Nuno M Reis1, Sarah H Needs2, Sophie M Jegouic2,3, Kirandeep K Gill1, Sirintra Sirivisoot4, Scott Howard1, Jack Kempe1, Shaan Bola1, Kareem Al-Hakeem1, Ian M Jones3, Tanapan Prommool5, Prasit Luangaram5, Panisadee Avirutnan4,5,6, Chunya Puttikhunt4,5,6, Alexander D Edwards2. 1. Department of Chemical Engineering and Centre for Biosensors, Biodevices and Bioelectronics (C3Bio), University of Bath, Claverton Down, Bath BA2 7AY, United Kingdom. 2. Reading School of Pharmacy, University of Reading, Whiteknights Campus, Reading, RG6 6AD United Kingdom. 3. School of Biological Sciences, University of Reading, Whiteknights Campus, Reading, RG6 6AJ, United Kingdom. 4. Dengue Hemorrhagic Fever Research Unit, Office for Research and Development, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, 10700, Thailand. 5. Molecular Biology of Dengue and Flaviviruses Research Team, Medical Molecular Biotechnology Research Group, National Center for Genetic Engineering and Biotechnology, National Science and Technology Development Agency, Pathum Thani, 73170, Thailand. 6. Siriraj Center of Research Excellence in Dengue and Emerging Pathogens, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, 10700, Thailand.
Abstract
A range of biosensing techniques including immunoassays are routinely used for quantitation of analytes in biological samples and available in a range of formats, from centralized lab testing (e.g., microplate enzyme-linked immunosorbent assay (ELISA)) to automated point-of-care (POC) and lateral flow immunochromatographic tests. High analytical performance is intrinsically linked to the use of a sequence of reagent and washing steps, yet this is extremely challenging to deliver at the POC without a high level of fluidic control involving, e.g., automation, fluidic pumping, or manual fluid handling/pipetting. Here we introduce a microfluidic siphon concept that conceptualizes a multistep ″dipstick″ for quantitative, enzymatically amplified immunoassays using a strip of microporous or microbored material. We demonstrated that gravity-driven siphon flow can be realized in single-bore glass capillaries, a multibored microcapillary film, and a glass fiber porous membrane. In contrast to other POC devices proposed to date, the operation of the siphon is only dependent on the hydrostatic liquid pressure (gravity) and not capillary forces, and the unique stepwise approach to the delivery of the sample and immunoassay reagents results in zero dead volume in the device, no reagent overlap or carryover, and full start/stop fluid control. We demonstrated applications of a 10-bore microfluidic siphon as a portable ELISA system without compromised quantitative capabilities in two global diagnostic applications: (1) a four-plex sandwich ELISA for rapid smartphone dengue serotype identification by serotype-specific dengue virus NS1 antigen detection, relevant for acute dengue fever diagnosis, and (2) quantitation of anti-SARS-CoV-2 IgG and IgM titers in spiked serum samples. Diagnostic siphons provide the opportunity for high-performance immunoassay testing outside sophisticated laboratories, meeting the rapidly changing global clinical and public health needs.
A range of biosensing techniques including immunoassays are routinely used for quantitation of analytes in biological samples and available in a range of formats, from centralized lab testing (e.g., microplate enzyme-linked immunosorbent assay (ELISA)) to automated point-of-care (POC) and lateral flow immunochromatographic tests. High analytical performance is intrinsically linked to the use of a sequence of reagent and washing steps, yet this is extremely challenging to deliver at the POC without a high level of fluidic control involving, e.g., automation, fluidic pumping, or manual fluid handling/pipetting. Here we introduce a microfluidic siphon concept that conceptualizes a multistep ″dipstick″ for quantitative, enzymatically amplified immunoassays using a strip of microporous or microbored material. We demonstrated that gravity-driven siphon flow can be realized in single-bore glass capillaries, a multibored microcapillary film, and a glass fiber porous membrane. In contrast to other POC devices proposed to date, the operation of the siphon is only dependent on the hydrostatic liquid pressure (gravity) and not capillary forces, and the unique stepwise approach to the delivery of the sample and immunoassay reagents results in zero dead volume in the device, no reagent overlap or carryover, and full start/stop fluid control. We demonstrated applications of a 10-bore microfluidic siphon as a portable ELISA system without compromised quantitative capabilities in two global diagnostic applications: (1) a four-plex sandwich ELISA for rapid smartphone dengue serotype identification by serotype-specific dengue virus NS1 antigen detection, relevant for acute dengue fever diagnosis, and (2) quantitation of anti-SARS-CoV-2 IgG and IgM titers in spiked serum samples. Diagnostic siphons provide the opportunity for high-performance immunoassay testing outside sophisticated laboratories, meeting the rapidly changing global clinical and public health needs.
The COVID-19
pandemic exposed
a major gap in the point-of-care (POC) high-performance immunoassay
detection of antigenic proteins, antibodies, and inflammatory and
other important biomarkers, as explained in WHO’s coordinated
global research roadmap.[1] Laboratory immunoassays
such as microplate enzyme-linked immunosorbent assay (ELISA) achieve
analytical performance through multiple steps (reagent, washing, incubation,
and enzymatic amplification). Lateral flow immunoassays (LFAs)[2] perform immunochromatography through a heterogeneous
porous nitrocellulose substrate, offering low cost and simplicity
fulfilling all of WHO’s ASSURED[3] requirements, but cannot match the analytical performance (sensitivity
and quantitation) of multistep lab immunoassays such as microplate
ELISA and are therefore not fit for many modern clinical needs. Multistep
bioassays have been successfully automated yet rely on a ″wicking
flow″ driven by capillary forces in substrates such as paper
or nitrocellulose or a multilayer microfluidic device, resulting in
uneven flow,[2] limited control of incubation
times, complex device design, and crossover of reagents, all compromising
analytical performance. There remains a need for a portable multistep
analytical approach or device able to deliver a sequence of immunoassay
reagents and washes without access to fluid handling equipment (pipettors
and pumps), approaching the usability of LFAs, and (ideally) capable
of multiplexing a panel of measurements.A range of power-free
microfluidic[4−12] devices using capillary action or capillary forces (unavoidable
at small scales) to manipulate liquids have evolved from the LFA’s
nitrocellulose membrane. These include paper-based capillary microfluidic
and passive capillary microfluidic devices extensively reviewed elsewhere.[13] Capillary phenomena[14] have been combined with, e.g., electrochemical[15] and electrostatic[16,17] effects or more advanced
designs[18] to deliver features of fluid
control essential for a multistep high-performance microfluidic immunoassay,
including stop flow,[19] prevention of bubble
trapping,[20] retention valves,[21,22] and preprogrammed sequential delivery of reagents.[23−27] However, the dependence on capillary forces constrains all these
devices to the same limitations as rapid LFA tests, with the relationship
between flow rate and flow resistance changing throughout the liquid
wicking of the material. In some cases, a larger waste pad is readily
used as a capillary ″pump″,[28] governed by capillary forces and limited to small volumes of reagents
and the sample. The flow rate is a key parameter as it governs the
test time and, often, the analytical performance of the assay.[29,30] As capillary forces are very sensitive to capillary diameter, small
differences in porosity result in uneven superficial fluid velocities,
which are characteristic of nitrocellulose membrane tests. Sequential
fluid delivery, essential for autonomous multistep immunoassays, has
been achieved through the use of, e.g., retention, burst or capillary
valves (with fluids delivered according to their order of capillary
pressure),[21] matrix plugs (with sequential
drainage of fluid based on the properties of the porous matrix),[25] dissolvable barriers/vents (with the thickness
of the hydrogel layer determining the sequence of drainage of each
fluid),[27] and dissolvable sugars (with
the fluid sequence determined by the sugar concentration in each channel).[24] Such strategies resulted in a significant overlap
of reagents and can make the operation of devices challenging in terms
of using small wash volumes or delivering good signal-to-noise ratios
in enzyme-amplified immunoassays. With low analyte concentrations,
even low residual levels of unbound enzyme/conjugate are sufficient
to yield a significant background, compromising the analytical performance
by raising the limit of detection. Microfluidic or microwell approaches
fed manually into a ″funnel″ face similar challenges[31] and require large wash volumes to remove all
unbound enzyme from dead volume in the inlet well.Microscale
siphons have been mentioned in the literature for two
very distinct applications: control of cell culture media in a tissue
culture microwell/microchannel[32,33] and control of fluid
movement in centrifugal microfluidics.[34−36] With centrifugal microfluidic
devices, a ″siphon″ is incorporated horizontally and
fluid flow is triggered by raising the centrifugal force to overcome
the siphon crest (or neck). The system presented by Ozaki et al.[32] for cell culture used pumped recirculation with
fluid flow down a porous paper matrix driven by gravity, without overcoming
a raised neck at higher hydraulic height and therefore not using the
siphon effect. The microfluidic cell culture channel proposed by Jeong
et al.[33] required manual priming by a syringe
and the cotton-yarn ″siphon″ pump operated at a hydraulic
pressure lower than the cell culture feeding reservoir. Therefore,
there are no previous reports of a vertical siphon being exploited
to automate microfluidic multistep bioassays specifically enzyme-amplified
immunoassays or ELISA.We present a novel microfluidic siphon
concept consisting of a
strip of porous substrate (porous membrane, microcapillary, or microbored
material) configured with a ″swan beak″ (see side view
in Figure A). A homogeneous
flow of sequential reagents is driven exclusively by hydrostatic pressure.[37,38] The microfluidic siphon presents a set of distinctive features suitable
for high-performance, quantitative, and multiplexed bioassays, including
enzyme immunoassays/ELISA, differentiating from all power-free microfluidic
approaches and sequential fluid delivery strategies previously proposed.
We illustrated here the fluidic performance of a range of siphon devices,
in particular, a multibore microcapillary film, single-bore glass
microcapillaries, and a glass fiber porous matrix. In addition, we
demonstrated the application as ELISA quantitation in patient samples
using a highly sensitive fluorescence substrate, AttoPhos, cleaved
by alkaline phosphatase (AP) to produce bright green fluorescence
when an analyte is enzymatically detected (Figure S1).
Figure 1
Operation of microfluidic siphons for multistep bioassays. (A)
Liquid sample, reagents, and washing buffer (labeled R1–R6,
shown in different colors) are automatically aspirated upon contact
with the tip of the siphon with a hydrostatic liquid head, H (driving force for operation). The fluid goes through
a neck at liquid high above the inlet before being discharged into
a waste well/tray. (B) Microfluidic siphon is able to self-prime and
allows start–stop operation with a sequence of reagents without
emptying; insets show key steps for a sandwich ELISA with optical/fluorescence
interrogation. (C) Photographs of a mass-manufacturable plastic siphon
cassette that can be operated manually or using a robotic arm, loaded
with an 89 mm long MCF strip and a large shallow waste reservoir (shown
in side view), keeping H constant at 45 mm. (D) 3D
printed cartridges loaded with a 47 mm long MCF strip (H = 25 mm) for single sample analysis (top) or stacked for up to 12
samples in parallel (bottom).
Operation of microfluidic siphons for multistep bioassays. (A)
Liquid sample, reagents, and washing buffer (labeled R1–R6,
shown in different colors) are automatically aspirated upon contact
with the tip of the siphon with a hydrostatic liquid head, H (driving force for operation). The fluid goes through
a neck at liquid high above the inlet before being discharged into
a waste well/tray. (B) Microfluidic siphon is able to self-prime and
allows start–stop operation with a sequence of reagents without
emptying; insets show key steps for a sandwich ELISA with optical/fluorescence
interrogation. (C) Photographs of a mass-manufacturable plastic siphon
cassette that can be operated manually or using a robotic arm, loaded
with an 89 mm long MCF strip and a large shallow waste reservoir (shown
in side view), keeping H constant at 45 mm. (D) 3D
printed cartridges loaded with a 47 mm long MCF strip (H = 25 mm) for single sample analysis (top) or stacked for up to 12
samples in parallel (bottom).
Experimental Section
Microcapillary Film (MCF)
Strips
We used a 10-bore
MCF, with a mean hydraulic diameter (dc) of 204 ± 13.8 μm, mass-manufactured from Teflon FEP
material through a melt-extrusion process,[39] with a cross section and dc for each
individual microcapillary shown in Figure S3A,B, respectively. The procedures for coating the strips with a high-molecular
polyvinyl alcohol (PVOH) solution, design, and operation of MCF-based
stacking siphons are fully described in the Supporting Information.
Glass Capillary Siphons
Hydrophilic,
142.4 mm long
single-bore circular glass capillaries with measured OD 1.0 mm and
ID 578 μm sourced from World Precision Instruments (Sarasota,
Florida, US). Capillary siphons were produced having a total length, L constant at 142.4 mm and net hydrostatic liquid height, H (Figure A) and H/L ratio varied by bending
the glass capillary at different points along the length with a Bunsen
burner.
Glass Fiber Porous Matrix Siphons
A grade 8964 lateral
flow conjugate pad (Ahlstrom, Pont-Evêque, France) was cut
into 7 mm wide and 80 mm long strips, and ends were placed in microwells
with the inlet 50 mm (vertical) higher than the outlet, i.e., H = 50 mm.
Dengue ELISA NS1 Viral Antigen Quantitation
and Serotype Identification
MCF strips (47 mm long) were
prepared as follows: a 1.4 m length
of MCF was coated with a panel of serotype-specific capture antibodies
against NS1 produced in-house (Table S1) at 60 μg/mL for DENV1 and 40 μg/mL for DENV2–4
in duplicate capillaries; these concentrations were selected as sufficient
to yield a stable half-monolayer of antibody on the surface of Teflon
FEP microcapillaries as characterized previously.[40] A positive control capillary was coated with 1 μg/mL
recombinant DENV2 NS1, and a negative capillary was filled with PBS
alone. These multiplexed MCF strips were incubated at RT for 1 h,
and the antibody solution replaced with 0.1 mg/mL PVOH (MW 146,000–186,000
g/mol, Sigma Aldrich, UK); the strips were coated for a further 2
h and finally blocked by filling up with 4% w/v BSA (GE Healthcare
Life Sciences, UK) and incubated overnight at 4 °C before cutting
into 47 mm test strips that were fitted into siphon holders. All assay
steps used a 60 μL reagent, sample, or wash volume; unless otherwise
stated, wash steps were left for 30 s to allow the full discharge
of the wash volume. Samples were either neat patient plasma or uninfected
control plasma spiked with recombinant DENV NS1. Neat plasma was incubated
for 10 min followed by wash with 0.1% w/v Tween 20 (Sigma Aldrich,
Dorset UK) in PBS (PBS-T). Biotinylated pan-serotype DENV NS1 antibody
manufactured in-house was added at 10 μg/mL, diluted in 4% BSA
and 10% rat serum (Sigma Aldrich, UK), and incubated for 5 min followed
by a PBS-T wash. Streptavidin-AP conjugate (Sigma Aldrich, Dorset,
UK) at 1:1000 dilution in 4% BSA was incubated for 5 min followed
by three sequential washes with PBS-T and then addition of the fluorescence
substrate.
MCF test strips were coated in all 10 capillaries
with 40 μg/mL
of SARS-CoV-2 S1 protein produced in-house by incubating for 1 h at
RT. The antigen solution was replaced with PVOH 0.1 mg/mL for 2 h
at RT and then a blocking buffer (Superblock + 10% FBS + 5% GS) and
subsequently stored at 4 °C in the blocking buffer prior to cutting
47 mm long test strips fitted into siphon holders. The seroreactivity
against the ″S1″ antigen was measured using a conventional
endpoint titer protocol, with plasma derived from whole blood volunteer
samples stabilized in 4% sodium citrate (Sigma Aldrich, UK) serially
diluted in the Seablock blocking buffer (ThermoFisher Scientific,
UK). The patient samples were incubated for 10–20 min in the
siphon devices followed by three washes with PBS-T totaling 5 min.
The secondary antibody, anti-human IgG conjugated with alkaline phosphatase
(ThermoFisher Scientific UK) at 1 μg/mL in the Seablock blocking
buffer, was incubated for a total of 10–20 min before being
washed three times with PBS-T allowing a 5 min total wash duration
and then addition of fluorescence substrate.
Imaging of the Fluorescent
Immunoassay Siphon Strips
For imaging both SARS-CoV-2 and
DENV immunoassay strips, the AttoPhos
AP Fluorescent Substrate System (S1000, Promega, UK) was added as
the final substrate to the siphons and the fluorescent image was captured
using both a Canon S120 digital camera and an iPhone 6S smartphone
after 15 min incubation for DENV immunoassay or 5 min for SARS-CoV-2
immunoassay. Samples were illuminated using an open-source blue LED
transilluminator, and images were taken through an amber acrylic emission
filter (IOrodeo, Pasadena, CA, USA) with 12–16 test strips
imaged simultaneously in a single image alongside the reference strip,
representing 160 measurements per photo. The fluorescence signal for
each individual capillary was determined from the peak green channel
pixel intensity using Image J (NIH, USA). The relative fluorescent
intensity was calculated relative to a reference MCF strip loaded
with 5 μM (DENV NS1) or 2 μM (anti-SARS-CoV2) fluorescein
to normalize fluorescence signal as reported previously.[37,38]
Anti-SARS-CoV2 ELISA Titer Antibody Quantitation in Microtiter
Plates
For the COVID-19 assay in microtiter plates, 10 μg/mL
of SARS-CoV2 S1 antigen diluted in carbonate buffer (Sigma-Aldrich,
UK) was coated onto the plate (Nunc Immunosorp) in duplicates and
incubated overnight at 4 °C. The plate was washed three times
with PBS-Tween 20 and blocked three times with Superblock. Patient
samples were diluted as described above and incubated for 1 h at RT.
Following four washes with PBS-Tween 20, the secondary antibody was
diluted as described above and incubated for 1 h at RT. Finally, following
four washes with PBS-T, the chromogenic substrate 3,3′,5,5′-tetramethylbenzidine
(TMB) (Europa Bioproducts, UK) was added and color development was
stopped by the addition of an equal volume of 0.25 M sulfuric acid.
Absorbance was read at 440 nm against a reference read at 700 nm.
For each set of duplicates, the average and the standard deviation
were calculated.
Similarly to the COVID-19 assay in MCF, various
levels of recombinant
human IgG and IgM anti-COVID-19 [CR3022] antibodies (Absolute Antibodies,
UK) were spiked into the control human serum. The recombinant antibodies
were spiked into the prepandemic human serum diluted 1:20 in Seablock
at 50 μg/mL (high IgG/M) or 5 μg/mL (low IgG/M). A ″no
IgG/M″ control consisted of the control prepandemic human serum
diluted 1:20 in Seablock only. As previously, the samples were serially
diluted in the Seablock blocking buffer before being incubated in
the MCF strips.
Results and Discussion
Microfluidic Siphon Concept
We have delivered the vision
of a pumping-free, microfluidic siphon able to sequentially aspirate
a sample plus multiple reagents, separated by incubation steps and
washings, from a preloaded reagent tray or microwell strip, located
at a vertical, z-position (i.e., also called hydrostatic
head) higher than the receiving waste well, as depicted in side view
diagrams in Figure A,B. The microfluidic siphons require a very minor level of intervention
from the operator in moving/depositing reagents in trays, which can
be done by an unskilled person in contrast to, e.g., a lab-based ELISA.
The full procedure for the operation of a four-channel MCF siphon
cassette is shown in the film footages provided in the Supporting Information. As the flow is entirely
governed by the net hydrostatic liquid head (H),
the siphon can virtually operate continuously with any porous (hydrophobic
and hydrophilic) material of any length L (we have
successfully tested siphons with H and L values as large as 1 m; results are not shown). However, it is essential
that the porous strip is hydrophilic to self-prime the siphon and
initiate flow without input from the operator, just as with a capillaric
microfluidic device. This first priming stage is essential to raise
the liquid height of the sample or first reagent from the tray up
and over the neck of the siphon, which happens by capillary action.We incorporated the siphon in mass-manufacturable plastic cassettes
similar to LFA cases (Figure C,D). In contrast to LFAs and capillaric circuits, the microfluidic
siphon operates vertically, being solely powered by gravity once the
first liquid overcomes the neck. Note that for devices operated through
capillary action including LFAs, there is no interest in operating
those vertically as capillary forces are quickly offset by fluid resistance
and gravity, impairing the performance of the device. The height of
the siphon neck is minimized to reduce material waste and flow resistance
but must be sufficient to be immersed into the well or droplets in
the reagent tray. The neck height should also not exceed the equilibrium
height given the Laplace pressure of the hydrophilic channel, which
for polyvinyl alcohol (PVOH) coated 10-bore, Teflon FEP melt-extruded
MicroCapillary Film (MCF) strips[41] used
here we measured as 58 ± 3.0 mm (Supporting Information, Table S2 and Figure S3).
Fluidic Measurement
Figure A shows the
sequential discharge of the dye
and buffer from a plastic tray (top image) through a siphon using
MCF strips with L = 60 mm.[41] A collage (Figure A bottom) illustrates the dye color at the strip center over time,
confirming the distinct sequential delivery of the dye and wash, with
no detectable fluid overlap. Washing efficiency was quantified by
breakthrough curves with a dye (Figure A middle and Supporting Information) showing siphon operation with only 100 μL of fluid, with
normalized concentration rapidly rising from 0 to 1 (for dye) and
falling from 1 to 0 (washing). Further imaging of colored and fluorescent
dye reagents using the siphon cassette shown in Figure C confirmed that reagent delivery can be
automated (see Supplementary Methods in the Supporting Information). The dye–wash sequences in Figure S2 show that reagents are rapidly equilibrated
in individual microcapillaries through simple molecular diffusion.
Figure 2
Fluidic
operation of 10-bore microfluidic siphons. (A) Sequential
flow of reagents from 100 μL droplets through a 10-bore MCF
siphon (H = 30 mm, L = 60 mm) into
a transparent 1 mL plastic cuvette preloaded with 300 μL of
buffer stained with blue food dye (to facilitate image analysis and
processing). A transparent washing buffer droplet approaches (0 s)
the siphon inlet, then the tray is moved to a dark pink droplet (50
s), and the procedure was repeated for five further steps. The horizontal
montage (middle) shows an RGB time sequence of the siphon imaged at
20 mm from the outlet of the strip, with the sequence of liquid ″reagents″
passing through the 10 individual parallel microcapillaries along
time. Breakthrough curves (bottom) with the respective normalized
concentration (C/C0)
of dye and washing showing that the flow of reagents is very reproducible.
Note that the total time shown for each reagent in the x axis corresponds to the time taken to fully discharge the volume
of the droplet plus the time allowed for stop-flow ″incubation″
on each step. The siphon shown was designed with a linear (axial)
neck distance of 30.0 mm curved at 180° with a diameter of curvature
equal to ∼19.1 mm. (B) Liquid volume discharged through a 10-bore
MCF siphon (H = 40 mm, L = 70 mm)
measured over time (top) by imaging the volume in the waste cuvette,
and the first derivative (bottom) yielding the instantaneous discharge
flow rate (Q) plotted against time. The vertical
dashed lines represent the time at which a fresh 100 μL droplet
stained with 0.05% (w/w) indigo carmine was moved to interface the
inlet of the siphon (a total of 75 s allowed for each reagent/washing).
Fluidic
operation of 10-bore microfluidic siphons. (A) Sequential
flow of reagents from 100 μL droplets through a 10-bore MCF
siphon (H = 30 mm, L = 60 mm) into
a transparent 1 mL plastic cuvette preloaded with 300 μL of
buffer stained with blue food dye (to facilitate image analysis and
processing). A transparent washing buffer droplet approaches (0 s)
the siphon inlet, then the tray is moved to a dark pink droplet (50
s), and the procedure was repeated for five further steps. The horizontal
montage (middle) shows an RGB time sequence of the siphon imaged at
20 mm from the outlet of the strip, with the sequence of liquid ″reagents″
passing through the 10 individual parallel microcapillaries along
time. Breakthrough curves (bottom) with the respective normalized
concentration (C/C0)
of dye and washing showing that the flow of reagents is very reproducible.
Note that the total time shown for each reagent in the x axis corresponds to the time taken to fully discharge the volume
of the droplet plus the time allowed for stop-flow ″incubation″
on each step. The siphon shown was designed with a linear (axial)
neck distance of 30.0 mm curved at 180° with a diameter of curvature
equal to ∼19.1 mm. (B) Liquid volume discharged through a 10-bore
MCF siphon (H = 40 mm, L = 70 mm)
measured over time (top) by imaging the volume in the waste cuvette,
and the first derivative (bottom) yielding the instantaneous discharge
flow rate (Q) plotted against time. The vertical
dashed lines represent the time at which a fresh 100 μL droplet
stained with 0.05% (w/w) indigo carmine was moved to interface the
inlet of the siphon (a total of 75 s allowed for each reagent/washing).The microfluidic siphon delivered a constant discharge
flow rate
throughout >90% of the reagent’s volume (Figure B), only reducing at the start
and end due
to fluid acceleration upon contact with the liquid reagent, and deceleration
as the reagent volume is exhausted. As a reference, for the devices
mentioned in Figure (10-bore), the siphon yielded a flow rate of ∼2.5 μL/s,
taking ∼40 s to fully discharge 100 μL of fluid.As flow in the siphon stops between each reagent or wash, the fluid
is fully retained within the membrane or microcapillary by the inlet
Laplace pressure, without emptying, allowing self-controlled stop-flow
incubation with the sample, reagent, or wash. Tests with siphons of
varying L revealed that the liquid is retained in
the microcapillaries or microporous membrane even for values of H greater than the maximum measured equilibrium capillary
rise[41] (see Table S2 and Figure S3). In practice, the cost of the device is reduced
with small L; therefore, minimizing H is desirable. For incubation times of 5–30 min (typical of
microfluidic immunoassays[39]), no evaporation
or air bubble entry was detected.Through measuring the discharge
flow rate of the siphon, Q, in a wide range of single-bore
glass capillaries (with
a measured mean internal diameter of 578 μm and mean contact
angle of 50.9°; see the Supporting Information) and 10-bore MCF strips (with a measured mean hydraulic diameter d = 204 ± 13.8 μm and mean contact
angle of 67 ± 1.3°) with varying H/L ratios (range 0.15–0.85), we were able to demonstrate
that the siphon concept is fully scalable (Figure ), with flow rate, Q, solely
governed by gravity, fluid properties, and the diameter of the porous
channel following a simple pressure balance model (derivation provided
in the Supporting Information):where N is
the number of parallel microcapillaries, μ is the kinematic
fluid viscosity (0.001002 kg m–1 s–1 for water), dc is mean hydraulic diameter
of the microcapillary (or equivalent mean porous diameter for the
membrane), ρ is the fluid density (998.2 kg m–3 for water), g is the acceleration due to gravity
(9.81 m[2] s–1), and L is the total length of the strip. Note that the model
shown in eq assumes
a steady-state operation with the liquid head pressure and pressure
drop resistance balanced and therefore is not suitable for predicting
the instantaneous fluid velocity during, e.g., liquid priming or emptying
of microcapillaries. The air–liquid meniscus presented during
the priming of the siphon would add an extra force associated with
surface tension that would need to be included in the pressure balance.
Also, eq assumes that
the liquid head pressure remains constant during operation of the
siphon, which was not the case in the siphon experiment shown in Figure . As can be seen
in the bottom panel of Figure B, the discharge flow rate reduced slightly as the liquid
level in the waste reservoir increased. The pressure balance that
yielded eq can also
be easily adapted to consider time variations in the liquid pressure
head, yet during operation of several siphons such as the ones shown
in Figures C,D, we
learned the importance of having a design that keeps the liquid head
pressure approximately constant, which can be easily implemented using
a shallow waste reservoir. This way, variations in the liquid head
height remain insignificant compared to the liquid head pressure of
the siphon.
Figure 3
Normalized mean discharge flow rates from 10-bore MCF and single-bore
glass capillary siphons at varying H/L ratios. The siphons were initially primed by dipping the head into
Ultra-Pure water until all capillaries were filled and the bottom
of the siphon submerged 2 mm below the liquid level that was kept
constant. The head of the siphon was then dipped into a 150 μL
droplet of DI water stained with 0.05% (w/w) indigo carmine, and the
time taken to fully uptake the droplet was recorded. The neck distance
was kept constant for the MCF siphons at precisely 19 mm (corresponding
to a bending diameter of 11 mm), and the L of the
strip varied in the range 40–100 mm. For the glass capillaries, L was kept constant at 142.4 mm and the H/L ratio varied by bending the glass capillary at
different points along the length. As a simplification, it was assumed
that the liquid height in the reagent wells and waste wells remained
constant.
Normalized mean discharge flow rates from 10-bore MCF and single-bore
glass capillary siphons at varying H/L ratios. The siphons were initially primed by dipping the head into
Ultra-Pure water until all capillaries were filled and the bottom
of the siphon submerged 2 mm below the liquid level that was kept
constant. The head of the siphon was then dipped into a 150 μL
droplet of DI water stained with 0.05% (w/w) indigo carmine, and the
time taken to fully uptake the droplet was recorded. The neck distance
was kept constant for the MCF siphons at precisely 19 mm (corresponding
to a bending diameter of 11 mm), and the L of the
strip varied in the range 40–100 mm. For the glass capillaries, L was kept constant at 142.4 mm and the H/L ratio varied by bending the glass capillary at
different points along the length. As a simplification, it was assumed
that the liquid height in the reagent wells and waste wells remained
constant.Further experiments with MCF strips
configured at varying neck
bending diameters of 4, 7, and 11 mm and a fixed ratio H/L = 0.69 (results not shown) showed undetectable
differences in resistance forces in the siphon neck, with a mean discharge
flow rate of 3.63 ± 0.07 μL s–1.
Siphon
Effect in a Microporous Material
We demonstrated
that the siphon effect is also achievable in strips of a porous substrate
such as a porous glass fiber matrix used in LFA conjugate pads (Figure ), with time scales
comparable to the ones observed for the MCF siphons. A pair of strips
imaged in Figure A
(top) was operated in parallel but with an offset in reagent sequence
to illustrate the dye and wash sequence. Breakthrough curves (bottom
of Figure A) confirmed
that the membrane siphon cycled through the full saturation of dye,
to undetectable absorbance with small reagent volumes (200 μL
for a 7 mm wide×80 mm long strip). Unlike the conventional LFA
format with the horizontal fluid movement governed by a balance between
capillary forces and resistance, the siphon membrane configuration
is uniquely able to deliver constant fluid velocity at each step (Figure B). These results
suggest that reconfiguration of the standard horizontal LFA into a
siphon format (vertical, with a swan beak) could overcome both the
uneven flow and varying flow resistance, acknowledged as two core
limitations of LFAs. For multistep operation (Figure A), it enables combining continuous fluid
flow (set by H) with stop–start flow features
of the siphon, therefore offering a new level of fluidic control to
LFA in porous materials.
Figure 4
Operation of glass fiber porous matrix siphons.
(A) Time snapshots
(top) showing flow of dyed fluid (dye 1: bromophenol blue, 0.1 mg/mL;
dye 2: phenol red, 0.1 mg/mL) in PBS-T, with the waste microwell emptied
after each step in two parallel siphon membranes, designed with lateral
flow conjugate pad, 7 mm wide, L = 80 mm, and H = 50 mm (more details provided in the methods section).
Photographs of strip 2 (middle) at end of discharge of each reagent
showing consistent and homogeneous concentration. Liquid breakthrough
curves (bottom) with normalized concentration vs time for each reagent
step through strip 1, as imaged at 36 mm from the outlet, with well-defined
concentration plateaus for both the dyed fluid and washing buffer.
(B) Position of the liquid meniscus tracked over time for a glass
fiber membrane (L = 80 mm, width = 7 mm) held in
vertical or horizontal position vs siphon design (same membrane as
strip 1 shown in (A)), in PBS stained with indigo carmine (3 mg/mL),
showing the movement of the fluid in the siphon format is linear in
contrast to LFA membranes formatted in vertical or horizontal positions.
Operation of glass fiber porous matrix siphons.
(A) Time snapshots
(top) showing flow of dyed fluid (dye 1: bromophenol blue, 0.1 mg/mL;
dye 2: phenol red, 0.1 mg/mL) in PBS-T, with the waste microwell emptied
after each step in two parallel siphon membranes, designed with lateral
flow conjugate pad, 7 mm wide, L = 80 mm, and H = 50 mm (more details provided in the methods section).
Photographs of strip 2 (middle) at end of discharge of each reagent
showing consistent and homogeneous concentration. Liquid breakthrough
curves (bottom) with normalized concentration vs time for each reagent
step through strip 1, as imaged at 36 mm from the outlet, with well-defined
concentration plateaus for both the dyed fluid and washing buffer.
(B) Position of the liquid meniscus tracked over time for a glass
fiber membrane (L = 80 mm, width = 7 mm) held in
vertical or horizontal position vs siphon design (same membrane as
strip 1 shown in (A)), in PBS stained with indigo carmine (3 mg/mL),
showing the movement of the fluid in the siphon format is linear in
contrast to LFA membranes formatted in vertical or horizontal positions.In addition to enabling the use of a sequence of
reagents and start–stop
flow with a good level of control over the flow rate and incubation
times (not possible in conventional LFAs), we further observed that
operation of a porous membrane formatted according to the siphon design
shown in Figure A,
having a net H and a neck at a higher liquid height
than the inlet, yielded a constant linear movement of the meniscus.
This is not possible to see in the same membrane material operated
in a vertical or horizontal format (Figure B), as in those cases fluid movement is dominated
by resistance and/or capillary forces that vary as the fluid permeates
the material.
Application of MCF Siphons as Quantitative
ELISA to Global Diagnostics
To use the microfluidic siphon
for a high-performance POC device
in a format suited to global healthcare needs, we developed 3D printed
prototype cassettes (shown in Figure D) that enabled simultaneous operation of parallel
siphon strips. The design can easily be mass-manufactured and has
a 9 mm pitch compatible with conventional 96-well microplates and
multichannel pipettes. Combined with the multibore MCF material, these
stacked siphon holders offer high-throughput instrument-free testing,
in addition to the multiplexing capabilities previously reported for
this mass-manufactured material.[42] Although
microporous membranes remain a gold standard in POC testing, the two
proof-of-concept ELISA applications presented here require quantitative
capabilities that are difficult to realize with a porous matrix in
an LFA. The optical transparency, multiplexing, and ″open fluidic″
features make the MCF strips especially fit for operation with siphon
cassettes.We demonstrated the clinical value of these siphon
cassettes by carrying out a four-plex sandwich ELISA for serotype-specific
quantitation of dengue NS1 protein with clinical human samples in
Thailand and singleplex quantitation of SARS-CoV-2 IgM and IgG antibodies
in simulated samples and convalescent COVID-19 human samples. Enzyme
AttoPhos fluorescence signal was recorded with a smartphone camera
and inexpensive LED illumination,[37,38] allowing the
near-patient or out-of-lab high-performance immunoassays needed for
diagnostics and surveillance of important infections such as dengue
and SARS-CoV-2.Viral infections can be detected by measurement
of antibodies against
viral antigens (antibody tests) or of viral components directly (antigen
tests). With 4 billion people at risk of dengue virus (DENV) infection
and an estimated 100 million infections annually,[43,44] access to decentralized diagnostic testing remains vital, and patients
are typically tested first by LFAs that combine antibody and antigen
detection; however, current point-of-care NS1 LFAs are not serotype-specific,[45] so centralized testing remains essential for
viral surveillance. We adapted a serotyping microplate ELISA that
uses serotype-specific anti-NS1 antibody clones[46] to siphon MCF devices. Multiplex capture strips with duplicate
pairs of capillaries were coated internally through passive adsorption[40] with serotype-selective capture antibodies (Table S1) against DENV NS1 (Figure A). Capture of viral antigen
was detected using a pan-NS1 detection antibody. A positive control
was included with a single capillary coated with recombinant NS1,
alongside an uncoated negative control capillary. Only capillaries
corresponding to a spiked recombinant NS1 showed positive fluorescence
(Figure B), with full
response curves showing clinically relevant sensitivity for all four
serotypes (Figure C) with no difference in quantitation between the
buffer and plasma. The four serotypes showed different analytical
performance, with 1 ng/mL being readily detectable for serotypes 2
and 3, whereas serotypes 1 and 4 showed a weaker signal at the lower
NS1 concentrations, yet all gave a strong signal at a 10 ng/mL threshold
for patients with acute infection. Notably, the data obtained with
siphon devices showed a limit of detection similar to that reported
for serotyping NS1 microplate ELISA.[47] Differences
in analytical performance between serotypes are expected as the capture
antibodies were selected primarily for serotype specificity and not
for highest analytical sensitivity. A different set of serotype-specific
anti-NS1 pairs showed a different pattern of analytical sensitivity
in laboratory lateral flow devices, demonstrating the influence of
antibody clone pairs on sensitivity.[48] As
dengue patients have from 10 ng/mL to 50 μg/mL NS1 in plasma
during acute infection,[49] the siphon MCF
devices should be sufficiently sensitive for serotype-specific NS1
dengue diagnostics. PCR-serotyped pairs of patient samples from each
serotype, plus matched negative controls (total n = 10), were tested using these devices, and NS1 was clearly detectable
in each of them using both a digital camera and a smartphone (top
two rows in Figure D). This demonstrated that multiplex viral antigen detection is feasible
using siphon microfluidics imaged with a midrange smartphone (iPhone
6S). A large-scale clinical study of these devices with >250 samples
found similar performance for the MCF siphon to microplate ELISA (manuscript
submitted).
Figure 5
Enzyme-linked immunosorbent assay (ELISA) in MCF siphons enabling
rapid smartphone diagnosis of dengue fever. (A) Configuration of serotype-specific
NS1 capture antibodies and control capillaries within MCF test strips.
(B) Image showing DENV NS1 quantitation spiked into buffer vs plasma,
illustrating that only pairs of capillaries coated with capture antibodies
corresponding to the NS1 serotype spiked show positive fluorescence.
Profile plots across the MCF test strip (left-hand side) illustrate
how fluorescence intensity was quantified for each individual capillary.
(C) Quantitation of NS1 detection in plasma vs buffer for the four
DENV serotypes. Mean fluorescence normalized to a fixed fluorescein
reference sample was plotted from the duplicate capillaries, with
error bars indicating 1 standard deviation. Data presented in (B)
and (C) are representative of more than three independent standard
curves in buffer vs plasma. (D) Digital camera and smartphone images
of 10 multiplex DENV NS1 MCF tests using siphon fluidics of undiluted
plasma samples compare two dengue fever patients for each of the four
DENV serotypes vs two negative controls (n = 10).
The serotype determined by PCR and NS1 serotype detected by siphon
MCF were recorded below images, and the NS1 concentration indicated
was calculated by comparison of fluorescent intensity in the patient
sample to reference dilutions of recombinant NS1 as shown in (C).
Enzyme-linked immunosorbent assay (ELISA) in MCF siphons enabling
rapid smartphone diagnosis of dengue fever. (A) Configuration of serotype-specific
NS1 capture antibodies and control capillaries within MCF test strips.
(B) Image showing DENV NS1 quantitation spiked into buffer vs plasma,
illustrating that only pairs of capillaries coated with capture antibodies
corresponding to the NS1 serotype spiked show positive fluorescence.
Profile plots across the MCF test strip (left-hand side) illustrate
how fluorescence intensity was quantified for each individual capillary.
(C) Quantitation of NS1 detection in plasma vs buffer for the four
DENV serotypes. Mean fluorescence normalized to a fixed fluorescein
reference sample was plotted from the duplicate capillaries, with
error bars indicating 1 standard deviation. Data presented in (B)
and (C) are representative of more than three independent standard
curves in buffer vs plasma. (D) Digital camera and smartphone images
of 10 multiplex DENV NS1 MCF tests using siphon fluidics of undiluted
plasma samples compare two dengue fever patients for each of the four
DENV serotypes vs two negative controls (n = 10).
The serotype determined by PCR and NS1 serotype detected by siphon
MCF were recorded below images, and the NS1 concentration indicated
was calculated by comparison of fluorescent intensity in the patient
sample to reference dilutions of recombinant NS1 as shown in (C).While viral antigens detected by sandwich immunoassay
are valuable
for some infections such as dengue, for other infections, anti-viral
antibodies are more important. These can be detected using indirect
ELISA by immobilizing viral antigens, incubating with the diluted
sample, and probing for human immunoglobulins (e.g., IgG, IgM, or
IgA). To quantify anti-viral antibody titer, a series of serial dilutions
of the patient sample must be tested, with the end point titer reflecting
the highest dilution at which a clear positive signal is retained.
During many acute viral infections, the signal is retained beyond
10,000-fold dilution, illustrating the very high affinity binding
characteristic of seroconversion. To establish quantitative indirect
ELISA in siphon devices, MCF strips coated with the SARS-CoV-2 spike
protein were tested with serial dilutions of recombinant IgM and IgG
antibodies spiked into a prepandemic serum (Figure ). After using these simulated samples to
establish assay conditions that minimized assay background while allowing
both high and low levels of IgG/M to be detected, we compared antibody
titer quantitation in microplates vs siphon MCF devices (Figure B,C). Volunteer plasma
samples previously identified as positive or negative for polyvalent
anti-SARS-CoV-2 spike protein by conventional microplate ELISA[50] were tested in parallel for IgG in siphon MCF
vs microplates (Figure B,C).
Figure 6
Smartphone enzyme-linked immunosorbent assay detection of IgM and
IgG anti-viral antibodies to SARS-CoV-2 with MCF siphons in simulated
acute vs post infection samples. (A) Detection of antiviral COVID-19
IgG antibodies in human plasma samples serially diluted and tested
in a set of eight MCF siphon devices each coated with recombinant
SARS-CoV2 spike protein within all 10 capillaries. (B) Comparison
of quantitation by indirect ELISA of IgG antibodies against SARS-CoV2
spike protein in positive vs negative plasma sample in MCF siphon
devices vs conventional microtiter plate, shown in (C) for each dilution
the mean and standard deviation plotted represent 10 replicate MCF
capillaries vs triplicate microtiter plate wells. Similar titration
curves were seen in three replicate SARS-CoV2 indirect ELISA experiments.
(D) Different levels of recombinant human IgM and IgG anti-SARS-CoV2
were spiked into prepandemic control serum to represent COVID-19 patients
with low and high levels of IgM or IgG, respectively. These simulated
patient samples were then serially diluted and tested using siphon
MCF test strips coated with spike protein from SARS-CoV-2. (E) Smartphone
imaging of 16 siphon MCF test strips quantifying IgG anti-viral antibody
in seven serial dilutions of two human plasma samples. Top set is
anti-SARS-CoV2 positive sample, with top dilution of 1:20 followed
by six threefold serial dilutions and a blank. Lower row is the negative
sample diluted and tested alongside the positive. A reference sample
of fixed fluorescent dye concentration is visible to the left. Black
MCF holders also visible. Images shown in (A) and fluorescence intensity
values presented in (B) were taken in the middle section of the strips
shown in (E), with no detectable change in greyscale pixel intensity
found along the length of the strip. Note that error bars corresponding
to 1 standard deviation were included in plots shown in (B), (C),
and (D) yet in some cases were too small to be shown.
Smartphone enzyme-linked immunosorbent assay detection of IgM and
IgG anti-viral antibodies to SARS-CoV-2 with MCF siphons in simulated
acute vs post infection samples. (A) Detection of antiviral COVID-19
IgG antibodies in human plasma samples serially diluted and tested
in a set of eight MCF siphon devices each coated with recombinant
SARS-CoV2 spike protein within all 10 capillaries. (B) Comparison
of quantitation by indirect ELISA of IgG antibodies against SARS-CoV2
spike protein in positive vs negative plasma sample in MCF siphon
devices vs conventional microtiter plate, shown in (C) for each dilution
the mean and standard deviation plotted represent 10 replicate MCF
capillaries vs triplicate microtiter plate wells. Similar titration
curves were seen in three replicate SARS-CoV2 indirect ELISA experiments.
(D) Different levels of recombinant human IgM and IgG anti-SARS-CoV2
were spiked into prepandemic control serum to represent COVID-19 patients
with low and high levels of IgM or IgG, respectively. These simulated
patient samples were then serially diluted and tested using siphon
MCF test strips coated with spike protein from SARS-CoV-2. (E) Smartphone
imaging of 16 siphon MCF test strips quantifying IgG anti-viral antibody
in seven serial dilutions of two human plasma samples. Top set is
anti-SARS-CoV2 positive sample, with top dilution of 1:20 followed
by six threefold serial dilutions and a blank. Lower row is the negative
sample diluted and tested alongside the positive. A reference sample
of fixed fluorescent dye concentration is visible to the left. Black
MCF holders also visible. Images shown in (A) and fluorescence intensity
values presented in (B) were taken in the middle section of the strips
shown in (E), with no detectable change in greyscale pixel intensity
found along the length of the strip. Note that error bars corresponding
to 1 standard deviation were included in plots shown in (B), (C),
and (D) yet in some cases were too small to be shown.We observed a clear difference between positive and negative
samples
(Figure A,D,E), with
signal decaying in a similar pattern with dilution, indicating that
siphon devices are suitable for serological COVID-19 testing including
antibody titer measurement. Furthermore, the smartphone could record
the full set of 16 × 10 microcapillary siphon devices comprising
positive plus negative plasma samples, each serially diluted seven
times, plus two negatives (i.e., 160 × 1 μL microcapillary
data points) in a single photo (Figure E).Note that both NS1 sandwich immunoassay and
indirect SARS-CoV-2
ELISAs herein reported were carried out in a siphon cassette and MCF
strips with a fixed H/L ratio. As
the discharge flow rate of the siphon and superficial fluid velocity
are linearly dependent on the H/L ratio, the reported ELISA performance should be carefully extrapolated
to other H/L ratios through a more
extensive validation. During this study, we noticed that the length
of the siphon strip and washing volumes can impact on the signal-to-noise
(and consequently performance) of bioassays; this will be the subject
of future publications.
Conclusions
Sequential siphon flow
has been realized for a range of microbored/microporous
material, relying entirely on gravity in contrast to all POC biosensing
devices developed to date. Siphon devices can quickly be reconfigured
with standard microplate ELISA reagents for singleplex or multiplex
format to measure protein biomarkers, antigens, or antibodies without
compromising analytical performance. This represents a major milestone
in decentralizing high-performance diagnostic testing, removing the
need of expensive automation, complex fluid handling, or skilled workforce.
The siphon concept can transform POC testing with a new generation
of high-performance power-free immunoassays transferring ELISA outside
laboratories, with results digitally recorded by a smartphone camera
to meet many rapidly changing global public health needs. Demonstration
of improved analytical capabilities with porous membrane siphons and
other H/L ratios is yet to be tested
and will be the subject of future publications.
Authors: Ana I Barbosa; Poonam Gehlot; Kalpita Sidapra; Alexander D Edwards; Nuno M Reis Journal: Biosens Bioelectron Date: 2015-03-05 Impact factor: 10.618