Giuseppe Boriani1, Marco Proietti2,3,4, Matteo Bertini5, Igor Diemberger6, Pietro Palmisano7, Stefano Baccarini8, Francesco Biscione9, Nicola Bottoni10, Antonio Ciccaglioni11, Alessandro Dal Monte12, Franco Alberto Ferrari13, Saverio Iacopino14, Marcello Piacenti15, Daniele Porcelli16, Stefano Sangiorgio17, Luca Santini18, Michele Malagù5, Giuseppe Stabile19, Jacopo Francesco Imberti1, Davide Caruso20, Massimo Zoni-Berisso20, Roberto De Ponti21, Renato Pietro Ricci22.
Abstract
BACKGROUND: The incidence of infections associated with cardiac implantable electronic devices (CIEDs) and patient outcomes are not fully known. AIM: To provide a contemporary assessment of the risk of CIEDs infection and associated clinical outcomes.Entities:
Keywords: cardiac implantable electronic device; implantable defibrillator; infection; outcome; pacemaker
Year: 2022 PMID: 35055406 PMCID: PMC8780465 DOI: 10.3390/jpm12010091
Source DB: PubMed Journal: J Pers Med ISSN: 2075-4426
Figure 1RI-AIAC registry enrolment rates.
Baseline characteristics of enrolled patients.
| 78 (70–84) | |
| 1720 (64.3) | |
| 194 (7.3) | |
| 38 (1.4) | |
| 73 (2.7) | |
| 622 (23.3) | |
| 41 (1.5) | |
| 550 (20.6) | |
| 740 (27.7) | |
| 72 (2.7) | |
| 821 (30.7) | |
| 18 (0.7) | |
| 12 (0.4) | |
| 44 (1.6) | |
| 2636 (98.5) | |
| 2 (1,2) | |
| 1 (0–4) | |
| 1 (0–2) | |
| 1 (0–2) |
CIED = Cardiac Implantable Electronic Device; CRT-D = Cardiac Resynchronization Therapy Defibrillator; CRT-P = Cardiac Resynchronization Therapy Pacing; eGFR = Estimated Glomerular Filtration Rate; ICD = Implantable Cardioverter Defibrillator; IQR = Interquartile Range; PADIT = Prevention of Arrhythmia Device Infection Trial; SD = Standard Deviation.
Figure 2Kaplan–Meier curves for clinical outcomes occurrence. (Upper Figure) and (Cumulative risk for CIED infection; Lower Figure) Cumulative risk for composite clinical event; Legend: CIED = Cardiac Implantable Electronic Device.
Baseline characteristics according to Occurrence of Infection during Follow-Up.
| No Infection | Infection |
| |
|---|---|---|---|
| 77 (69–83) | 74 (66–77) | 0.041 | |
| 0.538 | |||
| 1601 (64.2) | 19 (67.9) | 0.686 | |
| 0.715 | |||
| <0.001 | |||
| 0.131 | |||
| 0.027 | |||
| 184 (7.4) | 3 (10.7) | 0.502 | |
| 34 (1.4) | 2 (7.1) | 0.010 | |
| 62 (2.5) | 1 (3.6) | 0.714 | |
| 553 (22.2) | 6 (21.4) | 0.926 | |
| 32 (1.3) | 2 (7.1) | 0.007 | |
| 499 (20.0) | 11 (39.3) | 0.012 | |
| 689 (27.6) | 11 (39.3) | 0.170 | |
| 62 (2.5) | 1 (3.6) | 0.714 | |
| 759 (30.4) | 9 (32.1) | 0.844 | |
| 17 (0.7) | 0 | 1.000 | |
| 10 (0.4) | 0 | 1.000 | |
| 37 (1.5) | 2 (7.1) | 0.016 | |
| 2455 (98.4) | 28 (100) | 0.499 | |
| 0.120 | |||
| 2 (1–2) | 2 (1–3) | 0.086 | |
| 1 (0–4) | 4 (1–5) | 0.008 | |
| 1 (0–2) | 1 (0–2) | 0.136 | |
| 1 (0–2) | 2 (1–3) | 0.080 |
CIED = Cardiac Implantable Electronic Device; CRT-D = Cardiac Resynchronization Therapy Defibrillator; CRT-P = Cardiac Resynchronization Therapy Pacing; eGFR = Estimated Glomerular Filtration Rate; ICD = Implantable Cardioverter Defibrillator; IQR = Interquartile Range; PADIT = Prevention of Arrhythmia Device Infection Trial; SD = Standard Deviation.
Logistic regression analysis for CIED infection occurrence.
| Univariate Analysis | Multivariate Analysis | Score Point | |||
|---|---|---|---|---|---|
| OR (95% CI) |
| OR (95% CI) |
| ||
|
| 5.57 (1.27–24.40) | 0.023 | 1.68 (0.30–8.87) | 0.579 | - |
|
| 5.92 (1.35–26) | 0.019 | 3.44 (0.74–15.94) | 0.116 | - |
|
| 2.59 (1.21–5.56) | 0.015 | 2.22 (1.02–4.84) | 0.045 | 1 |
|
| 5.11 (1.17–22.32) | 0.030 | 3.96 (0.85–18.57) | 0.080 | 1 |
CI = Confidence Interval; CIED = Cardiac Implantable Electronic Device; OR = Odds Ratio.
Logistic regression analysis for clinical events occurrence.
| Univariate Analysis | Multivariate Analysis | Score Points | |||
|---|---|---|---|---|---|
| OR (95% CI) |
| OR (95% CI) |
| ||
|
| 3.06 (1.58–5.94) | 0.001 | 2.9 (1.45–5.77) | 0.002 | 1 |
|
| 2.55 (1.83–3.57) | <0.001 | 2.03 (1.43–2.88) | <0.001 | 1 |
|
| 1.62 (1.25–2.11) | <0.001 | |||
|
| 1.58 (1.09–2.28) | 0.015 | 1.43 (0.98–2.08) | 0.063 | 1 |
|
| 2.47 (1.2–5.08) | 0.014 | 2.60 (1.25–5.40) | 0.010 | 1 |
|
| 3.21 (1.41–7.32) | 0.006 | 3.29 (1.29–8.35) | 0.012 | 1 |
CI = Confidence Interval; CIED = Cardiac Implantable Electronic Device; eGFR = Estimated Glomerular Filtration Rate; OR = Odds Ratio.
Predictive ability of clinical scores.
| Infection | Clinical Events | |||
|---|---|---|---|---|
| C-Index (95% CI) |
| C-Index (95% CI) |
| |
|
| 0.64 (0.53–0.76) | 0.010 | 0.51 (0.47–0.56) | 0.612 |
|
| 0.56 (0.46–0.66) | 0.261 | 0.63 (0.59–0.68) | <0.001 |
|
| 0.58 (0.47–0.68) | 0.159 | 0.62 (0.58–0.67) | <0.001 |
|
| 0.64 (0.52–0.75) | 0.015 | - | - |
|
| - | - | 0.67 (0.63–0.71) | <0.001 |
* Adjusted for sex; † adjusted for age and sex; CI= Confidence Interval; OR= Odds Ratio; PADIT = Prevention of Arrhythmia Device Infection Trial; RI-AIAC = Ricerca sulle Infezioni Associate a ImpiAnto o sostituzione di CIED.
Figure 3RI-AIAC scores distribution; (Upper Figure RI-AIAC Infection Score; Lower Figure) RI-AIAC Event Score; RI-AIAC = Ricerca sulle Infezioni Associate a ImpiAnto o sostituzione di CIED.
Logistic regression of clinical scores.
| Infection | Clinical Events | |||
|---|---|---|---|---|
| OR (95% CI) |
| OR (95% CI) |
| |
|
| 1.28 (1.1–1.50) | 0.002 | 1.01 (0.94–1.09) | 0.748 |
|
| 1.55 (0.73–3.27) | 0.251 | 1.36 (1.2–1.55) | <0.001 |
|
| 1.69 (0.89–3.21) | 0.112 | 1.32 (1.17–1.47) | <0.001 |
|
| 2.38 (1.6–3.55) | <0.001 | - | - |
|
| - | - | 1.56 (1.38-1.75) | <0.001 |
* Adjusted for sex; † adjusted for age and sex; CI = Confidence Interval; OR = Odds Ratio; PADIT = Prevention of Arrhythmia Device Infection Trial; RI-AIAC = Ricerca sulle Infezioni Associate a ImpiAnto o sostituzione di CIED.
Predictive ability of clinical scores in external validation cohort.
| Infection | Clinical Events | SE | SP | |||
|---|---|---|---|---|---|---|
| C-Index (95% CI) |
| C-Index (95% CI) |
| |||
|
| 0.53 (0.38–0.67) | 0.746 | 0.49 (0.44–0.53) | 0.600 | - | - |
|
| 0.64 (0.5–0.79) | 0.065 | 0.65 (0.6–0.7) * | <0.001 | 70.3% | 51.2% |
|
| 0.62 (0.46–0.77) | 0.131 | 0.56 (0.51–0.62) † | 0.025 | - | - |
|
| 0.58 (0.42–0.74) | 0.292 | - | - | - | - |
|
| - | - | 0.68 (0.63–0.72) | <0.001 | 95.9% | 16.3% |
* No significant difference compared to RI-AIAC Event (De Long–De Long p = 0.2; † significantly less predictive than RI-AIAC Event (De Long–De Long p < 0.001); CI = Confidence Interval; PADIT = Prevention of Arrhythmia Device Infection Trial; RI-AIAC = Ricerca sulle Infezioni Associate a ImpiAnto o sostituzione di CIED; SE = Sensitivity; SP = Specificity.