Weifen Gong1, Haoyu Chen1, Fan Yang1, Shibin Lin1, Chao Li1, Geng Wang1.
Abstract
Purpose: To evaluate the ocular biometric parameters in patients with constant and intermittent exotropia by the measurement of swept-source optical coherence tomography (SS-OCT) optical biometer OA-2000 and comparing it with the normal control subjects. Design: Cross-sectional case-control study. Participants: Fifty-five constant and 24 intermittent patients with exotropia with central fixation and 77 orthotropic normal control participants aged 4-18 years old.Entities:
Keywords: OA-2000; anisometropia; concomitant exotropia; intereye differences; ocular biometric parameters
Year: 2022 PMID: 35059408 PMCID: PMC8763704 DOI: 10.3389/fmed.2021.724122
Source DB: PubMed Journal: Front Med (Lausanne) ISSN: 2296-858X
Comparisons of the absolute values of inter-eye differences between concomitant exotropia and control participants.
|
|
|
|
| |
|---|---|---|---|---|
| AnisoSE (D) | 0.878 (0.931) | 0.577 (0.625) | 2.369 (0.019 | 0.034 |
| AnisoAL (mm) | 0.395 (0.404) | 0.208 (0.265) | 3.423 (0.001 | < 0.001 |
| AnisoACD (mm) | 0.060 (0.059) | 0.032 (0.030) | 3.782 (<0.001 | 0.001 |
| AnisoPD (mm) | 0.557 (0.476) | 0.340 (0.357) | 3.229 (0.002 | 0.001 |
| AnisoLT (mm) | 0.060 (0.073) | 0.031 (0.035) | 3.136 (0.002 | 0.003 |
XT, concomitant exotropia; NCP, normal control participants; AnisoSE, absolute values of intereye differences for spherical equivalent; AnisoAL, absolute values of intereye differences for axial length; AnisoACD, absolute values of intereye differences for anterior chamber depth; AnisoPD, absolute values of intereye differences for pupil diameter; AnisoLT, absolute values of intereye differences for lens thickness.
Results with equal variances not assumed and reanalyzed with Mann-Whitney U-test.
Correlation coefficients of the ocular biometric parameters between the two eyes within concomitant exotropia and control participants.
|
|
| |||||
|---|---|---|---|---|---|---|
|
|
|
|
| |||
| SE | 0.877 | 0.876 | <0.001 | 0.943 | 0.944 | <0.001 |
| AL | 0.914 | 0.915 | <0.001 | 0.971 | 0.972 | <0.001 |
| ACD | 0.955 | 0.956 | <0.001 | 0.987 | 0.987 | <0.001 |
| PD | 0.767 | 0.826 | <0.001 | 0.944 | 0.945 | <0.001 |
| LT | 0.924 | 0.924 | <0.001 | 0.972 | 0.974 | <0.001 |
XT, concomitant exotropia; NCP, normal control participants; ICC, intraclass correlation coefficient; SE, spherical equivalent; AL, axial length; ACD, anterior chamber depth; PD, pupil diameter; LT, lens thickness.
Figure 1Bland-Altman plots (A,B) and scatter plots (C,D) of SE in concomitant exotropia (A,C) and normal control participants (B,D) of the left and right eyes. OS, left eye; OD, right eye; SE, spherical equivalent; LOA, 95% limits of agreement.
Figure 5Bland-Altman plots (A,B) and scatter plots (C,D) of LT in concomitant exotropia (A,C) and normal control participants (B,D) of the left and right eyes. OS, left eye; OD, right eye; LT, lens thickness; LOA, 95% limits of agreement.
Summary of the inter-ocular agreement of biometric parameters between concomitant exotropia and control participants.
|
|
| |||||||
|---|---|---|---|---|---|---|---|---|
|
|
|
|
| |||||
| SE (D) | 0.0522 ± 1.2828 | −2.4622 | 2.5666 | 5.0288 | −0.0974 ± 0.8484 | −1.7603 | 1.5655 | 3.3258 |
| AL (mm) | 0.0429 ± 0.5650 | −1.0644 | 1.1502 | 2.2146 | 0.0364 ± 0.3360 | −0.6222 | 0.6950 | 1.3172 |
| ACD (mm) | −0.0182 ± 0.0827 | −0.1804 | 0.1439 | 0.3243 | 0.0070 ± 0.0429 | −0.0771 | 0.0911 | 0.1682 |
| PD (mm) | −0.3770 ± 0.6300 | −1.6112 | 0.8579 | 2.4691 | −0.0609 ± 0.4908 | −1.0230 | 0.9011 | 1.9241 |
| LT (mm) | 0.0101 ± 0.0950 | −0.1760 | 0.1963 | 0.3723 | −0.0004 ± 0.0474 | −0.0933 | 0.0925 | 0.1858 |
XT, concomitant exotropia; NCP, normal control participants; SE, spherical equivalent; AL, axial length; ACD, anterior chamber depth; PD, pupil diameter; LT, lens thickness.