Literature DB >> 35035048

Finite Element Analysis of effect of cusp inclination and occlusal contacts in PFM and PEEK implant-supported crowns on resultant stresses.

Githanjali Manchikalapudi1, Sreeramulu Basapogu2.   

Abstract

BACKGROUND: Effect of prosthesis design on occlusal overload and long-term implant stability cannot be overstated. In Porcelain Fused to Metal (PFM) crowns, low cusp inclination and occlusal contacts limited to central fossa ensure axially directed forces on an implant but often pose esthetic and functional challenges. It is theorized that resilient Polyetheretherketone (PEEK) crowns have shock absorption capacity for favorable stress distribution. This study compared two implant crown materials and evaluated the effect of cusp inclination and occlusal contact distribution on resultant stresses.
METHODS: Thirty 3D finite element models of implant-supported PFM and PEEK crowns, generated using Solidedge 3D CAD solid modeling software (v19, Siemens PLM Software Inc.,US), were used to study the effect of 3 cups inclinations (0°, 15°, 30°) under five load conditions, with 300N force distributed over one, two, or three contact areas and exported to ANSYS (v18.1, ANSYS Inc. Pennsylvania, US) for stress analysis.
RESULTS: Maximum stress in both PFM and PEEK models was at the neck of the implant under Load 3(300N distributed over three contact areas: central fossa, buccal cusp tip, marginal ridge). Minimum stress in all models was under Load 1(300N applied at one contact area in central fossa). Maximum stresses were recorded for 30° cusp inclination in PFM models.
CONCLUSION: In both PFM and PEEK crown models, contact areas placed away from the implant axis generated greater implant and peri-implant stresses and had more effect on resultant stresses than that of increase in cusp inclination. The effect of cusp inclination on the resultant stresses was dependent on the crown material.
© 2021 Director General, Armed Forces Medical Services. Published by Elsevier, a division of RELX India Pvt. Ltd.

Entities:  

Keywords:  Dental implant; Dental occlusion; Dental stress analysis; Finite element analysis; Polyetheretherketone; Porcelain-metal

Year:  2021        PMID: 35035048      PMCID: PMC8737102          DOI: 10.1016/j.mjafi.2020.11.014

Source DB:  PubMed          Journal:  Med J Armed Forces India        ISSN: 0377-1237


  28 in total

1.  In vivo registration of force development with ceramic and acrylic resin occlusal materials on implant-supported prostheses.

Authors:  Roger Bassit; Håkan Lindström; Bo Rangert
Journal:  Int J Oral Maxillofac Implants       Date:  2002 Jan-Feb       Impact factor: 2.804

Review 2.  Biological complications with dental implants: their prevention, diagnosis and treatment.

Authors:  N P Lang; T G Wilson; E F Corbet
Journal:  Clin Oral Implants Res       Date:  2000       Impact factor: 5.977

3.  Effects of occlusal inclination and loading on mandibular bone remodeling: a finite element study.

Authors:  Chaiy Rungsiyakull; Pimdeun Rungsiyakull; Qing Li; Wei Li; Michael Swain
Journal:  Int J Oral Maxillofac Implants       Date:  2011 May-Jun       Impact factor: 2.804

Review 4.  Finite-element modeling and analysis in nanomedicine and dentistry.

Authors:  Andy H Choi; Richard C Conway; Besim Ben-Nissan
Journal:  Nanomedicine (Lond)       Date:  2014-08       Impact factor: 5.307

5.  The effect of direct and indirect force transmission on peri-implant bone stress - a contact finite element analysis.

Authors:  Annike Rand; Meike Stiesch; Michael Eisenburger; Andreas Greuling
Journal:  Comput Methods Biomech Biomed Engin       Date:  2017-06-16       Impact factor: 1.763

Review 6.  The Role of Occlusion in Implant Therapy: A Comprehensive Updated Review.

Authors:  Rachel A Sheridan; Ann M Decker; Alexandra B Plonka; Hom-Lay Wang
Journal:  Implant Dent       Date:  2016-12       Impact factor: 2.454

Review 7.  Occlusal overload and bone/implant loss.

Authors:  Ignace Naert; Joke Duyck; Katleen Vandamme
Journal:  Clin Oral Implants Res       Date:  2012-10       Impact factor: 5.977

8.  The effect of platform switching on stress distribution in implants and periimplant bone studied by nonlinear finite element analysis.

Authors:  Siyu Liu; Chunbo Tang; Jinhua Yu; Wenyong Dai; Yidong Bao; Dan Hu
Journal:  J Prosthet Dent       Date:  2014-05-29       Impact factor: 3.426

Review 9.  Mechanical complications of dental implants.

Authors:  M S Schwarz
Journal:  Clin Oral Implants Res       Date:  2000       Impact factor: 5.977

10.  Biomechanical considerations on tooth-implant supported fixed partial dentures.

Authors:  Konstantinos X Michalakis; Pasquale Calvani; Hiroshi Hirayama
Journal:  J Dent Biomech       Date:  2012-10-29