Maurits A den Boer1,2, Szu-Hsueh Lai1,2, Xiaoguang Xue3, Muriel D van Kampen3, Boris Bleijlevens3, Albert J R Heck1,2. 1. Biomolecular Mass Spectrometry and Proteomics, Bijvoet Center for Biomolecular Research and Utrecht Institute of Pharmaceutical Sciences, Utrecht University, Padualaan 8, 3584 CH Utrecht, The Netherlands. 2. Netherlands Proteomics Center, Padualaan 8, 3584 CH Utrecht, The Netherlands. 3. Genmab, Uppsalalaan 15, 3584 CT Utrecht, The Netherlands.
Abstract
Qualitative and quantitative mass analysis of antibodies and related macromolecular immune complexes is a prerequisite for determining their identity, binding partners, stoichiometries, and affinities. A plethora of bioanalytical technologies exist to determine such characteristics, typically based on size, interaction with functionalized surfaces, light scattering, or direct mass measurements. While these methods are highly complementary, they also exhibit unique strengths and weaknesses. Here, we benchmark mass photometry (MP), a recently introduced technology for mass measurement, against native mass spectrometry (MS) and size exclusion chromatography multi-angle light scattering (SEC-MALS). We examine samples of variable complexity, namely, IgG4Δhinge dimerizing half-bodies, IgG-RGY hexamers, heterogeneously glycosylated IgG:sEGFR antibody-antigen complexes, and finally megadalton assemblies involved in complement activation. We thereby assess the ability to determine (1) binding affinities and stoichiometries, (2) accurate masses, for extensively glycosylated species, and (3) assembly pathways of large heterogeneous immune complexes. We find that MP provides a sensitive approach for characterizing antibodies and stable assemblies, with dissociation correction enabling us to expand the measurable affinity range. In terms of mass resolution and accuracy, native MS performs the best but is occasionally hampered by artifacts induced by electrospray ionization, and its resolving power diminishes when analyzing extensively glycosylated proteins. In the latter cases, MP performs well, but single-particle charge detection MS can also be useful in this respect, measuring masses of heterogeneous assemblies even more accurately. Both methods perform well compared to SEC-MALS, still being the most established method in biopharma. Together, our data highlight the complementarity of these approaches, each having its unique strengths and weaknesses.
Qualitative and quantitative mass analysis of antibodies and related macromolecular immune complexes is a prerequisite for determining their identity, binding partners, stoichiometries, and affinities. A plethora of bioanalytical technologies exist to determine such characteristics, typically based on size, interaction with functionalized surfaces, light scattering, or direct mass measurements. While these methods are highly complementary, they also exhibit unique strengths and weaknesses. Here, we benchmark mass photometry (MP), a recently introduced technology for mass measurement, against native mass spectrometry (MS) and size exclusion chromatography multi-angle light scattering (SEC-MALS). We examine samples of variable complexity, namely, IgG4Δhinge dimerizing half-bodies, IgG-RGY hexamers, heterogeneously glycosylated IgG:sEGFR antibody-antigen complexes, and finally megadalton assemblies involved in complement activation. We thereby assess the ability to determine (1) binding affinities and stoichiometries, (2) accurate masses, for extensively glycosylated species, and (3) assembly pathways of large heterogeneous immune complexes. We find that MP provides a sensitive approach for characterizing antibodies and stable assemblies, with dissociation correction enabling us to expand the measurable affinity range. In terms of mass resolution and accuracy, native MS performs the best but is occasionally hampered by artifacts induced by electrospray ionization, and its resolving power diminishes when analyzing extensively glycosylated proteins. In the latter cases, MP performs well, but single-particle charge detection MS can also be useful in this respect, measuring masses of heterogeneous assemblies even more accurately. Both methods perform well compared to SEC-MALS, still being the most established method in biopharma. Together, our data highlight the complementarity of these approaches, each having its unique strengths and weaknesses.
With the continued
advancement of antibody-based formats as biopharmaceuticals,
analytical techniques providing robust and accurate characterization
of these products and related macromolecular immune complexes become
increasingly important. Antibody functioning strongly depends on noncovalent
protein–protein interactions, with their unique structural
organization bridging molecular recognition with the recruitment of
effector functions.[1] Structurally, standard
IgG-based antibodies are homo-heterodimers consisting of two heavy
chains (HCs) and two light chains (LCs) that are connected through
several disulfide bridges. Target recognition is enabled by two variable
antigen–binding (Fab) arms, which engage in highly specific
epitope–paratope interactions.[2,3] Effector functions,
on the other hand, are primarily mediated by the constant (Fc) tail,
which recruits and directs immune cells by binding to Fc receptor
proteins,[4,5] but can also initiate humoral immune responses
such as the classical complement pathway.[6,7] Furthermore,
the Fc tail can facilitate the formation of functional oligomers—linked
covalently in IgA and IgM or by noncovalent interactions in surface-bound
IgGs.[8] The ability to accurately characterize
antibodies and their interactions with antigens and receptors is thus
of crucial importance, both for fundamental research as well as in
the optimization of antibody engineering and drug development.Affinities and kinetics of antibody–antigen interactions
are typically assessed by biosensors that quantify interactions with
functionalized surfaces. The most prevalent of such approaches is
surface plasmon resonance,[9,10] but bio-layer interferometry,[11] quartz crystal microbalances,[12] and Förster resonance energy transfer microscopy[13,14] provide accurate readouts down to sub-nanomolar Kd values. However, these techniques are generally limited
to binary interactions, preventing their use in studying oligomerization
and the formation of larger immune complexes of multiple stoichiometries
and compositions.Low-resolution biophysical methods based on
size, charge, diffusion,
or light scattering are widely used in academia and industry to study
antibody oligomerization and complex formation. Oligomeric distributions
and aggregation states can be evaluated using dynamic light scattering.[15−17] Analytical ultracentrifugation, on the other hand, can provide quantitative
data at a higher resolution,[18,19] allowing the technique
to be used to study protein–protein interactions as well,[20] but arduous experimental procedures make this
approach impractical for routine use.[21] Moreover, also capillary electrophoresis-related techniques have
been applied for the characterization of antibodies and their interactions.[22,23]Size exclusion chromatography (SEC) remains the long-standing
industry
standard for analyzing the quaternary structure of antibody products.
SEC utilizes a porous matrix as a stationary phase to enable size-based
separation, followed by detection to provide sensitive, highly reproducible,
and quantitative data. As SEC alone does not provide an accurate means
to assess masses, it often includes average molecular mass determination
by coupling to a multi-angle light scattering (MALS) detector, which
acts by measuring the light scattering generated by particles, being
proportional to their molar mass and concentration. This makes SEC-MALS
a very versatile tool for studying antibodies, although it is also
hampered by some issues because dilution and shear forces during chromatographic
separation can affect equilibria. The technique also requires optimization
regarding running conditions, as protein species can adsorb to the
matrix.Direct mass measurement by native mass spectrometry
(MS)[24,25] represents a relatively newer component
of the analytical toolbox
for antibody analysis. Compared to the techniques described above,
native MS yields higher mass resolution and accuracy that can be used
to assess, for instance, microheterogeneity of monoclonal antibodies[26,27] and their derivatives,[28−30] the formation of antibody–antigen
complexes,[31−33] and larger megadalton particle immune complexes.[34,35] As noncovalent interactions are retained, native MS can also be
used to probe binding equilibria of protein–ligand and protein–protein
interactions when instrumental parameters are carefully optimized
to avoid bias and artifacts.[36−40] Direct online coupling of SEC or capillary zone electrophoresis
to mass spectrometers can further enable separation and structural
characterization of protein assemblies.[41,42]Typically,
in native MS, mass analysis relies on resolving charge
states of the same species in the m/z space of the mass spectrum, which becomes more difficult when the
analytes become heavier and more heterogeneous. In such cases, single-particle
charge detection-MS (CD-MS)[43] may be very
useful. This technique, which recently was also demonstrated on commercial
Orbitrap-based instruments,[35,44] makes it possible to
directly assess the mass of single ions by measuring their m/z in parallel with their charge z, as inferred from their single-ion intensity value.Mass photometry (MP) was recently introduced as a technology that
enables mass analysis of proteins and protein complexes under native
buffering conditions.[45] MP makes use of
interferometric scattering microscopy to detect and quantify light
scattering caused by single particles.[46−50] When particles in solution bind nonspecifically to
a glass surface, their scattering signal interferes with the measured
reflectivity of the glass/water interface. Because the optical properties
and density of proteins are quite uniform, this reflectivity change
is proportional to the molecular mass, allowing MP to provide a direct
mass measurement for each particle.[46,51] MP has already
been shown to be able to provide quantitative data, allowing the binding
affinities and kinetics of antibody–antigen and antibody–receptor
interactions to be explored.[45,52,53] However, because typical experimental conditions are limited to
low nanomolar range concentrations, the methods outlined in these
studies apply only to relatively strong and slowly dissociating protein
assemblies.Here, we compare MP side by side with native MS
and SEC-MALS, employing
these techniques for the analysis of a variety of antibody formats
and heavily glycosylated macromolecular immune complexes. We expand
the affinity range of quantitative MP experiments by modeling the
dissociation of weaker interactions, and we compare the techniques
in terms of their ability to assess a wide range of binding equilibria.
We assess the pros and cons of each of these approaches, considering
dynamic range, robustness, mass resolution, and mass accuracy, and
we highlight their strengths in resolving extensively glycosylated
species.
Materials and Methods
More detailed descriptions of
the methods are provided in the Supporting Information Methods.
Protein Samples
Anti-EGFR antibodies in IgG4Δhinge,
IgG1, and IgG1-RGY format and sEGFR were recombinantly expressed and
purified by Genmab.[34,40,54,55] Human C1q was obtained from Complement Technology.
Samples were buffer-exchanged to the appropriate solution. Protein
complexes were assembled by mixing subcomponents at the desired molar
ratios, followed by incubation at room temperature for at least 30
min. For quantitative experiments, incubation after preparing a dilution
series was proceeded for at least 4 h.
Native MS and CD-MS
For native MS, proteins in 150
mM aqueous ammonium acetate pH 7.5 were measured by direct infusion
from a static nano-electrospray ionization (ESI) source. Quantitative
experiments were performed on a modified LCT time-of-flight instrument
(Waters), measuring samples in triplicate. All other experiments were
performed on a Q Exactive Plus UHMR Orbitrap instrument (Thermo Fisher
Scientific). For CD-MS, dilute samples were measured at low pressure
and high resolution (1 s transient) for accurate determination of
both m/z and z of
single ions.
SEC-MALS
SEC-MALS experiments were
performed on a Waters
HPLC with an in-line UV detector (Waters 2487 Dual Absorbance), a
MALS detector (MiniDAWN, Wyatt Technology), and an RI detector (Optilab,
Wyatt Technology). Proteins were separated on an SRT SEC-500 column
(Sepax Technologies) using 100 mM sodium phosphate, 100 mM sodium
sulfate, and pH 6.8 as mobile phase at 0.35 mL/min. Data were processed
by ASTRA software (Wyatt) based on MALS-RI for antibody mass determination
or MALS-UV-RI (Protein Conjugate Analysis) for the analysis of glycan
contributions and larger complexes.
Mass Photometry
MP experiments were performed by measuring
the samples in PBS on a Refeyn OneMP mass photometer (Refeyn).
Triplicate measurements of 12,000 frames were combined into a single
mass histogram. When measuring protein complexes, high concentration
solutions were jump-diluted to nM range measurement concentrations
in approximately 5–30 s. For quantitative experiments, a dilution
series was measured in triplicate in recordings of 6000 frames. For
these experiments, dissociation upon jump dilution was modeled to
infer complex abundance in the original solution.
Results and Discussion
We started our MP analysis by characterizing the monomer–dimer
equilibrium of hinge-deleted human IgG4 (IgG4Δhinge) molecules,
providing a simple and small one-component system, for which we reported
earlier data from SEC and native MS.[40] Deletion
of the hinge region removes the disulfide bonds that bridge the two
HCs, meaning that the two-halves of the antibody interact solely via
noncovalent interactions. This results in an equilibrium between antibody
half molecules (HLs) and HL dimers (HL)2. Previous work
from our group assessed the effects of specific mutations in the CH3
domain on this equilibrium by native MS and SEC,[40] providing a panel of highly similar samples, with Kd values spanning 6 orders of magnitude (10–10 to 10–4 M). Because MP experiments
are typically performed at concentrations of only a few nM, we used
jump dilution to quickly dilute concentrated samples just before starting
the measurement. Assuming that the koff of the interaction is low enough, the observed distribution of protein
assemblies should then reflect that of the original concentrated sample.
Jump-Diluted
IgG4Δhinge Dimers Dissociate during MP Analysis
The
distinct light scattering caused by single particles of different
masses as measured by MP can be converted into masses and shown in
histograms. The mass histograms of “wt” IgG4Δhinge
jump-diluted from a 16 μM solution to a measurement concentration
of 4 nM reveal two distinct distributions centered at the expected
masses, namely, 73 kDa HL and 146 kDa (HL)2 (Figure A; see Table S1 for an overview of all measured masses). The particle
counts constituted about 74% of (HL)2 dimers (mass abundance
of 85%), which is lower than expected for this relatively strong interaction
(Kd = 50 nM by native MS[40]). Upon further inspection, we observed that the abundance
of the dimer already decreased during the first seconds of the MP
measurement with a pattern resembling exponential decay (Figure B). This indicates
that the (HL)2 dimer readily starts to dissociate upon
jump dilution as the sample re-equilibrates to the measurement concentration,
meaning that simply summing the data of the measurement window will
lead to an underestimation of the actual dimer abundance.
Figure 1
Qualitative
and quantitative characterization of IgG4Δhinge
mutants by MP. (A) Mass histogram showing particle counts of “wt”
IgG4Δhinge in PBS, jump-diluted from 16 μM and measured
at 4 nM, with normal distributions fitted for HL (bright red, 26%)
and (HL)2 (dark red, 74%). This histogram corresponds to
the first 80 s after jump dilution. Masses are indicated as the mean
of a fitted normal distribution. (B) Monomer–dimer distribution
during an extended experiment in triplicate (shades of red) revealed
that the abundance of “wt” (HL)2 decreased
during the analysis time window. Data were split into bins of 100
events, and an exponential decay function was fitted to the dimer
abundance within the bin to determine the koff. (C) Determined koff was used to estimate
the ratio between HL and (HL)2 at the instant of jump dilution
for a dilution series of the “wt” measured in triplicate,
revealing the apparent Kd of each measurement,
followed by the calculation of a Kd value
for the whole dilution series. (D) Fractional dimer abundances and Kd values resulting from a dilution series of
four IgG4Δhinge mutational variants, demonstrating that MP can
assess affinities over a broad dynamic range.
Qualitative
and quantitative characterization of IgG4Δhinge
mutants by MP. (A) Mass histogram showing particle counts of “wt”
IgG4Δhinge in PBS, jump-diluted from 16 μM and measured
at 4 nM, with normal distributions fitted for HL (bright red, 26%)
and (HL)2 (dark red, 74%). This histogram corresponds to
the first 80 s after jump dilution. Masses are indicated as the mean
of a fitted normal distribution. (B) Monomer–dimer distribution
during an extended experiment in triplicate (shades of red) revealed
that the abundance of “wt” (HL)2 decreased
during the analysis time window. Data were split into bins of 100
events, and an exponential decay function was fitted to the dimer
abundance within the bin to determine the koff. (C) Determined koff was used to estimate
the ratio between HL and (HL)2 at the instant of jump dilution
for a dilution series of the “wt” measured in triplicate,
revealing the apparent Kd of each measurement,
followed by the calculation of a Kd value
for the whole dilution series. (D) Fractional dimer abundances and Kd values resulting from a dilution series of
four IgG4Δhinge mutational variants, demonstrating that MP can
assess affinities over a broad dynamic range.
Modeling for Dissociation Expands the Affinity Range of IgG4Δhinge
Mutants Assessable by MP
To obtain a more accurate representation
of the oligomer distributions in solution before jump dilution, we
adjusted our data processing approach by modeling (HL)2 dissociation. When the sample concentration is diluted by several
orders of magnitude, especially when [HL] ≪ Kd, we can assume that the initial decrease in (HL)2 abundance is driven primarily by koff. Thus, by determining koff of the interaction,
we can fit an exponential decay function to the measured [(HL)2] over time to estimate [(HL)2] before jump dilution
(see Supporting Information Methods). We
evaluated this method using a panel of four IgG4Δhinge mutants
spanning a broad affinity range, measuring a dilution series to determine
their Kd (Figure C,D). Modeling for dissociation allowed us
to determine affinities well into the μM range, substantially
improving the dynamic range of quantitative MP experiments. Still,
a few issues remained. We found that the most consistent results were
obtained for relatively strong interactions with low koff values, such as those of “wt” IgG4Δhinge
(Kd = 2.3 nM). For the even stronger interactions
of the R409K mutant (Kd = 9.4 pM), the
equilibrium was still mostly geared toward the dimer at concentrations
assessable by MP, reducing the precision of Kd determination for this mutant. Weaker interactions such as
those of the D399S (Kd = 7.8 μM)
and L368A (Kd = 54 μM) mutants could
also be measured, although higher dissociation rates (0.012 and 0.029
s–1) reduced the accuracy of the model. Nonetheless,
MP experiments led to the same affinity ranking of the mutants as
did native MS, although with some discrepancies between the obtained
absolute Kd values (Table S2). An important difference is that native MS is performed
with a volatile buffering solution (i.e., aqueous ammonium acetate),
while MP enabled the use of PBS as a more physiological buffer. Furthermore,
standard native MS is also somewhat restrained to high nM to low μM
concentrations, reducing the accuracy by which (sub) nM range Kd values can be assessed. MP and native MS are
highly complementary in this sense, as they each have their own distinctive
preferred concentration range.
MP Outperforms Native MS
in the Mass Assessment of Heavily Glycosylated
Antibody–Antigen Assemblies
We next characterized
the interactions between antibodies and their antigen by MP and native
MS. Several therapeutic antibodies target glycosylated receptor proteins,
some of which are notoriously hard to analyze by native MS due to
their high degree of microheterogeneity. Methods that can accurately
mass measure and quantify these antigens and their interaction with
mAbs are therefore of great use to both fundamental and biopharmaceutical
research. Here, we analyze an IgG1 mAb targeting epidermal growth
factor receptor (EGFR), whereby we used the soluble domain (sEGFR).
This protein with a mass of 69,409 Da in its non-glycosylated form
is very heterogeneous, harboring 11 typical N-glycosylation motifs
that can be variably occupied.[56]Starting with the IgG1 alone, although the average mass obtained
by MP was in good agreement with native MS, the latter provided unparalleled
mass accuracy and resolution, enabling baseline resolution of individual
glycoforms (Figure ). However, native mass analysis of extensively glycosylated sEGFR
alone was difficult, being unable to resolve charge states because
of the presence of a plethora of proteoglycoforms. This obstacle was
overcome by using charge detection MS (CD-MS), recently developed
as a methodology for Orbitrap instruments, which provides an extra
dimension of data by measuring the charge of the ions independently.
CD-MS measured a mass of 88 kDa for sEGFR, in close agreement with
an earlier reported value derived by tandem MS experiments.[30] Similarly, MP readily provided a mass of 86
kDa, with SEC-MALS-UV-RI also measuring a mass of 91 kDa for sEGFR
(Figure S3). Next, when the anti-sEGFR
mAb was incubated together with sEGFR, MP presented further advantages.
In native MS, additional ion signals were observed for (IgG1)1:(sEGFR)1 (m/z 7000–8500), although poor resolution hampered mass determination,
while the full (IgG1)1:(sEGFR)2 complex (m/z 8500–10,000) could not be resolved
at all. However, both CD-MS and jump dilution MP enabled the reliable
measurement of the average masses for all co-occurring complexes,
clearly revealing the stoichiometry. However, binding occupancy was
somewhat lower in CD-MS, likely because of re-equilibration upon dilution
before the somewhat longer measurements. SEC-MALS was similarly able
to discern the full (IgG1)1:(sEGFR)2 complex,
although the resolution was substantially lower (Figure S3). Combining them, these data already show that MP
and CD-MS have advantages for mass analysis of heterogeneous antibody-antigen
complexes.
Figure 2
MP and CD-MS may overcome certain limitations of native MS in the
mass measurements of highly heterogeneous antibody–antigen
complexes. (A) MP provides an average mass for IgG1 (upper panel)
and sEGFR (middle) and is not hampered by the high micro-heterogeneity
of the latter. When 2 μM IgG1 was incubated with 5 μM
of sEGFR to form (IgG1)1:(sEGFR)1 and (IgG1)1:(sEGFR)2 complexes, jump dilution MP could resolve
these highly heterogeneous species (lower). (B) Although native MS
on samples at the same concentrations provided superior mass resolution
and accuracy for free IgG1 (upper), resolving individual glycoforms
(zoom), the high microheterogeneity of sEGFR, measured separately
(middle) and in antibody–antigen complexes (lower), resulted
in unresolved features. In these experiments, overlapping charge states
prevented mass measurements of these species. (C) More accurate masses
could be obtained by native CD-MS, measuring in two dimensions m/z and z (insets) for
sEGFR (upper) and all co-occurring species involving IgG1 and sEGFR
(lower). For these experiments, the same native MS samples were diluted
20-fold, leading to re-equilibration and thus a lower binding occupancy.
MP and CD-MS may overcome certain limitations of native MS in the
mass measurements of highly heterogeneous antibody–antigen
complexes. (A) MP provides an average mass for IgG1 (upper panel)
and sEGFR (middle) and is not hampered by the high micro-heterogeneity
of the latter. When 2 μM IgG1 was incubated with 5 μM
of sEGFR to form (IgG1)1:(sEGFR)1 and (IgG1)1:(sEGFR)2 complexes, jump dilution MP could resolve
these highly heterogeneous species (lower). (B) Although native MS
on samples at the same concentrations provided superior mass resolution
and accuracy for free IgG1 (upper), resolving individual glycoforms
(zoom), the high microheterogeneity of sEGFR, measured separately
(middle) and in antibody–antigen complexes (lower), resulted
in unresolved features. In these experiments, overlapping charge states
prevented mass measurements of these species. (C) More accurate masses
could be obtained by native CD-MS, measuring in two dimensions m/z and z (insets) for
sEGFR (upper) and all co-occurring species involving IgG1 and sEGFR
(lower). For these experiments, the same native MS samples were diluted
20-fold, leading to re-equilibration and thus a lower binding occupancy.
MP, SEC, and Native MS Analyses of the Monomer–Hexamer
Equilibrium of Soluble IgG1-RGY Hexamers Produce Consistent Results
We next evaluated the performance of MP in the characterization
of larger and more complex antibody-based systems, involved in immune
activation through the complement pathway.[8,34] Target-bound
IgG can initiate complement activity by forming a hexameric binding
platform for recognition of complex C1q. Although these IgG oligomers
are thought to only form by clustering on antigenic surfaces in vivo, this process can be mimicked in solution using
the engineered IgG-RGY platform, a triple mutant that readily forms
hexamers in equilibrium with monomers.[8,34,55]MP mass histograms and SEC-MALS chromatograms
of IgG1-RGY revealed as expected two species corresponding to the
monomer (denoted (IgG1)1) and hexamer ((IgG1)6) (Figure A,B). While
the MP mass of the monomer was in good agreement with the native MS
data (Figure C), we
noticed that the mass of the hexamer was consistently off by about
+70 kDa, possibly due to its non-globular shape as a flat disc. In
SEC-MALS, we observed peak trailing for the hexamer, potentially driven
by shear force-induced dissociation or re-equilibration during separation.
As reported previously,[8,34] native mass spectra of IgG1-RGY
also showed two well-resolved distributions for (IgG1)1 and (IgG1)6, but uniquely also intermediate oligomers
at lower abundance. Possibly, the ESI process could result in partial
dissociation of the hexamers, as this process is distinct from gas-phase-based
tandem MS (Figure S4).
Figure 3
MP enables qualitative
and quantitative characterization of the
monomer–hexamer equilibrium of IgG1-RGY. (A) MP mass histogram
(top) of 2 μM IgG1-RGY in PBS jump-diluted to 10 nM showing
monomeric ((IgG1)1) and hexameric ((IgG1)6)
species. The mass of the hexamer was consistently measured about 70
kDa too high. The relative abundance of the IgG1-RGY hexamer was measured
over a dilution series spanning a concentration range of 0.1 to 8
μM (bottom), with error bars indicating the standard deviation
over three technical replicate measurements. (B) SEC-MALS chromatogram
of the same 2 μM IgG1-RGY sample (top) and the fractional abundance
of the hexamer as measured by SEC-UV over a dilution series (bottom),
revealing a similar monomer to hexamer ratio. The quantitative data
in the lower panel are adapted from the work of van Kampen et al.[58] (C) Native mass spectrum (top) of 2 μM
IgG1-RGY measured in 150 mM NH4OAc pH 7.5, revealing two
distinct ion series for the monomer and hexamer, with ions originating
from intermediate oligomeric states observed at lower abundance. While
generally in good agreement with the other methods, hexamer abundances
measured by MS (bottom) were less consistent and higher than expected,
particularly at the lowest measured concentrations.
MP enables qualitative
and quantitative characterization of the
monomer–hexamer equilibrium of IgG1-RGY. (A) MP mass histogram
(top) of 2 μM IgG1-RGY in PBS jump-diluted to 10 nM showing
monomeric ((IgG1)1) and hexameric ((IgG1)6)
species. The mass of the hexamer was consistently measured about 70
kDa too high. The relative abundance of the IgG1-RGY hexamer was measured
over a dilution series spanning a concentration range of 0.1 to 8
μM (bottom), with error bars indicating the standard deviation
over three technical replicate measurements. (B) SEC-MALS chromatogram
of the same 2 μM IgG1-RGY sample (top) and the fractional abundance
of the hexamer as measured by SEC-UV over a dilution series (bottom),
revealing a similar monomer to hexamer ratio. The quantitative data
in the lower panel are adapted from the work of van Kampen et al.[58] (C) Native mass spectrum (top) of 2 μM
IgG1-RGY measured in 150 mM NH4OAc pH 7.5, revealing two
distinct ion series for the monomer and hexamer, with ions originating
from intermediate oligomeric states observed at lower abundance. While
generally in good agreement with the other methods, hexamer abundances
measured by MS (bottom) were less consistent and higher than expected,
particularly at the lowest measured concentrations.Quantifying the abundance of the hexamer in a dilution series
of
IgG1-RGY by all three techniques resulted in highly comparable data,
although with some subtle differences. In agreement with earlier studies,[57] longer MP recordings showed that hexamers re-equilibrate
only very slowly upon jump dilution (Figure S5), meaning that such MP experiments should directly provide an accurate
representation of the monomer–hexamer equilibrium. To characterize
the equilibrium of IgG1-RGY by MP, we measured a dilution series and
compared results to SEC and native MS. MP measurements proved to be
quite consistent between replicates and could be performed down to
nM range concentrations that cannot be assessed by native MS or SEC
(Figure A). While
SEC also proved to provide very robust data, hexamer abundances were
fractionally lower, potentially due to dissociation during separation.
Finally, although native MS performed well at higher concentrations,
variability increased at lower concentrations. Nonetheless, each of
the three techniques revealed that about half of the IgGs are incorporated
into hexamers at a concentration of 1 μM, consistent with previously
reported data.[34]
Characterization of Complement
Component C1q by MP Exposes Shortcomings
of Native MS and SEC-MALS
We next characterized complement
component C1q, the recognition complex of the classical complement
pathway, revealing striking differences between the three tested techniques.
C1q is a 464 kDa 18-membered protein complex that consists of three
pairs of triple helices (A2B2C2)
that are joined in a stem and end with six globular headpieces.[59,60] Although each headpiece has a low affinity for the Fc of IgGs,[61,62] clustering into oligomers allows for multivalent binding with increased
avidity,[63] making the six-armed C1q complex
highly compatible with IgM[64] and IgG hexamers.[8] As demonstrated previously, C1q behaves anomalously
in SEC, eluting at a short retention time that suggests a mass of
>1 MDa[34] (Figure S6A). Because of its open structure, C1q may have a much larger
hydrodynamic
radius than globular proteins of similar mass, producing a bias in
size-based separation. Concordantly, the coupled MALS system revealed
that this elution peak did correspond to the free C1q complex with
a mass of about 444 kDa. Curiously, the native mass spectra of C1q
were consistently marked by the presence of three main ion series
corresponding to two-armed, four-armed, and complete six-armed C1q
species (Figure S6B), decreasing in abundance
with the size of the complex. Partial C1q complexes did not display
asymmetrical charge partitioning characteristic for collisional dissociation
in the gas phase, suggesting that, similar to intermediate oligomers
of IgG-RGY, smaller complexes are formed by the electrospray process.
Lastly, MP measurements of C1q revealed predominantly particles of
the intact six-armed C1q complex (A6B6C6), with a minor contribution being made by a two-armed A2B2C2 complex (Figure S6C). We thus conclude that MP seems the most unbiased tool
for the analysis of C1q.
MP and CD-MS Tackle Mass Heterogeneity When
Analyzing Immune
Complexes of IgG-RGY Hexamers Bound to Highly Glycosylated Antigens
and C1q
Having demonstrated that IgG1-RGY hexamers and C1q
can be measured accurately by MP, we next sought to characterize immune
activation complexes involving antigen-bound IgG1-RGY hexamers and
C1q. When IgG1-RGY was first incubated with C1q, MP revealed the formation
of a 1.43 MDa complex corresponding to (IgG1)6:(C1q)1 (Figure A).
Similar to measurements of the IgG1-RGY hexamers, this mass is about
60 kDa higher than expected, possibly due to the non-globular shapes
of both complexes. Next, we assembled larger complement activation
complexes associated with highly heterogeneous sEGFR antigens, whereby
sEGFR was incubated with preformed (IgG1)6:(C1q)1 complexes (see Figure S7 for the analysis
of IgG1-RGY with sEGFR separately). MP revealed the presence of particles
with an average mass of 2.35 MDa, likely corresponding to (sEGFR)12:(IgG1)6:(C1q)1. However, as the resulting
peak was quite broad, we could not yet exclude the possibility of
other (co-occurring) stoichiometries. Similar to MP, SEC-MALS-UV-RI
measurement of IgG1-RGY incubated with C1q led to the detection of
(IgG1)6:(C1q)1 with a fairly adequate mass of
1.3 MDa (Figure B).
Although larger complexes of ∼1.9 MDa were observed upon addition
of sEGFR, here, the width of this peak and the accuracy of the mass
measurement was insufficient for determining the exact stoichiometry,
possibly also due to dissociation of sEGFR during separation. In agreement
with earlier reports,[34] native MS of IgG1-RGY
with C1q alone revealed the presence of (IgG1)6:(C1q)1 complexes, for which an accurate mass of 1361 kDa could be
established (Figure C). Contrary to MP, however, relatively more (IgG1)6 and
uniquely also (IgG1)6:(A4B4C4)1 complexes were observed, potentially dissociation
products of the full complex formed in the MS source region. While
larger but unresolved ions could be detected by native MS after the
addition of sEGFR, an inability to resolve charge states prevented
mass determination. For these samples, CD-MS enabled confident assignment
of a mass of 2.42 MDa, corresponding to the complete (sEGFR)12:(IgG1)6:(C1q)1 assembly (Figure D).
Figure 4
MP and CD-MS successfully
determine the mass and stoichiometry
of highly heterogeneous (sEGFR)12:(IgG1)6:C1q
immune complexes. (A) MP measurements of IgG1-RGY incubated with C1q
reveal the formation of (IgG1)6:(C1q)1 complexes,
with nearly all IgG hexamers occupied. When incubating C1q with pre-formed
(IgG1)6:(sEGFR)12, MP resolves a 2.35 MDa complex,
likely corresponding to (sEGFR)12:(IgG1)6:(C1q)1. (B) SEC-MALS-UV-RI analysis similarly reveals the formation
of ∼1.3 MDa (IgG1)6:(C1q)1 (with (IgG1)6 measured as the 0.76 MDa “protein” and C1q
as a 0.49 MDa “modifier”). When sEGFR was added, SEC-MALS-UV-RI
revealed the formation of larger complexes of around 1.9 MDa (1.5
MDa for (sEGFR)12:(IgG1)6 with a 0.41 MDa modifier).
(C) Measurement of the same samples by native MS reveals an accurate
mass for (IgG1)6:(C1q)1, but the technique struggles
with complexes involving sEGFR. Larger ion species were detected in
such experiments, but they could not be charge-resolved. (D) Single-particle
measurements of the distribution around m/z 21,000 by CD-MS (top) revealed a mass of 2.42 MDa (bottom)
corresponding to the expected mass of the full (sEGFR)12:(IgG1)6:(sEGFR)12 complex (bottom).
MP and CD-MS successfully
determine the mass and stoichiometry
of highly heterogeneous (sEGFR)12:(IgG1)6:C1q
immune complexes. (A) MP measurements of IgG1-RGY incubated with C1q
reveal the formation of (IgG1)6:(C1q)1 complexes,
with nearly all IgG hexamers occupied. When incubating C1q with pre-formed
(IgG1)6:(sEGFR)12, MP resolves a 2.35 MDa complex,
likely corresponding to (sEGFR)12:(IgG1)6:(C1q)1. (B) SEC-MALS-UV-RI analysis similarly reveals the formation
of ∼1.3 MDa (IgG1)6:(C1q)1 (with (IgG1)6 measured as the 0.76 MDa “protein” and C1q
as a 0.49 MDa “modifier”). When sEGFR was added, SEC-MALS-UV-RI
revealed the formation of larger complexes of around 1.9 MDa (1.5
MDa for (sEGFR)12:(IgG1)6 with a 0.41 MDa modifier).
(C) Measurement of the same samples by native MS reveals an accurate
mass for (IgG1)6:(C1q)1, but the technique struggles
with complexes involving sEGFR. Larger ion species were detected in
such experiments, but they could not be charge-resolved. (D) Single-particle
measurements of the distribution around m/z 21,000 by CD-MS (top) revealed a mass of 2.42 MDa (bottom)
corresponding to the expected mass of the full (sEGFR)12:(IgG1)6:(sEGFR)12 complex (bottom).
Comparing Advantages and Disadvantages Reveals
That MP, Native
MS, and SEC-MALS Are Highly Complementary
Reflecting on the
analyses performed in this study on a wide variety of systems, we
can compare the advantages and disadvantages of the approaches (see Table S1 for an overview of all the measured
masses and Table S3 for a qualitative comparison
between the techniques). Analytical SEC-MALS is the most established
method, providing reproducible, robust, and quantitative measurements.
While accurate for smaller proteins, masses of larger multicomponent
systems were underestimated, often by as much as 10–20%, making
SEC-MALS suboptimal for large complexes that may dissociate by shear
stress or dilution effects during column separation. MP represents
a relatively new approach that is a fast and comparatively straightforward
technique to measure more accurate masses in native-like buffering
solutions, enabling it to tackle multicomponent systems more effectively
than SEC-MALS. Still, mass resolution limits the technology mostly
to antibody–protein interactions, as mass differences induced
by small molecule or peptide binding are mostly too small to resolve.
Sample consumption is low, however, and unlike conventional native
MS, the technique is not hampered by extensive protein glycosylation.
While MP could reliably measure molecular masses and quantify strongly
interacting and slowly dissociating protein complexes, jump dilution
to nM range concentrations induced dissociation of weaker interactions.
Another consideration for using MP is that mass measurements may be
affected by the shape of the particles, as we consistently measured
a higher-than-expected mass for IgG1-RGY hexamers. Native MS provided
superior mass resolution and accuracy, resolving individual proteoglycoforms
in samples of moderate complexity and uniquely providing unambiguous
stoichiometries for protein complexes. After careful optimization
of instrumental parameters, reliable quantification could be achieved
when measuring at conventionally used concentrations (around 10–6 M). Certain protein complexes, however, proved to
be sensitive to dissociation induced in solution by the ESI process,
and performance of conventional native MS deteriorated when analyzing
extensively glycosylated proteins. However, when charge states could
not be resolved in the m/z domain,
this challenge was effectively overcome by CD-MS, providing unparalleled
mass accuracy for highly heterogeneous protein assemblies.
Conclusions
Here, we compared MP with native MS and SEC-MALS for the qualitative
and quantitative analysis of antibodies and related immune complexes.
Single-molecule and solution-based MP provides a relatively straightforward
way to assess protein complexes and can fill gaps between the two
other techniques in terms of mass accuracy and resolution, while also
being able to quantitatively assess strong and stable protein–protein
interactions. Among the main benefits of MP are its high sensitivity
(nM) and ability to measure in-solution using a wide variety of buffer
solutions. However, here we show that the mass resolving power of
MP is still somewhat limited and that some protein complexes may dissociate
due to dilution required for measurements. Among the main benefits
of native MS are its comparatively very high mass resolution, but
only when charge states can be resolved. When high heterogeneity makes
this impossible, single-molecule CD-MS can be used to infer charge
states in an alternative manner, providing lower resolving power than
conventional native MS, but generally still higher than MP. Native
MS is, however, a gas-phase technique that requires volatile buffers
and has a low tolerance for salts and detergents. SEC and SEC-MALS
are well-established technologies for mass assessment of proteins
within the biopharmaceutical laboratories, benefiting from their proven
robustness and ease of use. However, for assessing protein assemblies,
the resolving power of SEC-MALS is relatively low, and quantification
may be somewhat hampered by dilution of the sample during separation.
Overall, our data show that the tested approaches are highly complementary,
each having its unique preferred use cases. With robust commercial
instruments now becoming available, these newer techniques may become
more accepted. Furthermore, combining techniques, such as SEC coupled
to native MS, may overcome some of their weaknesses, while taking
advantage of their strengths.
Authors: Andrey Dyachenko; Guanbo Wang; Mike Belov; Alexander Makarov; Rob N de Jong; Ewald T J van den Bremer; Paul W H I Parren; Albert J R Heck Journal: Anal Chem Date: 2015-05-27 Impact factor: 6.986
Authors: Gavin Young; Nikolas Hundt; Daniel Cole; Adam Fineberg; Joanna Andrecka; Andrew Tyler; Anna Olerinyova; Ayla Ansari; Erik G Marklund; Miranda P Collier; Shane A Chandler; Olga Tkachenko; Joel Allen; Max Crispin; Neil Billington; Yasuharu Takagi; James R Sellers; Cédric Eichmann; Philipp Selenko; Lukas Frey; Roland Riek; Martin R Galpin; Weston B Struwe; Justin L P Benesch; Philipp Kukura Journal: Science Date: 2018-04-27 Impact factor: 47.728