Literature DB >> 34990461

Empirical modeling of the percent depth dose for megavoltage photon beams.

Xiao-Jun Li1, Yan-Cheng Ye1, Yan-Shan Zhang1, Jia-Ming Wu1,2,3.   

Abstract

INTRODUCTION: This study presents an empirical method to model the high-energy photon beam percent depth dose (PDD) curve by using the home-generated buildup function and tail function (buildup-tail function) in radiation therapy. The modeling parameters n and μ of buildup-tail function can be used to characterize the Collimator Scatter Factor (Sc) either in a square field or in the different individual upper jaw and lower jaw setting separately for individual monitor unit check. METHODS AND MATERIALS: The PDD curves for four high-energy photon beams were modeled by the buildup and tail function in this study. The buildup function was a quadratic function in the form of [Formula: see text] with the main parameter of d (depth in water) and n, while the tail function was in the form of e-μd and was composed by an exponential function with the main parameter of d and μ. The PDD was the product of buildup and tail function, PDD = [Formula: see text]. The PDD of four-photon energies was characterized by the buildup-tail function by adjusting the parameters n and μ. The Sc of 6 MV and 10 MV can then be expressed simply by the modeling parameters n and μ.
RESULTS: The main parameters n increases in buildup-tail function when photon energy increased. The physical meaning of the parameter n expresses the beam hardening of photon energy in PDD. The fitting results of parameters n in the buildup function are 0.17, 0.208, 0.495, 1.2 of four-photon energies, 4 MV, 6 MV, 10 MV, 18 MV, respectively. The parameter μ can be treated as attenuation coefficient in tail function and decreases when photon energy increased. The fitting results of parameters μ in the tail function are 0.065, 0.0515, 0.0458, 0.0422 of four-photon energies, 4 MV, 6 MV, 10 MV, 18 MV, respectively. The values of n and μ obtained from the fitted buildup-tail function were applied into an analytical formula of Sc = nE(S)0.63μE to get the collimator to scatter factor Sc for 6 and 10 MV photon beam, while nE, μE, S denotes n, μ at photon energy E of field size S, respectively. The calculated Sc were compared with the measured data and showed agreement at different field sizes to within ±1.5%.
CONCLUSIONS: We proposed a model incorporating a two-parameter formula which can improve the fitting accuracy to be better than 1.5% maximum error for describing the PDD in different photon energies used in clinical setting. This model can be used to parameterize the Sc factors for some clinical requirements. The modeling parameters n and μ can be used to predict the Sc in either square field or individual jaws opening asymmetrically for treatment monitor unit double-check in dose calculation. The technique developed in this study can also be used for systematic or random errors in the QA program, thus improves the clinical dose computation accuracy for patient treatment.

Entities:  

Mesh:

Year:  2022        PMID: 34990461      PMCID: PMC8735605          DOI: 10.1371/journal.pone.0261042

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


I. Introduction

The measurement of x-ray dose at the central axis in radiation oncology is usually tabulated and used for the clinical dose calculation. Percentage depth dose (PDD) and tissue-phantom ratio (TPR) are dominated by the scattering effect of depth and field size [1]. They consist mainly in two parts, primary fluence (adjusted for inverse square, beam hardening effects) and another part, which represents the effects of attenuation. PDD and the other quantities such as TPR and field size factor (Scp) are typically measured for simple square-shaped fields for each therapy machine and modality [2, 3]. The collection of clinical data for the implantation of treatment planning system for dose accuracy calculation is time consuming and repetitive. The achievement of this study provides a lot of help in double check of the measurement results to reduce the time of measurements, and the confidence to use interpolation. It would be helpful to know the smallest number of measurements needed to characterize the high-energy x-ray beam [4]. Several protocal provide a general framework and describes a large number of tests and procedures that should be considered by the users of RTPSs. However, the workload for the implementation of the recommendations from those documents are enormous and requires far more personnel and instrumentation resources than is available in most facilities, particularly within smaller hospitals. These hospitals are not always able to perform complete characterization, algorithm validation and software testing of dose calculation algorithms used in RTPS [5]. These may include the Sc factors, also known as head scatter factors, which account for the variation in beam output with field size from changes in direct and indirect radiation from the head of the linear accelerator. The term Scp contains both the collimator and phantom scatter that is defined as the ratio of dose for the field of interest to that of a reference field for the same delivered monitor units measured under full scatter conditions in a large water tank at the reference depth [6]. In this study, we proposed a simple mathematic equation to model the PDD by using the buildup-tail function. The quantity of interest is the parameters extracted from the well modeled PDD buildup-tail function for characterizing the Sc either in a square field or in the different individual upper jaw and lower jaw setting separately for specific patient treatment monitor unit calculation.

II. Materials and methods

II. A Experiment design and steps

The experiment was conducted in the following steps: The PDD for four high-energy photon energies of 4 MV, 6 MV, 10 MV, 18 MV were modeled in this study. The measurements data were 6 MV, 10 MV (Elekta infinity, Stockholm, Sweden, and Varian trueBeam, Palo Alto, Ca), while 4 MV and 18 MV were published data [7, 8]. Measurement of photon beam PDD was conducted by two linear accelerators with two-photon energies of 6 MV and 10 MV at SSD = 100cm with different field sizes. The quantity percentage depth dose is defined as the quotient, expressed as a percentage, of the absorbed dose at any depth d to the absorbed dose at a reference depth (usually at the depth of dose maximum) along the central axis of the beam. The Sc for radiation beams of two-photon energies of 6 MV and 10 MV were measured. High energy photon PDD curves were modeled by the buildup and tail function generated in this study. The PDD of four-photon energies were modeled by the buildup-tail function by adjusting the parameters n and μ to get the best fitting. The Sc of 6 MV and 10 MV can then be expressed simply by the modeling parameters n and μ. The details of each step are described in the following sections.

II. B Percent depth dose measurement

PDDs were acquired with a PTW MP3-T water phantom (PTW ionization Freiburg Gmbh) at WuWei Heavy Ion Center, Wuwei Cancer Hospital, Gansu, China (WHICH). PDDs were measured with a PTW Semiflex parallel-plate ionization chamber (PTW ionization Freiburg Gmbh, type 31010, volume 0.125 cm3). For the acquisition of PDDs the chamber position was automatically corrected to the effective point of measurement. In both photon energies water phantom measurements were performed at 100 cm source-to-surface distance (SSD) for square field sizes of 5 x 5, 10 x 10, 15 x 15, 20 x 20, and 40 x 40 cm2 with a step size of 0.1 cm. PDDs were normalized to 100% at dmax depth. Since the parallel-plate chamber has a small plate separation and it is explicit that the point of measurement is the front surface of the cavity. The depth curve measured by parallel-plate chamber was then compared with the PDD curve measured by films.

II. C The comparison of depth dose curve converted via parallel-plate chamber ionization curve with GAF EBT 3film

We used Gafchromic EBT3 films (Ashland Specialty Ingredients GP, NJ USA; Lot # 04022001) for the depth dose curves measurement in for comparing the PDD measured by plane-parallel ion chamber. The film processing and dose profile measurements followed the international protocols [9]. A pre-exposure technique was used for the calibration curve derivation [10]. This was performed by giving each film an initial dose of 2 Gy with a photon energy of 6 MV to homogenize the film polymer density in Gafchromic film experiment. The Hurter-Driffield calibration curve (H-D curve) [11] and the PDD films data scanned with a red filter [12] were analyzed by the VeriSoft imaging procession software (PTW-Freiburg, Germany, VeriSoft version 3.1) for further comparing with the depth dose curves measured by the parallel-plate chamber. Absolute output and machine quality assurance were performed before conducting the measurements of percent ionization depth by the parallel-plate chamber as well as PDD curve by Gafchromic films.

II.D Measurement of Sc

The measurement of Sc for Elekta infinity and Varian trueBeam with the high-energy photon energy of 6 and 10 MV were conducted using a PTW Semiflex chamber, type31010 (PTW, Freiburg), coupled to either a PTW UNIDOS or a Scanditronix-Wellhofer Dose1 (IBADosimetry, Schwarzenbruck) electrometer. The chamber and phantom fulfill the suitability criteria for this type of measurement according to the recommendation by the AAPM TG-74 report. The chamber was placed in an aluminum(ρ = 2.7 gcm3) mini phantom with 3.9 cm of the material above the chamber, a measurement depth beyond dmax equivalent to a depth of 10 cm water for sufficiently avoiding contaminant electrons. The phantom and chamber axis were vertically aligned to the beam central axis, and the chamber reference point was set at 100 cm source-to-chamber distance. All measurements were normalized to the reference field reading of 10 x10 cm2. Sc values were measured for a selection of square fields from 3 x 3 to 40 x 40 cm2 with the chamber in mini-phantom.

II. E percent depth dose numerical equation

There are two home-generated numerical equations for describing the PDD curves of high energy photon beam, buildup function, and tail function. The buildup function was a quadratic function in the form of with two main parameters of d (depth in water) and n, while the tail function was in the form of e− and was composed by an exponential function with main parameters of d and μ. The modeled PDD was the product of buildup and tail function to be, PDDb-t = . PDDb-t is described separately as buildup function and tail function in the following, where d is the depth in water along the central axis in unit cm, n is a beam hardening factor with unitless, a scalar. where d is the depth in water along the central axis in unit cm, μ is a linear attenuation coefficient factor in unit cm-1 for adjusting the slope of the tail. The tail function in the form of e− was composed of an exponential function with main parameters of d and μ. The empirical function of percentage depth dose is the combination of these two functions, denoted as PDDb-t (the abbreviation of PDDbuildup-tail) The PDD of four photon energies were modeled by the buildup-tail function by adjusting the parameters n and μ to get the best fitting. All PDD of four high-energy photon beams with different field sizes at SSD = 100cm were adjusted by the main parameters of n and μ to get the best fitting.

III. Results

III. A. The best fitting of percent depth dose was conducted by empirical function in four-photon energies

The dose variations of PDD of high energy photon beams measured by films and by ion chamber were less than 0.5%. The PDD with different energies adopted in this study was already measured by the water phantom during commissioning and was spot checked in this experiment. By adjusting the main parameters of n and μ, the best fitting for four-photon PDD curves in every energy at the field sizes of 10 cm x 10 cm were listed in Fig 1A–1D. The comparison of four high-energy photon measured and modeled PDD with two published and two facility data is listed in Table 1.
Fig 1

The best fitting of photon PDD curve at energy 4, 6, 10 and 18 MV from a–d), representatively. The measurements data were 6 MV (b) and 10 MV (c) while 4 MV (a) and 18 MV (d) were published data.

Table 1

The comparison of four high-energy photon measured and modeled PDD with two facility and two published data.

4 MV photon (published data)6 MV photon facility data)10 MV photon (facility data)18 MV photon (published data)
depth (cm)measured PDDmodeled PDDerror (%)measured PDDmodeled PDDerror (%)measured PDDmodeled PDDerror (%)measured PDDmodeled PDDerror (%)
0.138.8929.46-24.2440.1223.28-41.9738.9817.00-56.3932.2010.94-66.03
198.6798.750.0998.0997.12-0.9993.9190.70-3.4276.5578.092.01
1.598.8098.920.12100.08100.080.0097.3995.82-1.6289.5591.592.27
298.4498.590.1599.6299.56-0.07100.0299.80-0.2296.0797.391.38
2.596.1796.320.1698.0897.82-0.2799.2299.240.0299.0199.590.58
393.7893.930.1696.1095.66-0.4597.9698.230.27100.00100.000.00
3.591.2191.360.1693.8993.33-0.5996.0196.520.5399.3399.480.15
488.6588.800.1791.5990.95-0.7094.0494.630.6398.4298.430.01
4.586.1086.250.1789.2588.56-0.7892.0492.620.6397.2197.09-0.12
583.6183.760.1886.9286.19-0.8490.0190.570.6195.4095.570.18
5.581.1481.290.1884.6283.86-0.8987.9888.500.5993.5893.950.39
678.7878.930.1982.3581.58-0.9385.9786.440.5592.0992.270.20
6.576.4476.590.2080.1279.35-0.9783.8784.400.6490.1590.560.46
774.1974.340.2077.9477.17-0.9981.3882.401.2588.3488.840.57
7.571.9972.140.2175.8275.04-1.0279.1480.431.6386.5787.120.64
869.8570.000.2173.7472.97-1.0477.2778.501.5884.7285.410.82
8.567.7867.930.2271.7170.96-1.0575.4476.601.5483.0583.710.80
965.7565.900.2369.7468.99-1.0773.6474.741.5081.3182.040.89
9.563.8163.960.2467.8167.08-1.0871.8972.931.4579.6880.390.89
1061.8962.040.2465.9465.22-1.0970.1671.141.4078.0778.760.88
10.560.0660.210.2564.1263.41-1.1068.5069.431.3576.5277.160.83
1158.2658.410.2662.3461.65-1.1066.8567.721.3174.8375.581.01
11.556.5456.690.2760.6159.94-1.1165.2666.081.2673.3174.030.99
1254.8454.990.2758.9358.27-1.1263.6964.461.2171.7472.521.07
12.553.2053.350.2857.3056.66-1.1262.1762.891.1670.2871.041.08
1351.5851.730.2955.7155.08-1.1360.6861.351.1168.8169.561.09
13.549.9650.110.3054.1653.55-1.1359.2159.841.0667.4068.141.10
1448.3548.500.3152.6652.06-1.1457.8058.381.0165.9966.721.11
14.546.7346.880.3251.2050.61-1.1456.3956.930.9664.6565.351.09
1545.1145.260.3349.7749.20-1.1455.0555.550.9163.3063.991.08
15.543.5043.650.3448.3947.84-1.1553.7254.180.8661.9962.681.10
1641.8842.030.3647.0446.50-1.1552.4452.860.8160.6861.361.13
16.540.2640.410.3745.7445.21-1.1551.1751.560.7659.4360.101.14
1738.6538.800.3944.4643.95-1.1549.9450.300.7158.1858.841.15
17.537.0337.180.4143.2342.73-1.1648.7449.070.6656.9857.631.15
1835.4235.570.4242.0341.54-1.1647.5647.860.6155.7856.421.15
18.533.8033.950.4440.8640.38-1.1646.4346.690.5654.6155.261.19
1932.1832.330.4739.7239.26-1.1645.3045.530.5153.4554.101.23
19.530.5730.720.4938.6238.17-1.1644.2244.420.4652.3152.991.30
2028.9529.100.5237.5437.10-1.1643.1543.320.4151.1751.881.37
20.527.3427.490.5536.5036.07-1.1642.1142.270.3650.0950.811.42
2125.7225.870.5835.4835.07-1.1741.1041.230.3149.0249.741.47
21.524.1024.250.6234.4934.09-1.1740.1140.210.2647.9948.711.50
2222.4922.640.6733.5333.14-1.1739.1439.230.2146.9747.691.53
22.520.8721.020.7232.6032.22-1.1738.2038.260.1646.0146.711.52
2319.2519.400.7831.6931.32-1.1737.2837.320.1145.0545.721.50
The best fitting of photon PDD curve at energy 4, 6, 10 and 18 MV from a–d), representatively. The measurements data were 6 MV (b) and 10 MV (c) while 4 MV (a) and 18 MV (d) were published data. Fig 1A–1D represents the fitting results of PDD curves of photon energy from 4 to 18 MV, representatively. Table 2 lists the best fitting parameters n and μ for four-photon energies.
Table 2

This table shows the best fitting parameters n and μ for four-photon energies.

parameters4 MV6 MV10MV18MV
n0.170.2080.4951.2
μ0.06050.05150.04580.0422

III. B. The Sc expressed by the parameters n and μ generated in the empirical buildup-tail function of photon energy 6 and 10 MV

The Sc of photon energy 6 and 10 MV can be expressed in a certain acceptable deviation within 0.8% by the parameters n and μ as a function of field size in empirical buildup-tail function by the equation: nE and μE denote the parameters n and μ in empirical buildup-tail function at photon energy E, while FS stand for field size of interest. Fig 2A and 2B show the best fitting of the Sc by the Sc,E equation.
Fig 2

The Sc of photon energy 6 MV (a) and 10 MV (b) can be expressed in a certain acceptable deviation within 0.8% by using the parameters n and μ modeled in empirical Buildup-tail function by the equation of Sc,E = nE · (FS)0.63μE, nE and μE denote the parameters n and μ in empirical Buildup-tail function at photon energy E, FS denotes the field size.

The Sc of photon energy 6 MV (a) and 10 MV (b) can be expressed in a certain acceptable deviation within 0.8% by using the parameters n and μ modeled in empirical Buildup-tail function by the equation of Sc,E = nE · (FS)0.63μE, nE and μE denote the parameters n and μ in empirical Buildup-tail function at photon energy E, FS denotes the field size. The measured Sc of two-photon energies can be characterized by the parameters n and μ generated in the buildup-tail function and the comparison between characterized and measured was listed in Table 3.
Table 3

This table shows the deviation of calculated and measured Sc at different square field sizes of Varian at photon energy 6 and 10 MV.

Field Size (cm x cm)6 MV Sc modelingFF Varian 6 MV measurementFF 6 MV modeling/meas. error (%)10 MV Sc modelingFF Varian 10 MV measurementFF 10 MV modeling/meas. error (%)
40.9709±0.0050.9690±0.0040.8005±0.0020.9739±0.0030.9670±0.0040.0072±0.003
50.9778±0.0040.9727±0.0050.5204±0.0030.9802±0.0030.9742±0.0030.0062±0.003
60.9836±0.0050.9810±0.0030.2612±0.0020.9854±0.0040.9808±0.0040.0047±0.002
70.9885±0.0030.9875±0.0030.1007±0.0020.9898±0.0040.9869±0.0040.0029±0.002
80.9928±0.0020.9921±0.0030.0692±0.0030.9936±0.0030.9923±0.0050.0013±0.003
90.9966±0.0040.9960±0.0040.0590±0.0030.9970±0.0040.9970±0.0040.0000±0.004
101.0000±0.0051.0000±0.0050.0000±0.0041.0000±0.0031.0000±0.0040.0000±0.003
121.0059±0.0061.0080±0.003-0.2051±0.0021.0053±0.0041.0071±0.003-0.0018±0.003
141.0110±0.0041.0130±0.005-0.1997±0.0031.0098±0.0031.0120±0.003-0.0022±0.004
161.0154±0.0051.0170±0.005-0.1607±0.0021.0137±0.0041.0160±0.003-0.0023±0.003
181.0193±0.0051.0210±0.004-0.1710±0.0031.0171±0.0041.0201±0.003-0.0029±0.004
201.0227±0.0041.0236±0.003-0.0836±0.0021.0202±0.0031.0234±0.004-0.0031±0.003
221.0259±0.0051.0270±0.003-0.1060±0.0031.0230±0.0031.0261±0.003-0.0030±0.003
241.0288±0.0061.0290±0.003-0.0183±0.0021.0256±0.0041.0288±0.003-0.0031±0.003
261.0315±0.0041.0310±0.0040.0472±0.0021.0280±0.0041.0310±0.004-0.0029±0.002
281.0340±0.0051.0330±0.0030.0939±0.0021.0302±0.0051.0331±0.004-0.0028±0.003
301.0363±0.0051.0345±0.0030.1728±0.0031.0322±0.0041.0349±0.003-0.0026±0.004
351.0415±0.0041.0385±0.0040.2873±0.0031.0368±0.0041.0394±0.003-0.0025±0.003
401.0460±0.0061.0399±0.0040.5871±0.0031.0408±0.0041.0429±0.003-0.0020±0.003
Table 3 shows the measured Sc in a range of 0.9709 to 1.046 for Varian 6 MV flattened photon beam and 0.9739 to 1.0408 for 10 MV flattened photon beam at square field sizes from 4 cm x 4 cm to 40 cm x 40 cm. The deviation of Sc was characterized by Eq 2 and the measurements were between 0.8% to -0.2% for 6 MV, while for 10 MV were within 0.1%.

IV. Discussion

To let readers better understand the origin of the home-generated Buildup-tail function. The author describes the derivation of this function in the following. The buildup-tail model originated from the proportion function y(x) = . When x increases from -∞ to 0, the curve of y is located in the region of the (-,-) quadrant. When x goes from 0 to +∞, the curve of y is located in the region of the (+, +) quadrant. Let be , then the curve of y(x) falls in the (-,+) and the (+,+) quadrants. The curve of y(x) = has a left and right tail of the dose-profile-shape pattern. Let y(x) = = . When x = 0, y(x) becomes infinite which does not happen in real dose-profiles. Therefore, we insert n into y(x) to be: f(x) = , where n>0., let tail(x) = , where x is the depth in water in unit of cm, n is a scalar of spread factor in real number. On other hand, the function tail(x) = demonstrates an ascending values tail(x)with an increasing depth of x in water. When introduces an exponential function e to tail(x), the combination becomes PDDb-t function , namely, PDDb-t = , where d is the depth in water in the unit of cm, n>0 and is a harden factor of real number scalar. When x = 0, n plays an important role to avoid primary(x) become infinite, while μ is the linear attenuation factor to fine turn the growth of the value. Finally, the buildup-tail model can be expressed as follows: The PDD can be fitted by using the buildup-tail modeling by adjusting the main parameters of n and μ in all photon energy for the standard PDD curves in Fig 1A–1D. The random variations of modeled PDD with measured PDD had a maximum deviation within 1.5% as shown in Table 1. The parameters n and μ represent the photon beam hardening factor of the buildup function and the beam penetration ability of the tail function, respectively. The more photon energy, the more n and less μ it has for the PDD curve fitting as shown in Table 2. Fig 2 show the parameter n and μ in describing the measured Sc in the energy of 6 and 10 MV photon beam in Fig 2A and 2B, respectively. The Sc of photon energy 6 and 10 MV can be expressed with a certain deviation within 0.8% by using the parameters n and μ generated in empirical buildup-tail function by Eq 2, Where nE and μE denote the parameters n and μ as a function of field size, FS, in empirical buildup-tail function at photon energy E. A high energy photon beam usually has a high penetration ability, namely, has a small attenuation coefficient μ and also has a large n to own a less surface dose and the deeper dmax. The combination of a small μ and a large n, large μ, and small n can characterize a high and low energy photon beam percentage depth dose, respectively (Table 2). The concept of in-air output ratio Sc was introduced to characterize how the incident photon fluence per monitor unit varies with collimator settings, and the author believe the Sc of a larger photon energy is more sensitive in accordance with the field size, namely, the larger collimator setting, the more Sc it is. This might be owing to the strong side scatter of the higher energy of photons. As we can see in Fig 2B, the Sc of 10 MV increases gradually when the field sizes open widely, on the contrary, Fig 2A, the Sc of 6 MV becomes slightly saturation when the field sizes open widely. This phenomenon causes the curvature of Sc of photon energy 10 MV to be easy fitted by the parameters n and μ as a function of field size than in a photon energy of 6 MV. Table 3 shows the deviation of characterized and measured Sc of Varian photon energy at 6 and 10 MV. Since parameter n represents photon beam hardening factor in buildup function, in other words, the larger n (n = 4.95) the high beam quality, therefore you can see the larger n, the less surface dose, and the deeper dmax (to compare n = 4.95 and n = 0.0495 in Fig 3).
Fig 3

The parameter n represents the photon beam hardening factor in the buildup, the larger n (n = 4.95) the high beam quality, thus accompany the less surface dose and the deeper dmax (to compare n = 4.95 and n = 0.0495 in Fig 2).

The parameter μ represents the attenuation coefficient of the photon beam in the medium. The larger μ (μ = 0.458) the more attenuation, thus, the steeper curves and to have a shortened range (to compare μ = 0.458 and μ = 0.00458).

The parameter n represents the photon beam hardening factor in the buildup, the larger n (n = 4.95) the high beam quality, thus accompany the less surface dose and the deeper dmax (to compare n = 4.95 and n = 0.0495 in Fig 2).

The parameter μ represents the attenuation coefficient of the photon beam in the medium. The larger μ (μ = 0.458) the more attenuation, thus, the steeper curves and to have a shortened range (to compare μ = 0.458 and μ = 0.00458). On the other hand, μ represents the attenuation coefficient, meanwhile, tail function represents the beam penetration ability of high energy photon beam. As shown in Fig 3, the larger μ (μ = 0.458) the more attenuation when photon penetrating in the medium, therefore, you can see the larger μ, the steeper curves and to have a shortened range (to compare μ = 0.458 and μ = 0.00458). It can be seen that the change of Sc at square field size with a range of 0.951 to 1.048 for the Elekta 6 MV flattened beam. It is very interesting that a weaker variation can be seen when the upper jaw and lower jaw collimator setting was reversed. For example, the Sc of upper jaw x lower jaw setting at 10 cm x 15 cm and 15 cm x 10 cm was 1.0049, 1.0104 in Table 4, respectively.
Table 4

This figure shows Sc of Elekta 6 MV photon beam with different upper and low jaw setting with field sizes from 4 cm x 4 cm to 40 cm x 40 cm.

Upper jaw setting(Y, cm)Lower jaw setting (X, cm)
45101520253040
40.95110.95630.96710.9720.97390.97450.97510.9752
50.95870.9610.97730.97960.98010.98050.98140.9823
100.9730.975911.00491.00631.00771.00861.0091
150.97980.97961.01041.01631.01811.02081.02221.0231
200.98560.98911.01671.02221.02581.02671.02891.0308
250.98720.99091.02041.02671.02851.03081.03261.0344
300.99030.99321.02311.02891.03311.03441.03761.0394
400.99520.99551.02851.03581.03811.04261.04491.048
The Sc of Elekta 6 MV photon beam with different upper and low jaw setting from 4 cm x 4 cm to 40 cm x 40 cm was listed in Table 4. The orthogonal equal upper and lower jaw setting (square field) from the upper left to the lower right of 4 cm x 4 cm to 40 cm x 40 cm follow the Eq 2 while Sc can be calculated for the different upper jaw and lower jaw setting separately following the equation below, The Sc of different upper and lower jaw settings can be calculated separately by using Eq 3. In Table 5a, the Sc can be calculated sequentially by using Eq 3 for various lower settings with upper jaw fixed at 4 cm, 5 cm, 20 cm, 30 cm, and 40 cm, while in Table 5b, the Sc was calculated sequentially for various upper setting with lower jaw fixed at 4 cm, 5 cm, 20 cm, 30 cm, and 40 cm. The Sc calculation by separating the individual upper or lower jaws setting does a lot of favor in individual treatment monitor unit double-checking at jaws opened asymmetrically [13].
Table 5

a. shows the Sc can be calculated sequentially by using Eq 3 for various lower settings with upper jaw fixed at 4 cm, 5 cm, 20 cm, 30 cm, and 40 cm.

b. shows the Sc was calculated sequentially for various upper settings with lower jaw fixed at 4 cm, 5 cm, 20 cm, 30 cm, and 40 cm.

Field Size (cm x cm)Sc (square field)Sc calculatederrorupper fix at 4 cm, lower jaw increasedupper fix at 5 cm, lower jaw increasedupper fix at 20 cm, lower jaw increasedupper fix at 30 cm, lower jaw increasedupper fix at 40 cm, lower jaw increased
upper jaw (Y, cm)lower jaw (X, cm)Sc (4 cm X lower jaw)Sc calculatederrorSc (5 cm X lower jaw)Sc calculatederrorSc (20 cm X lower jaw)Sc calculatederrorSc (30 cm X lower jaw)Sc calculatederrorSc (40 cm X lower jaw)Sc calculatederror
440.9510.9560.5500.9510.9520.5500.9590.9620.1040.9860.9950.0100.9900.9960.0060.9951.0050.010
550.9610.9650.3840.9560.9550.3840.9610.964-0.1730.9890.9980.0090.9931.0000.0070.9961.0090.013
10101.0000.991-0.8880.9670.963-0.8880.9770.972-0.4631.0171.006-0.0101.0231.011-0.0111.0291.020-0.008
15151.0161.007-0.9230.9720.967-0.9230.9800.977-0.4821.0221.011-0.0111.0291.018-0.0111.0361.027-0.009
20201.0261.018-0.7330.9740.971-0.7330.9800.980-0.3331.0261.015-0.0111.0331.023-0.0101.0381.032-0.006
25251.0311.027-0.3510.9750.973-0.3510.9810.983-0.1271.0271.018-0.0091.0341.027-0.0081.0431.035-0.007
30301.0381.035-0.2980.9750.975-0.2980.9810.9850.0301.0291.020-0.0091.0381.030-0.0081.0451.038-0.006
40401.0481.046-0.1740.9750.979-0.1740.9820.9890.3661.0311.023-0.0071.0391.034-0.0051.0481.043-0.004
Field Size (cm x cm)Sc (square field)Sc calculatederrorlower fix at 4 cm, upper jaw increasedlower fix at 5 cm, upper jaw increasedlower fix at 20 cm, upper jaw increasedlower fix at 30 cm, upper jaw increasedlower fix at 40 cm, upper jaw increased
upper jaw (Y, cm)lower jaw (X, cm)Sc (upper jaw x 4 cm)Sc calculatederror (%)Sc (upper jaw x 5 cm)Sc calculatederror (%)Sc (upper jaw x 20 cm)Sc calculatederror (%)Sc (upper jaw x 30 cm)Sc calculatederror (%)Sc (upper jaw x 40 cm)Sc calculatederror (%)
440.9510.9560.5500.9510.940-1.1640.9560.946-1.0420.9740.9871.2970.9750.9810.6340.9750.9820.699
550.9610.9650.3840.9590.946-1.2890.9610.953-0.8640.9800.9931.3320.9810.9880.6600.9820.9880.530
10101.0000.991-0.8880.9730.966-0.6960.9760.973-0.3261.0061.0140.7671.0091.0090.0031.0091.005-0.429
15151.0161.007-0.9230.9800.978-0.1790.9800.9850.5121.0181.0260.8181.0221.021-0.1201.0231.015-0.791
20201.0261.018-0.7330.9860.9870.0940.9890.9930.4101.0261.0350.9291.0291.0300.0901.0311.022-0.821
25251.0311.027-0.3510.9870.9930.6030.9911.0000.9011.0291.0421.3401.0331.0370.4011.0341.028-0.613
30301.0381.035-0.2980.9900.9990.83810.9931.0051.2191.0331.0481.4421.0381.0420.4651.0391.033-0.640
40401.0481.046-0.1740.9951.0071.21140.9961.0141.8601.0381.0571.8291.0451.0510.6281.0481.040-0.744

a. shows the Sc can be calculated sequentially by using Eq 3 for various lower settings with upper jaw fixed at 4 cm, 5 cm, 20 cm, 30 cm, and 40 cm.

b. shows the Sc was calculated sequentially for various upper settings with lower jaw fixed at 4 cm, 5 cm, 20 cm, 30 cm, and 40 cm.

V. Conclusions

The dosimetric quantities Sc, even the quantity Sp by dividing Scp with Sc, and hence required for planning system measurement, or a monitor unit check methodology, were easily and accurately parameterized for our flattened beams using a simple mathematical expression in this study. The data reproduced may be used as expectation values for comparison when commissioning similar beams, as there are scant published data on Sc of all photon beams from these accelerator types. In this study, we presented a study with an empirical method to model the PDD curve for high energy photon beam by using the buildup and tail function in radiation therapy. The modeling parameters n and μ can also be used to predict the Sc in either square field or with jaws opening asymmetrically for individual treatment monitor unit double check in the patient’s dose calculation. The achievement of this study provides a lot of help in double check of the measurement results to reduce the time of measurements, and the confidence to use interpolation. It would be helpful to know the smallest number of measurements needed to characterize the high-energy x-ray beams. (XLSX) Click here for additional data file. 5 Oct 2021 PONE-D-21-29815Empirical Modeling of the Percent Depth Dose for Megavoltage Photon BeamsPLOS ONE Dear Dr. Wu, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please submit your revised manuscript by Nov 19 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript: A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'. A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'. An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. We look forward to receiving your revised manuscript. Kind regards, Suhairul Hashim, PhD Academic Editor PLOS ONE Journal Requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf 2. We note that the grant information you provided in the ‘Funding Information’ and ‘Financial Disclosure’ sections do not match. When you resubmit, please ensure that you provide the correct grant numbers for the awards you received for your study in the ‘Funding Information’ section 3. Thank you for stating the following in the Acknowledgments Section of your manuscript: "This work was supported by the Funding: Key R&D plan of Science and Technology Program of Gansu Province, China. (18YF1FH153)." We note that you have provided funding information that is not currently declared in your Funding Statement. However, funding information should not appear in the Acknowledgments section or other areas of your manuscript. We will only publish funding information present in the Funding Statement section of the online submission form. Please remove any funding-related text from the manuscript and let us know how you would like to update your Funding Statement. Currently, your Funding Statement reads as follows: "This work was supported by the Funding: Key R&D plan of Science and Technology Program of Gansu Province, China. (18YF1FH153)." Please include your amended statements within your cover letter; we will change the online submission form on your behalf. 4. In your Data Availability statement, you have not specified where the minimal data set underlying the results described in your manuscript can be found. PLOS defines a study's minimal data set as the underlying data used to reach the conclusions drawn in the manuscript and any additional data required to replicate the reported study findings in their entirety. All PLOS journals require that the minimal data set be made fully available. For more information about our data policy, please see http://journals.plos.org/plosone/s/data-availability. "Upon re-submitting your revised manuscript, please upload your study’s minimal underlying data set as either Supporting Information files or to a stable, public repository and include the relevant URLs, DOIs, or accession numbers within your revised cover letter. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories. Any potentially identifying patient information must be fully anonymized. Important: If there are ethical or legal restrictions to sharing your data publicly, please explain these restrictions in detail. Please see our guidelines for more information on what we consider unacceptable restrictions to publicly sharing data: http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions. Note that it is not acceptable for the authors to be the sole named individuals responsible for ensuring data access. We will update your Data Availability statement to reflect the information you provide in your cover letter. 5. PLOS requires an ORCID iD for the corresponding author in Editorial Manager on papers submitted after December 6th, 2016. Please ensure that you have an ORCID iD and that it is validated in Editorial Manager. To do this, go to ‘Update my Information’ (in the upper left-hand corner of the main menu), and click on the Fetch/Validate link next to the ORCID field. This will take you to the ORCID site and allow you to create a new iD or authenticate a pre-existing iD in Editorial Manager. Please see the following video for instructions on linking an ORCID iD to your Editorial Manager account: https://www.youtube.com/watch?v=_xcclfuvtxQ Additional Editor Comments: Please address the following concerns by the reviewers to improvise the current manuscript. [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Yes ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: Yes ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: No ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: No Reviewer #2: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: This manuscript reports the development of an empirical method to model the high-energy photon beam percent depth dose (PDD) curve. It is an interesting and potentially beneficial for generating physical parameters used for monitor unit checking in radiation therapy. The accuracy of the mathematical formulation produced is very impressive. The manuscript in general is well written. The issue with the manuscript however, is with regard to the formatting and to a lesser extent, the language used. The authors should be more critical with the use of acronyms, and font formatting. For example: • The acronym PDD should be used after it has been defined when it first appeared in the manuscript. • Letter ‘c’ for Sc should be subscripted. ‘Sc’ should be used instead of ‘collimator scatter’ after it has been defined. • Scalar quantity ‘n’ should be italised. Some of the results are presented in the ‘Discussion’ section, which should go under the ‘Results’ section. More discussions on the methods and comparison with other methods/studies are desirable. The manuscript can also be benefited with proofreading exercise to polish up the language. Some other specific comments as follow: Abstract: Page 3, line 54-55. Please acronymise percentage depth dose. Please check throughout the manuscript. Page 3, line 57. Remove ‘collimator scatter’. Sc has been defined earlier. Please check throughout the manuscript. Page 4, line 73. Insert ‘to’ between ‘sizes’ and ‘within’. Page 4, line 77. Add ‘setting’ after ‘clinical’. Page 4, line 77. Please subscript the letter ‘c’ in Sc. Check throughout manuscript. Introduction: Page 6, line 100. Do not capitalise letter ‘m’ in measurement. Page 6, line 103. Change ‘on’ to ‘of’ Page 6 and 7, line 108-119. The authors presented the physics and a number of physical parameters of high energy x-ray which can be found in common radiation physics textbook, which I strongly feel is unnecessary. The authors should instead provide and explain the problem statement/study background or brief literature review of similar study, leading to the aims/objectives of the manuscript. Page 7, line 127-133. The detailed explanation on the mathematical formula is not required here, as it is provided in the method. I suggest to just include the objective(s) of the current study, and its importance. Method: Page 9, line 159. Change ‘Percent depth curve measurement’ to ‘Percent depth dose measurement’ Page 9, Line 163. Superscript number ‘3’ on cm3. Page 10, line 181 to 183. Please clarify the statement ‘This was performed by giving each film a priming dose of 2 Gy, to homogenize the film density using WHICH facility with a dose of 1 Gy at the photon energy of 6 MV’. I don’t understand what is mean by priming the film to 2 Gy, with a dose of 1 Gy? Page 10, line 190. Please specify the version and manufacturer of the software used. Page 10, line 191. Gafchromic film is specified as GAF film in some places, and back to Gafchromic at some places. Please be consistent. The explanation of measurement using Gafchromic film in my opinion, is excessive. The method used for calibrating the film is widely published and the authors can just cite the reference(s), instead of explaining the procedures in great detail. Results: Page 14, line 246-247. It was stated that the measurement using ion chamber coincide with the film measurement. Please provide quantitative values on the agreement. Page 14, line 249. Change ‘at the commissioning’ to ‘during commissioning’. Page 15-16, Figure 1(a)-(d). Despite the excellent agreement between the fitting and measurement, it is imperative to provide the values of goodness of fit (or other quantitate deviation measurement) as a quantitative measure on the fitting performance. Page 17, line 305-309. The expression of Sc should be provided in the methods section. Is ‘FS’ the dimension of field size? This should be stated. Page 18, caption for Figure 2. It was stated that ‘the Sc of photon energy 6 MV (a) and 10 MV 321 (b) can be expressed perfectly by using the parameters n and μ’. I disagree to that because albeit from a reasonably good fit, it is not perfect. There are some slight disagreements at certain field sizes. Page 19, Table 2. I noted the deviations for 6 MV are in general larger than 10 MV. Any explanations for this? Discussion: Some of the results presented here should go under the ‘Results’ section. For instance, page 21, lines Page 21, lines 383-390. Page 24, line 440. Change ‘interest’ to ‘interesting’. Page 24, line 440-442. It was stated that ‘It is very interest that a weaker variation can be seen when the upper jaw and lower jaw collimator setting was reversed’. Any plausible reason/explanation for this? Some tables here (Figure 3 and 4) are results and should go under the results section. More discussion on the utility of the empiral mathematical formula is desirable. Are there any similar empirical methods/mathematical formula reported in the literature? What are the benefits of this method for generating physical parameters for MU checking? What is the future work/recommendations as I believe more work is required to verified the accuracy/feasibility of using such formula in clinical setting? Reviewer #2: Title: Empirical Modeling of the Percent Depth Dose for Megavoltage Photon Beams Manuscript number: PONE-D-21-29815 General comments: This paper describes and empirical method to model the PDD of the high-energy photon beam using a mathematical – buildup-tail function. The study was interesting because the proposed mathematical model could be useful to predict the collimator scatter factor, Sc for PDDs. Please clarify if the measured PDD in Fig 1a-d, were PDD from published literature or from ion chamber measurements. If they were from published data, how did the author managed to acquire such detail of individual measurement points? There are numerous language and grammatical errors throughout the manuscript, unnecessary capitalization at random places, subscripts for Sc etc should be carefully pick-up and corrected by the authors. Specific comments: Introduction – Could the authors present the background and the need for this study? Because, typically PDDs and Sc are measured from the actual beam, because the beams are different for every linac, and depending on the detector used to measure it. Line 100 – measurement – do not capitalize “M” Line 149 and many other places – Sc - to make sure the c is subscripted Line 153 – four photon energies Line 159 – Percent depth dose curve Line 163 – please specify the company, state and country of the PTW. The volume is cm3. Line 182 – What is the reason to prime the film for 2Gy? Was it 1Gy or 2Gy? Please clarify? If the films are primed with 1-2 Gy, then, how would that affect the calibration curves? Where is the 0 Gy point? Line 186 – Why is the gafchromic calibration curve called a Hurter-Diffield calibration curve? There were description of the use of gaf film – but I fail to see the purpose of the film measurements, since the authors only used the ion chamber measured data. Line 200 – 214 – Please explain the linac used to measure the Sc. Line 222 – no need to capitalize percent depth dose Line 237 – four photon energies Line 240 – please write in full sentence. Figure 1a – d – The two modeled and published PDD appears to match very well, but, at the region of dose buildup and end of tail, the dose descrepencies are obvious. Perhaps an additional graph on the error/difference between the 2 curves can be added to the figures to show the differences? Figure 2 a and b – again, what are the errors of the measured values, i.e. error bars? Where were these values obtained from? Table 1 – These parameters are specific for the 4 PDDs that the authors had modeled on. Please discuss how these values are going to be applicable for any PDDs? Line 306 – no need to capitalize buildup Table 2 – shows the error between the measured and predicted Sc. Perhaps, displaying this in a graph format would be better and summarize the average, SD or max deviation at what field size. Unless the numerical value has any value for any PDD or field sizes of the reader. Line 359 – no need to capitalize buildup Interesting explanation of the equations. However, one of the limitation of the model, correct me if I am wrong, is unable to predict the dose at the surface, at d = 0 cm and the discrepancy of the model at the dose buildup region could be quite big. Can the authors comment on the errors at these regions? Line 427 – no figure 3 caption. Table 3 – should this be in the results? ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. 14 Oct 2021 Please check the point-to-point respond for the reviewer's file, and the color highlighted file for further procession. Thank you so much. Submitted filename: point-to-point reply to reviewers comments 20211011.pdf Click here for additional data file. 23 Nov 2021 Empirical Modeling of the Percent Depth Dose for Megavoltage Photon Beams PONE-D-21-29815R1 Dear Dr. Wu, We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements. Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication. An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. Kind regards, Suhairul Hashim, PhD Academic Editor PLOS ONE Additional Editor Comments (optional): Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: All comments have been addressed Reviewer #2: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: The authors have addressed my earlier concerns. I can now recommend to accept the paper for publication. Reviewer #2: The authors has adequately address the questions and the manuscript is much improved. I recommend for the acceptance of the manuscript. ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No 20 Dec 2021 PONE-D-21-29815R1 Empirical Modeling of the Percent Depth Dose for Megavoltage Photon Beams Dear Dr. Wu: I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. If we can help with anything else, please email us at plosone@plos.org. Thank you for submitting your work to PLOS ONE and supporting open access. Kind regards, PLOS ONE Editorial Office Staff on behalf of Dr. Suhairul Hashim Academic Editor PLOS ONE
  8 in total

1.  Modeling scatter-to-primary dose ratio for megavoltage photon beams.

Authors:  Ken Kang-Hsin Wang; Timothy C Zhu
Journal:  Med Phys       Date:  2010-10       Impact factor: 4.071

2.  Effect of correction methods of radiochromic EBT2 films on the accuracy of IMRT QA.

Authors:  O A García-Garduño; J M Lárraga-Gutiérrez; M Rodríguez-Villafuerte; A Martínez-Dávalos; T Rivera-Montalvo
Journal:  Appl Radiat Isot       Date:  2015-10-22       Impact factor: 1.513

3.  The characterization of unflattened photon beams from a 6 MV linear accelerator.

Authors:  Jason Cashmore
Journal:  Phys Med Biol       Date:  2008-03-11       Impact factor: 3.609

4.  Report of AAPM Therapy Physics Committee Task Group 74: in-air output ratio, Sc, for megavoltage photon beams.

Authors:  Timothy C Zhu; Anders Ahnesjö; Kwok Leung Lam; X Allen Li; Chang-Ming Charlie Ma; Jatinder R Palta; Michael B Sharpe; Bruce Thomadsen; Ramesh C Tailor
Journal:  Med Phys       Date:  2009-11       Impact factor: 4.071

5.  Monitor unit calculations for external photon and electron beams: Report of the AAPM Therapy Physics Committee Task Group No. 71.

Authors:  John P Gibbons; John A Antolak; David S Followill; M Saiful Huq; Eric E Klein; Kwok L Lam; Jatinder R Palta; Donald M Roback; Mark Reid; Faiz M Khan
Journal:  Med Phys       Date:  2014-03       Impact factor: 4.071

6.  Head scatter data for several linear accelerators (4-18 MV).

Authors:  K R Kase; G K Svensson
Journal:  Med Phys       Date:  1986 Jul-Aug       Impact factor: 4.071

7.  An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan.

Authors:  David Lewis; Andre Micke; Xiang Yu; Maria F Chan
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.506

8.  Report of AAPM Task Group 235 Radiochromic Film Dosimetry: An Update to TG-55.

Authors:  Azam Niroomand-Rad; Sou-Tung Chiu-Tsao; Michael P Grams; David F Lewis; Christopher G Soares; Leo J Van Battum; Indra J Das; Samuel Trichter; Michael W Kissick; Guerda Massillon-Jl; Paola E Alvarez; Maria F Chan
Journal:  Med Phys       Date:  2020-10-30       Impact factor: 4.506

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.