Abstract
PURPOSE: This study examined the effect of different distances and curing times on the microhardness (VHN) of nanofilled resin-based composite (RBC) restorations polymerized with high-intensity LED LCUs.Entities:
Keywords: Composite; Efficiency; High intensity; LED; Microhardness; Polymerization
Year: 2021 PMID: 34938047 PMCID: PMC8665174 DOI: 10.1016/j.sdentj.2021.05.007
Source DB: PubMed Journal: Saudi Dent J ISSN: 1013-9052
| Resin-based composite restorative material | ||||||
|---|---|---|---|---|---|---|
| Composite (shade) | ||||||
| Tetric-N-Ceram | Nanofilled, Shade A2 | 10–20 s with high irradiance LCUs (>1000 mW/cm2) | Dimethacrylates, additives, catalysts, stabilizer sand pigments, barium glass, ytterbium trifluoride, mixed oxide and prepolymerized filler (prepolymers) (56% vol.) | Ivoclar-Vivadent, AG, 9494 Schaan/Liechtenstein | Z017Y9 | |
| Light curing units (LCUs) | ||||||
| (LCU) | ||||||
| Blue phase G2 | blue/violet polywave LED | 1570 mW/cm2 | 1390 mW/cm2 | 1210 mW/cm2 | Ivoclar Vivadent AG FL-9494 Schaan/Liechtenstein, Austria. | 222,788 |
| Elipar S10TM | Blue/Single-peak LCU | 1120 mW/cm2 | 1010 mW/cm2 | 890 mW/cm2 | 3 M ESPE, D-82229 Seefeld, Germany. | 939,123,009,611 |
| Planmeca Lumion | Blue/Single-peak LCU | 1220 mW/cm2 | 1070 mW/cm2 | 700 mW/cm2 | Mectron S.p.A., via Loreto 15′A, 16,042 Carasco-GE-Italy. | 309PG403 |
| Light curing radiometer | ||||||
| Name | ||||||
| Bluphase - meter | 4,5 DC, 3x Battery | Ivoclar Vivadent AG FL-9494 Schaan/Liechtenstein, Austria. | 00,735 | |||
| Microhardness tester | ||||||
| Name | ||||||
| NOVA | Weight: 37.5 kg | INNOVATEST, Europe BV, Borgharenweg 140 6222 AA, Maastricht, The Netherlands | 13,002,178,648 | |||
Fig. 1The histogram suggests an approximately normal statistical distribution of the VHN values.
Mean and standard deviation of Vickers hardness number VHN (s.d.) of the top/bottom surfaces for each LCU at different distances/curing times.
| Distance/curing time | LCU | VHN (top) kg/mm2 | VHN (bottom) kg/mm2 |
|---|---|---|---|
| 0 mm/20 sec | B | 61.26A (1.63) | 51.76D (0.93) |
| E | 55.81B (2.19) | 52.62C (1.43) | |
| P | 45.75E (1.06) | 36.30F (1.10) | |
| 2 mm/20 sec | B | 59.45A (1.01) | 54.39B (1.61) |
| E | 52.37C (1.12) | 45.13D (1.00) | |
| P | 42.46E (2.73) | 37.04F (1.35) | |
| 2 mm/40 sec | B | 56.85A (2.75) | 51.66C (0.93) |
| E | 55.73A (1.50) | 52.81B (1.44) | |
| P | 45.56D (1.46) | 36.46E (0.94) | |
| 4 mm/20 sec | B | 58.47A (0.75) | 48.08C (2.39) |
| E | 52.01B (0.85) | 43.08D (0.58) | |
| P | 39.70E (1.90) | 37.17F (2.02) | |
| 4 mm/60 sec | B | 61.11A (1.10) | 51.58C (0.83) |
| E | 54.99B (1.55) | 52.68C (1.60) | |
| P | 46.71D (1.55) | 39.90E (2.16) |
B: Bluephase GC; E: Elipar S10; P: Planmica Lumion. Means followed by the same letter indicate no significant difference at the 95% confidence level (p < 0.05).
Fig. 2The chart represents cluster interaction between all the variables (LCUs; curing time 20, 40, 60 sec; distances 0 mm, 2 mm, 4 mm).