Literature DB >> 27199309

Methodology of method comparison studies evaluating the validity of cardiac output monitors: a stepwise approach and checklist.

L J Montenij1, W F Buhre2, J R Jansen3, C L Kruitwagen4, E E de Waal5.   

Abstract

The validity of each new cardiac output (CO) monitor should be established before implementation in clinical practice. For this purpose, method comparison studies investigate the accuracy and precision against a reference technique. With the emergence of continuous CO monitors, the ability to detect changes in CO, in addition to its absolute value, has gained interest. Therefore, method comparison studies increasingly include assessment of trending ability in the data analysis. A number of methodological challenges arise in method comparison research with respect to the application of Bland-Altman and trending analysis. Failure to face these methodological challenges will lead to misinterpretation and erroneous conclusions. We therefore review the basic principles and pitfalls of Bland-Altman analysis in method comparison studies concerning new CO monitors. In addition, the concept of clinical concordance is introduced to evaluate trending ability from a clinical perspective. The primary scope of this review is to provide a complete overview of the pitfalls in CO method comparison research, whereas other publications focused on a single aspect of the study design or data analysis. This leads to a stepwise approach and checklist for a complete data analysis and data representation.
© The Author 2016. Published by Oxford University Press on behalf of the British Journal of Anaesthesia. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Keywords:  cardiac output; trends; validation studies

Mesh:

Year:  2016        PMID: 27199309     DOI: 10.1093/bja/aew094

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  29 in total

1.  Accuracy and precision of non-invasive cardiac output monitoring by electrical cardiometry: a systematic review and meta-analysis.

Authors:  M Sanders; S Servaas; C Slagt
Journal:  J Clin Monit Comput       Date:  2019-06-07       Impact factor: 2.502

2.  Goal-directed therapy: hit early and personalize!

Authors:  Bernd Saugel; Frederic Michard; Thomas W L Scheeren
Journal:  J Clin Monit Comput       Date:  2017-06-26       Impact factor: 2.502

3.  Real-time estimation of mean arterial blood pressure based on photoplethysmography dicrotic notch and perfusion index. A pilot study.

Authors:  Jona Joachim; Maxime Coutrot; Sandrine Millasseau; Joaquim Matéo; Alexandre Mebazaa; Etienne Gayat; Fabrice Vallée
Journal:  J Clin Monit Comput       Date:  2020-02-20       Impact factor: 2.502

4.  A comparison of volume clamp method-based continuous noninvasive cardiac output (CNCO) measurement versus intermittent pulmonary artery thermodilution in postoperative cardiothoracic surgery patients.

Authors:  Julia Y Wagner; Annmarie Körner; Leonie Schulte-Uentrop; Mathias Kubik; Hermann Reichenspurner; Stefan Kluge; Daniel A Reuter; Bernd Saugel
Journal:  J Clin Monit Comput       Date:  2017-05-24       Impact factor: 2.502

5.  Clinical evaluation of a cutaneous zero-heat-flux thermometer during cardiac surgery.

Authors:  Christophe Verheyden; Arne Neyrinck; Annouschka Laenen; Steffen Rex; Elke Van Gerven
Journal:  J Clin Monit Comput       Date:  2021-09-24       Impact factor: 1.977

6.  Transpulmonary thermodilution in patients treated with veno-venous extracorporeal membrane oxygenation.

Authors:  Gregor Loosen; Alice Marguerite Conrad; Michael Hagman; Nils Essert; Manfred Thiel; Thomas Luecke; Joerg Krebs
Journal:  Ann Intensive Care       Date:  2021-07-02       Impact factor: 6.925

7.  Bioreactance and fourth-generation pulse contour methods in monitoring cardiac index during off-pump coronary artery bypass surgery.

Authors:  Laura Anneli Ylikauma; Pasi Petteri Ohtonen; Tiina Maria Erkinaro; Merja Annika Vakkala; Janne Henrik Liisanantti; Jari Uolevi Satta; Tatu Sakari Juvonen; Timo Ilari Kaakinen
Journal:  J Clin Monit Comput       Date:  2021-05-26       Impact factor: 1.977

8.  Performance of a capnodynamic method estimating effective pulmonary blood flow during transient and sustained hypercapnia.

Authors:  Thorir Svavar Sigmundsson; Tomas Öhman; Magnus Hallbäck; Eider Redondo; Fernando Suarez Sipmann; Mats Wallin; Anders Oldner; Caroline Hällsjö Sander; Håkan Björne
Journal:  J Clin Monit Comput       Date:  2017-05-11       Impact factor: 2.502

Review 9.  The haemodynamic dilemma in emergency care: Is fluid responsiveness the answer? A systematic review.

Authors:  Mohammed H Elwan; Ashraf Roshdy; Eman M Elsharkawy; Salah M Eltahan; Timothy J Coats
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2017-03-06       Impact factor: 2.953

10.  Evaluation of respiratory rate monitoring using a microwave Doppler sensor mounted on the ceiling of an intensive care unit: a prospective observational study.

Authors:  Hiroyuki Tanaka; Masashi Yokose; Shunsuke Takaki; Takahiro Mihara; Yusuke Saigusa; Takahisa Goto
Journal:  J Clin Monit Comput       Date:  2021-06-30       Impact factor: 2.502

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.