Sun Kyung Jeon1, Jeong Min Lee1,2,3, Ijin Joo1, Jeong Hee Yoon1.
Abstract
PURPOSE: This study aimed to assess the inter-platform reproducibility of ultrasound attenuation examination in patients with nonalcoholic fatty liver disease (NAFLD).Entities:
Keywords: Fatty liver; Liver; Nonalcoholic fatty liver disease; Ultrasonography
Year: 2021 PMID: 34933319 PMCID: PMC8942738 DOI: 10.14366/usg.21167
Source DB: PubMed Journal: Ultrasonography ISSN: 2288-5919
Fig. 1.Measurement of attenuation coefficients (AC) using three ultrasound attenuation examination platforms.
AC values were measured in each patient using Attenuation Imaging (Canon Medical System) (A), Tissue Attenuation Imaging (Samsung Medison) (B), and Ultrasound-Guided Attenuation Parameter (GE Healthcare) (C), respectively.
Patient characteristics
| Variable | Value (n=46) |
|---|---|
| Age (year) | 52.3±12.4 (24-80) |
| Sex | |
| Male | 23 (50.0) |
| Female | 23 (50.0) |
| BMI (kg/m2) | 26.2±3.0 (19.0-34.6) |
| Skin-to-liver capsule distance (mm) | 19.9±4.1 (11-32) |
| Aspartate aminotransferase (IU/L) | 31.2±17.1 (14-80) |
| Alanine aminotransferase (IU/L) | 42.1±31.4 (8-145) |
| Visual hepatic steatosis grade | |
| No steatosis | 7 (15.2) |
| Mild steatosis | 14 (30.4) |
| Moderate steatosis | 17 (37.0) |
| Severe steatosis | 8 (17.4) |
| AC measured using ultrasound attenuation examination | |
| ATI | 0.75±0.12 (0.50-0.98) |
| TAI | 0.80±0.11 (0.61-1.08) |
| UGAP | 0.74±0.09 (0.52-0.93) |
Values are presented as mean±standard deviation (range) or number (%).
BMI, body mass index; AC, attenuation coefficient; ATI, Attenuation Imaging (Canon Medical System); TAI, Tissue Attenuation Imaging (Samsung Medicine); UGAP, Ultrasound-Guided Attenuation Parameter (GE Healthcare).
Mean AC values obtained using different ultrasound attenuation examination platforms
| AC value (dB/cm/MHz) | P-value[ | |||||
|---|---|---|---|---|---|---|
| Mean±SD | Range | ATI vs. TAI | ATI vs. UGAP | TAI vs. UGAP | ||
| ATI | 0.75±0.12 | 0.50-0.98 | <0.001 | <0.001 | >0.99 | <0.001 |
| TAI | 0.80±0.11 | 0.61-1.08 | ||||
| UGAP | 0.74±0.09 | 0.52-0.93 | ||||
AC, attenuation coefficient; SD, standard deviation; ATI, Attenuation Imaging (Canon Medical System); TAI, Tissue Attenuation Imaging (Samsung Medicine); UGAP, UltrasoundGuided Attenuation Parameter (GE Healthcare).
P-values were calculated using repeated-measures analysis of variance with Bonferroni post-hoc analysis.
Fig. 2.Distribution of mean attenuation coefficients obtained using three ultrasound attenuation examination platforms.
AC, attenuation coefficient; ATI, Attenuation Imaging; TAI, Tissue Attenuation Imaging; UGAP, Ultrasound-Guided Attenuation Parameter.
Inter-platform reproducibility of AC values between three ultrasound attenuation examination platforms
| Mean bias (dB/cm/MHz) | BALA (dB/cm/MHz) | ICC | CV (%) | Pearson r | |
|---|---|---|---|---|---|
| ATI vs. TAI | -0.05 (-0.08 to -0.03) | -0.22 to 0.11 | 0.797 (0.495 to 0.904) | 9.47 (7.91 to 10.80) | 0.733 (0.562 to 0.844) |
| ATI vs. UGAP | 0.01 (-0.02 to 0.03) | -0.17 to 0.18 | 0.794 (0.628 to 0.886) | 8.27 (6.79 to 9.54) | 0.675 (0.479 to 0.807) |
| TAI vs. UGAP | 0.06 (0.04 to 0.08) | -0.08 to 0.20 | 0.790 (0.273 to 0.915) | 8.23 (6.26 to 9.82) | 0.778 (0.631 to 0.872) |
Numbers in parentheses are 95% confidence intervals.
AC, attenuation coefficient; BALA, Bland-Altman 95% limits of agreement; ICC, intraclass correlation coefficient; CV, coefficient of variation; ATI, Attenuation Imaging; TAI, Tissue Attenuation Imaging; UGAP, Ultrasound-Guided Attenuation Parameter.
Fig. 3.Bland-Altman plots of the ultrasound attenuation examination platforms.
The Bland-Altman plots demonstrate differences in attenuation coefficient (AC) values between Attenuation Imaging (ATI) and Tissue Attenuation Imaging (TAI) (A), ATI and Ultrasound-Guided Attenuation Parameter (UGAP) (B), and TAI and UGAP (C). The solid blue line in the middle represents the mean AC values obtained from each pair of the three platforms, and the dotted brown lines indicate ±1.96 standard deviations (SDs), with the associated 95% confidence intervals indicated by the green bars.
Inter-session pairwise repeatability of AC values obtained using different ultrasound attenuation examination platforms
| AC value (dB/cm/MHz) | P-value[ | Mean bias | BALA | ICC | CV (%) | Pearson r | ||
|---|---|---|---|---|---|---|---|---|
| 1st session | 2nd session | |||||||
| ATI | 0.75±0.12 | 0.74±0.12 | 0.085 | 0.00 | -0.03 to 0.03 | 0.962 (0.931 to 0.979) | 4.9 (0.9 to 7.0) | 0.931 (0.878 to 0.961) |
| TAI | 0.80±0.11 | 0.81±0.11 | 0.126 | -0.01 | -0.10 to 0.08 | 0.957 (0.922 to 0.976) | 3.9 (1.2 to 5.3) | 0.920 (0.858 to 0.955) |
| UGAP | 0.74±0.09 | 0.75±0.09 | 0.385 | 0.00 | -0.07 to 0.06 | 0.962 (0.931 to 0.979) | 3.4 (2.6 to 4.0) | 0.926 (0.869 to 0.959) |
Numbers in parentheses are 95% confidence intervals.
AC, attenuation coefficient; BALA, Bland-Altman limits of agreement; ICC, intraclass correlation coefficient; CV, coefficient of variation; ATI, Attenuation Imaging; TAI, Tissue Attenuation Imaging; UGAP, Ultrasound-Guided Attenuation Parameter.
P-values were calculated using the paired Student t-test (two-tailed).