Literature DB >> 34941918

Quantification of stroke volume in a simulated healthy volunteer model of traumatic haemorrhage; a comparison of two non-invasive monitoring devices using error grid analysis alongside traditional measures of agreement.

Sam D Hutchings1,2, Jim Watchorn1, Rory McDonald2, Su Jeffreys2, Mark Bates3, Sarah Watts3, Emrys Kirkman3.   

Abstract

INTRODUCTION: Haemorrhage is a leading cause of death following traumatic injury and the early detection of hypovolaemia is critical to effective management. However, accurate assessment of circulating blood volume is challenging when using traditional vital signs such as blood pressure. We conducted a study to compare the stroke volume (SV) recorded using two devices, trans-thoracic electrical bioimpedance (TEB) and supra-sternal Doppler (SSD), against a reference standard using trans- thoracic echocardiography (TTE).
METHODS: A lower body negative pressure (LBNP) model was used to simulate hypovolaemia and in half of the study sessions lower limb tourniquets were applied as these are common in military practice and can potentially affect some haemodynamic monitoring systems. In order to provide a clinically relevant comparison we constructed an error grid alongside more traditional measures of agreement.
RESULTS: 21 healthy volunteers aged 18-40 were enrolled and underwent 2 sessions of LBNP, with and without lower limb tourniquets. With respect to absolute SV values Bland Altman analysis showed significant bias in both non-tourniquet and tourniquet strands for TEB (-42.5 / -49.6 ml), rendering further analysis impossible. For SSD bias was minimal but percentage error was unacceptably high (35% / 48%). Degree of agreement for dynamic change in SV, assessed using 4 quadrant plots showed a seemingly acceptable concordance rate for both TEB (86% / 93%) and SSD (90% / 91%). However, when results were plotted on an error grid, constructed based on expert clinical opinion, a significant minority of measurement errors were identified that had potential to lead to moderate or severe patient harm.
CONCLUSION: Thoracic bioimpedance and suprasternal Doppler both demonstrated measurement errors that had the potential to lead to clinical harm and caution should be applied in interpreting the results in the detection of early hypovolaemia following traumatic injury.

Entities:  

Mesh:

Year:  2021        PMID: 34941918      PMCID: PMC8699736          DOI: 10.1371/journal.pone.0261546

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

Haemorrhage remains the leading cause of preventable death following traumatic injury [1]. Although the diagnosis of severe haemorrhage is often easy in retrospect, it can be challenging to make contemporaneously, especially as patients typically present to medical providers in austere settings far removed from the monitoring facilities available in the operating theatre or intensive care unit. Detection of haemorrhage induced hypovolaemia is usually based on assessment of so called traditional vital signs, particularly blood pressure. However, the use of arterial blood pressure to estimate the degree of blood loss is problematic and potentially inaccurate [2-4] and there is a poor association between recorded blood pressure and tissue perfusion parameters [5]. A monitoring system that could detect changes in flow and volume based parameters may allow better detection of early hypovolaemia in patients at risk of blood loss. Obtaining precise effect and response data as it relates to the physiology of haemorrhagic shock is complicated by the uncontrolled nature of the clinical environment, an alternative approach is to study healthy volunteers using lower body negative pressure (LBNP), a technique which produces central hypovolaemia, and therefore simulates the early stages of haemorrhage. Progressive application of LBNP is a well-established technique used to simulate haemorrhage as well as a range of other conditions including syncope associated with orthostasis [6, 7]. A subject is asked to lie supine with the lower portion of the body (from the waist down) sealed in a chamber. The pressure within the chamber is reduced to sub-atmospheric levels (application of LBNP), which results in blood pooling in the veins of the legs and pelvis. Since the capacity of the pelvic veins is much greater than those of the legs, it is the pelvic veins that make the greatest contribution to this blood pooling. As the LBNP is gradually augmented (made more negative with respect to atmospheric), progressively greater amounts of blood are trapped in the lower body and there is a corresponding decrease in venous return to the heart, simulating the effects of haemorrhage. Using this technique it is possible to elicit both the initial compensatory response to haemorrhage (tachycardia while blood pressure is maintained) and the later depressor phase (reflex bradycardia and hypotension) leading to pre-syncope and then syncope. This pattern closely mirrors the cardiovascular changes seen with actual haemorrhage [8-11]. We used an LBNP model in order to compare the performance of two non-invasive monitoring devices that utilised thoracic electrical bioimpedance and supra-sternal Doppler to measure stroke volume. Trans-thoracic echocardiography derived stroke volume was used as the standard or reference measurement. As the application of lower limb tourniquets to prevent catastrophic haemorrhage is not uncommon following major trauma, and may produce an effect on arterial waveforms and hence influence the output of some monitoring devices, we applied lower limb tourniquets during half of the study sessions. In order to assess the clinical relevance, as well as the mathematical agreement between the measurements produced we utilised an error grid analysis approach alongside more traditional measures of agreement.

Materials and methods

Ethical approval

The study received ethical approval from the Ministry of Defence Research Ethics Committee (538/MODREC/14). Subjects provided written informed consent prior to study enrolment.

Study cohort

21 healthy volunteers aged between 18 and 40. The sample size was based on 75% power to detect a 10% difference in SV between candidate devices, with an alpha of 0.05.

Recruitment screening and exclusion criteria

Volunteers filled out a medical questionnaire and were interviewed by a study investigator prior to enrolment. They also underwent 12-lead ECG and cardiac echocardiography examination. Volunteers were not eligible for enrolment if they had a history, or any symptoms or signs, of cardiovascular disease. Individuals judged to be at risk of thromboembolic complications (family history, recent lower limb or pelvic trauma or current usage of oral contraceptive medication) were excluded.

Lower body negative pressure protocol

Volunteers lay supine with the lower portion of the body, from the waist down, sealed in a chamber. The pressure within the chamber was reduced to sub-atmospheric levels using a vacuum generator. A continuous, real time, measure of reconstructed brachial artery blood pressure (rBAP) was made using a Finometer® PRO (Finapres Medical Systems BV, Netherlands) via a cuff placed on the right middle finger. The rBAP was continuously normalised to heart level (reference zero) using a hydrostatic height sensor affixed to the volunteer’s chest at the mid-axial line, and to the finger cuff. Cardiac stroke volume (SVFin) was derived continuously using the built-in Modelflow® technology, which has been shown to be reliable in reporting trend changes in cardiac output [12], and hence stroke volume. The cardiovascular data (rBAP and SVFin) were exported continuously from the Finometer® PRO via a digital to analogue converter and recorded on the computerized data acquisition system (MacLab 8/s). Heart rate was derived from the rBAP trace using the LabChart®7 PRO software. Each subject rested supine in the LBNP chamber for 20 min after instrumentation to allow a steady state to be reached. Thereafter two baseline readings of all cardiovascular variables were made at 5 min intervals. Average data from the final 3 minutes of the second baseline period was used to calculate each subject’s baseline SVFin for that session. Negative pressure was then applied within the chamber to trap blood in the lower body causing central hypovolaemia. By sequentially applying steps of negative pressure, a progressive haemorrhage was simulated. The initial two steps were to an absolute pressure level of 10 and 20 mmHg below atmospheric, and thereafter to attain a target SVFin of 80%, 65–70% and 55–64% of the baseline recorded in the subject. LBNP was then reduced in one or two steps to a final level of 0 mmHg (no suction). Each step lasted approximately 5 min. Stroke volume measurements from the three monitoring devices (trans-thoracic echocardiography, thoracic bioreactance and supra-sternal Doppler) were recorded approximately 2 minutes into each step. Sessions involving tourniquet application were conducted in an identical manner to those without tourniquet, except that bilateral tourniquets (Conical Leg Tourniquets, SCT2x2, Braun & Co Ltd, UK) were inflated around the upper thighs to a pressure of 100 mmHg above the subject’s systolic arterial pressure immediately after the second baseline measurement. Two minutes after tourniquet inflation all cardiovascular measurements were made before commencing the LBNP protocol described above. The tourniquets remained inflated without further adjustment of pressure until the end of the LBNP protocol.

Trans Thoracic Echocardiography (TTE)

Serial focussed trans thoracic echocardiography studies were performed (Sparq, S41 transducer, Philipps UK) by a single operator, accredited with the British Society of Echocardiography. Measurement of the left ventricular outflow tract (LVOT) diameter was obtained using a parasternal long axis window prior to the commencement of LBNP. Serial measurements of SVTTE were obtained by applying continuous wave Doppler to the LVOT in a 5 chamber view and calculating the velocity time integer (VTi) of the resultant signal.

Supra Sternal Doppler (SSD)

Serial measurements were made of the blood flow in the ascending aorta / aortic arch using supra sternal Doppler (Ultra Sonic Cardiac Output Monitor, USCOM 1A, USCOM, Australia). A single trained user conducted all measurements. The software automatically produces a value for SVSSD based on subject demographics (age, weight, height and gender) and a calculation of VTi. Supra-sternal Doppler measurements were taken immediately following TTE.

Thoracic electrical bio-impedance (TEB)

Serial stroke volume measures were taken using a system that utilises trans-thoracic bio-impedance (NICOM®, Cheetah Medical) This technique relies on the phase shift of electric current between four surface electrodes placed on the thorax. A continuous reading for SVTEB was produced and recorded synchronously with SVTTE measurements.

Data analysis and statistical methods

Distribution of data was assessed using D’Agostino & Pearson omnibus normality test. Parametric continuous data is presented as mean ± standard deviation (SD). Statistical analysis was performed using Graphpad Prism v. 6 and NCSS v. 11. p values of <0.05 was taken as statistically significant. Changes in cardiovascular parameters were analysed by linear mixed model ANOVA with repeated measures over time, using baseline values as a covariate. Data are presented as mean±SEM. Comparison between absolute values of reference stroke volume (SVTTE) and those recorded by supra-sternal Doppler (SVSSD) and thoracic electrical bioimpedance (SVTEB) was performed using Bland-Altman analysis [13] adapted for repeated measurements of a range of stroke volumes in the series of volunteers; the percentage error was calculated by dividing the limit of agreement (1.96 SD) by the mean value for SVTTE. Changes in stroke volume from baseline values recorded by trans-thoracic echocardiography (ΔSVTTE), supra-sternal Doppler (ΔSVSSD) and thoracic bioimpedance (ΔSVTEB) was compared using four quadrant plots with a 15% central exclusion zone applied [14, 15]. Correlation between ΔSVTTE and ΔSVTBI / ΔSVSSD was assessed using Pearson’s correlation coefficient. In order to assess the clinical significance of differences in SV measurements recorded by the subject devices we utilised an error grid analysis approach [16]. Firstly, we distributed a questionnaire to intensive care and anaesthesia specialists at our institution and in the United Kingdom Defence Medical Services (S1 File). We asked them to envisage a clinical scenario in which they were managing a previously fit patient at risk from hypovolaemia following traumatic injury and to consider that they were using a device to monitor stroke volume and that the therapeutic intervention at their disposal was administration of blood products to treat hypovolaemia. We firstly asked them to categorise the percentage fall in SV which they would regard as requiring i) no current action ii) action indicated iii) action essential. We then asked the clinicians to consider, in the light of these responses, the potential for harm resulting from a divergence in SV change between the actual value and the value recorded by a measurement device. We asked respondents to categorise this potential for harm as: None, Mild, Moderate or Severe. Based on these responses we constructed an error grid using the following methodology: A numerical weighting factor was applied to the potential risk of harm as follows: None 0, Mild 2, Moderate 5, Severe 10. This weighting factor is necessarily subjective. A table was created from the combined questionnaire returns which compared actual SV reduction against measured SV reduction (in the range 0 to -60%). Average risk weighted numerical scores, based on the questionnaire responses, were entered into each cell of this table. This produced a numeric range from 0 (indicating no respondent thought there was no potential for harm) to 150 (indicating that all respondents felt that there was a severe risk of harm). (Supplementary material–error grid returns). An MS Excel spreadsheet was created detailing all potential values for SV reduction recorded by device against actual SV reduction in a range from 0 to 60% and values were then converted into a percentage and a Red–Yellow–Green colour characteristic applied to all cells such that the risk of harm was illustrated graphically from Green (0) to Red (150). Supplementary material–error grid spreadsheet. An image of the colour coded spreadsheet was exported and polygons drawn over areas of similar colour in order to create 4 zones representing no risk (green), mild risk (yellow), moderate risk (orange) and severe risk (red). (Supplementary material—error grid polygons). The image was imported into a graphing program (Desmos, https://www.desmos.com). Values for TTE derived SV change were plotted on the x axis and values derived from the assessed measurement devices were plotted on the y axis. The number of recorded values falling into each area of risk was recorded for each measurement device.

Results

Participant demographics

Healthy volunteers were predominantly male 13/21 (62%), aged 27±5 years and had a body mass index of 23±3 kg/m2.

Cardiovascular response to LBNP

The overall cardiovascular response to LBNP are shown in Fig 1, and the corresponding statistical output is shown in Table 1. There was a significant reduction in stroke volume, rise in heart rate and fall in pulse pressure as LBNP was progressively increased. The overall pattern of response was similar in both the tourniquet and non- tourniquet strands with the exception of blood pressure, which fell slightly in the non-tourniquet strand and rose in the tourniquet strand.
Fig 1

Cardiovascular response to progressive LBNP.

Effects of progressive LBNP (-10, -20 mmHg, then titrated to target SVFPleth, stroke volume measured by finger Plethysmography, of 80, 65–70 and 55–64% baseline SVFPleth, LBNP1-5 respectively) on mean arterial blood pressure (MBP), heart rate (HR), arterial pulse pressure (PP), stroke volume measured by trans-thoracic echo (SVTTE). Data presented as mean±SEM. p values indicated on each panel show probability of differences between groups, and differences in the pattern of changes, being due to chance (ANOVA). Changes over time were statistically significant (p <0.001) for each variable.

Table 1

Summary repeated measures ANOVA output for the cardiovascular parameters and LBNP levels shown in Fig 1.

ParameterEffect typeTermF Df(Num, Den)F-ValueP-Value
SVFPleth (% baseline)MainLBNP sequence6, 205265.27<0.001
MainStrand1, 2155.030.026
InteractionLBNP*Strand6, 2051.810.098
SV TTE MainLBNP sequence6, 195122.96<0.001
MainStrand1, 2124.720.031
InteractionLBNP*Strand6, 1950.940.465
HR MainLBNP sequence6, 20455.13<0.001
MainStrand1, 20815.43<0.001
InteractionLBNP*Strand6, 2041.180.316
PP MainLBNP sequence6, 20047.09<0.001
MainStrand1, 2179.710.002
InteractionLBNP*Strand6, 2001.470.191
MAP MainLBNP sequence6, 2035.32<0.001
MainStrand1, 2107.320.007
InteractionLBNP*Strand6, 2022.090.056
LBNP MainLBNP sequence6, 204284.32<0.001
MainStrand1, 21099.88<0.001
InteractionLBNP*Strand6, 20421.46<0.001

Abbreviations for the parameters are as listed in Fig 1. The main effects of LBNP sequence (Time) and Strand (Tourniquet / No tourniquet), and the interaction of the two main effects (LBNP*Time) are shown. F Df(Num, Den), F degrees of freedom (Numerator, Denominator). SVFPleth stroke volume finger plehysmography, SVTTE stroke volume echocardiography, HR heart rate, PP pulse pressure, MAP mean arterial pressure, LBNP lower body negative pressure.

Cardiovascular response to progressive LBNP.

Effects of progressive LBNP (-10, -20 mmHg, then titrated to target SVFPleth, stroke volume measured by finger Plethysmography, of 80, 65–70 and 55–64% baseline SVFPleth, LBNP1-5 respectively) on mean arterial blood pressure (MBP), heart rate (HR), arterial pulse pressure (PP), stroke volume measured by trans-thoracic echo (SVTTE). Data presented as mean±SEM. p values indicated on each panel show probability of differences between groups, and differences in the pattern of changes, being due to chance (ANOVA). Changes over time were statistically significant (p <0.001) for each variable. Abbreviations for the parameters are as listed in Fig 1. The main effects of LBNP sequence (Time) and Strand (Tourniquet / No tourniquet), and the interaction of the two main effects (LBNP*Time) are shown. F Df(Num, Den), F degrees of freedom (Numerator, Denominator). SVFPleth stroke volume finger plehysmography, SVTTE stroke volume echocardiography, HR heart rate, PP pulse pressure, MAP mean arterial pressure, LBNP lower body negative pressure. SVTTE generally correlated well with SVFIN (Pearson’s R2 > 0.7 in 100% of subjects in the non-tourniquet stream and 88% of subjects in the tourniquet stream).

Comparison of absolute values of SVTTE and SVTEB / SVSSD

Bland Altman plots comparing SVTTE and SVTEB are shown in Fig 2. 144 paired measurements were made in the non-tourniquet strand and 138 in the tourniquet strand. Due to the fact that bias obviously increased with increasing mean SV, limits of agreement are not shown for this comparison; percentage error between SVTTE and SVTEB was not calculated for the same reason.
Fig 2

Bland-Altman plot comparing stroke volume (SV) values recorded using thoracic bioimpedence (TBI) and trans-thoracic echocardiography (TTE).

Values indicate bias±SD and 95% limits of agreement.

Bland-Altman plot comparing stroke volume (SV) values recorded using thoracic bioimpedence (TBI) and trans-thoracic echocardiography (TTE).

Values indicate bias±SD and 95% limits of agreement. Bland Altman plots comparing SVTTE and SVSSD are shown in Fig 3. 152 paired measurements were made in the non-tourniquet strand with a bias of 0.1ml and limits of agreement between 36.7 ml and -36.5ml. 139 paired measurements were made in the tourniquet strand with a bias of -4.6ml and limits of agreement between 22.4 ml and -31.6 ml. Percentage error between SVTTE and SVSSD was 35% in the non-tourniquet strand and 48% in the tourniquet strand.
Fig 3

Bland-Altman plot comparing stroke volume (SV) values recorded using supra-sternal Doppler (SSD) and trans-thoracic echocardiography (TTE).

Values indicate bias±SD and 95% limits of agreement.

Bland-Altman plot comparing stroke volume (SV) values recorded using supra-sternal Doppler (SSD) and trans-thoracic echocardiography (TTE).

Values indicate bias±SD and 95% limits of agreement.

Comparison of change from baseline values for SVTTE and SVTEB / SVSSD

Four quadrant plots comparing changes in SV from baseline for TTE (ΔSVTTE) and SSD (ΔSVSSD) are shown in Fig 4. In the non-tourniquet strand 119 paired measurements were made, ΔSVTTE and ΔSVSSD were positively correlated (r = 0.65, p<0.0001) with a 91% concordance rate. In the tourniquet strand 113 paired measurements were made, ΔSVTTE and ΔSVSSD were positively correlated (r = 0.75, p<0.0001) with a concordance rate of 90%.
Fig 4

4 quadrant plot showing changes in stroke volume from baseline (ΔSV) recorded using trans thoracic bioimpedence (TBI) and trans-thoracic echocardiography (TTE).

15% exclusion zone shown as dotted central square. Line of identity (y = x) shown as dotted line and regression line shown as solid line. r value indicates degree of correlation (Pearson’s co-efficient).

4 quadrant plot showing changes in stroke volume from baseline (ΔSV) recorded using trans thoracic bioimpedence (TBI) and trans-thoracic echocardiography (TTE).

15% exclusion zone shown as dotted central square. Line of identity (y = x) shown as dotted line and regression line shown as solid line. r value indicates degree of correlation (Pearson’s co-efficient). Four quadrant plots comparing changes in SV from baseline for TTE (ΔSVTTE) and TEB (ΔSVTEB) are shown in Fig 5. In the non-tourniquet strand 121 paired measurements were made, ΔSVTTE and ΔSVTEB were positively correlated (r = 0.41, p<0.0001) with a 86% concordance rate. In the tourniquet strand 112 paired measurements were made, ΔSVTTE and ΔSVTEB were positively correlated (r = 0.78, p<0.0001) with a 93% concordance rate.
Fig 5

4 quadrant plot showing changes in stroke volume from baseline (ΔSV) recorded using supra-sternal Doppler (SSD) and trans-thoracic echocardiography (TTE).

15% exclusion zone shown as dotted central square. Line of identity (y = x) shown as dotted line and regression line shown as solid line. r value indicates degree of correlation (Pearson’s co-efficient).

4 quadrant plot showing changes in stroke volume from baseline (ΔSV) recorded using supra-sternal Doppler (SSD) and trans-thoracic echocardiography (TTE).

15% exclusion zone shown as dotted central square. Line of identity (y = x) shown as dotted line and regression line shown as solid line. r value indicates degree of correlation (Pearson’s co-efficient).

Error grid analysis

Responses from 15 specialists were used to construct an error grid. Responses are provided in the (S2 File). The resultant spreadsheet used to create the error grid and the polygon overlay template created from the spreadsheet are provided in the (S3 File and S1 Fig). Error grids comparing ΔSVTTE and ΔSVTEB are shown in Fig 6. In the non-tourniquet strand 121 matched values were plotted. The risk of harm as a result of a measurement error was classed as: none 75 (62%), mild 18 (15%), moderate 22 (18%) and severe 6 (5%). In the tourniquet strand 113 matched values were plotted. The risk of harm as a result of a measurement error was classed as: none 80 (71%), mild 17 (15%), moderate 16 (14%). No values indicated a risk of severe harm in this strand. The overall risk of severe harm when measurements for both streams were combined was 2.6%. 1 subject in the non-tourniquet strand had 3 measurement values indicative of severe harm, 1 subject had 2 such values and a third subject had a single value.
Fig 6

Error grid showing changes in stroke volume from baseline (ΔSV) recorded using trans thoracic electrical bioimpedence (TEB) and trans-thoracic echocardiography (TTE).

Coloured zones represent the perceived degree of clinical harm resulting from a measurement error: Green = none, Yellow = mild, Orange = moderate, Red = severe.

Error grid showing changes in stroke volume from baseline (ΔSV) recorded using trans thoracic electrical bioimpedence (TEB) and trans-thoracic echocardiography (TTE).

Coloured zones represent the perceived degree of clinical harm resulting from a measurement error: Green = none, Yellow = mild, Orange = moderate, Red = severe. Error grids comparing ΔSVTTE and ΔSVSSD are shown in Fig 7. In the non-tourniquet strand 121 matched values were plotted. The risk of harm as a result of a measurement error was classed as: none 79 (65%), mild 22 (18%), moderate 17 (14%) and severe 3 (2%). In the tourniquet strand 113 matched values were plotted. The risk of harm as a result of a measurement error was classed as: none 86 (76%), mild 14 (12%), moderate 11 (10%) and severe 2 (2%). The overall risk of severe harm when measurements for both streams were combined was 4.4%. 4 subjects, 2 in each strand, each had 1 measurement value indicative of severe harm.
Fig 7

Error grid showing changes in stroke volume from baseline (ΔSV) recorded using suprasternal Doppler (SSD) and trans-thoracic echocardiography (TTE).

Coloured zones represent the perceived degree of clinical harm resulting from a measurement error: Green = none, Yellow = mild, Orange = moderate, Red = severe.

Error grid showing changes in stroke volume from baseline (ΔSV) recorded using suprasternal Doppler (SSD) and trans-thoracic echocardiography (TTE).

Coloured zones represent the perceived degree of clinical harm resulting from a measurement error: Green = none, Yellow = mild, Orange = moderate, Red = severe.

Discussion

The LBNP model achieved the aim of simulated hypovolaemia, producing a graded, sustained fall in stroke volume followed by a return to baseline when suction was discontinued. Most previous studies have used fixed degrees of applied LBNP in order to produce simulated hypovolaemia but this is limited by the inter-individual response and can produce a non-uniform reduction in stroke volume. Our approach was to target the degree of LBNP to a reduction in stroke volume recorded using finger plethysmography, a device that produces a continuous measure of SV. In this way we were able to produce a more standardised fall in SV across all of the study subjects. Although finger plethysmography was not a candidate device for comparison in the current study, principally because of its lack of utility in an urgent or emergency care setting, the SV values produced by this device correlated well to the reference standard. The application of lower limb tourniquets were a feature of this study for two principal reasons. Firstly, we wished to assess whether the presence of a potentially increased afterload caused by the presence of the tourniquets produced divergent results for SV measurement in the two strands of the study. In fact, our results show very little discernible difference between the two strands and it is likely that the addition of lower limb tourniquets did not produce a noticeable difference in the measured SV. Secondly we wanted, as far as possible, to reproduce the nociceptive response to trauma which has been shown to alter the haemodynamic response to haemorrhage [2]. Our results demonstrate that both non-invasive stroke volume monitors produced divergent values from the reference standard produced by echocardiography. Consensus opinion holds that a percentage error (PE) of 30% is the upper limit of acceptability in a clinical cardiac output monitoring device [15], however, both subject devices in the present study recorded errors well above this limit. In the case of the thoracic bioimpedance device there was also a very noticeable degree of bias which increased progressively at higher stroke volumes. Previous studies that have compared thoracic bio-impedence derived cardiac output to a reference standard have also shown that they are unreliable in producing accurate values for absolute measures of stroke volume. A meta-analysis examining the use of thoracic bioimpedance in nine studies showed only one instance where the PE value of < 30% with many reporting values well in excess of this threshold and up to 73% [17]. More recent studies conducted in pregnant women have reported that thoracic bioimpedance produces acceptable limits of agreement when compared to echocardiography but these studies do not induce any dynamic change in stroke volume in the study subjects, limiting the translatability of the results [18, 19]. The findings of the current study show that although supra-sternal Doppler was more accurate than thoracic electrical bio-imepdence it still showed a potentially unacceptable degree of error with a PE of 35% and 48%. These findings are in keeping with a meta-analysis of 6 studies that compared supra-sternal Doppler derived cardiac output to a reference standard and showed a mean PE of 42.7% [20]. Both devices performed better at assessing the change in SV rather than the absolute value, with concordance rates of around 90%, a value which has been suggested indicates an acceptable performance [15]. There was also a significant correlation between ΔSVTTE and both ΔSVTEB/SSD, however, the observed correlation was not particularly strong, most notably in the bioimpedance non–tourniquet strand. It is worth commenting that the use of concordance rates in this setting may also give a falsely reassuring picture as to device performance. Although the majority of changes in values are in the same direction and hence the overall concordance rate is high there are a number of highly divergent readings. These discrepancies are more clearly identified by using an error grid methodology. The use of error grid analysis was originally described as a way of assessing the clinical acceptability of point of care blood glucose testing [21]. Although error grid methodology has been used to assess clinically relevant differences in blood pressure measurements [16] and has been recommended as a potentially useful technique in cardiac output comparison studies [14], there are few published studies in this area. The major advantage of this methodology is in highlighting the clinical relevance of differences in measurements. In keeping with previous studies we used consensus expert opinion from experienced specialists to construct an error grid, with a clinical scenario written to reflect the key research question. Although the majority of measurements recorded by both candidate devices fell into the no risk zone a significant number produced errors that were perceived by the consensus panel to have the potential for clinical risk. In a very small minority of cases this harm was perceived to have been severe. The potential for both the commission (i.e. administration of potentially unnecessary blood products or fluid) and omission (failure to provide blood product or fluid resuscitation) of therapy was reflected in the results for both devices. The pattern of error in the supra-sternal Doppler group appeared to be different to that of the thoracic bioimpedance group; the former only having single isolated errors whilst the latter had clusters of multiple errors in each subject. The results of the current study highlight the fundamental difficulty in using non-invasive stroke volume monitors in the initial management of patients with suspected blood loss. Firstly, whilst most clinicians would accept that a trend in a physiological variable is more useful as a gauge of the degree of blood loss or response to treatment, such a trend necessarily requires repeated measures. At the onset of a clinical scenario, clinicians must make an assessment based on the data to hand, and are therefore reliant on a one off measurement. It is this aspect that makes the measurement of systolic blood pressure such a seductive target as population baseline values are widely understood, even by individuals with limited clinical training and experience. However, as discussed and demonstrated in the current study, blood pressure is often a poor measure of the degree of blood loss. What is needed is a quick and precise way of ascertaining stroke volume and hence blood flow and then indexing that value to the size of the patient in order to provide a more accurate measure of intra vascular volume status. The results of the current study do not demonstrate that either of the candidate devices is capable of this. Although, both devices were better at detecting trends in stroke volume changes, with apparently acceptable concordance rates, the use of error grid analysis clearly highlights the presence of a small number of highly aberrant results, produced by both devices, which if acted upon could have the potential to produce clinical harm. Our study has several limitations. Firstly although the LBNP technique is well described and efficacious in producing a change in stroke volume over a clinically relevant range by reducing venous return, our study is experimental and not clinical which may limit the translation of these data into clinical practice. However, controlled experiments of flow monitoring in actual traumatic haemorrhage are highly unlikely to be achievable. An additional limitation is the use of healthy subjects with no cardiovascular disease and the absence of cardiovascular changes induced by vasoactive drugs and large infusions of fluid. Finally, whilst the thoracic bioimpedance device is essentially non operator dependent both supra-sternal Doppler and our chosen reference standard for stroke volume assessment rely on a user acquiring an optimised Doppler signal. We mitigated this limitation by using a single user for both techniques across the entire study and ensuring that both users were trained and experienced in the technique. In conclusion we found that two flow monitoring devices, based on different physical principles, both had poor accuracy in measuring absolute stroke volume when compared to a reference standard. Both devices were better able to detect trend changes in stroke volume in a simulated hypovolaemia model but in a small minority of cases produced measurement errors that had the potential to produce significant clinical harm. If such devices are used for the early detection of hypovolaemia following haemorrhage, the values produced should be interpreted with caution and not used to determine therapy in isolation.

Error grid questionnaire.

Background material and information sent to respondents in order to provide material to construct error grids. (DOCX) Click here for additional data file.

Error grid questionnaire returns.

Error grid questionnaire results from 15 respondents. Respondents first provided information on what fall in stroke volume from baseline would either require no action (Code A), possible action (Code B) or essential action (Code C)–these results are shown on the ACTUAL column of each table. Respondents were then asked to quantify the harm from a divergent measurement using the same range of stroke volumes–shown as DEVICE in the table. Harm was quantified numerically as None (0), Mild (2), Moderate (5) or Severe (10). Cumulative results are shown in a single table in which a range of actual versus measured falls in stroke volume are shown along with the perceived degree of harm from measurement error. The range of harm is from 0 (0 respondents thought that harm could occur) to 150 (all respondents thought that severe harm was likely). (DOCX) Click here for additional data file.

Error grid results.

Excel spreadsheet where each cell is a comparator between actual fall in stroke volume and measured fall in stroke volume. The range of harm from 0–150 has been recalculated as a percentage value and colour coded using the colour scales function. (XLSX) Click here for additional data file.

Error grid polygons.

Smoothed polygon created from the Excel spreadsheet presented in File S3. Colours indicate degree of harm Red (Severe), Orange (Moderate), Yellow (Mild), Green (None). (PNG) Click here for additional data file. 6 Sep 2021 PONE-D-21-19494Detection of hypovolaemia in a simulated healthy volunteer model of traumatic haemorrhage; a comparison of two non-invasive monitoring devices using error grid analysis alongside traditional measures of agreementPLOS ONE Dear Dr. Hutchings, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. ACADEMIC EDITOR: I appreciated the importance of your work. The reviewers were positive about your manuscript, but one of them requested revisions. Please address all the comments pointed out by the reviewer. You will find that the comments help improve the manuscript. Please submit your revised manuscript by Oct 21 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript: A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'. A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'. An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. We look forward to receiving your revised manuscript. Kind regards, Kenta Matsumura Academic Editor PLOS ONE Journal Requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf 2. Please include additional information regarding the survey or questionnaire used in the study and ensure that you have provided sufficient details that others could replicate the analyses. For instance, if you developed a questionnaire as part of this study and it is not under a copyright more restrictive than CC-BY, please include a copy, in both the original language and English, as Supporting Information. If the original language is written in non-Latin characters, for example Amharic, Chinese, or Korean, please use a file format that ensures these characters are visible. 3. Please state whether you validated the questionnaire prior to testing on study participants. Please provide details regarding the validation group within the methods section. 4. PLOS requires an ORCID iD for the corresponding author in Editorial Manager on papers submitted after December 6th, 2016. Please ensure that you have an ORCID iD and that it is validated in Editorial Manager. To do this, go to ‘Update my Information’ (in the upper left-hand corner of the main menu), and click on the Fetch/Validate link next to the ORCID field. This will take you to the ORCID site and allow you to create a new iD or authenticate a pre-existing iD in Editorial Manager. Please see the following video for instructions on linking an ORCID iD to your Editorial Manager account: https://www.youtube.com/watch?v=_xcclfuvtxQ 5. Please include captions for your Supporting Information files at the end of your manuscript, and update any in-text citations to match accordingly. Please see our Supporting Information guidelines for more information: http://journals.plos.org/plosone/s/supporting-information. [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: No ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: I Don't Know Reviewer #2: No ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: In this report, hypovolaemia which is simulated by lower body negative pressure (LBNP) method is studied by detecting a stroke volume (SV). The SV is measured by two devices, trance-thoracic electrical bioimpedance (TEB) and supra-sternal Doppler (SSD), and is compared between two devices against a reference standard measured with using trans-thoracic echocardiography (TTE). Results show that these two devices have poor accuracy in measuring absolute stroke volume when compared to a reference standard. Authors describe that both devices are more useful for detecting trend changes in the SV than the absolute value; the obtained value with two device should be interpreted with caution. Measurement reliability is most important for medical workers using these device, and is essentially judged only by accumulation of a large amount of data. Hence this paper is an important contribution and I recommend that it be accepted for publication. Reviewer #2: I found the topic interesting, and the study appears to be well executed. However, what the authors did in this study was not detection of hypovolaemia but examination of agreement of SV of different devices using LBNP technique. Although data itself is potentially useful, the authors need to change the title, introduction, and discussion to reflect what they did in this study. Introduction: Please summarize hemodynamic changes induced by LBNP referring to previous studies (e.g., Guo, J Appl Physiol,100: 1785–1792, 2006. Levine, Circulation 90: 298–306, 1994). Based on the above, the authors should mention the weak points of former works (identification of the gaps) and describe the current investigation's novelties to justify the paper deserves to be published in PLOS ONE. Methods: "21 subjects": Was the alpha level set to 0.05? Please define "SSD" Body mass index —> "k"g/m2 Please add references regarding the Bland-Altman plot. "In order to…error grid methodology).": Is this necessary? If so, please summarize the results in the Results section. Results: Please add the results of a series of repeated ANOVAs, such as F (20, 400) = 34.56, partial eta squared = 0.63, p = 0.0024, by main effect and interaction, to the main text. Discussion: The authors will need to revise this section according to the above changes. ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. 29 Oct 2021 Journal Requirements 1. The manuscript has been reformatted in accordance with the journal requirements 2. A copy of the questionnaire is included in the supporting information (S1 File). 3. The questionnaire was not validated prior to the study 4. ORCID ID included in editorial manager 5. Supporting information citations re formatted within manuscript Responses to Reviewer 2 Reviewer #2: I found the topic interesting, and the study appears to be well executed. However, what the authors did in this study was not detection of hypovolaemia but examination of agreement of SV of different devices using LBNP technique. Although data itself is potentially useful, the authors need to change the title, introduction, and discussion to reflect what they did in this study. The title has been amended to include the phrase assessment of stroke volume. Introduction: Please summarize hemodynamic changes induced by LBNP referring to previous studies (e.g., Guo, J Appl Physiol,100: 1785–1792, 2006. Levine, Circulation 90: 298–306, 1994). Based on the above, the authors should mention the weak points of former works (identification of the gaps) and describe the current investigation's novelties to justify the paper deserves to be published in PLOS ONE. Section added to the introduction providing more granularity on the LBNP technique – suggested references cited Methods: "21 subjects": Was the alpha level set to 0.05? Yes alpha 0.05 – added to manuscript Please define "SSD" “Supra sternal Doppler” – clarification added Body mass index —> "k"g/m2 Corrected Please add references regarding the Bland-Altman plot. Actioned "In order to…error grid methodology).": Is this necessary? If so, please summarize the results in the Results section. I’m afraid I don’t understand this comment. Error grid methodology is a core part of the paper and the results are clearly presented. Results: Please add the results of a series of repeated ANOVAs, such as F (20, 400) = 34.56, partial eta squared = 0.63, p = 0.0024, by main effect and interaction, to the main text. The requested data has been added in Table 1 1 Dec 2021 PONE-D-21-19494R1Quantification of stroke volume in a simulated healthy volunteer model of traumatic haemorrhage; a comparison of two non-invasive monitoring devices using error grid analysis alongside traditional measures of agreementPLOS ONE Dear Dr. Hutchings, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. The revised version of your study satisfied most of the concerns raised by the reviewers. However, the reviewer raised additional comments that would strengthen your analysis. Thus, I would like to invite you to submit a re-revised version of the paper for consideration again. If your response is comprehensive, I will likely review the changes myself rather than send the paper out for re-review. Please submit your revised manuscript by Jan 15 2022 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript: A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'. A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'. An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. We look forward to receiving your revised manuscript. Kind regards, Kenta Matsumura Academic Editor PLOS ONE [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author Reviewer #2: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #2: Partly ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #2: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #2: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #2: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #2: Thank you for your revision. I am satisfied with most responses. But, I suggest that the error grid protocol be incorporated into the main text, not as supplementary material, and mention its novelty in the introduction. These would strengthen your study. ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #2: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. 2 Dec 2021 We have incorporated the error grid methodology in the main methods section of the manuscript and added a paragraph to the introduction expanding on the role of error grid methodology in device comparison studies. 6 Dec 2021 Quantification of stroke volume in a simulated healthy volunteer model of traumatic haemorrhage; a comparison of two non-invasive monitoring devices using error grid analysis alongside traditional measures of agreement PONE-D-21-19494R2 Dear Dr. Hutchings, We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements. Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication. An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. Kind regards, Kenta Matsumura Academic Editor PLOS ONE Additional Editor Comments (optional): Reviewers' comments: 10 Dec 2021 PONE-D-21-19494R2 Quantification of stroke volume in a simulated healthy volunteer model of traumatic haemorrhage; a comparison of two non-invasive monitoring devices using error grid analysis alongside traditional measures of agreement Dear Dr. Hutchings: I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. If we can help with anything else, please email us at plosone@plos.org. Thank you for submitting your work to PLOS ONE and supporting open access. Kind regards, PLOS ONE Editorial Office Staff on behalf of Dr. Kenta Matsumura Academic Editor PLOS ONE
  20 in total

1.  A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose.

Authors:  J L Parkes; S L Slatin; S Pardo; B H Ginsberg
Journal:  Diabetes Care       Date:  2000-08       Impact factor: 19.112

Review 2.  A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output.

Authors:  Lester A Critchley; Anna Lee; Anthony M-H Ho
Journal:  Anesth Analg       Date:  2010-08-24       Impact factor: 5.108

3.  The microcirculatory response to compensated hypovolemia in a lower body negative pressure model.

Authors:  Sebastiaan A Bartels; Rick Bezemer; Dan M J Milstein; Matthijs Radder; Alexandre Lima; Thomas G V Cherpanath; Michal Heger; John M Karemaker; Can Ince
Journal:  Microvasc Res       Date:  2011-08-02       Impact factor: 3.514

Review 4.  Haemorrhage control in severely injured patients.

Authors:  Russell L Gruen; Karim Brohi; Martin Schreiber; Zsolt J Balogh; Veronica Pitt; Mayur Narayan; Ronald V Maier
Journal:  Lancet       Date:  2012-09-22       Impact factor: 79.321

5.  Cerebral autoregulation is preserved during orthostatic stress superimposed with systemic hypotension.

Authors:  Hong Guo; Nancy Tierney; Frederic Schaller; Peter B Raven; Scott A Smith; Xiangrong Shi
Journal:  J Appl Physiol (1985)       Date:  2006-01-19

Review 6.  Haemodynamic changes in trauma.

Authors:  E Kirkman; S Watts
Journal:  Br J Anaesth       Date:  2014-08       Impact factor: 9.166

7.  Cerebral versus systemic hemodynamics during graded orthostatic stress in humans.

Authors:  B D Levine; C A Giller; L D Lane; J C Buckey; C G Blomqvist
Journal:  Circulation       Date:  1994-07       Impact factor: 29.690

8.  A critical reappraisal of the ATLS classification of hypovolaemic shock: does it really reflect clinical reality?

Authors:  M Mutschler; U Nienaber; T Brockamp; A Wafaisade; H Wyen; S Peiniger; T Paffrath; B Bouillon; M Maegele
Journal:  Resuscitation       Date:  2012-07-24       Impact factor: 5.262

Review 9.  Lower body negative pressure as a model to study progression to acute hemorrhagic shock in humans.

Authors:  William H Cooke; Kathy L Ryan; Victor A Convertino
Journal:  J Appl Physiol (1985)       Date:  2004-04

10.  Comparison of bioreactance and echocardiographic non-invasive cardiac output monitoring and myocardial function assessment in primagravida women.

Authors:  A Doherty; A El-Khuffash; C Monteith; L McSweeney; C Breatnach; E Kent; E Tully; F Malone; P Thornton
Journal:  Br J Anaesth       Date:  2017-04-01       Impact factor: 9.166

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.