 | 30.004 |
| 12 | 4 | N,E,S,W | 120.0045 | 60.007 |
| 12 | 6 | N,E,S,W | 120.0045 | 89.903 |
| 12 | 8 | N,E,S,W | 120.0045 | 119.981 |
| 12 | 10 | N,E,S,W | 120.0045 | 149.994 |
| 12 | 12 | N,E,S,W | 120.0045 | 179.999 |
| 14 | 0 | N,E,S,W | 140.004 | 0.014 |
| 14 | 2 | N,E,S,W | 140.004 | 30.004 |
| 14 | 4 | N,E,S,W | 140.004 | 60.007 |
| 14 | 6 | N,E,S,W | 140.004 | 89.903 |
| 14 | 8 | N,E,S,W | 140.004 | 119.981 |
| 14 | 10 | N,E,S,W | 140.004 | 149.994 |
| 14 | 12 | N,E,S,W | 140.004 | 179.999 |
| 16 | 0 | N,E,S,W | 160.0005 | 0.014 |
| 16 | 2 | N,E,S,W | 160.0005 | 30.004 |
| 16 | 4 | N,E,S,W | 160.0005 | 60.007 |
| 16 | 6 | N,E,S,W | 160.0005 | 89.903 |
| 16 | 8 | N,E,S,W | 160.0005 | 119.981 |
| 16 | 10 | N,E,S,W | 160.0005 | 149.994 |
| 16 | 12 | N,E,S,W | 160.0005 | 179.999 |
Notes: The radiographic inclination and anteversion angles were determined using a high-precision coordinate measuring machine (CMM). Radiographic inclination was represented by rotation around the base. Radiographic anteversion was represented by rotation around the fan. These verified angles were taken to represent the true angles.
Figure 8Testing process.
Accuracy Testing Results
| Device | |||||
|---|---|---|---|---|---|
| 1 | 2 | 3 | 4 | 5 | |
| Mean absolute error (°) | 1.34 | 0.86 | 1.56 | 1.00 | 1.30 |
| Standard deviation of absolute error (°) | 0.84 | 0.52 | 0.76 | 0.49 | 0.60 |
| Upper 95% CI of absolute error (°) | 1.54 | 0.99 | 1.75 | 1.13 | 1.44 |
| Maximum absolute error (°) | 4.79 | 2.40 | 3.69 | 3.20 | 2.69 |
Note: For each device, the statistics were calculated using the absolute error at each of 250 device orientations.
Abbreviation: CI, confidence interval.
Figure 9Accuracy of Navbit Sprint Navigation System.