Effect of distance on the power density from two light guides.

R B Price1, T Dérand, M Sedarous, P Andreou, R W Loney.   

Abstract

PURPOSE: This study determined the effect of distance on the power density from standard and Turbo light guides (Demetron/Kerr, Danbury, Connecticut).
MATERIALS AND METHODS: Power density was measured from 0 to 10 mm away from the tip of standard 8-mm curved light guides and 13/8-mm Turbo curved light guides. To determine the effect of distance on power density, a polynomial regression line was fitted. The Kolmogorov-Smirnov (K-S) statistic and the Wilcoxon rank sum (WR) tests were used to determine if there was a difference in the rate at which the power density decreased for the standard and Turbo light guides as the distance from the tip increased. Photographs of the light dispersion from each tip were also taken.
RESULTS: At 0 mm, the mean (+/- SD) power density from the two standard light guides was 743 +/- 6.1 mW/cm2 and from the four Turbo light guides was 1128 +/- 22.1 mW/cm2. As the distance from the tip of the light-guide tip increased, the power density decreased, but the rate of decrease was greater from the Turbo light guides than from the standard light guides. At 6 mm the power density from the standard light guides fell to 372 mW/cm2 (50% of the original value) and the power density from the Turbo light guides fell to 263 mW/cm2 (23% of the original value). Both the K-S statistic and the WR sum test indicated that the distribution of light intensities was significantly different from the two light guides (WR p-value = .0246, K-S p-value < .0001). The two estimated polynomials intersected at 3.66 mm, and the 95% prediction intervals intersected at about 2.8 and 4.8 mm. Therefore, beyond 5 mm away from the tip of the light guide, the standard light guides gave higher power density readings than the Turbo light guides. Photographs showed that the light dispersed at a wider angle from the Turbo light guides than from the standard light guide. CLINICAL SIGNIFICANCE: The design of the light guide of a light curing unit affects light dispersion, power density, and ultimately the dentist's ability to properly cure composite. For these reasons, manufacturers should report the power density at the tip of the light guide and 6 mm from the tip of the light guide, since significant differences exist between light guide designs.

Mesh:

Year:  2000        PMID: 14743527     DOI: 10.1111/j.1708-8240.2000.tb00241.x

Source DB:  PubMed          Journal:  J Esthet Dent        ISSN: 1040-1466


  14 in total

1.  In vitro comparison of mechanical properties and degree of cure of bulk fill composites.

Authors:  Pascal Czasch; Nicoleta Ilie
Journal:  Clin Oral Investig       Date:  2012-03-14       Impact factor: 3.573

2.  Effect of power density of curing unit, exposure duration, and light guide distance on composite depth of cure.

Authors:  Anders Lindberg; Anne Peutzfeldt; Jan W V van Dijken
Journal:  Clin Oral Investig       Date:  2005-04-07       Impact factor: 3.573

3.  Influence of light polymerization modes on degree of conversion and crosslink density of dental composites.

Authors:  Eduardo Moreira da Silva; Laiza Tatiana Poskus; José Guilherme Antunes Guimarães; Alexandre de Araújo Lima Barcellos; Carlos Eduardo Fellows
Journal:  J Mater Sci Mater Med       Date:  2007-08-01       Impact factor: 3.896

4.  Influence of irradiation distance on the mechanical performances of resin composites polymerized with high-irradiance light curing units.

Authors:  Soram Oh; Hyun Ju Kim; Hyun-Jung Kim; Sibel A Antonson; Sun-Young Kim
Journal:  Biomater Res       Date:  2022-05-20

5.  Power output from 12 brands of contemporary LED light-curing units measured using 2 brands of radiometers.

Authors:  Cristiane Maucoski; Richard B Price; Cesar A Arrais; Braden Sullivan
Journal:  PLoS One       Date:  2022-07-08       Impact factor: 3.752

6.  The light-curing unit: An essential piece of dental equipment.

Authors:  Richard B Price; Jack L Ferracane; Reinhard Hickel; Braden Sullivan
Journal:  Int Dent J       Date:  2020-07-21       Impact factor: 2.607

7.  Comparison of flowable bulk-fill and flowable resin-based composites: an in vitro analysis.

Authors:  Frank Engelhardt; Sebastian Hahnel; Verena Preis; Martin Rosentritt
Journal:  Clin Oral Investig       Date:  2016-01-09       Impact factor: 3.573

8.  Impact of light intensity on the polymerization kinetics and network structure of model hydrophobic and hydrophilic methacrylate based dental adhesive resin.

Authors:  Farhana Abedin; Qiang Ye; Kyle Camarda; Paulette Spencer
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2015-09-04       Impact factor: 3.368

9.  Five second photoactivation? A microhardness and marginal adaptation in vitro study in composite resin restorations.

Authors:  Carlos Alberto Kenji Shimokawa; Paula Mendes Acatauassú Carneiro; Tamile Rocha da Silva Lobo; Victor Elias Arana-Chavez; Michel Nicolau Youssef; Míriam Lacalle Turbino
Journal:  Int Dent J       Date:  2016-04-08       Impact factor: 2.607

10.  Influence of light-curing distance on degree of conversion and cytotoxicity of etch-and-rinse and self-etch adhesives.

Authors:  Florian J Wegehaupt; Nancy Lunghi; Georgios N Belibasakis; Thomas Attin
Journal:  BMC Oral Health       Date:  2016-07-07       Impact factor: 2.757

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.