Hyuntae Cho1, Seung-Ro Lee2, Yunju Baek3.
Abstract
Anemia is a condition in which red blood cells or the hemoglobin in the blood is lower than in healthy people. Red blood cells transport and supply oxygen needed to various organs in the human body. Anemia is caused by hypoxemia due to the lack of red blood cells and causes other serious health problems, such as heart problems, pregnancy complications, severe fatigue, or death. There are many causes of anemia, and it can be diagnosed by measuring hematocrit or hemoglobin levels in the blood. Even though there are various diagnostic devices on the market, these devices are inconvenient because their systems are bulky, heavy, expensive, or inaccurate. This study proposed a new anemia diagnostic system based on the impedance measurement of red blood cells. The proposed system consists of a test strip that collects a blood sample from the finger and a hemoglobin meter that measures the impedance of the blood and converts it into the concentration of hemoglobin. The proposed test strip that does not contain enzymes or reagents was designed in accordance with class 1 approval by the Food and Drug Administration (FDA). The hemoglobin meter was designed to include a hardware block, an algorithm block and a calibration block through empirical work. We also compared it to reference impedance to prove the accuracy of the hemoglobin meter. The experimental results with human blood indicated the superiority of the anemia diagnostic system. As a result, the overall standard deviation of impedance measurements was less than 1%, and the coefficient of variance of the proposed system was 1.7%, which was better than that of other commercial systems.Entities:
Keywords: anemia-measuring method; blood diagnostics; hematocrit; hemoglobin; in vitro system; red blood cells; test strip
Mesh:
Substances:
Year: 2021 PMID: 34884044 PMCID: PMC8659443 DOI: 10.3390/s21238043
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Criteria for low blood hemoglobin concentration.
| Male Adult | <13 g/dL |
| Female Adult | <12 g/dL |
| Age 6~16 Years | <12 g/dL |
| Age 6 Months~6 Years | <11 g/dL |
| Pregnancy | <11 g/dL |
Figure 1Classification of automatic hemoglobin meters for POCT.
Figure 2The designed strips: (a) structure of the strip, (b) electrodes of the test strip, (c) interface of the test strip, and (d) appearance of the test strip.
Figure 3The manufactured test strip: (a) electrode structure and (b) appearance.
Figure 4Hemoglobin meter: (a) block diagram and (b) prototypes (50 × 75 mm).
Figure 5Impedance measurement circuit.
Figure 6Flow chart of the system operation.
Figure 7Impedance curve according to hemoglobin.
Impedance measurement accuracy.
|
| 680 | 0.98 k | 1.96 k | 20.18 k | 32.8 k | 50.58 k |
|
| 695.3 | 0.998 k | 1.99 k | 20.34 k | 32.8 k | 49.91 k |
|
| 3.91 | 5.44 | 10.6 | 137.85 | 211.75 | 454.74 |
|
| 0.58 | 0.55 | 0.54 | 0.68 | 0.65 | 0.9 |
Figure 8Experimental results using human blood: (a) average concentration and (b) CV.