| Literature DB >> 35082482 |
Marina Delgado João1, Jorge Vasco Costa1, Tiago Monteiro1, Nuno Franqueira1, Fernando Faria Correia1, Fernando Vaz1.
Abstract
PURPOSE: To evaluate the effect of neodymium-doped yttrium aluminum garnet (Nd:YAG) laser capsulotomy on intraocular lens (IOL) position and anterior segment parameters with a new swept-source anterior segment optical coherence tomography (AS-OCT) device (Anterion®, Heidelberg Engineering GmbH). PATIENTS AND METHODS: A total of 50 eyes from 50 consecutive patients were included. All patients had visually significant posterior capsular opacification (PCO) after uneventful phacoemulsification surgery with manual capsulorhexis and single-piece C-loop acrylic IOL implantation (AcrySof® SA60AT) and were treated with Nd:YAG laser. Anterior segment images were captured with Anterion® in non-dilated conditions before and one month after the procedure. In the "Metrics App", we collected data of the anterior chamber angle (ACA) 3 and 9 o'clock, the anterior chamber depth (ACD), the anterior chamber volume (ACV), and the central corneal thickness (CCT). We also collected demographic and clinical data [age, gender, months from surgery to Nd: YAG capsulotomy, pre- and post-capsulotomy corrected distance visual acuity (CDVA) and spherical equivalent, and axial length].Entities:
Keywords: capsulotomy; intraocular lens; neodymium-doped yttrium aluminum garnet laser; posterior capsular opacification; swept-source anterior segment optical coherence tomography
Year: 2022 PMID: 35082482 PMCID: PMC8785224 DOI: 10.2147/OPTH.S349649
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Figure 1Pre-procedure Anterion® printout showing cornea, anterior chamber, lens, and pupil parameters.
Anterior Segment Optical Coherence Tomography Parameters Comparison Before and One Month After Nd:YAG Laser Posterior Capsulotomy
| Parameters | Before Capsulotomy | One Month After Capsulotomy | |
|---|---|---|---|
| Anterior chamber depth (mm) | 3.96±0.55 | 3.97±0.55 | 0.10 |
| Anterior chamber angle 3 o’clock (o) | 49.23±12.34 | 47.81±9.92 | 0.809 |
| Anterior chamber angle 9 o’clock (o) | 47.55±11.23 | 46.43±10.55 | 0.667 |
| Anterior chamber volume (mm3) | 198.09±43.43 | 201.31±46.27 | 0.09 |
| Central corneal thickness (µm) | 552.79±39.90 | 553.26±41.42 | 0.615 |
Figure 2Box plots of anterior segment optical coherence tomography parameters before and after Nd:YAG posterior capsulotomy.
Visual and Refractive Results Before and One Month After Nd:YAG Laser Posterior Capsulotomy
| Parameters | Before Capsulotomy | One Month After Capsulotomy | |
|---|---|---|---|
| Spherical equivalent (diopters) | −0.15±0.84 | −0.25±0.93 | 0.42 |
| Distance corrected visual acuity (logMAR) | 0.37±0.21 | 0.12±0.27 | 0.015 |
Figure 3Scatter plot showing the anterior chamber depth variation after the Nd:YAG posterior capsulotomy according to axial length.