Yilin Liu1, Li Zhan1, Jesse W Shen1, Bàrbara Baro2, Andrea Alemany3,4, James Sackrison5, Oriol Mitjà3,4,6, John C Bischof1,7. 1. Department of Mechanical Engineering, University of Minnesota, Minneapolis, Minnesota 55455, United States. 2. ISGlobal, Hospital Clínic, Universitat de Barcelona, Barcelona 08036, Spain. 3. Fight AIDS and Infectious Diseases Foundation, Badalona 08916, Spain. 4. Hospital Universitari Germans Trias i Pujol, Badalona 08916, Spain. 5. 3984 Hunters Hill Way, Minnetonka, Minnesota 55345, United States. 6. Lihir Medical Centre - International SOS, Lihir Island, New Ireland 633, Papua New Guinea. 7. Department of Biomedical Engineering, University of Minnesota, Minneapolis, Minnesota 55455, United States.
Abstract
The SARS-CoV-2 global pandemic created an unprecedented need for rapid, sensitive, and inexpensive point-of-care (POC) diagnostic tests to treat and control the disease. Many POC SARS-CoV-2 lateral flow immunoassays (LFAs) have been developed and/or commercialized, but with only limited sensitivity (μM-fM). We created an advanced LFA based on gold nanospheres (GNSs) with comprehensive assay redesign for enhanced specific binding and thermal contrast amplification (TCA) on GNSs for signal amplification, which enabled fM-aM detection sensitivity for SARS-CoV-2 spike receptor-binding domain (RBD) proteins within 30 min. The advanced LFA can visually detect RBD proteins down to 3.6 and 28.6 aM in buffer and human nasopharyngeal wash, respectively. This is the first reported LFA achieving sensitivity comparable to that of the PCR (aM-zM) by visual reading, which was much more sensitive than traditional LFAs. We also developed a fast (<1 min) TCA reading algorithm, with results showing that this TCA could distinguish 26-32% visual false negatives for clinical commercial LFAs. When our advanced LFAs were applied with this TCA, the sensitivities were further improved by eightfold to 0.45 aM (in buffer) and 3.6 aM (in the human nasopharyngeal wash) with a semiquantitative readout. Our proposed advanced LFA with a TCA diagnostic platform can help control the current SARS-CoV-2 pandemic. Furthermore, the simplicity and speed with which this assay was assembled may also facilitate preparedness for future pandemics.
The SARS-CoV-2 global pandemic created an unprecedented need for rapid, sensitive, and inexpensive point-of-care (POC) diagnostic tests to treat and control the disease. Many POC SARS-CoV-2 lateral flow immunoassays (LFAs) have been developed and/or commercialized, but with only limited sensitivity (μM-fM). We created an advanced LFA based on gold nanospheres (GNSs) with comprehensive assay redesign for enhanced specific binding and thermal contrast amplification (TCA) on GNSs for signal amplification, which enabled fM-aM detection sensitivity for SARS-CoV-2 spike receptor-binding domain (RBD) proteins within 30 min. The advanced LFA can visually detect RBD proteins down to 3.6 and 28.6 aM in buffer and human nasopharyngeal wash, respectively. This is the first reported LFA achieving sensitivity comparable to that of the PCR (aM-zM) by visual reading, which was much more sensitive than traditional LFAs. We also developed a fast (<1 min) TCA reading algorithm, with results showing that this TCA could distinguish 26-32% visual false negatives for clinical commercial LFAs. When our advanced LFAs were applied with this TCA, the sensitivities were further improved by eightfold to 0.45 aM (in buffer) and 3.6 aM (in the human nasopharyngeal wash) with a semiquantitative readout. Our proposed advanced LFA with a TCA diagnostic platform can help control the current SARS-CoV-2 pandemic. Furthermore, the simplicity and speed with which this assay was assembled may also facilitate preparedness for future pandemics.
The past decades have
witnessed a recurrence of multiple pandemics
such as severe acute respiratory syndrome (SARS), Middle East respiratory
syndrome coronavirus (MERS), H1N1 and other influenza viruses, the
Ebola virus, the Zika virus, and, most recently, the SARS-CoV-2 virus.[1,2] A recurring issue in the response to these pandemics is the lack
of rapid, accurate, and affordable diagnostic tests to efficiently
prevent and control the spread of the disease.[3] The SARS-CoV-2 pandemic has caused significant global morbidity
and mortality. As of 13 July 2021, there have been over 186 million
confirmed cases and over 4 million cumulative deaths worldwide due
to SARS-CoV-2.[4] To control the spread,
multiple testing methods have been developed and applied, such as
the RT-PCR, rapid tests (i.e., lateral flow immunoassays,
LFAs), the CRISPR-based assay, and diagnostic imaging (e.g., computed tomography on patients’ lungs).[5−7] The reverse
transcription-polymerase chain reaction (RT-PCR) is one of the most
widely used methods due to its ultrahigh sensitivity, but it usually
has a long turnaround time and requires expensive reagents, equipment,
and professional training.[3,8] While its claimed assay
time can be less than an hour, the actual time to diagnose a patient
suspected with SARS-CoV-2 infection was >24 h or even several days
during outbreaks due to the need to ship samples to laboratories and
the limited available testing resources (i.e., facilities,
reagents, and professionals).[3] In addition,
many underdeveloped countries lack sufficient resources and budget
to carry out population-wide RT-PCR tests.[9,10]Serological and antigen rapid tests were developed and commercialized
to address these difficulties and help relieve the diagnostic burden.
While these rapid LFA tests are faster, cheaper, and easier to use
than the RT-PCR, their performance results have been inconsistent,
with a notable lack of sensitivity and no quantification. Some cohort
studies showed good sensitivity of commercial serological and antigen
rapid tests when comparing results with confirmatory RT-PCR results,[11−14] but unsatisfactory diagnostic performance was also reported. The
Coris SARS-CoV-2 Ag Respi-Strip test, for example, showed sensitivity
as low as 30.2% for clinical samples,[15] and the BinaxNOW antigen rapid test cards were reported to have
an analytical sensitivity approximately equivalent to the cycle threshold
of 29–30 from a generic qRT-PCR, which may not be sufficient
to detect all relevant infections.[16] A
similar issue of low sensitivities of rapid antigen and antibody tests
was reported elsewhere, and these tests were less recommended for
diagnosing acute SARS-CoV-2 infection.[17−20] In short, the commercial rapid
antigen and antibody tests remain of limited usefulness for detecting
all relevant SARS-CoV-2 infections, and thus cannot fully serve the
immediate response to and management of the pandemic spreading worldwide.[3]To better detect SARS-CoV-2 infections
at the point-of-care (POC),
many research efforts have been put to improving the sensitivity of
commercial LFAs.[21−23]Table shows the recent literature applying either signal amplification
methods (e.g., fluorescence and surface-enhanced
Raman scattering) or assay improvement (e.g., novel
affinity molecules) to LFAs for better detection of SARS-CoV-2. Most
had assay time between 10 and 20 min and detection sensitivity in
the nM–fM (i.e., 10–9–10–15 M) range, still much lower than RT-PCR (aM–zM,
or 10–18–10–21 M). Although
it is recently reported that LFA read by scanning electron microscopy
(SEM) could achieve sensitivity comparable to RT-PCR,[24] the testing platform is too complicated and expensive for
POC use.
Table 1
Detection Sensitivity and Assay Time
of the Reported SARS-CoV-2 Antigen and Antibody Rapid Testsa,b
limit of detection (LoD)
analytes
LFA improvement
sample matrix
mass concentration
molarity
assay time
(min)
refs
IgG
fluorescence
diluted
serum by buffer
NA
10
(30)
IgG/IgM
fluorescence
buffer
NA. 104-fold
better than visual reading with colloidal gold
Assume that the molecular weights
for SARS-CoV-2 spike and nucleocapsid proteins are 175 and 114 kDa,
respectively, when they were not notified in the literature.
NA: not
applicable; SERS: surface-enhanced
Raman scattering; ACE2: angiotensin-converting enzyme 2; S-protein:
spike protein; N-protein: nucleocapsid protein; RBD: receptor-binding
domain; and TCA: thermal contrast amplification.Assume that the molecular weights
for SARS-CoV-2 spike and nucleocapsid proteins are 175 and 114 kDa,
respectively, when they were not notified in the literature.Meanwhile, assays with sample enrichment
and LFA readout have been
under rapid development, aiming for more accurate POC tests. The reverse
transcript loop-mediated isothermal amplification LFA (RT-LAMP-LFA)
system, for example, can detect SARS-CoV-2 RNA at concentrations of
≥2 copies/μL (3 aM) in 40 min.[25] Similar work used various isothermal amplification methods coupled
with an LFA readout, whose detection sensitivity was aM–zM
with assay time from approximately 30 min–1 h.[26−29] These are promising for future POC use to replace traditional PCR
tests, especially for population screening in epidemics. However,
challenges for these isothermal amplification LFA platforms still
exist and need to be addressed for future POC use.[21] Compared with LFAs, for example, these assays need more
expensive reagents, multiple steps, and isothermal incubators, and
have less tolerance to contamination.[21]To overcome these challenges, we established an advanced LFA
based
on gold nanospheres (GNSs) with thermal contrast amplification (TCA)
to detect the spike SARS-CoV-2 receptor-binding domain (RBD) antigen
for acute infection. To break through the fM detection limit of current
LFAs, we carried out comprehensive assay optimization along with signal
amplification (i.e., TCA). Specifically, we use high-affinity
antibodies, optimal running buffer, large GNSs, and an improved conjugation
method with increased antigen-binding sites for enhanced specific
binding (SB), along with TCA reading where the GNSs captured at test
lines were laser-excited to generate the amplified photothermal signals.
As a result, the advanced LFAs in this work provided 3.6 and 28.6
aM detection limits readable with the naked eye to test the RBD proteins
in buffer and human nasopharyngeal wash, respectively. This is the
first report on an advanced LFA achieving fM–aM analytical
sensitivity that can be read with the naked eye, which was much more
sensitive than conventional LFAs and even comparable to PCR’s
sensitivity range.[21]Table shows its advantage in detection sensitivity
to SARS-CoV-2 over other LFAs from the literature. Meanwhile, a fast
(<1 min) TCA reading was developed and identified 26–32%
visual false negatives from clinical commercial LFAs. When we applied
this TCA to our advanced LFAs, the detection limits were further lowered
to 0.45 aM (in buffer) and 3.6 aM (in human nasopharyngeal wash) in
a semiquantitative readout format. The advanced LFAs with TCA can
help prevent the recurrence of the current SARS-CoV-2 pandemic and
may also better prepare the world to diagnose and control the spread
of other diseases and prevent future pandemics.
Results and Discussion
TCA: Faster
Reading Without Loss of Accuracy
TCA can
improve the sensitivities of LFAs although the throughput of the traditional
reading algorithm needs improvement. With TCA, the captured gold nanoparticles
at a test line were excited by laser irradiation and showed stronger
photothermal effects than the background membrane, enabling the detection
of subvisual positives (Figure A).[38−43] As a result, TCA reading improved the sensitivities of clinical
commercial LFAs to diagnose group A Streptococcus and influenza A and B in previous cohort studies.[43,44] TCA reading also showed a semiquantitative readout in diagnosing
C-reactive proteins and HIV p24 proteins,[41,42] whose numerical thermal signals increased with increasing antigen
load (Figure B). However,
the old discrete algorithm for TCA reading (i.e.,
point-by-point reading across a test line, as shown in Figure C), used in the previous studies,[38−43] took about 15–20 min for one reading of an LFA.
Figure 1
Building and
characterizing a fast TCA reader by continuous reading.
(A) Schematic working mechanism of TCA, where gold nanoparticles are
excited under laser irradiation and their elevated temperature is
recorded by an infrared (IR) camera. (B) Readout formats by visual
with qualitative results (±) vs TCA with semiquantitative
(numeric) results, where strong, medium, or weak signal can correlate
to the antigen load. (C) Schematic of discrete (slow) and continuous
(fast) reading algorithms in a TCA reader. The thermal signal of a
test line was obtained by calculating the area under the temperature
curve, recorded by an IR camera when scanning across the test line.
(D) Thermal signals of test lines from calibration LFAs, read by both
continuous and discrete reading algorithms in the TCA reader. In discrete
reading, the heating time per point included 5.5 and 10.5 s. For continuous
scans, the scan velocity included 0.1 and 0.25 mm/s. The test lines
were printed with various concentrations of gold nanospheres (GNSs),
which were characterized by the peak absorbance (unit: OD, optical
density) for a 1 cm path length of light through their solutions.
(E) Measured time consumption per reading by both discrete and continuous
reading algorithms. For discrete readings, the heating time per point
ranged from 1.5 to 10 s. In continuous reading, the scan velocity
ranged from 0.1 to 2 mm/s. For all reading conditions, the number
of replicates was three. The statistical significance is indicated
with asterisks: ns: p > 0.05; *p < 0.05.
Building and
characterizing a fast TCA reader by continuous reading.
(A) Schematic working mechanism of TCA, where gold nanoparticles are
excited under laser irradiation and their elevated temperature is
recorded by an infrared (IR) camera. (B) Readout formats by visual
with qualitative results (±) vs TCA with semiquantitative
(numeric) results, where strong, medium, or weak signal can correlate
to the antigen load. (C) Schematic of discrete (slow) and continuous
(fast) reading algorithms in a TCA reader. The thermal signal of a
test line was obtained by calculating the area under the temperature
curve, recorded by an IR camera when scanning across the test line.
(D) Thermal signals of test lines from calibration LFAs, read by both
continuous and discrete reading algorithms in the TCA reader. In discrete
reading, the heating time per point included 5.5 and 10.5 s. For continuous
scans, the scan velocity included 0.1 and 0.25 mm/s. The test lines
were printed with various concentrations of gold nanospheres (GNSs),
which were characterized by the peak absorbance (unit: OD, optical
density) for a 1 cm path length of light through their solutions.
(E) Measured time consumption per reading by both discrete and continuous
reading algorithms. For discrete readings, the heating time per point
ranged from 1.5 to 10 s. In continuous reading, the scan velocity
ranged from 0.1 to 2 mm/s. For all reading conditions, the number
of replicates was three. The statistical significance is indicated
with asterisks: ns: p > 0.05; *p < 0.05.To boost throughput, we developed
a fast and continuous reading
algorithm (Figure C), allowing times of <1 min per reading with comparable performance
to the traditional discrete reading type. Characterization of the
fast-reading algorithm was conducted by reading calibration LFAs (detailed
in Section S1 in the Supporting Information).
Thermal signals and time consumption are shown in Figure D,E. In Figure D, both reading algorithms showed semiquantitative
thermal signals corresponding to concentrations of GNSs at test lines. Figure D also shows that
the continuous reading, with either a 0.1 or 0.25 mm/s scan velocity,
had a similar limit of detection (LoD) for GNSs (2–3 optical density, OD) to the discrete reading with 5.5 s heating
time per point. Although the discrete heating with 10.5 s heating
time had a twofold lower LoD (2–4 OD), its reading
time was >15 min, significantly longer than that of the continuous
reading (as low as <1 min), as shown in Figure E. While increasing scan velocity can further
reduce reading time, scans that are too quick may miss some information
from real LFAs, which are expected to be more complex than calibration
LFAs. Therefore, in this study, a 0.25 mm/s scan velocity was used
to continuously read both commercial and lab prototype LFAs, with
reading time <1 min.
Clinical Commercial SARS-CoV-2 LFAs: Reduction
of False Negatives
by TCA
To benchmark the improved performance of our new SARS-CoV-2
antigen LFA, we performed a baseline test of selected existing clinical
commercial SARS-CoV-2 LFAs by visual and TCA reading. Results showed
a reduction in false negatives to diagnose clinical samples by TCA
over visual reading on PanBio SARS-CoV-2 Ag Rapid LFAs from Abbott
and CLINITEST Rapid SARS-CoV-2 Antigen LFAs from Siemens. Comparison
of visual and thermal results in Figure B,C showed that TCA detected about 32 and
26% visual false negatives from Abbott and Siemens SARS-CoV-2 antigen
LFAs, respectively. Thermal signals of the visually false negatives
identified by TCA are plotted in Figure S1. This improved sensitivity of commercial SARS-CoV-2 antigen LFAs
by TCA is also consistent with our previous cohort studies for rapid
diagnosis of influenza[43] and group A: Streptococcus.[44]
Figure 2
TCA reading of commercial
SARS-CoV-2 antigen LFAs and illustration
of specific and nonspecific binding in LFAs. (A) Schematic flowchart
for testing clinical SARS-CoV-2 samples by PCR, commercial visual
SARS-CoV-2 antigen LFAs, and thermal contrast amplification (TCA)
LFA reading. (B, C) Statistics of visual and thermal results of (B)
Abbott and (C) Siemens SARS-CoV-2 antigen LFAs tested with clinical
samples. PCR results were taken as true negative and true positive
results for comparison. (D) Schematic of specific binding occurring
with nonspecific binding at test lines. Legends for B and C include
TN: true negative; FN: false negative; TP: true positive; and FP:
false positive.
TCA reading of commercial
SARS-CoV-2 antigen LFAs and illustration
of specific and nonspecific binding in LFAs. (A) Schematic flowchart
for testing clinical SARS-CoV-2 samples by PCR, commercial visual
SARS-CoV-2 antigen LFAs, and thermal contrast amplification (TCA)
LFA reading. (B, C) Statistics of visual and thermal results of (B)
Abbott and (C) Siemens SARS-CoV-2 antigen LFAs tested with clinical
samples. PCR results were taken as true negative and true positive
results for comparison. (D) Schematic of specific binding occurring
with nonspecific binding at test lines. Legends for B and C include
TN: true negative; FN: false negative; TP: true positive; and FP:
false positive.TCA also slightly increased false
positives for both Abbott and
Siemens LFAs, as shown in Figures B,C and S1. This is mainly
due to amplified noise from nonspecific binding (NSB) at test lines
by TCA.[43] It is known that NSB usually
exists along with SB in LFAs, corresponding to noise and signal, respectively.
As shown in Figure D, NSB can arise from different side reactions, including the hydrophobic
and electrostatic interactions of GNS conjugates with capture antibodies
and/or with the membrane, and the physical capture of GNS aggregates
in the membrane.[42,45,46] The accumulation of NSB can therefore cause false positives in visual
reading. This issue can be even more apparent after signal amplification,
thus setting a limit to sensitivity improvement for various signal
amplification methods.[21] To reduce NSB
and false positives, extensive assay optimization is needed to fit
with signal amplification. With a redesign of the assay, for example,
the LFA with TCA achieved ELISA-level sensitivity (∼pM) in
detecting HIV p24 proteins.[42]In
short, while it creates some false positives, TCA can reduce
visual false negatives in commercial LFAs. To dramatically improve
sensitivity (≤fM) and suppress possible false positives from
cross-reactions between built-up reagents, we developed advanced ultrasensitive
SARS-CoV-2 antigen LFAs by comprehensive assay optimization for maximal
SB and minimal NSB with signal amplification (TCA).
LFA Redesign
I: Antibody and Buffer Optimization
To
develop the high-performing SARS-CoV-2 antigen LFA, we optimized the
antibody (i.e., antibody pair and amount) and buffer
for maximal SB and minimal NSB. For quantitative optimization, NSB
was measured by the noise from the test line when testing a blank
buffer (i.e., negative controls), while SB was characterized
by the signal from the test line when testing an antigen-loaded buffer
(i.e., positive controls). First, the antibody pair
was optimized for maximal SB. Figure A compares the signals of SB between LFAs using two
antibody pairs with a controlled amount of GNSs per LFA and with a
controlled SARS-CoV-2 spike receptor-binding domain (RBD) antigen
concentration in buffer. The first antibody pair gained a much higher
SB signal than the second pair, while both assays showed invisible
test lines for blank buffer testing. To optimize the running buffer,
we, therefore, proceeded with the first pair. NSB was first minimized
to preclude any visible false positives, and the specific-to-nonspecific
binding ratio (SB/NSB) was then maximized.[42] The testing ranges of pH (7–9), bovine serum albumin (BSA,
0.5–2%), and Triton X-100 (0.5–2%) to optimize the running
buffer were selected based on our previous LFA studies using ∼100
nm GNSs.[41,42]Figure B shows that NSB was quite sensitive to buffer pH and
that minimal NSB noise occurred at pH 9. Figure C shows that 0.5–1% BSA offered good
assay performance, whereas a higher concentration (2%) led to stronger
NSB. Figure D suggests
that 0.5% Triton was optimal with the highest SB/NSB while further
increasing Triton reduced SB/NSB. Based on the abovementioned optimization,
the optimal template of running buffer for the first antibody pair
was 60 mM Tris–HCl buffer (pH 9), 0.5% BSA, 0.5% Triton X-100,
0.15 M NaCl, and 0.05% ProClin 300. Finally, the concentration of
the capture antibody precoated at test lines was optimized. Figure E suggests that increasing
the capture antibody’s concentration significantly increased
the SB/NSB ratio since the reaction rate of SB increased with the
concentration of the reactant. There might be a maximal SB/NSB value
at a higher concentration of the capture antibody. However, note that,
considering the assay cost and the concentration limit of the as-received
antibody (2.4 mg/mL), a 2 mg/mL concentration was used for the precoating
capture antibody, although a higher concentration may increase the
SB/NSB.
Figure 3
Optimizing antibody and buffer for SARS-CoV-2 spike receptor-binding
domain (RBD) protein LFA. (A) Comparison of SB of two antibody pairs
(Hytest and MyBioSource; see Materials for
details) to the RBD protein in an LFA format. The first pair was used
in the following assay optimization work. (B) Effects of buffer pH
on NSB. The concentrations of BSA, Triton X-100, NaCl, and ProClin
300 in buffer were 0.5%, 0.5%, 0.15 M, and 0.05%, respectively. (C)
Effects of the BSA concentration in running buffer on NSB. The pH
and the concentrations of Triton X-100, NaCl, and ProClin 300 in buffer
were 9, 1%, 0.15 M, and 0.05%, respectively. (D) Effects of the Triton
X-100 concentration in running buffer on the SB/NSB ratio. The pH
and the concentrations of Triton X-100, NaCl, and ProClin 300 in buffer
were 9, 1%, 0.15 M, and 0.05%, respectively. (E) Optimizing concentration
of the capture antibody at test lines by comparing SB/NSB ratios.
The SB and NSB in panel (E) were thermal signals from thermal contrast
amplification (TCA) reading of test lines, while those in (A–D)
were grayscale intensities, calculated as the area above the grayscale
curves across test lines plotted in ImageJ. NSB or SB was noise or
signal of a test line when an LFA was run with blank or antigen-loaded
buffers, respectively. The antigen concentrations were controlled
as the same for each set of studies in panels (A), (D), and (E), respectively.
Optimizing antibody and buffer for SARS-CoV-2 spike receptor-binding
domain (RBD) protein LFA. (A) Comparison of SB of two antibody pairs
(Hytest and MyBioSource; see Materials for
details) to the RBD protein in an LFA format. The first pair was used
in the following assay optimization work. (B) Effects of buffer pH
on NSB. The concentrations of BSA, Triton X-100, NaCl, and ProClin
300 in buffer were 0.5%, 0.5%, 0.15 M, and 0.05%, respectively. (C)
Effects of the BSA concentration in running buffer on NSB. The pH
and the concentrations of Triton X-100, NaCl, and ProClin 300 in buffer
were 9, 1%, 0.15 M, and 0.05%, respectively. (D) Effects of the Triton
X-100 concentration in running buffer on the SB/NSB ratio. The pH
and the concentrations of Triton X-100, NaCl, and ProClin 300 in buffer
were 9, 1%, 0.15 M, and 0.05%, respectively. (E) Optimizing concentration
of the capture antibody at test lines by comparing SB/NSB ratios.
The SB and NSB in panel (E) were thermal signals from thermal contrast
amplification (TCA) reading of test lines, while those in (A–D)
were grayscale intensities, calculated as the area above the grayscale
curves across test lines plotted in ImageJ. NSB or SB was noise or
signal of a test line when an LFA was run with blank or antigen-loaded
buffers, respectively. The antigen concentrations were controlled
as the same for each set of studies in panels (A), (D), and (E), respectively.
LFA Redesign II: Larger GNSs Improve Visual
and Thermal Contrast
and Specific Binding
Larger GNSs (120 nm) were used as labels
for stronger visual and thermal contrast of a positive test line and
higher SB in the assay. The large GNS can improve visual and thermal
contrast by its large absorption cross section within the visible
range (380–700 nm). The visual reading of a positive test line
depends on the scattering contrast of visible light between the test
line and background regions of the membrane, while the thermal contrast
relies on the absorption contrast at 532 nm (i.e.,
laser’s wavelength in TCA reader). The bare membrane scatters
strongly but absorbs weakly within the visible range. When loaded
with visible-light absorbers (e.g. GNSs), however,
strong scattering reduction and absorption enhancement for the membrane
were observed, which induced visible and thermal contrast, respectively.[47] By Mie theory calculation,[48]Figure A shows that larger GNSs (120 nm) had a much stronger absorption
of visible light than smaller ones (30 nm), indicating larger visual
and thermal contrast for the controlled GNS amount. Their TEM images
are shown in Figure B. In addition, 120 nm GNSs can achieve more SB at test lines than
smaller GNSs due to increased antibody loading capacity, as seen with
the 100 nm GNSs used in our previous study.[41] The reaction rate constant for each GNS in SB is proportional to
the available binding sites on its surface.[49] Compared with smaller GNSs, larger GNSs with larger surface areas
can load more antibodies and therefore have more antigen-binding sites,
thus producing more SB in the assay. Based on this consideration,
120 nm GNSs were used throughout the design of the advanced SARS-CoV-2
antigen LFAs below, instead of the smaller ones (20–40 nm)
more prevalent in traditional LFAs.
Figure 4
(A) Absorption cross-sectional curves
of 120 and 30 nm GNSs in
the visible range (380–700 nm). (B) TEM images of 120 and 30
nm GNSs.
(A) Absorption cross-sectional curves
of 120 and 30 nm GNSs in
the visible range (380–700 nm). (B) TEM images of 120 and 30
nm GNSs.
Physical adsorption (Figure A) has been widely
used to conjugate GNSs with detection antibodies due to its simplicity
and ease of scaling up. For even better LFA performance, we explored
the covalent conjugation method (Figure A), hypothesizing that this could lead to
more active antigen-binding sites from coated antibodies and therefore
a higher more SB.
Figure 5
Comparing the covalent vs physical adsorption
conjugation methods. (A) Schematic of physical adsorption and covalent
conjugation methods. (B) Schematic of characterizing active antigen-binding
sites from GNS conjugates prepared by covalent and physical adsorption
conjugation methods. The GNS–antibody conjugates first reacted
with R-phycoerythrin (R-PE)–antigen (i.e.,
mouse IgG) conjugates to label antigen-binding sites with fluorophores
(i.e., R-PE). The R-PE was then released from the
GNS–antibody–antigen–RPE complex by dithiothreitol
(DTT) cleavage of the disulfide bond between R-PE and antigen, which
was provided by a cross-linker, succinimidyl 3-(2-pyridyldithio)propionate
(SPDP). The released R-PEs were then measured by a fluorimeter and
indicated the binding site amount from GNS conjugates. (C) Normalized
fluorescent intensities of released R-PEs post DTT cleavage on the
GNS–antibody–antigen–RPE complexes, and normalized
R-PE–antigen consumption rates (= bond R-PE-antigen/total
addition) in the reaction between GNS conjugates and R-PE–antigen
conjugates. The intensities were normalized by GNS amounts from the
two conjugation types (∼15% difference). (D) Thermal signals
of the test lines of mouse IgG LFAs using physical adsorption conjugation.
(E) Thermal signals of the test lines of mouse IgG LFAs using covalent
conjugation. The data shown in panels C–E were all based on
the model antibody and antigen, i.e., goat antimouse
IgG antibody (No. M8642, Sigma-Aldrich) and mouse IgG (No. I5381 Sigma-Aldrich).
Except for the difference in conjugation methods, other parameters
were controlled as the same, including the detection antibody and
amount (5 μg per mL of stock GNS solution), LFA strips, running
buffer, and TCA reading parameters. The statistical significance is
indicated with asterisks: ns: p > 0.05; *p < 0.05; and **p < 0.01.
Comparing the covalent vs physical adsorption
conjugation methods. (A) Schematic of physical adsorption and covalent
conjugation methods. (B) Schematic of characterizing active antigen-binding
sites from GNS conjugates prepared by covalent and physical adsorption
conjugation methods. The GNS–antibody conjugates first reacted
with R-phycoerythrin (R-PE)–antigen (i.e.,
mouse IgG) conjugates to label antigen-binding sites with fluorophores
(i.e., R-PE). The R-PE was then released from the
GNS–antibody–antigen–RPE complex by dithiothreitol
(DTT) cleavage of the disulfide bond between R-PE and antigen, which
was provided by a cross-linker, succinimidyl 3-(2-pyridyldithio)propionate
(SPDP). The released R-PEs were then measured by a fluorimeter and
indicated the binding site amount from GNS conjugates. (C) Normalized
fluorescent intensities of released R-PEs post DTT cleavage on the
GNS–antibody–antigen–RPE complexes, and normalized
R-PE–antigen consumption rates (= bond R-PE-antigen/total
addition) in the reaction between GNS conjugates and R-PE–antigen
conjugates. The intensities were normalized by GNS amounts from the
two conjugation types (∼15% difference). (D) Thermal signals
of the test lines of mouse IgG LFAs using physical adsorption conjugation.
(E) Thermal signals of the test lines of mouse IgG LFAs using covalent
conjugation. The data shown in panels C–E were all based on
the model antibody and antigen, i.e., goat antimouse
IgG antibody (No. M8642, Sigma-Aldrich) and mouse IgG (No. I5381 Sigma-Aldrich).
Except for the difference in conjugation methods, other parameters
were controlled as the same, including the detection antibody and
amount (5 μg per mL of stock GNS solution), LFA strips, running
buffer, and TCA reading parameters. The statistical significance is
indicated with asterisks: ns: p > 0.05; *p < 0.05; and **p < 0.01.To test this hypothesis, we conducted binding site characterization
by fluorescence measurement (Figure B), showing more active antigen-binding sites on covalent
GNS conjugates than physical adsorption ones (Figure C). To achieve this, a model antigen, mouse
IgG, was labeled by a fluorophore (R-phycoerythrin, R-PE) through
a cross-linker, which provided a cleavable disulfide bond. After antigen–antibody
binding between GNS–antibody conjugates and R-PE–antigen
conjugates, the fluorophores were released from the GNS complexes
and measured as a representation of antigen-binding sites by the two
methods (Figure B).
The excess free R-PE–antigen conjugates were separated from
GNSs and measured. As compared in Figure C, the fluorescent intensity (normalized
by GNS amount) standing for antigen-binding sites from covalent conjugation
was about 46% higher than that from physical adsorption conjugation.
A similar difference was seen in the R-PE–antigen consumption
rate (= bond R-PE-antigen/total addition) for the antigen–antibody
reaction in Figure C. This increase in antigen-binding sites might be due to better
coating efficiency and/or oriented antibodies in covalent conjugation
compared with physical adsorption.To further verify the hypothesis
in an LFA format, model mouse
IgG LFAs were developed; these showed better analytical performance
by covalent conjugation than physical adsorption. As shown in Figure D,E, the physical
adsorption had a higher background thermal noise (∼0.8 °C)
than the covalent one (∼0.6 °C). With TCA reading, the
LoD for mouse IgG in buffer by covalent conjugation (12.5 pg/mL, 0.08
pM) was about eightfold lower than that by physical adsorption (100
pg/mL, 0.7 pM). The LFA samples are shown in Figure S2A.Above all, compared to physical adsorption conjugation,
covalent
conjugation had more active antigen-binding sites on the GNS surface
and improved analytical sensitivity (i.e., lower
LoD) of LFAs. Therefore, in developing SARS-CoV-2 antigen LFAs, covalent
conjugation was used, aiming for better sensitivity.
New SARS-CoV-2
TCA LFA: fM–aM Detection Sensitivity of
Spike RBD
Based on the abovementioned optimization, we developed
the new SARS-CoV-2 spike RBD protein LFA, which demonstrated fM–aM
detection sensitivity for the RBD antigen in buffer and human nasopharyngeal
wash fluids. When testing serial dilutions of RBD protein in running
buffer (in Figure A), the visual LoD was 2–3 fg/mL (3.6 aM). With
TCA reading, subvisual positives were detected, which further lowered
the LoD by eightfold down to 2–6 fg/mL (0.45 aM).
To assess the impact of working in clinical samples, the RBD protein
was spiked into the human nasopharyngeal wash, which was collected
during the 2017–2018 influenza season without SARS-CoV-2, and
then compared with buffer solutions. Here, the LoDs increased by eightfold
for both visual and TCA readout, which was 1 fg/mL (28.6 aM) and 2–3 fg/mL (3.6 aM), respectively. The LFA samples are
shown in Figure S2B. This reduced sensitivity
in nasopharyngeal wash compared to clean buffer was probably due to
the abundant proteins present in the wash, which might have interacted
with and covered certain binding sites from detection and capture
antibodies in the assay; the SB could thus have been lowered to a
certain extent by the nasopharyngeal wash. Fortunately, there were
limited cross-reactions between the antibodies with influenza and
other nontargeted proteins from the nasopharyngeal wash because the
background noise from testing the unmodified nasopharyngeal wash sample
was on a similar level to that from clean buffer, as shown in Figure B. This also indicates
good specificity of the chosen antibody pair. In addition, the semiquantitative
thermal signals from TCA in Figure A,B, which were proportional to GNS concentrations
at test lines, could be used to indicate the concentrations of RBD
analyte in samples.
Figure 6
Analytical performance of the SARS-CoV-2 spike receptor-binding
domain (RBD) protein LFA with TCA. (A) Thermal signals of test lines
to test RBD proteins in running buffer. (B) Thermal signals of test
lines to test RBD proteins in the pooled human nasopharyngeal wash.
The covalent conjugation was used to prepare those LFAs. (C) Comparing
detection sensitivity and assay time of the advanced LFA from this
work with other literature work summarized in the previous perspective.[21] Panel C was reproduced from ref (21). Copyright 2021 American
Chemical Society. The statistical significance is indicated with asterisks:
ns: p > 0.05; *p < 0.05; **p < 0.01; and ***p < 0.001.
Analytical performance of the SARS-CoV-2 spike receptor-binding
domain (RBD) protein LFA with TCA. (A) Thermal signals of test lines
to test RBD proteins in running buffer. (B) Thermal signals of test
lines to test RBD proteins in the pooled human nasopharyngeal wash.
The covalent conjugation was used to prepare those LFAs. (C) Comparing
detection sensitivity and assay time of the advanced LFA from this
work with other literature work summarized in the previous perspective.[21] Panel C was reproduced from ref (21). Copyright 2021 American
Chemical Society. The statistical significance is indicated with asterisks:
ns: p > 0.05; *p < 0.05; **p < 0.01; and ***p < 0.001.In terms of SARS-CoV-2 diagnosis, our advanced
SARS-CoV-2 spike
RBD protein LFA showed much lower LoDs both visually and thermally
over literature work, as shown in Table . This increased sensitivity by visual reading
was mainly due to the high affinity of the antibodies (high SB), optimal
running buffer (high SB and low NSB), improved antibody–GNS
conjugation methods leading to more antigen-binding sites (high SB),
and the use of 120 nm GNSs (strong visual and thermal contrast, high
SB). To gain a broader perspective of diagnosis, its performance was
compared with other diagnostic platforms summarized in our previous
perspective,[21] as shown in Figure C. Although its assay time
was about 30 min, the advanced LFA in this study showed significant
improvement in sensitivity over conventional LFAs and also broke through
the detection limits of most previously published signal-amplified
LFAs (≥fM). A deeper understanding of this high detection sensitivity
can be achieved through kinetic analysis of the reaction and flow
in assay development, which is detailed in Section S5 in the Supporting Information. Briefly, the analysis shows
it was the SB of this advanced LFA that was significantly enhanced
through those assay optimization steps described above, especially
when testing low-concentrated analytes.
Conclusions
To
achieve ultrasensitive diagnostics, an advanced LFA based on
GNSs with a TCA diagnostic platform was developed and reported for
rapid and highly sensitive testing of the SARS-CoV-2 antigen. In the
assay design, comprehensive assay optimization for enhanced SB coupled
with signal amplification (TCA) on GNSs was carried out, including
high-affinity antibody pair, optimal buffer condition, use of large
(120 nm) GNSs, and covalent conjugation. The improved covalent conjugation
increased active antigen-binding sites by ∼46% compared with
the traditional physical adsorption. The advanced LFA with only visual
reading could detect 3.6 and 28.6 aM SARS-CoV-2 spike RBD protein
in buffer and human nasopharyngeal wash, respectively. This advanced
LFA showed lower visual and thermal LoDs over other literature work
in diagnosing SARS-CoV-2. As for the broad view of diagnostics, its
detection sensitivity was also much better than conventional LFAs
and even comparable to PCR. A fast (<1 min) TCA reading was also
developed, which could distinguish 26–32% visual false negatives
in clinical commercial LFAs. When applying this TCA to the advanced
LFA, another eightfold improvement in sensitivities was achieved,
reaching 0.45 aM (in buffer) and 3.6 aM (in the human nasopharyngeal
wash) with a semiquantitative readout. Future work is required to
further validate both the sensitivity and specificity of the advanced
LFA in a cohort or clinical study before commercialization and large-scale
production.
Materials and Methods
Materials
Chloroauric
acid (No. G4022), hydroquinone
(No. H9003), poly(ethylene glycol) 2-mercaptoethyl ether acetic acid
(SH-PEG2100-COOH, average Mn 2100, No. 757829), bovine
serum albumin (BSA, No. A7906), ovalbumin (No. A5503), N-hydroxysuccinimide (NHS), potassium phosphate dibasic (No. 795496),
potassium phosphate monobasic (No. 795488), Tween 20 (No. P1379),
Triton X-100 (laboratory grade), trizma hydrochloride (No. T3252),
trizma base (No. T1503), MES monohydrate (No. 69892), goat antimouse
IgG antibody (No. M8642), goat antihuman IgG antibody (No. I1011),
and mouse IgG (No. I5381) were purchased from Sigma-Aldrich. Sodium
citrate (ACS grade) and sucrose (ACS grade) were produced by Macron
Fine Chemicals. N-(3-Dimethylaminopropyl)-N′-ethylcarbodiimide hydrochloride (EDC, No. PG82079),
pyridyl disulfide-derivative R-phycoerythrin (R-PE, No. P806, 1.0
average pyridyl disulfide residues per molecule), succinimidyl 3-(2-pyridyldithio)propionate
(SPDP, No. 21857), and 10X phosphate-buffered saline (PBS) were purchased
from Thermo Fisher Scientific. Dithiothreitol (DTT, No. BP172–5)
was purchased from Fisher Scientific. A customized 5X PBS-EDTA buffer
(No. BM-682, 1X buffer: 10.1 mM Na2HPO4, 1.8 mM KH2PO4, 137 mM NaCl,
2.7 mM KCl, 10 mM EDTA disodium dihydrate) was produced by Boston
Bioproducts. N-Hydroxysulfosuccinimide sodium salt
(sulfo-NHS, No. 12831) was purchased from Chem Impex. The chimeric
antibodies to SARS-CoV-2 spike RBD protein (Nos. RBD5324 and RBD5308)
were purchased from Hytest (Finland). The recombinant SARS-CoV-2 spike
RBD protein (35 kDa) was purchased from Dr. Fang Li’s lab at
the University of Minnesota. The polyclonal antimouse IgG antibody
and mouse IgG protein were gifted from James Sackrison (from Scantibodies,
Inc.). The glass fiber conjugate pad (No. GFCP203000) and nitrocellulose
membrane (No. HF 13502XSS) were purchased from EMD Millipore. The
wicking pad (No. CF5) was purchased from Cytiva.
TCA: Developing
the Fast-Reading Algorithm
To increase
the throughput of the TCA reading, we developed a continuous reading
algorithm, which allowed it to finish one reading as fast as <1
min. During continuous reading, the laser scanned continuously across
a test line (see Figure D). The laser and the IR sensor were turned on and kept still while
the LFA strip being read was moved at a constant speed controlled
by a linear motor. After this reading, a continuous temperature curve
was obtained and the area under this curve (AUC) was calculated as
its thermal signal. According to the IUPAC method, the detection limit
was determined as the lowest analyte concentration in a serial dilution
study, whose thermal signal was larger than the summation of mean
and 3 times the standard deviation of the thermal signals from blank
samples (or negative controls). Meanwhile, the ANOVA analysis of signals
from blank and low-analyte samples should show a p-value <0.05.
Clinical Commercial SARS-CoV-2 LFAs for TCA
Reading
Clinical samples of nasopharyngeal swabs in viral
transport media
were collected by healthcare workers during mass testing of unexposed
asymptomatic residents in northeast Spain (Metropolità Nord).[50] RT-qPCR tests were performed on these fresh
samples stored at 2–8 °C within 24 h, followed by commercial
rapid SARS-CoV-2 antigen LFAs from five commercial brands within the
next 12 h.[50] The visual readout (±)
of rapid tests was compared to PCR results, which were viewed as true
results.[50] More details about the clinical
sample collection and testing procedures are provided in the recent
publication.[50]To understand how
TCA helps improve the sensitivity of commercial SARS-CoV-2 LFAs, some
of the commercial LFAs were left to dry after testing and were later
transported from Spain to the University of Minnesota (UMN) for TCA
reading, as shown in Figure A. They included 54 LFAs from the PanBio SARS-CoV-2 Ag Rapid
test by Abbott and 53 from the CLINITEST Rapid SARS-CoV-2 Antigen
Test by Siemens. These LFAs were selected to consist of mostly visual
false negatives, a small portion of true positives and true negatives,
and none or a few false positives. To dry these LFAs, the sample,
conjugation, and wicking pads were removed to stop any further flow
and reaction of reagents for most LFAs, except for a few which had
very clear backgrounds after assay completion. The dried LFAs were
kept and shipped in a bag with desiccants for further TCA reading
with a continuous reading algorithm. An LFA was determined as thermally
positive or negative via TCA reading by comparing its thermal signal
to a cutoff threshold, which was set as the summation of the mean
and 3 times the standard deviation of thermal signals from four PCR-negative
samples (true negatives). The correctness (true or false) of thermal
and visual results (±) was determined by comparison to PCR results
(±) obtained during the clinical study in Spain.
LFA Development
I: Screening the SARS-CoV-2 Antibody Pair
The antibody pairs
from Hytest and MyBioSource were tested in a
lateral flow format and their SB signals were quantified. The capture
antibodies and secondary antibodies to detection antibodies were pipetted
onto the nitrocellulose membrane as test and control dots at a 1 mg/mL
concentration. The membrane was dried in a vacuum overnight prior
to assembly and cutting. Corresponding detection antibodies were conjugated
with GNSs by the physical adsorption at preoptimized pH with a controlled
concentration. The pH of the running buffer was optimized to 9 for
both antibody pairs to minimize false positives when testing the blank
buffer. To quantify the SB signal, the same positive controls (i.e., the same concentration of SARS-CoV-2 spike RBD protein
in buffer) were tested by LFAs using the two antibody pairs. The positive
signals were analyzed by plotting grayscale curves across test lines
using ImageJ software and the areas above the grayscale curve were
calculated for quantitative SB comparison.
LFA Development II: Synthesizing
and Characterizing GNSs
The 120 nm GNSs were synthesized
by the seed-mediated growth method.[51] Briefly,
15 nm seeds were synthesized according
to the method described by Frens et al.[52] First, 1 mL fresh 3.3% (w/v) sodium citrate was added to 100 mL
of a boiling 0.25 mM HAuCl4 solution under vigorous stirring.
Boiling and stirring were continued for another 10–15 min.
The seed solution was cooled to room temperature and diluted to 100
mL with milli-Q ultrapure water for further use or stock. The 120
nm GNSs were then synthesized by adding 1 mL of 0.25 mM HAuCl4, 1 mL of 15 mM sodium citrate, 0.3 mL of seed solution, and
1 mL of 0.25 mM hydroquinone into 97.66 mL of water under vigorous
stirring in quick succession. Stirring was kept for at least 2 h at
room temperature to complete the growth of GNSs. The GNSs were stabilized
by adding Tween 20 at a final concentration of 0.05% (v/v) before
storage at 2–8 °C. After synthesis, GNSs were characterized
by an ultraviolet–visible–near-infrared (UV–vis–NIR)
spectrophotometer (Cary 5000 UV–vis–NIR) and a transmission
electron microscope (TEM, Tecnai G2). The concentration of GNSs was
determined by Beer’s law, where the molar extinction coefficient
(ε, M–1·cm–1) of GNSs
was estimated as[51]
LFA Development
III: Conjugating GNSs with Detection Antibodies
Physical Adsorption Conjugation
The stock GNSs were
centrifuged once and resuspended in ultrapure water, whose pH was
adjusted to near the isoelectric point of detection antibodies by
adding 0.2 M K2CO3. For model mouse IgG assay,
a goat antimouse IgG antibody (M8642, Sigma-Aldrich) was added into
GNS solutions at 5 μg per mL GNS at a stock concentration, immediately
followed by vortexing and incubation on a rotator at room temperature
for 1.5 h. To block the free GNS surface, BSA was added to a final
concentration of 1% (w/v), followed by rotating incubation for 0.5
h at room temperature. The GNS–antibody conjugates were centrifuged
and washed once by 5 mM PB buffer. Finally, the conjugates were recovered
in resuspension buffer (10 mM PB, pH 7.4, 3% (w/v) sucrose, 0.5% (w/v)
BSA, 0.05% (v/v) Proclin 300) and stored at 4 °C until further
use.
Covalent Conjugation
The stock GNSs were centrifuged
once and resuspended in ultrapure water at a fivefold concentration
of its stock solution. SH-PEG2100-COOH was added into the GNS solution
until a final concentration of 0.75% (w/v). The mixture was stirred
vigorously overnight at room temperature to complete the reaction.
The PEGylated GNSs were washed three times and resuspended in 10 mM
MES buffer for covalent conjugation with antibodies through EDC/NHS
chemistry. In 600 μL of a GNS solution, 6 μL of a fresh
EDC solution (10 mg/mL) and 12 μL of a fresh sulfo-NHS solution
(10 mg/mL) were added and vortexed vigorously, followed by rotation
for 30 min at room temperature. The EDC-activated GNSs were washed
once to remove excess EDC and sulfo-NHS, resuspended in 5 mM PB buffer
(pH 7.2), and incubated with antibodies for 1.5 h at room temperature
under rotating. BSA was then added to a final concentration of 1%
(w/v) to block the GNS surface, followed by rotation for 0.5 h at
room temperature. To ensure sufficient blocking of excess active carboxyl
groups on the GNS surface, 12 μL of 12.5% (wt) hydroxylamine
was added and incubated for 10 min at room temperature. Finally, the
conjugates were washed twice, recovered in resuspension buffer (same
as physical adsorption), and stored at 4 °C until further use.
Overall, 5 and 10 μg per mL stock GNS of goat antimouse IgG
antibody (M8642, Sigma-Aldrich) and antispike RBD antibody (RBD5324,
Hytest), respectively, were used in covalent conjugation. The antispike
RBD antibody was predialyzed to remove sodium azide before the conjugation
steps.For both conjugation types, the characterization of antigen-binding
sites is described in Section S4 in the
Supporting Information.
LFA Development IV: Fabricating
and Running LFAs
The
method to fabricate and perform LFA generally followed our previous
studies.[41,42] Briefly, capture antibodies were precoated
onto the nitrocellulose membrane by a dispenser (ClaremontBio, Automated
Lateral Flow Reagent Dispenser). For the model mouse IgG LFA, an antimouse
IgG antibody and mouse IgG (both from Scantibodies, Inc.) were dispensed
at 1 and 0.2 mg/mL as test and control lines, respectively. For the
SARS-CoV-2 spike LFA, an antispike antibody (No. RBD5308, Hytest)
and a goat antihuman IgG antibody (I1011, Sigma-Aldrich) were dispensed
at 2 and 0.2–0.4 mg/mL as test and control lines, respectively.
Post precoating, the membrane was dried overnight in vacuum at room
temperature to immobilize proteins. The conjugate pad, membrane, and
absorbent pad were assembled onto an adhesive backing card, with 1–2
mm overlapping between adjacent components to facilitate solution
migration. The assembly was laminated and cut into strips 3 mm wide.
To run the assay, the LFA strips were dipped into a 96-well plate
filled with 140 μL of the sample or buffer solutions. The visual
results were recorded after 20 min (mouse IgG LFA) or 30 min (SARS-CoV-2
spike RBD LFA) until completed migration of most GNSs before TCA reading.
The number of replication of LFAs for the same condition was three.
SARS-CoV-2 TCA LFA: Testing RBD in the Human Nasopharyngeal
Wash
To acquire LFAs’ analytical sensitivity for an
antigen in human samples, the recombinant SARS-CoV-2 spike RBD protein
was spiked into the human nasopharyngeal wash. This wash was collected
in the 2017–2018 flu season, pooled from multiple deidentified
patients, and stored at −80 °C prior to use.[43] Although some of those patients were diagnosed
with influenza A and B infections, the nasopharyngeal wash was void
of SARS-CoV-2 infections and thus appropriate for the dilution study
of the SARS-CoV-2 spike RBD protein. The recombinant RBD protein was
serially diluted by the stock nasopharyngeal wash at different concentrations.
Per 1 mL of the RBD-diluted nasopharyngeal wash, 100 μL of running
reagents (0.59M Trizma base, 0.39 M Trizma HCl, 5.4% BSA, 8.7% Triton
X-100, 1.1% proclin 300) was added and mixed by shaking before each
assay test. To run LFAs, the strips were dipped in a 96-well plate
filled with 140 μL of the nasopharyngeal wash mixed with running
reagents. The assay took about 30 min before visual reading followed
by a TCA scan. The number of replication of LFAs for each condition
was three.
Authors: Bàrbara Baro; Pau Rodo; Dan Ouchi; Antoni E Bordoy; Emilio N Saya Amaro; Sergi V Salsench; Sònia Molinos; Andrea Alemany; Maria Ubals; Marc Corbacho-Monné; Pere Millat-Martinez; Michael Marks; Bonaventura Clotet; Nuria Prat; Oriol Estrada; Marc Vilar; Jordi Ara; Martí Vall-Mayans; Camila G-Beiras; Quique Bassat; Ignacio Blanco; Oriol Mitjà Journal: J Infect Date: 2021-04-18 Impact factor: 6.072
Authors: Emily R Adams; Mark Ainsworth; Rekha Anand; Monique I Andersson; Kathryn Auckland; J Kenneth Baillie; Eleanor Barnes; Sally Beer; John I Bell; Tamsin Berry; Sagida Bibi; Miles Carroll; Senthil K Chinnakannan; Elizabeth Clutterbuck; Richard J Cornall; Derrick W Crook; Thushan de Silva; Wanwisa Dejnirattisai; Kate E Dingle; Christina Dold; Alexis Espinosa; David W Eyre; Helen Farmer; Maria Fernandez Mendoza; Dominique Georgiou; Sarah J Hoosdally; Alastair Hunter; Katie Jefferey; Dominic F Kelly; Paul Klenerman; Julian Knight; Clarice Knowles; Andrew J Kwok; Ullrich Leuschner; Robert Levin; Chang Liu; César López-Camacho; Jose Martinez; Philippa C Matthews; Hannah McGivern; Alexander J Mentzer; Jonathan Milton; Juthathip Mongkolsapaya; Shona C Moore; Marta S Oliveira; Fiona Pereira; Elena Perez; Timothy Peto; Rutger J Ploeg; Andrew Pollard; Tessa Prince; David J Roberts; Justine K Rudkin; Veronica Sanchez; Gavin R Screaton; Malcolm G Semple; Jose Slon-Campos; Donal T Skelly; Elliot Nathan Smith; Alberto Sobrinodiaz; Julie Staves; David I Stuart; Piyada Supasa; Tomas Surik; Hannah Thraves; Pat Tsang; Lance Turtle; A Sarah Walker; Beibei Wang; Charlotte Washington; Nicholas Watkins; James Whitehouse Journal: Wellcome Open Res Date: 2020-06-11
Authors: Youcef Azeli; Alberto Fernández; Federico Capriles; Wojciech Rojewski; Vanesa Lopez-Madrid; David Sabaté-Lissner; Rosa Maria Serrano; Cristina Rey-Reñones; Marta Civit; Josefina Casellas; Abdelghani El Ouahabi-El Ouahabi; Maria Foglia-Fernández; Salvador Sarrá; Eduard Llobet Journal: Sci Rep Date: 2022-09-16 Impact factor: 4.996