The poor accuracy of D-dimer for the diagnosis of prosthetic joint infection but its potential usefulness in early postoperative infections following revision arthroplasty for aseptic loosening.

M Fernandez-Sampedro1, I Sanlés-González2, C García-Ibarbia3, N Fañanás-Rodríquez4, M Fakkas-Fernández5, M C Fariñas2.   

Abstract

BACKGROUND: D-dimer was introduced in 2018 as an alternative biomarker for C-reactive protein (CRP) in the diagnostic of prosthetic joint infection (PJI) criteria of the Musculoskeletal Infection Society. We assessed the accuracy of plasma D-dimer for the diagnosis of early, delayed, and late PJI according to Infectious Diseases Society of America (IDSA) criteria, and whether persistently high levels of D-dimer in cases of aseptic loosening (AL) may be predictive of subsequent implant-related infection.
METHODS: A prospective study of a consecutive series of 187 revision arthroplasties was performed at a single institution. Septic (n = 39) and aseptic revisions (n = 141) were classified based on IDSA criteria. Preoperative assessment of CRP, erythrocyte sedimentation rate (ESR) and D-dimer was performed. Receiver operating curves were used to determine maximum sensitivity and specificity of the biomarkers. The natural progress of D-dimer for AL cases was followed up either until the date of implant-related infection at any time during the first year or 1 year after revision in patients without failure. Clinical outcomes for those AL cases included infection-related failure that required a new surgery or need for antibiotic suppression.
RESULTS: Preoperative D-dimer level was significantly higher in PJI cases than in AL cases (p = 0.000). The optimal threshold of D-dimer for the diagnosis of PJI was 1167 ng/mL. For overall diagnosis of PJI, C-reactive protein (CRP) achieved the highest sensitivity (84.6%), followed by erythrocyte sedimentation rate (ESR) and D-dimer (82% and 71.8%, respectively). Plasma D-dimer sensitivity was lower for all PJI types. When combinations of 2 tests were studied, the combined use of ESR and CRP achieved the best accuracy for all types of PJI (76.9%). 4.25% of AL cases had implant failure due to implant-related infection during the first year after the index revision arthroplasty, only the cases with early failure maintained high D-dimer levels.
CONCLUSIONS: Plasma D-dimer did not offer an improvement over the individual or combined diagnosis for any type of PJI according to IDSA criteria. Persistently raised levels of D-dimer after revision arthroplasty in AL cases might be used to effectively diagnose early postoperative infection.
© 2022. The Author(s).

Entities:  

Keywords:  First-line screening test; PJI guidelines; Plasma D-dimer; Prosthetic joint infection

Mesh:

Substances:

Year:  2022        PMID: 35086474      PMCID: PMC8793264          DOI: 10.1186/s12879-022-07060-8

Source DB:  PubMed          Journal:  BMC Infect Dis        ISSN: 1471-2334            Impact factor:   3.090


Introduction

In recent years, several scientific societies have developed criteria to standardize definitions for prosthetic joint infection (PJI) [1-4]. However, none have been widely adopted and the diagnostic approach to patients with suspected PJI remains extremely variable from center to center, depending on local experience, technological equipment, and adherence to available guidelines. The main peripheral blood parameters used for the preoperative diagnosis of PJI include primarily 2 serological markers, serum C-reactive protein (CRP) and erythrocyte sedimentation rate (ESR). In the last 2 years, D-dimer has been recommended as a promising biomarker in suspected PJI, and has, in fact, been included in the 2018 Musculoskeletal Infection Society (MSIS) and International Consensus Meeting (ICM) criteria [5, 6]. Relevant studies that used MSIS or ICM criteria addressing the evaluation of D-dimer for the diagnosis of PJI reach different conclusions [7-12]. This may be due to the different methodologies used among the studies and, because the thresholds of serological markers may vary, depending on the heterogeneity of the definitions, the time of infection, different D-dimer assays, and/or the infecting organism, therefore more work is needed to further validate their role in the diagnosis of PJI [6] No published data are available on the role of D-dimer, taking into account the Infectious Diseases Society of America (IDSA) guidelines which describe the combined use of an abnormal ESR and CRP as the best combination for identifying patients with suspected PJI. In this study, we assessed the accuracy of plasma D-dimer in improving the preoperative diagnosis of early (less than 3 months after prosthesis implantation), delayed (between 3 and 12 months), and late PJI (> 12 months after the prosthesis implantation) according to IDSA criteria, and whether the persistence of high levels of D-dimer in cases of aseptic loosening (AL) may be predictive of subsequent implant-associated infection.

Materials and methods

In this prospective observational study, all consecutive patients aged 18 years or older undergoing a total or partial revision of knee or hip arthroplasty were enrolled between February 2013 and February 2015. The study was conducted in the Division of Orthopedics of the Hospital Universitario Marques of Valdecilla, a tertiary care hospital. Patients were identified as having PJI according to IDSA guidelines [1]. Patients with a non-infectious diagnosis were defined as cases of AL: pain in the thigh or hip region, knee pain and radiological signs of loosening (inadequate initial fixation, mechanical loss of fixation over time, or biologic loss of fixation caused by particle-induced osteolysis around the implant). All patients with previous arthroplasty revisions were performed for non-infectious causes. Since ESR, CRP and white blood cell count are part of our clinical routine; a preoperative determination was performed 2 weeks before the surgery. Based on receiver-operating-characteristic (ROC) analysis of the data, the cut-off level for CRP was > 1 mg/dl, while for ESR it was > 15 mm/h. D-dimer levels were evaluated in heparin plasma samples using the Tina-quant D-Dimer immunoturbidimetric assay (Roche Diagnostics©) on the Cobas c501 analyzer (Roche©) [13]. The histopathological criteria established by Morawietz and Krem (Additional file 1: Appendix S1) [14] were used to define a standardized evaluation of the periprosthetic membrane. Venous blood samples were obtained preoperatively on the day of surgery, and D-dimer was determined in all study patients. Data (demographics, comorbidities, type of implant, surgical procedure, microbiological results of cultures and antimicrobial treatment) of all patients were collected prospectively by the clinical researchers of the institution using a standardized data collection form. Patients with comorbidities that could increase plasma levels of D-dimer were excluded, e.g., arterial thromboembolic disease, myocardial infarction, stroke, acute limb ischemia, intracardiac thrombus within 4 weeks, venous thromboembolic disease, such as deep vein thrombosis, pulmonary embolism, disseminated intravascular coagulation, hypercoagulation disorder, periprosthetic fracture or joint dislocation within 2 weeks, active malignancy, infections in other regions of the body, skin ulcer or hematoma. Patients with AL were followed up after inclusion in the study for a minimum of 24 months. Clinical outcomes include infection-related failure that required a new surgery or need for antibiotic suppression. The natural progress of D-dimer in non-PJI patients was followed up either until the date of implant-related infection at any time during the first year or 1 year after revision in patients without failure.

Statistical analysis

Statistical analysis was performed using Statistical Package for the Social Sciences Statistics for Windows version 25 (IBM Corporation, Armonk, NY, USA). Patient records were anonymized prior to analysis. Statistical analysis was performed by a physicist external to the recruitment and clinical management of the patients. Frequencies were given for categorical variables and compared in contingency tables using the Chi-squared test. Means were calculated for quantitative variables and compared using Student’s t-test. A p-value of less than 0.05 was considered significant. The Youden index (J = sensitivity + specificity−1) based on its correspondence with the diagnostic of PJI was used to determine the optimal threshold value for D-dimer [15]. Based on the cut-off values, the sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of serum markers were calculated. The combination of different diagnostic test was performed using an “AND” combination. The combination schemes were both tests positive (test 1 positive and test 2 positive, composite diagnostic positive).

Ethics, registration, funding, and potential conflicts of interest

The local ethics committee approved the study protocol (CTR2010.163) and all subjects provided oral informed consent before participation. This work was supported by the Fundación Marqués de Valdecilla, through the project API 11/9. IDIVAL, Instituto de Investigación Valdecilla, Cantabria. Spain. The authors have no conflicts of interest to declare.

Results

Study population

A total of 187 cases in 185 patients were prospectively included during the study period. 7 cases (3.8%) were excluded: 3 due to revision for dislocation, 2 due to a recent periprosthetic fracture, 1 due to a recent history of deep vein thrombosis, and 1 due to the presence of a venous leg ulcer. Of the remaining 180 cases, 141 (78.3%) were diagnosed as AL, while the other 39 (21.7%) patients were defined as PJI. Of these, 7 (18%) were classified as early, 9 (23%) as delayed, and 23 (59%) as late PJI. Detailed demographic data of all patients are presented in Table 1. There were no statistically significant differences between the 2 groups (PJI and AL) in terms of age, gender, and comorbidities (mostly diabetes) (p > 0.05). Patients with PJI had a greater number of previous revisions than the non-PJI group (p < 0.013).
Table 1

Characteristics of study patients

CharacteristicsAL (n = 141)PJI (n = 39)P value
Age (year)-mean ± SD69.23 ± 11.1466.54 ± 14.130.220
Male n (%)60 (42.6%)21 (53.8%)0.123
Body mass index-mean ± SD30.58 ± 5.229.88 ± 6.410.542
Hip n (%)91 (64.5%)27 (69.2%)0.73
Knee n (%)50 (35.5%)12 (30.8%)0.73
Diabetes mellitus n (%)25 (17.7%)7 (17.9%)0.86
Immunosuppressive therapy n (%) MTX2 (2.1%)1 (2.6%)0.62
Systemic steroid therapy n (%)2 (1.4%)00.62
Prior revision arthroplasty n (%)38 (27%)20 (51.3%)0.011
CRP mean ± SD0.78 ± 1.573.47 ± 3.720.00011
ESR mean ± SD15.15 ± 15.4139.16 ± 28.400.000013
D-dimer mean ± SD942.36 ± 1085.781968.54 ± 1471.840.00025

PJI, Prosthetic joint infection; AL, Aseptic loosening; MTX, Methotrexate; CRP, C-reactive protein; ESR, erythrocyte sedimentation rate; SD, Standard Deviation

Characteristics of study patients PJI, Prosthetic joint infection; AL, Aseptic loosening; MTX, Methotrexate; CRP, C-reactive protein; ESR, erythrocyte sedimentation rate; SD, Standard Deviation Preoperative plasma D-dimer level was significantly higher in PJI cases than in AL cases (1968 ± 1471 ng/mL vs. 942 ± 1085 ng/mL; p = 0.000). Mean serum ESR and CRP values were also significantly higher among PJI patients; ESR was 39.16 ± 28.40 mm/h in the PJI group compared to 15.15 ± 15.41 mm/h in the AL group (p = 0.000), while mean CRP was 3.47 ± 3.72 mg/dL in PJI cases compared with 0.78 ± 1.57 mg/dL in AL cases (p = 0.000) (Table 1). The optimal threshold of D-dimer for the diagnosis of PJI was 1167 ng/mL, and this value demonstrated a sensitivity of 72% and specificity of 75% (75.3%), AUC = 0.774 (0.693–0.854).

Comparison of individual diagnostic parameters for the diagnosis of PJI

According to individual parameters for the overall diagnosis of PJI, CRP achieved the highest sensitivity (84.6%), followed by ESR and D-dimer (82% and 71.8%, respectively). In early PJI diagnosis, ESR showed the highest sensibility (100%) followed by both CRP and D-dimer (85.7%). For delayed PJI, both CRP and ESR had the highest sensibility (88.8%), followed by D-dimer (66.6%) and, in late PJI, the highest sensibility was obtained with CRP (82.6%), followed by D-dimer and ESR (73.9% and 69.5%, respectively) (Table 2).
Table 2

Diagnostic Accuracy of individual Preoperative Tests according to type of PJI

MarkersSpecificity (%)Sensitivity (%)PPV (%)NPV (%)AUC (95% CI)
D-dimer > 1167 ng/mL
 All PJI75.3 (69–83)71.8 (58–86)45.9 (33.3–58.4)90.4 (81.1–99.7)0.774 (0.693–0.854)
 Early75.3 (69–83)85.7 (60–100)15.4 (4–26.7)99 (88–100)0.863 (0.785–0.942)
 Delayed75.3 (69–83)66.6 (36–97)15 (4–26.7)97.1 (86.8–100)0.721 (0.555–0.887)
 Late75.3 (69–83)69.5 (51–88)32.7 (19.5–46.8)93.7 (83.9–10090.767 (0.665–0.87)
CRP > 1 mg/dL
 All PJI82.1 (76–88)84.6 (73–96)56.8 (44–69)95 (87–100)0.869 (0.797–0.940)
 Early82.1 (76–88)85.7 (60–100)19.3 (5–33)99.13 (91–100)0.828 (0.612–1)
 Delayed82.1 (76–88)88.8 (68–100)24.2 (9.6–38.8)99.1 (91–100)0.936 (0.885–0.986)
 Late82.1 (76–88)82.6 (67–98)43.2 (28.5–57.8)96.6 (88.5–100)0.863 (0.772–0.953)
ESR > 15 mm/h
 All PJI72.8 (65–80)82 (70–94)45.7 (34–57.4)93.5 (82.8–100)0.792 (0.7–0.885)
 Early72.8 (65–80)100 (100–100)15.5 (5–26)100 (88–100)0.897 (0.828–0.967)
 Delayed72.8 (65–80)88.8 (63–100)17.3 (6.4–28.3)99 (87.3–100)0.866 (0.765–0.968)
 Late72.8 (65–80)73.9 (56–92)30.9 (18.7–43.1)94.4 (18.7–43.1)0.767 (0.592–0.871)

PJI, Prosthetic joint infection; PPV, Positive predictive value; NPV, Negative predictive value; AUC, Area under the curve; CI, Confidence interval

Diagnostic Accuracy of individual Preoperative Tests according to type of PJI PJI, Prosthetic joint infection; PPV, Positive predictive value; NPV, Negative predictive value; AUC, Area under the curve; CI, Confidence interval

Comparison of grouped diagnostic parameters for the diagnosis of PJI

When evaluating potential combinations of 2 tests in the detection of PJI, we found that the overall combined sensitivity and specificity for ESR + CRP was 76.9% and 87.8%, respectively, 59% and 92.7% for CRP + D-dimer, and 61.5% and 88.3% for ESR + D-dimer. For each type of infection, ESR + CRP diagnosed 85.7% of early, 77.7% of delayed, and 73.9% of late cases. The combination of CRP + D-dimer diagnosed 57.1% of early, 44.4% of delayed, and 65.2% of late cases. Finally, the addition of ESR + D-dimer diagnosed 71.4% of early, 44.4% of delayed and 65.2% of late cases (Table 3). Additional file 2: Appendix S2 shows the diagnostic accuracy of individual and grouped preoperative tests in PJI.
Table 3

Diagnostic Accuracy of Combined Preoperative Tests according to type of PJI

MarkersSpecificity (%) (95% CI)Sensitivity (%) (95% CI)PPV (%) (95% CI)NPV (%) (95% CI)AUC (95% CI)
D-dimer + CRP
 All PJI patients92.7 (88.3–97)59 (43.5–74.4)69.7 (54–85.4)88.8 (83.6–94)0.885 (0.831–0.938)
 Early92.7 (88.3–97)57.1 (20.5–93.8)28.6 (4.9–52.2)97.7 (93–100)0.944 (0.902–0.985)
 Delayed92.7 (88.3–97)44.4 (12–76.9)28.6 (5–52.2)96.2 (91.6–100)0.914 (0.853–0.975)
 Late92.7 (88.3–97)65.2 (45.7–84.7)60 (40.8–79.2)94.1 (89.6–98.5)0.855 (0.782–0.928)
D-dimer + ESR
 All PJI patients88.3 (82.9–93.7)61.5 (46.3–76.8)60 (44.8–75.2)89 (83.5–94.4)0.812 (0.729–0.896)
 Early88.3 (82.9–93.7)71.4 (38–100)23.8 (5.6–42)98.4 (92–100)0.917 (0.856–0.978)
 Delayed88.3 (82.9–93.7)44.4 (12–76.9)20 (2.4–37.5)96 (89.9–100)0.872 (0.786–0.957)
 Late88.3 (82.9–93.7)65.2 (45.7–84.7)48.4 (30.8–66)93.8 (87.9–99.7)0.7358 (0.633–0.882)
CRP + ESR
 All PJI patients87.8 (82.4–93.3)76.9 (63.7–90.1)63.8 (50–77.6)93.2 (87.3–99)0.838 (0.757–0.919)
 Early87.8 (82.4–93.3)85.7 (59.8–100)26 (8.1–44)99,1 (92.77–100)0.903 (0.83–0.976)
 Delayed87.8 (82.4–93.3)77.7 (50.6–100)29.2 (11–47.3)98.4 (92–100)0.926 (0.869–0.984)
 Late87.8 (82.4–93.3)73.9 (56–91.8)50 (33.2–66.8)95.3 (89.2–100)0.783 (0.66–0.919)

PJI, Prosthetic joint infection; PPV, Positive predictive value; NPV, Negative predictive value; AUC, Area under the curve; CI, Confidence interval

Diagnostic Accuracy of Combined Preoperative Tests according to type of PJI PJI, Prosthetic joint infection; PPV, Positive predictive value; NPV, Negative predictive value; AUC, Area under the curve; CI, Confidence interval

Outcome

6 (4.25%) of the 141 AL cases in our cohort had implant failure due to implant-related infection. All failures occurred during the first year after the index revision arthroplasty and, although all of them showed high levels of circulating D-dimer at the time of diagnosis, only 2 cases of early failure (< 3 months) maintained high levels of D-dimer. In both, virulent microorganisms, Staphylococcus aureus and Enterobacter cloacae, were isolated, and these patients showed raised levels of ESR and CRP after postoperative day 1 that peaked at postoperative day 4, and then remained elevated until the time of the failure, whereas plasma D-dimer showed persistently high levels that peaked at the time of the implant failure (Table 4).
Table 4

Aseptic loosening patients with failure due to implant-related infection

Index revision surgeryFollowing revision due to infection
CaseESR mm/hCRP mg/dlD-dimer ng/mLTime to failure (months)ESR mm/hCRP mg/dlD-dimer ng/mLMicrobiology
#1140.521210NDND1323Staphylococcus epidermidis
#250.21181222.12583Methicillin- sensitive S. aureus
#340.576811711529Staphylococcus epidermidis
#480.28404740.42782Staphylococcus epidermidis
#5100.467023515.32498Enterobacter cloacae
#6100.14761080.11331Coagulase-negative staphylococci

Values in bold are elevated values within the reference range for each parameter

CRP, C-reactive protein; ESR, erythrocyte sedimentation rate; ND, Not done

Aseptic loosening patients with failure due to implant-related infection Values in bold are elevated values within the reference range for each parameter CRP, C-reactive protein; ESR, erythrocyte sedimentation rate; ND, Not done

Microbiology and serological markers

43 microorganisms were isolated from 39 PJI patients. Only 1 patient was culture-negative and a respiratory source of infection was identified. Serological markers (ESR, CRP and D-dimer) based on the infecting organisms isolated were then evaluated (Table 5).
Table 5

Values of ESR, CRP and D-dimer according to microorganisms in predicting infection

MicroorganismsESR (> 15 mm/h)CRP (> 1 mg/dl)D-dimer (> 1167 ng/mL)
S. epidermidis (n = 18)14/1815/1816/18
 Early2/22/22/2
 Delayed2/33/33/3
 Late10/1310/1311/13
Other CoNS (n = 11)a9/119/114/11
 Early2/22/22/2
 Delayed4/44/41/4
 Late3/53/51/5
Streptococcus spp. (n = 3)3/33/33/3
 Early000
 Delayed1/11/11/1
 Late2/22/22/2
Corynebacterium spp. (n = 3)3/33/33/3
 Early1/11/11/1
 Delayed000
 Late2/22/22/2
Gram-negative (n = 4)3/43/43/4
 Early3/32/33/3
 Delayed000
 Late0/11/10/1
Others (n = 4)b3/44/42/4
 Early1/11/11/1
 Delayed2/22/21/2
 Late0/11/10/1

ESR, Erythrocyte sedimentation rate; CRP, C-reactive protein

aOther CoNS (coagulase-negative Staphylococci): Staphylococcus lugdunensis (n = 4), Staphylococcus capitis (n = 3), Staphylococcus caprae (n = 2), Staphylococcus haemolyticus (n = 1), Staphylococcus hominis (n = 1)

bOthers: Enterococcus faecalis (n = 2), Peptostreptococcus asaccharolyticus (n = 1), methicillin-sensitive Staphylococcus aureus (n = 1)

Values of ESR, CRP and D-dimer according to microorganisms in predicting infection ESR, Erythrocyte sedimentation rate; CRP, C-reactive protein aOther CoNS (coagulase-negative Staphylococci): Staphylococcus lugdunensis (n = 4), Staphylococcus capitis (n = 3), Staphylococcus caprae (n = 2), Staphylococcus haemolyticus (n = 1), Staphylococcus hominis (n = 1) bOthers: Enterococcus faecalis (n = 2), Peptostreptococcus asaccharolyticus (n = 1), methicillin-sensitive Staphylococcus aureus (n = 1)

Discussion

To our knowledge, this is the first study that evaluates D-dimer levels in the diagnosis of PJI according to IDSA guidelines which highlights the use of both an abnormal ESR and CRP as the best combination for patients with suspected PJI. Our results show that plasma D-dimer did not improve the individual diagnosis of ESR or CRP for any type of PJI. The sensitivity of plasma D-dimer was lower for all PJI types, except for early infection in which it offered the second-best sensitivity, equal to ESR. When combinations of 2 tests were studied, the combined use of ESR and CRP achieved the best accuracy for all types of PJI, and the addition of D-dimer to ESR achieved the second-best sensitivity for early PJI. In the last 2 years, D-dimer has been recommended as a promising biomarker in suspected PJI. Indeed, D-dimer was introduced in 2018 as an alternative biomarker with a cut-off value of 860 ng/mL in the new MSIS and ICM validated diagnostic PJI criteria [5, 6]. Recently, 8 meta-analyses addressing the diagnostic accuracy of D-dimer in PJI highlighted the wide heterogeneity among studies, because of that, univariate meta-regression and subgroup analysis were performed and found that the type of sample (serum or plasma) could be the primary factor affecting the heterogeneity of sensitivity, while the country of origin of the study (China and USA) may be the main source of heterogeneity in terms of specificity [16-23]. Seven studies have investigated plasma D-dimer in the diagnosis of PJI, including 6 Asian studies and 1 European study (Germany) (Table 6) [24-30]. On the basis of our study conducted according to IDSA guidelines, plasma D-dimer achieved the second-best sensitivity and the third-best specificity when compared with the aforementioned studies. Our results are in close agreement with those published in Asian populations, in terms of both sensitivity and specificity, and contrast with the European study. In this regard, 2 studies on serum D-dimer that recruited their patients from an American population using MSIS criteria and a threshold value of 850 ng/mL showed conflicting results, with sensitivities and specificities of 89% and 93% vs 96% and 32%, respectively [7-10]. Therefore, it should be reevaluated whether racial differences are a major factor affecting the outcome of the different studies.
Table 6

Plasma D-dimer Studies

ReferencesCountryPJI/ALStudy designSiteAge (PJI/AL)BMI (PJI/AL)Cut-off ng/mLStandard ReferenceSensitivity (%)Specificity (%)Exluded hipercoagulation disorderExcluded inflammatory arthritis
Fu et al. [24]China15/15ProspectiveH/K65.6/65.425.7/24.7850MSIS66.6760NoYes
Li et al. [25]China76/363Retrospective (multicentre)H/K63.7/61.325.0/25.11250ICM64.565NoNo
Xu et al. [26]China129/189RetrospectiveH/KNANA1020MSIS68.2950.70NoNo
Chen et al. [30]China30/60Retrospective (Matched study)H/K61.5/67.726.0/24.91690MSIS66.776.7NoNo
Wu et al. [27]China35a27b/81RetrospectiveH/K62.6/69.124.7/24.5410MSIS75.867NoNo
Ackmann et al. [28]Europe (Germany)52/67RetrospectiveH/K70.5/6830.9/28.42750MSIS3894YesYes
Huang et al. [29]China47/38RetrospectiveH/K63.7/66.8NA1170MSIS6085YesYes
Present studyEurope (Spain)39/141ProspectiveH/K69.2/66.530.6/29.91167IDSA71.875.3YesNo

PJI, Prosthetic joint infection; AL, Aseptic loosening; MSIS, Musculoskeletal Infection Society; ICM, International Consensus Meeting; IDSA, Infectious Diseases Society of America; NA, not applicable; H, Hip; K, Knee

aPatients underwent first-stage and spacer insertion. bPatients underwent second-stage reimplantation

Plasma D-dimer Studies PJI, Prosthetic joint infection; AL, Aseptic loosening; MSIS, Musculoskeletal Infection Society; ICM, International Consensus Meeting; IDSA, Infectious Diseases Society of America; NA, not applicable; H, Hip; K, Knee aPatients underwent first-stage and spacer insertion. bPatients underwent second-stage reimplantation Among other concerns, the optimal cut-off value for D-dimer and its sensitivity and specificity remain unclear. Some authors considered that patients’ body mass index (BMI), age or sex may comprise a confounding factor resulting in higher levels of D-dimer, while women and older patients may have higher serum levels of D-dimer when used in the diagnosis of thrombosis and/or embolism [31]. In our cohort, the threshold for plasma D-dimer was similar to that described in the Asian population, particularly by Huang et al. [29], who also reported lower mean BMIs and lower median age than the Europeans. Recently, Grzelecki et al., found that the influence of the operated joint type may be a very important factor in the concentration of D-dimer and that plasma D-dimer has a relatively high value in the detection of knee PJI, but a moderate to low value for hip PJI [32]. It does seem logical that these characteristics could have unbalanced the optimal threshold of D-dimer for the diagnosis of PJI which ranged between 410 ng/mL to 2300 ng/mL. In short, pending more robust studies that provide new data and investigate their correlations with sample type, comorbidities, ethnic variability, joint type, PJ type, etc., each institution should optimize D-dimer thresholds. A further point to be discussed is whether plasma or serum is the most appropriate sample for performing D-dimer testing. The coagulant factor present in plasma but not in serum interferes with many biochemical parameters and the higher metabolite concentrations in serum offer more sensitive results in biomarker detection. Korte et al. demonstrated a close agreement of D-dimer concentrations determined from citrated plasma samples and serum for the exclusion of venous thromboembolism [33]. However, they added that the use of serum for D-dimer determination needs to be evaluated and validated in a clinical study. A recent meta-analysis of prosthesis-related infections showed that serum D-dimer outperformed plasma D-dimer in sensitivity and specificity (0.86 and 0.84 vs. 0.67 and 0.60, respectively) [34]. In this regard, 6 studies on serum D-dimer in PJI were published between 2017 and 2020, 2 from the USA and 4 from China. Five used MSIS criteria for the diagnosis of PJI and the other used ICM. The D-dimer threshold ranged from 850 to 1170 ng/mL [7-12]. Unfortunately, they reached different conclusions and these results raise the question of whether differences in the type of sample are a major factor affecting the substantial discrepancies between studies. In a prospective study that measured D-dimer levels before and after primary total hip or knee arthroplasty, the most significant changes in D-dimer levels were observed during the early postoperative period. Levels increased sharply and peaked on the first day after joint replacement surgery, decreasing to baseline levels on the following day [35]. In our AL cohort, 4.25% of the cases had implant failure due to infection. All of them had high levels of circulating D-dimer at the time of diagnosis. However, in the 2 cases with early failure, D-dimer levels remained raised and peaked at the time of failure. In these 2 cases, ESR and CRP levels peaked before the time failure was diagnosed. It would be of interest to study a larger number of cases with early failure to determine if there is any difference. Finally, coagulase-negative staphylococci (CoNS) (67.4%) were the most common pathogens isolated in our study, while S. epidermidis was the most prevalent species, particularly in late infections. Higher plasma D-dimer levels were associated with a slight tendency towards the detection of more S. epidermidis isolates than other inflammatory markers, but in contrast were associated with fewer PJI infections due to other CoNS isolates. We are unaware if small colony variants or biofilm formation by different CoNS play a role in these findings. No differences were found for other isolates such as low-virulence bacteria. Our study has several limitations that should be considered. First, this is a single-center prospective study with potential for uncontrolled selection biases; however, methodologies were applied in a standard fashion, something that does not occur when patients are included from multiple centers where the test result of D-dimer can be interpretated as positive or negative with different thresholds and, therefore any change in thresholds may have a substantial impact on its diagnostic value. Second, the sample size for PJI infections in the study was small due to the low incidence of PJI and may lack statistical power to detect some associations in different subtypes of PJI being this limitation is described in other studies [30]. Third, there is no gold standard for the diagnosis of PJI and it is possible that the lack the sensitivity in detecting PJI, specially in chronic or low-grade infections. Fourth, we excluded patients with conditions that may induce high expression of D-dimer and could result in high false positive rates. These pathologies accounted for almost 4% of the study patients and, this low prevalence, does not appear to have been a bias for the study. However, different D-dimer assays may point to differences in sensitivity and specificity between studies. This emphasizes the need for standardization of D-dimer assays. Fifth, prior revisions of PJI group differed significantly from AL group which may affect the validity of the results. It has been recognized that the risk of PJI increases with the number of previous joint arthroplasties. However, it has been unclear whether the increased risk of PJI in patients with prior joint arthroplasties is due to an increased number of comorbid conditions, a prolonged operating time, an increased number of blood transfusions, or higher frequency of postoperative wound complications. Therefore, the identification of patients at high risk for PJI would allow for improved preoperative risk assessment, increase the index suspicion, and identify patients for whom focused efforts at prevention are necessary. In conclusion, plasma D-dimer determined according to IDSA guidelines did not improve the individual or combined diagnosis of ESR or CRP for any type of PJI. The persistence of raised levels of plasma D-dimer after revision arthroplasty in AL cases might be used effectively in diagnosing early postoperative infection. Because of the remaining concerns surrounding this new serological biomarker in implant-associated infection, we urge caution in accepting serum and/or plasma D-dimer as a first-line screening test for PJI diagnosis, regardless of the guidelines adopted. This conclusion should be confirmed in appropriate clinical trials. Additional file 1: Appendix S1. Histopathological classification of the periprosthetic membrane. Additional file 2: Appendix S2. Diagnostic Accuracy of individual preoperative tests and combinated test in PJI. PPV Positive predictive value; NPV Negative predictive value; AUC Area under the curve; CI Confidence interval.
  34 in total

1.  Latex-enhanced immunoturbidimetry allows D-dimer determination in plasma and serum samples.

Authors:  W Korte; W Riesen
Journal:  Clin Chem       Date:  2000-06       Impact factor: 8.327

2.  Index for rating diagnostic tests.

Authors:  W J YOUDEN
Journal:  Cancer       Date:  1950-01       Impact factor: 6.860

3.  Synovial Fluid Viscosity Test is Promising for the Diagnosis of Periprosthetic Joint Infection.

Authors:  Jun Fu; Ming Ni; Wei Chai; Xiang Li; Libo Hao; Jiying Chen
Journal:  J Arthroplasty       Date:  2019-02-13       Impact factor: 4.757

4.  The diagnostic effect of α-defensin, D-dimer, and IL-6 in periprosthetic joint infection: A systematic review and diagnostic meta-analysis.

Authors:  Baozhong Tian; Liwen Cui; Weihai Jiang
Journal:  J Orthop Surg (Hong Kong)       Date:  2020 Sep-Dec       Impact factor: 1.118

5.  Circulating D-Dimer versus Fibrinogen in the Diagnosis of Peri-Prosthetic Joint Infection: A Meta-Analysis.

Authors:  Qingyu Zhang; Jinlei Dong; Dongsheng Zhou; Fanxiao Liu
Journal:  Surg Infect (Larchmt)       Date:  2020-04-27       Impact factor: 2.150

6.  Different diagnostic performance of plasma fibrinogen and D-dimer in periprosthetic joint infection: a propensity score matched study.

Authors:  Xi Chen; Wenwei Qian; Xisheng Weng; Jin Lin; Jin Jin; Yiou Wang; Shibai Zhu
Journal:  BMC Musculoskelet Disord       Date:  2021-05-07       Impact factor: 2.362

7.  Natural progress of D-dimer following total joint arthroplasty: a baseline for the diagnosis of the early postoperative infection.

Authors:  Yong Seuk Lee; Young-Kyun Lee; Seung Bum Han; Chang Hyun Nam; Javad Parvizi; Kyung-Hoi Koo
Journal:  J Orthop Surg Res       Date:  2018-02-13       Impact factor: 2.359

8.  D-dimer in the diagnosis of periprosthetic joint infection: a systematic review and meta-analysis.

Authors:  Guangxu Lu; Tong Li; Haoqi Ye; Shujin Liu; Peng Zhang; Wenliang Wang
Journal:  J Orthop Surg Res       Date:  2020-07-16       Impact factor: 2.359

9.  Diagnostic accuracy of D-dimer in periprosthetic joint infection: a diagnostic meta-analysis.

Authors:  Haitao Zhang; Xiaobo Sun; Pengfei Xin; Xingyang Zhu; Ke Jie; Houran Cao; Wenjun Feng; Yuqing Zeng; Yan Lv; Jinlun Chen; Jie Li; Jianchun Zeng; Yirong Zeng
Journal:  J Orthop Surg Res       Date:  2020-08-17       Impact factor: 2.359

10.  Meta-analysis of serum and/or plasma D-dimer in the diagnosis of periprosthetic joint infection.

Authors:  Cheng Li; Donara Margaryan; Cristina Ojeda-Thies; Carsten Perka; Andrej Trampuz
Journal:  J Orthop Surg Res       Date:  2020-08-06       Impact factor: 2.359

View more
  1 in total

Review 1.  Diagnoses Based on C-Reactive Protein Point-of-Care Tests.

Authors:  Miroslav Pohanka
Journal:  Biosensors (Basel)       Date:  2022-05-17
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.