| Literature DB >> 34963836 |
Einat Slonimsky1, Brad Zacharia2, Alex Mamourian3.
Abstract
BACKGROUND: Programmable shunts can be adjusted to optimize CSF diversion in patients with hydrocephalus without the need for re-operation. Currently, all shunts incorporate radiopaque markers so that their setting can be determined on skull X-ray images. The purpose of this study was to evaluate whether the shunt setting could also be determined ex vivo and in vivo using the data from a standard head CT scan since one is nearly always obtained when patients with VP shunts present with new symptoms that could be due to shunt malfunction. Materials andEntities:
Keywords: computed tomography; dual energy ct; hydrocephalus; metal artifacts; programmable shunts
Year: 2021 PMID: 34963836 PMCID: PMC8702386 DOI: 10.7759/cureus.19818
Source DB: PubMed Journal: Cureus ISSN: 2168-8184
Figure 1The phantom, scanned with two different shunts taped to the left and right parietal bones of the skull
A dried human skull was used as the phantom and scanned with CertasTM Plus shunt taped to the parietal bone to replicate the beam hardening that is expected to occur in patients from the skull immediately adjacent to the shunt valve.
CT scan parameters for various shunts for both shunt settings
| CertasTM Plus | Polaris® | proGAV 2.0 ® | Hakim® | ||
| Position 1 | Scan 1 | CTDIvol 75, 140kVp, 330 mAs | CTDIvol 75, 140kVp, 330 mAs | CTDIvol 75, 140kVp, 330 mAs | CTDIvol 75, 140kVp, 330 mAs |
| Scan 2 | CTDIvol 60, 120 kVp, 390 mAs | CTDIvol 60, 120 kVp, 390 mAs | CTDIvol 60, 120 kVp, 390 mAs | CTDIvol 60, 120 kVp, 390 mAs | |
| Scan 3 | CTDIvol 40 80 kVp 430 mAs 140 kVp 215 mAs | CTDIvol 40 80 kVp 430 mAs 140 kVp 215 mAs | CTDIvol 40 80 kVp 430 mAs 140 kVp 215 mAs | CTDIvol 40 80 kVp 430 mAs 140 kVp 215 mAs | |
| Position 2 | Scan 4 | CTDIvol 75, 140 kVp, 330 mAs | CTDIvol 75, 140 kVp, 330 mAs | CTDIvol 75, 140 kVp, 330 mAs | CTDIvol 75, 140 kVp, 330 mAs |
| Scan 5 | CTDIvol 60, 120 kVp, 390 mAs | CTDIvol 60, 120 kVp, 390 mAs | CTDIvol 60, 120 kVp, 390 mAs | CTDIvol 60, 120 kVp, 390 mAs | |
| Scan 6 | CTDIvol 40 80 kVp 430 mAs 140 kVp 215 mAs | CTDIvol 40 80 kVp 430 mAs 140 kVp 215 mAs | CTDIvol 40 80 kVp 430 mAs 140 kVp 215 mAs | CTDIvol 40 80 kVp 430 mAs 140kVp V 215 mAs |
Shunt parameter setting for each programable shunt on both scans
| CertasTM Plus | Polaris® | proGAV 2.0 ® | Hakim® | |
| Side | Right | Left | Right | Left |
| Position 1 | 4 | 2 | 15 | 6 |
| Position 2 | 7 | 4 | 170 | 70 |
Figure 2Volume reconstructions of the four shunts
(A) CertasTM Plus (Codman, Raynham, Massachusetts). (B) Polaris® (Sophysa, Orsay, France). (C) Hakim® (Codman, Raynham, Massachusetts). (D) proGAV 2.0® (Braun, Bethlehem, Pennsylvania). Note the difficulty in detection of detail within the valves on volume rendered images (C) and (D). While the two lobes on the Certas valve (A) and their angle relative to the reference indicator (long arrow) are evident, it is essential to recognize the slight difference in the appearance of the two lobes. A small tab is present on only one lobe, and this is indicated on the volume rendered by the short red arrow. Correct identification of each of the two lobes is critical for correct setting prediction. The two lobes of the Polaris valve (B), on the other hand, are identical and only a limited range of angles are available relative to the long axis of the valve. This arrangement is very likely why we found a high degree of accuracy in determining the valve setting among our 10 observers.
Evaluation of shunt setting based on volume rendered images from phantom-CT scans
*CI: confidence interval (0-5)
| Correct read of shunt setting | ||
| Setting 1 (CI*) | Setting 2 (CI*) | |
| Polaris®shunt | 10/10 (4.2) | 10/10 (4.2) |
| CertasTM Plus shunt | 9/10 (4.0) | 10/10 (4.4) |
| Hakim® shunt | 0/10 (1.4) | 0/10 (1.4) |
| proGAV 2.0® shunt | 1/10 (1.7) | 0/10 (1.6) |
Figure 3CertasTM Plus volume rendered image reconstruction from routine non-contrast head CT
The shunt setting is predicted compared with the manufacturing guide and was correctly predicted as four. The lobes that come closest to the center marker is the one with the extra metal dot, assisting in reading the setting correctly. The setting is compared with the manufacture guidelines in relation to the radiopaque marker (arrow).
Figure 4CertasTM Plus valve manufacture guide for interpretation of shunt setting on x-Ray
Figure 5Manufacturer selections for materials and design of their programmable shunts very likely influences the conspicuity of the valve setting on the CT volume rendered images
(A) Hakim® (Codman, Raynham, Massachusetts). (B) proGAV 2.0® (Braun, Bethlehem, Pennsylvania). (C) Polaris® (Sophysa, Orsay, France). (D) CertasTM Plus (Codman, Raynham, Massachusetts).