Rafiq M Lubken1,2, Max H Bergkamp1,2, Arthur M de Jong3,2, Menno W J Prins1,3,2,4. 1. Department of Biomedical Engineering, Eindhoven University of Technology, Eindhoven 5612 AZ, The Netherlands. 2. Institute for Complex Molecular Systems (ICMS), Eindhoven University of Technology, Eindhoven 5612 AZ, The Netherlands. 3. Department of Applied Physics, Eindhoven University of Technology, Eindhoven 5612 AZ, The Netherlands. 4. Helia Biomonitoring, Eindhoven 5612 AZ, The Netherlands.
Abstract
Studies on the dynamics of biological systems and biotechnological processes require measurement techniques that can reveal time dependencies of concentrations of specific biomolecules, preferably with small time delays, short time intervals between subsequent measurements, and the possibility to record over long time spans. For low-concentration biomolecules, these requirements are very challenging since low-concentration assays are typically slow and require new reagents in every assay. Here, we present a sensing methodology that enables rapid monitoring of picomolar and sub-picomolar concentrations in a reversible affinity-based assay, studied using simulations. We demonstrate that low-concentration biomolecules can be monitored with small time delays, short time intervals, and in principle over an endless time span.
Studies on the dynamics of biological systems and biotechnological processes require measurement techniques that can reveal time dependencies of concentrations of specific biomolecules, preferably with small time delays, short time intervals between subsequent measurements, and the possibility to record over long time spans. For low-concentration biomolecules, these requirements are very challenging since low-concentration assays are typically slow and require new reagents in every assay. Here, we present a sensing methodology that enables rapid monitoring of picomolar and sub-picomolar concentrations in a reversible affinity-based assay, studied using simulations. We demonstrate that low-concentration biomolecules can be monitored with small time delays, short time intervals, and in principle over an endless time span.
Entities:
Keywords:
affinity kinetics; biomolecules; fast response; high precision; low concentration; monitoring; sensing
Biological
systems and biotechnological
processes exhibit time dependencies that are at the most basic level
regulated by the dynamics of constituting biomolecules, such as small
molecules, hormones, proteins, and nucleic acids. This calls for measurement
technologies that allow the monitoring of biomolecular concentrations,
for instance, to serve fundamental research on biological and biomedical
dynamics,[1−6] to enable the development of patient monitoring strategies based
on real-time biomolecular data,[7−10] and to enable the development of closed loop control
strategies in biotechnological applications.[11−16] Desirable characteristics of a generic monitoring technology are
(1) precise and specific measurements, (2) small time delays between
sampling input and data output, (3) short time intervals between successive
measurements, and (4) a long total time span over which time-dependent
biomolecular concentration data can be recorded.It is a fundamental
challenge to develop a technology that can
rapidly monitor low-concentration biomolecules over long time spans.
Sensitive assays are available, such as ELISA and flow cytometry,[17−20] but in these assays, new reagents are needed for every sample that
is taken. The repeated consumption of reagents complicates applications
where biomolecular concentrations need to be monitored over long time
spans. On the other hand, sensing technologies that can operate without
consuming reagents, such as surface plasmon resonance,[21] redox cycling,[22] and
quartz crystal microbalance,[23] have not
been designed for monitoring biomolecules at low concentrations, such
as in the picomolar and sub-picomolar range.A generic principle
to measure specific biomolecules at low concentrations
is by using the biochemical affinity between specific binder molecules
(such as antibodies and aptamers) and the biomolecule that is to be
detected (the analyte). The specificity originates from molecular
interactions such as charge, hydrogen bonding, van der Waals forces,
and hydrophobic and steric effects.[24] To
be able to measure biomolecules at low concentrations with high precision,
binder molecules are needed that have strong interactions with the
analyte, which corresponds to high binding energies, low equilibrium
dissociation constants Kd, and low dissociation
rate constants koff.[25,26] However, this conflicts with the desire to have small time delays
because low dissociation rate constants would imply a need for long
incubation times to reach equilibrium.[24−26] Furthermore, low dissociation
rate constants result in a slow reversibility, which conflicts with
the wish to enable short time intervals between successive measurements.In this paper, a sensing methodology is presented that enables
rapid monitoring of low-concentration biomolecules in the picomolar
and sub-picomolar range, studied using simulations. The method is
based on the use of binder molecules with a high affinity in a limited-volume
assay, with a fully reversible detection principle and time-dependent
sampling of the analyte of interest. The system allows optimal trade-offs
between time characteristics and precision. We present the measurement
concept, time dependencies of sensor signals, and a comprehensive
analysis of the achievable time characteristics and precision as a
function of sensor design parameters. We demonstrate that the sensing
methodology enables precise quantification of low biomolecular concentrations,
with time delays and interval times that are much shorter than the
time dictated by the dissociation rate constant of the binder molecules.
Furthermore, due to the reversible detection method, measurements
can in principle be performed over an endless time span.
Basic Concept
of the Sensing Methodology
The basic concepts of the sensing
methodology are sketched in Figure . The sensing system
features time-dependent sampling of the analyte of interest, provided
by a time-controlled analyte exchange between a biological or biotechnological
system of interest and a measurement chamber (see Figure a). The measurement chamber
contains specific binder molecules from which signals are recorded.
The data are translated into concentration–time profiles, which
should resemble as close as possible the true concentration–time
profile of analyte molecules in the system of interest. During the
exchange of analyte molecules, various processes occur, such as mass
transport by advection and diffusion and association and dissociation
of analyte molecules to binder molecules (see Figure b). To illustrate the concept, a generic
rectangular measurement chamber is assumed with height H, width W, and length L. The bottom
surface is provided with binder molecules, where association and dissociation
of analyte molecules occurs. The rates of association and dissociation
depend on the association rate constant kon, the dissociation rate constant koff, the density Γb of binder molecules, and the analyte
concentration Ca at the sensor surface.
These processes result in a time-dependent density γab of analyte–binder complexes, also represented as a fractional
occupancy f of binder molecules occupied by analyte
molecules: f = γab/Γb. Variables γab and f are changing
as a function of analyte concentration and time. In an affinity-based
sensor, the observed sensor signal scales with f;
therefore, f is used in this paper as the sensor
readout parameter to determine the analyte concentration. Analyte
exchange between the system of interest and the measurement chamber
is facilitated by diffusion or a combination of diffusion and advection.
A net diffusive molar flux Ja (orange
gradient) is caused by concentration differences between the system
of interest and the measurement chamber and by concentration differences
within the measurement chamber. Advective mass transport of analyte
molecules into the measurement chamber is facilitated by a developed
laminar flow profile with flow rate Q and mean flow
speed vm (black arrows). Here, it is assumed
that diffusive transport occurs in both the longitudinal (x-direction) and the lateral direction (y-direction) and scales with the diffusion coefficient D, while advective transport occurs only in the longitudinal direction
and scales with the mean flow speed vm.
Figure 1
Concept of the sensing methodology for the rapid monitoring of
low analyte concentrations. (a) Sensing system for biomolecular monitoring.
Analyte molecules are exchanged between a biological or biotechnological
system of interest and a measurement chamber. The data result in a
concentration–time profile which should
correspond as close as possible to the true analyte concentration-time
profile in the system of interest. (b) Geometry of the measurement
chamber, with height H, width W,
and length L. A reaction rate at the sensor surface
is caused by the association and dissociation between analyte molecules
(orange) and binder molecules (brown), described by the association
rate constant kon, the dissociation rate
constant koff, the total binder density
Γb, the analyte concentration Ca near the sensor surface, and the density of analyte–binder
complexes γab. Analyte exchange is facilitated by
diffusion and advection, where diffusion occurs in both x- and y-direction with the diffusion coefficient D, resulting in a net molar flux Ja, and where advection occurs in the x-direction
only, with a developed flow profile with flow rate Q and a mean flow speed vm. (c) Time profile
of the sensor response for low analyte concentration (Ca,0 < Kd), for two conditions:
infinite-volume and limited-volume assays. For the infinite-volume
condition, the time-to-equilibrium τ is determined by koff, while for the limited-volume condition,
τ is determined by the effective binder concentration. (d) Biomolecular
monitoring using a limited-volume assay involves repeated cycles with
two phases. In phase 1, analyte molecules are exchanged between the
system of interest and the measurement chamber. In phase 2, analyte
molecules are not exchanged between the system of interest and the
measurement chamber. The time-dependent signal is recorded in the
middle of the measurement chamber at distance L/2
from the entrance. The limited-volume condition gives a time dependence
of the analyte concentration in the measurement chamber: a decrease
over time (depletion) or an increase over time (repletion), depending
on the input analyte concentration Ca,0 and the initial fractional occupancy finit of binder molecules by analyte molecules. The input analyte concentration Ca,0 is derived from the measured time-dependent
fractional occupancy f.
Concept of the sensing methodology for the rapid monitoring of
low analyte concentrations. (a) Sensing system for biomolecular monitoring.
Analyte molecules are exchanged between a biological or biotechnological
system of interest and a measurement chamber. The data result in a
concentration–time profile which should
correspond as close as possible to the true analyte concentration-time
profile in the system of interest. (b) Geometry of the measurement
chamber, with height H, width W,
and length L. A reaction rate at the sensor surface
is caused by the association and dissociation between analyte molecules
(orange) and binder molecules (brown), described by the association
rate constant kon, the dissociation rate
constant koff, the total binder density
Γb, the analyte concentration Ca near the sensor surface, and the density of analyte–binder
complexes γab. Analyte exchange is facilitated by
diffusion and advection, where diffusion occurs in both x- and y-direction with the diffusion coefficient D, resulting in a net molar flux Ja, and where advection occurs in the x-direction
only, with a developed flow profile with flow rate Q and a mean flow speed vm. (c) Time profile
of the sensor response for low analyte concentration (Ca,0 < Kd), for two conditions:
infinite-volume and limited-volume assays. For the infinite-volume
condition, the time-to-equilibrium τ is determined by koff, while for the limited-volume condition,
τ is determined by the effective binder concentration. (d) Biomolecular
monitoring using a limited-volume assay involves repeated cycles with
two phases. In phase 1, analyte molecules are exchanged between the
system of interest and the measurement chamber. In phase 2, analyte
molecules are not exchanged between the system of interest and the
measurement chamber. The time-dependent signal is recorded in the
middle of the measurement chamber at distance L/2
from the entrance. The limited-volume condition gives a time dependence
of the analyte concentration in the measurement chamber: a decrease
over time (depletion) or an increase over time (repletion), depending
on the input analyte concentration Ca,0 and the initial fractional occupancy finit of binder molecules by analyte molecules. The input analyte concentration Ca,0 is derived from the measured time-dependent
fractional occupancy f.Figure c sketches
two different sensor designs, namely, an infinite-volume assay and
a limited-volume assay. The graphs visualize the fractional occupancy f of binder molecules occupied by analyte molecules as a
function of time, with the corresponding characteristic time-to-equilibrium
τ, defined as the time needed to attain 63% of the difference
between the starting level and the equilibrium level of f (see Note S2). In an infinite-volume
assay, continuous analyte exchange is enabled between the system of
interest and the measurement chamber, where the system of interest
is assumed to be much larger than the measurement chamber. The continuous
analyte exchange could, for example, be facilitated by diffusive analyte
transport across a contact area between the system of interest and
the measurement chamber, while another configuration may involve a
continuous flow of sample fluid provided into the measurement chamber
from the system of interest. When the analyte exchange is effective
and gives negligible time delay, then the analyte concentration at
the sensor surface (Ca) is equal to the
input analyte concentration (Ca,0). In
the case of low analyte concentrations (Ca,0 < Kd), the infinite-volume assay
condition leads to a characteristic time-to-equilibrium τ ≅
1/koff (see Note S2). This implies that the time-to-equilibrium is determined by the
dissociation rate constant koff, so this
time is long when the binder molecules strongly bind to the analyte
molecules.The sensor design with a limited-volume assay has
very different
properties. Here, analyte exchange between the system of interest
and the measurement chamber is enabled for a limited time. After this
analyte exchange, the binder molecules in the measurement chamber
interact with only a limited sample volume and therefore with a limited
amount of analyte molecules. Due to this limited volume, we can now
define an effective volumetric concentration of binder molecules Cb,0 = Γb/H, which is based on the number of binder molecules in the measurement
chamber and the volume of the measurement chamber. When Cb,0 is high, with Cb,0 > Ca,0 and Cb,0 > Kd, then the time-to-equilibrium τ of the
assay becomes dominated by the high concentration of binder molecules.
When diffusional transport delays can be ignored, then the time-to-equilibrium
of the assay equals τ ≅ 1/(konCb,0) (see Table and Notes S1 and S2). Thus, the time-to-equilibrium of the limited-volume assay is determined
by the association rate constant and the effective volumetric concentration Cb,0 of binder molecules, which leads to equilibrium
timescales that are much shorter than the time-to-equilibrium of the
infinite-volume assay.
Table 1
Standard Parameter
Values Used in
the Finite-Element Simulationsa
parameter
value
description
input
H
200 μm
measurement chamber height
D
10–10 m2 s–1
diffusion coefficient of
the analyte molecule
Γb
10–9 mol m–2 (600 μm–2)
binder density
koff
10–4 s–1
dissociation rate constant
kon
106 M–1 s–1
association rate constant
Ca,0
0.1 pM
input analyte concentration
derived
τD = H2/D
400 s
characteristic diffusion
time
200 s
characteristic reaction
time for limited-volume assay with Cb,0 ≫ Ca,0 and Cb,0 ≫ Kd
Cb,0 = Γb/H
5 nM
effective volumetric binder
concentration
Kd = koff/kon
100 pM
equilibrium dissociation
constant
α = Γb/HKd = Cb,0/Kd
50
acceleration factor: reduction
factor of the time-to-equilibrium of
a limited-volume assay with τ(H, Γb), compared to an infinite-volume assay with τ(koff)
Da = τD/τR,LV = konΓbH/D
2
Damköhler number
Details
about the simulations are
described in Note S4. Additional standard
parameter values are given in Table S1 (see Note S1).
Details
about the simulations are
described in Note S4. Additional standard
parameter values are given in Table S1 (see Note S1).In monitoring applications, it is preferred to record measurements
with one and the same sensor over long time spans. To realize the
limited-volume assay principle in a monitoring application, the sensor
needs to be switched between two different conditions: an open condition
and a closed condition. In the open condition, analyte molecules are
exchanged effectively between the system of interest and the measurement
chamber, as sketched in Figure a,b (also see Note S5). In the
closed condition, analyte molecules are not exchanged between the
system of interest and the measurement chamber, causing a limited-volume
incubation in the measurement chamber, as sketched in the bottom graph
of Figure c. We refer
to the switching concept between the open and closed condition as
“time-controlled analyte exchange”. Figure d illustrates the operating
principle for a sensor where time-controlled analyte exchange is realized
by a modulated flow. Phase 1 is the exchange phase, where the measurement
chamber is supplied with the sample fluid so that the starting concentration
in the chamber equals Ca,0. Phase 2 is
the incubation phase, where the exchange process is stopped so that
the limited-volume assay condition is provided. During incubation
under the limited-volume condition, the analyte concentration in the
measurement chamber decreases over time (depletion) or increases over
time (repletion), depending on the initial fractional occupancy finit of binder molecules by analyte molecules.
When finit is low, the concentration of
analyte molecules in the measurement chamber decreases over time,
corresponding to the depletion of analytes. When finit is high, the concentration of analyte molecules in
the chamber increases over time, corresponding to the repletion of
analytes. For known finit, the supplied
analyte concentration Ca,0 can be derived
from the measured time-dependent fractional occupancy f(t) during the incubation phase. At least two measurements
need to be performed to determine the input analyte concentration Ca,0, for example, a measurement at the initial
value finit and a measurement at the final
value fend, as indicated in the graph.By sequentially applying cycles of open condition and closed condition,
discrete samples with a limited volume are serially measured and result
in time-dependent data that relate to the different samples supplied
to the sensor. Each former measurement causes a varying nonzero initial
fractional occupancy finit in the next
measurement. The values of finit and Ca,0 determine whether depletion or repletion
occurs during the incubation phase. In the case of depletion, a higher
input analyte concentration Ca,0 yields
a larger, positive change in fractional occupancy Δf = fend – finit since more analyte molecules are captured from solution.
In the case of repletion, a higher Ca,0 yields a smaller, negative change in fractional occupancy Δf since less analyte molecules are repleted from the sensor
surface into solution. An important property of the sensor is that
the interactions between binder and analyte molecules are reversible.
This gives the advantage that the limited-volume assay with time-controlled
analyte exchange can be used over an endless time span.In the
next sections, we will study how sensor design parameters
influence the time characteristics and precision of the sensing methodology.
The time characteristics are quantified by finite-element simulations
of mass transport in the sensor and reaction processes at the sensor
surface, and the precision is quantified by calculating the stochastic
variabilities in the measurements. The simulations and calculations
are verified by experiments using a sensing technique with a single-molecule
resolution, called biosensing by particle mobility (BPM, see Note S7).[27−29]
Experimental
Section
Finite-Element Analysis
Finite-element simulations
were performed by solving diffusion, advection, and reaction equations
simultaneously using COMSOL (COMSOL Multiphysics 5.5) and MATLAB (MATLAB
R2019a, COMSOL Multiphysics LiveLink for MATLAB) (see Note S4). From the simulations, the time-to-equilibrium
τ was determined by calculating the time at which the analyte–binder
complex density γab is at 63% of the difference between
the starting level and the equilibrium level of γab. The time-controlled analyte exchange (see Figure ) was simulated by instantaneously increasing/decreasing
the analyte concentration throughout the measurement chamber Ca to Ca,0, with
which a new measurement cycle starts. The density of analyte–binder
complexes γabstart at the start of a cycle was set to be equal to the density
of analyte–binder complexes γabend at the end of the preceding cycle.
Sensor signals are reported at distance L/2 in the
measurement chamber (see Figure d). Precisions are reported at a distance L/2 in the measurement chamber, where the signal is collected over
a signal collection area of 1 mm2 (Figure c–e).
Figure 3
Simulated response
of the biomolecular monitoring system using
time-controlled analyte exchange. (a) Analyte concentration Ca in the measurement chamber (brown line) and
the fractional occupancy f of binder molecules by
analyte molecules (orange line) as a function of time, for low finit and depletion of analyte in solution (left),
and for high finit and repletion of analytes
in solution (right). The dashed lines indicate time points where instantaneous
analyte exchange occurs, where the bulk analyte concentration was
set to Ca,0 = 0.1 pM after each period
of approximately 50 min. The insets highlight the kinetics of the
first cycle, showing a time-to-equilibrium of τ = 340 s. For
many cycles (n → ∞), both curves would
approach , which equals the equilibrium value when
an infinite volume is supplied (see Table S1). (b) Fractional occupancy f as a function of time
where cycles of analyte exchange and incubation are applied every
15 min with alternatingly Ca,0 = 0.15
pM and Ca,0 = 0.05 pM. The curve saturates
at feq,IV = 10·10–4, which equals the infinite-volume equilibrium value for the average
concentration value Ca,0 = 0.1 pM (see Table S1). Dashed lines: continuous supply of Ca,0 = 0.05 pM yields feq,IV = 5·10–4 and Ca,0 = 0.15 pM yields feq,IV = 15·10–4 (see Table S1). The right panel shows zoom-in images of three sections
of the solid curve, each representing four cycles of instantaneous
analyte exchange and subsequent incubations of 15 min. In zoom-in
1 (t = 0–1 h), all curve segments show depletion
behavior. In zoom-in images 2 (t = 12–13 h)
and 3 (t = 42–43 h), depletion is seen for Ca,0 = 0.15 pM, since finit < feq,IV(Ca,0 = 0.15 pM), and repletion is seen for Ca,0 = 0.05 pM, since finit > feq,IV(Ca,0 = 0.05 pM). For all curve segments, the time-to-equilibrium τ
= 340 s. The vertical scale bars indicate Δf = 10–4.
Figure 5
Analytical performance
of the limited-volume assay, derived from
simulations of a single measurement cycle. (a) Fractional occupancy
at the end of the incubation fend as a
function of analyte concentration Ca,0 for different initial fractional occupancies finit. The right y-axis indicates the number
of surface-bound analyte molecules at the end of the cycle γabend. (b) Absolute
change in fractional occupancy Δf as a function
of Ca,0 for various finit. The right y-axis indicates Δγab. A positive Δf and Δγab indicate depletion; negative values indicate repletion.
Note that the lines for finit = 10–3 and lower are overlapping. The inset shows the same
data on a linear–linear scale. (c) Coefficient of variation
CVC with which the analyte concentration Ca,0 can be determined as a function of analyte concentration Ca,0 for various initial fractional occupancies finit. CVC scales as for low finit and high Ca,0; CVC scales
as 1/Ca,0 for high finit and low Ca,0. The horizontal
dashed line indicates the 10% CV level that defines the LoQ (limit
of quantification). (d) CVC as a function of measurement
chamber height H (top) and binder density Γb (bottom) for various initial fractional occupancies finit and Ca,0 =
0.1 pM. The arrows on the x-axes indicate the standard
parameter values for H and Γb, as
listed in Table .
(e) CVC as a function of the observed time-to-equilibrium
τ when varying the measurement chamber height H (left) or binder density Γb (right) for various
initial fractional occupancies finit and Ca,0 = 0.1 pM. The sketches above the graphs
visualize a measurement chamber with an increasing height or a decreasing
binder density. The arrows on the x-axes indicate
the obtained time-to-equilibrium using the standard parameter values
for H and Γb, as listed in Table .
Fluid Cell Assembly
Glass slides (25 × 75 mm,
#5, Menzel-Gläser) were cleaned by 40 min sonication in isopropanol
(VWR, absolute) and twice by 10 min sonication in Milli-Q (Thermo
Fisher Scientific, Pacific AFT 20). Subsequently, the glass slides
were dried under nitrogen flow. A polymer mixture of PLL(20)-g[3.5]-PEG(2) (SuSoS) and PLL(15)-g[3.5]-PEG(2)-N3 (Nanosoft Polymers) was prepared at a final concentration
of 0.45 and 0.05 mg/mL in Milli-Q, respectively. The glass slides
were treated by oxygen plasma (Plasmatreat GmbH) for 1 min. A custom-made
fluid cell sticker (Grace Biolabs), with an approximate volume of
20 μL, was attached to the glass slide and immediately filled
with the polymer mixture. After 2 h of incubation, the polymer mixture
was removed and the fluid cell was immediately filled with 0.5 nM
dsDNA tether solution (221 bp, with DBCO at one end and biotin at
the other end) in 0.5 M NaCl in phosphate-buffered saline (PBS).[30] After overnight incubation, the solution in
the fluid cell was exchanged by 2 μM DBCO-functionalized dsDNA
solution[30] in 0.5 M NaCl in PBS and incubated
for several days until use.
Particle Functionalization
A total
of 2 μL of
streptavidin-functionalized particles (10 mg/mL, Dynabeads MyOne Streptavidin
C1, Thermo Fisher Scientific) was incubated with 1 μL of biotinylated
ssDNA binder molecules (10 μM, IDT, HPLC purification) and 4
μL of PBS for 70 min. The particles were magnetically washed
in 0.05 vol % Tween-20 (Sigma-Aldrich) in PBS and resuspended in 0.5
M NaCl in PBS to a final concentration of 0.1 mg/mL and sonicated
using an ultrasonic probe (Hielscher).
BPM Assay
A total
of 25 μL of particle solution
was added to the fluid cell and incubated for 10 min. After incubation,
the fluid cell was reversed causing unbound particles to sediment.
After washing with 40 μL of 0.5 M NaCl in PBS, 40 μL of
mPEG-biotin (500 μM, PG1-BN-1k, Nanocs) in 0.5 M NaCl in PBS
was added to the fluid cell. After 15 min of incubation, the fluid
cell was washed twice with 40 μL of PBS. A mixture of ssDNA
analyte molecules (IDT, standard desalting) and free binder molecules
in PBS was added to the flow cell at the required concentration, immediately
after preparation. The sample was observed under a white light source
using a microscope (Leica DMI5000M) with a dark-field illumination
setup at a total magnification of 10× (Leica objective, N plan
EPI 10×/0.25 BD). A field of view of approximately 1100 ×
700 μm2 with a few thousand particles was imaged
using a CMOS camera (FLIR, Grasshopper3, GS3-U3-23S6M-C) with an integration
time of 5 ms and a sampling frequency of 30 Hz. The particles were
tracked by applying a phasor-based localization method.[31] The particle activity was determined from the x- and y-trajectories of all particles,
by applying a maximum-likelihood multiple-windows change point detection
algorithm.[32] The particle activity at equilibrium
and the time-to-equilibrium were extracted by fitting the measured
particle activity over time using the equation given in Box S1.
Results and Discussion
Timescales
of the Limited-Volume Assay
Figure shows simulation results of
the time-to-equilibrium of the limited-volume assay, for sensor designs
with different measurement chamber heights, different binder densities,
and different flow rates, assuming standard parameter values, as listed
in Table . Figure a shows how the time-to-equilibrium
τ depends on the measurement chamber height H, for a sensor with instantaneous analyte exchange (see Note S6 for the influence of analyte exchange
on the sensor performance). The arrow on the x-axis
indicates the height as listed in Table . The data show that the time-to-equilibrium
increases with the measurement chamber height. At small H, this increase is caused by a decrease in the effective volumetric
binder concentration, while at large H, this increase
is caused by diffusive transport limitations. The inset shows the
same data, plotted as a function of the Damköhler number (Da = τD/τR,LV = konΓbH/D, see Table ); low Da means that the kinetics are limited by
the reaction, and high Da means that the kinetics
are limited by diffusion. To achieve a fast time-to-equilibrium, the
sensor should be designed with a large Cb,0, so a small H.
Figure 2
Time-to-equilibrium τ of a limited-volume
assay for sensor
designs with different heights, binder densities, and flow rates of
analyte exchange. (a) Time-to-equilibrium τ as a function of
measurement chamber height H (orange line) for instantaneous
analyte exchange. For small H, the observed τ
is reaction-dominated (black dotted line), while for increasing H, the observed τ becomes diffusion-dominated. The
inset shows the same data, where τ is normalized to τR and plotted as a function of Damköhler number Da. The sketch above the graph visualizes a measurement
chamber with an increasing H. (b) Time-to-equilibrium
τ as a function of the binder density Γb (orange
line) for an instantaneous analyte exchange. For low Γb, the observed τ is reaction-dominated (τ = τR, black dotted line), while for increasing Γb, the observed τ becomes diffusion-dominated. The inset shows
the same data, where τ is normalized to the characteristic diffusion
time τD and plotted as a function of Da. The sketch above the graph visualizes a measurement chamber with
an increasing Γb. (c) Time-to-equilibrium τ
as a function of flow rate Q for three aspect ratios
λ = L/H, for time-controlled
analyte exchange by advection where the flow duration texch equals the characteristic advection time τA. For small Q, the observed τ is limited
by the advective transport of analyte molecules from the inlet toward
the point of sensing at distance L/2 from the inlet.
For increasing Q, this transport process becomes
faster causing the observed τ to be dominated by reaction and/or
diffusion at high flow rates. The inset shows the same data (Da = 2) supplemented with Da = 0.2 (reaction-limited)
and Da = 20 (diffusion-limited), where τ is
normalized to τR and plotted as a function of the
longitudinal Péclet number PeL.
The dotted lines show the τ/τR value at high Q and are equal to the values found in panel a. The sketch
above the graph visualizes a measurement chamber with an increasing Q. In all panels, the black arrows on the x-axis indicate the standard parameter values for H, Γb, and Q, as listed in Table .
Time-to-equilibrium τ of a limited-volume
assay for sensor
designs with different heights, binder densities, and flow rates of
analyte exchange. (a) Time-to-equilibrium τ as a function of
measurement chamber height H (orange line) for instantaneous
analyte exchange. For small H, the observed τ
is reaction-dominated (black dotted line), while for increasing H, the observed τ becomes diffusion-dominated. The
inset shows the same data, where τ is normalized to τR and plotted as a function of Damköhler number Da. The sketch above the graph visualizes a measurement
chamber with an increasing H. (b) Time-to-equilibrium
τ as a function of the binder density Γb (orange
line) for an instantaneous analyte exchange. For low Γb, the observed τ is reaction-dominated (τ = τR, black dotted line), while for increasing Γb, the observed τ becomes diffusion-dominated. The inset shows
the same data, where τ is normalized to the characteristic diffusion
time τD and plotted as a function of Da. The sketch above the graph visualizes a measurement chamber with
an increasing Γb. (c) Time-to-equilibrium τ
as a function of flow rate Q for three aspect ratios
λ = L/H, for time-controlled
analyte exchange by advection where the flow duration texch equals the characteristic advection time τA. For small Q, the observed τ is limited
by the advective transport of analyte molecules from the inlet toward
the point of sensing at distance L/2 from the inlet.
For increasing Q, this transport process becomes
faster causing the observed τ to be dominated by reaction and/or
diffusion at high flow rates. The inset shows the same data (Da = 2) supplemented with Da = 0.2 (reaction-limited)
and Da = 20 (diffusion-limited), where τ is
normalized to τR and plotted as a function of the
longitudinal Péclet number PeL.
The dotted lines show the τ/τR value at high Q and are equal to the values found in panel a. The sketch
above the graph visualizes a measurement chamber with an increasing Q. In all panels, the black arrows on the x-axis indicate the standard parameter values for H, Γb, and Q, as listed in Table .Figure b shows
how the time-to-equilibrium depends on the binder density Γb, for a sensor with instantaneous analyte exchange. The arrow
indicates the density, as listed in Table . For small Γb, the time-to-equilibrium
is long and determined by the dissociation rate constant (τ
≅ 1/koff). For Γb > HKd ≅ 20 μm–2, the time-to-equilibrium decreases, until it stabilizes due to diffusive
transport limitations (τ ≅ τD = H2/D). The inset shows the same
data plotted as a function of Da. To achieve a fast
time-to-equilibrium, the sensor should be designed with a large Cb,0, so a large Γb.Figure c shows
how analyte exchange by advection contributes to the time-to-equilibrium
per measurement cycle. The exchange phase involves a temporary flow
of fluid into the measurement chamber, with flow rate Q and duration texch (see Note S6 for the influence of analyte exchange
on the sensor performance). In the simulations, texch was chosen to be equal to the characteristic advection
time τA = HLW/Q (see Table ), which
means that a total fluid volume equal to the volume of the measurement
chamber is displaced. The time-to-equilibrium τ, which now includes
a contribution texch related to the exchange,
is shown as a function of flow rate, for several values of the chamber
aspect ratio λ = L/H. The
arrow indicates the flow rate, as listed in Table . For small Q, the observed
τ is limited by texch, that is,
the advective transport time of analyte molecules from the inlet toward
the point of sensing at a distance L/2 from the inlet,
as sketched in Figure d. For increasing λ, that is, increasing L with a fixed H, the time-to-equilibrium increases
since τA (and thus also texch) increases. For increasing Q, the time-to-equilibrium
decreases, until it stabilizes at a level where the reaction and diffusion
times determine the observed τ. The inset shows the same data
(Da = 2), supplemented with Da =
0.2 (reaction-limited) and Da = 20 (diffusion-limited),
plotted as a function of the longitudinal Péclet number (PeL = τD/τA = Q/λDW, see Table S1); low PeL means that the analyte exchange is limited by advection, and high PeL means that the analyte exchange is limited
by diffusion. A low PeL causes a long
time-to-equilibrium due to slow mass transport by advection. Increasing PeL results in a decrease in the time-to-equilibrium
due to rapid filling of the chamber, until it stabilizes at a τ
value equal to the value indicated in Figure a. Figure c shows the flow rate required to minimize the influence
of the exchange process on the time-to-equilibrium. In the following
sections, exchange with a high PeL is
assumed, that is, rapid filling of the measurement chamber without
influence of the flow rate on the time-to-equilibrium.
Limited-Volume
Assay with Time-Controlled Analyte Exchange
Figure shows simulation results for a limited-volume assay
with time-controlled analyte exchange. The analyte exchange is assumed
to be instantaneous and the incubation phase includes mass transport
by diffusion and reaction kinetics within the measurement chamber
itself but no analyte exchange between the system of interest and
the measurement chamber. Figure a shows data for repeated incubations with Ca,0 = 0.1 pM. The analyte concentration Ca in the measurement chamber (brown line) and
the fractional occupancy f of the binders by analyte
molecules (orange line) are plotted as a function of time, for conditions
of analyte depletion (left) and analyte repletion (right). The time-to-equilibrium
τ of each incubation equals approximately 340 s (see Figure ), having contributions
from reaction (τR = 200 s) and diffusion (τD = 400 s). The contribution from the reaction to the time-to-equilibrium
is much smaller than 1/koff = 104 s, the value that would have been observed in the case of an infinite-volume
assay (cf. Figure c). In the absence of diffusion limitations, the acceleration that
can be achieved with a limited-volume assay compared to an infinite-volume
assay equals , which clarifies how the speed of the assay
is directly related to the ratio between effective volumetric binder
concentration and the equilibrium dissociation constant.Simulated response
of the biomolecular monitoring system using
time-controlled analyte exchange. (a) Analyte concentration Ca in the measurement chamber (brown line) and
the fractional occupancy f of binder molecules by
analyte molecules (orange line) as a function of time, for low finit and depletion of analyte in solution (left),
and for high finit and repletion of analytes
in solution (right). The dashed lines indicate time points where instantaneous
analyte exchange occurs, where the bulk analyte concentration was
set to Ca,0 = 0.1 pM after each period
of approximately 50 min. The insets highlight the kinetics of the
first cycle, showing a time-to-equilibrium of τ = 340 s. For
many cycles (n → ∞), both curves would
approach , which equals the equilibrium value when
an infinite volume is supplied (see Table S1). (b) Fractional occupancy f as a function of time
where cycles of analyte exchange and incubation are applied every
15 min with alternatingly Ca,0 = 0.15
pM and Ca,0 = 0.05 pM. The curve saturates
at feq,IV = 10·10–4, which equals the infinite-volume equilibrium value for the average
concentration value Ca,0 = 0.1 pM (see Table S1). Dashed lines: continuous supply of Ca,0 = 0.05 pM yields feq,IV = 5·10–4 and Ca,0 = 0.15 pM yields feq,IV = 15·10–4 (see Table S1). The right panel shows zoom-in images of three sections
of the solid curve, each representing four cycles of instantaneous
analyte exchange and subsequent incubations of 15 min. In zoom-in
1 (t = 0–1 h), all curve segments show depletion
behavior. In zoom-in images 2 (t = 12–13 h)
and 3 (t = 42–43 h), depletion is seen for Ca,0 = 0.15 pM, since finit < feq,IV(Ca,0 = 0.15 pM), and repletion is seen for Ca,0 = 0.05 pM, since finit > feq,IV(Ca,0 = 0.05 pM). For all curve segments, the time-to-equilibrium τ
= 340 s. The vertical scale bars indicate Δf = 10–4.Figure b shows
the response of a limited-volume assay with time-controlled analyte
exchange for an analyte concentration that varies in time. As an example,
the sensor is incubated with a time series of analyte concentrations
that alternate around 0.1 pM (orange line): the analyte concentration
is alternatingly Ca,0 = 0.05 pM or Ca,0 = 0.15 pM. The infinite-volume equilibrium
fractional occupancy feq,IV is given for Ca,0 = 0.15 pM and Ca,0 = 0.05 pM by the dashed black lines (see Table S1). The panels on the right show zoom-in images of the sensor
response at three different time periods (starting at t = 0, 12, and 42 h). In all cases, the time-to-equilibrium is τ
= 340 s = 5.7 min. Incubation with Ca,0 = 0.15 pM gives depletion behavior at all times (since finit < feq,IV, top black
line); for Ca,0 = 0.05 pM, depletion behavior
is seen at t < 10 h and repletion at t > 10 h (when finit > feq,IV, bottom black line).Figure shows an
experimental study on how the time-to-equilibrium in a limited-volume
assay depends on the total binder concentration in the measurement
chamber. Here, the total binder concentration has two contributions,
namely, a contribution from surface-bound binders and a contribution
from binders supplemented in solution. For detection, we made use
of BPM, which is a biomolecular monitoring principle with single-molecule
resolution. In the BPM sensor, the particles are transducers that
record the binding of analyte molecules to specific binder molecules
(see Note S7). Figure a shows a schematic representation of a measurement
chamber with binder molecules present in the two forms: immobilized
and nonimmobilized. Immobilized binder molecules are present with
an effective volumetric concentration Cb,0. Binder molecules supplemented free in solution have concentration Cb,suppl. In the absence of supplemented binder
molecules (top), the total binder concentration in the measurement
chamber equals Cb,tot = Cb,0 = Γb/H. In the presence
of supplemented binder molecules (bottom), the total binder concentration
equals Cb,tot = Γb/H + Cb,suppl. Since the time-to-equilibrium
of the reaction scales according to τR,LV ∝
1/Cb,tot (see Table ), an increasing supplemented binder concentration Cb,suppl results in a smaller τ. Figure b shows the measured
time-to-equilibrium τ (left) and the signal change ΔS (right) as a function of Cb,suppl, for an analyte concentration of 200 pM (see the Experimental Section). The data show that the time-to-equilibrium
decreases for increasing Cb,suppl. The
measured signal change decreases with increasing supplemented binder
concentration because only surface-captured analyte molecules generate
a measurable signal. The dashed lines in Figure b represent model fits (see the caption),
demonstrating a good correspondence between model and experimental
results. We conclude that the measurements of Figure prove the basic concept of the sensing methodology
proposed in this paper, namely, that a limited-volume design with
time-controlled analyte exchange allows one to control the response
time by tuning the concentration of binder molecules in the measurement
chamber. Once the optimal binder concentration is known, the sensor
can be made with binders immobilized in the measurement chamber. This
will be a topic for future research.
Figure 4
Experimental study of a limited-volume
assay with varying binder
concentrations using BPM. (a) Sketch of the measurement chamber in
a BPM measurement (see Note S7) without
(top) and with (bottom) supplemented binders with concentration Cb,suppl. For simplicity, the particles of the
BPM sensor are not shown in the sketch. (b) Experimentally observed
time-to-equilibrium τ (left) and normalized signal change ΔS (right) as a function of supplemented binder concentration Cb,suppl in a BPM measurement. The model assay
is based on DNA–DNA interactions, with ssDNA analytes at a
concentration of 200 pM (see Note S7).
Left: the dashed line shows the fitted curve τ = p1/(p2 + Cb,suppl) + p3, where p1 = 1/kon (kon is assumed to be equal for all binders), p2 = Γb/H,
and p3 is the delay contributed by diffusion
(see τD, black line, see also Figure b) and experimental steps (see the Experimental Section). Assuming H = 200 μm (see Table ), the fit gives Γb=(3 ± 1)·10–10 mol m–2, which is comparable to
the standard parameter value, as listed in Table . The fitted association rate constant is kon=(1.5 ± 0.4)·105 M–1 s–1, which is in the range of values
reported in the literature for comparable DNA–DNA hybridization
reactions.[33,34] Right: In the depletion condition
(finit < feq,IV), the fractional occupancy scales according to f ∝ 1/Cb,tot = Ca,0/(Cb,0 + Cb,suppl). The dashed line shows the fitted curve ΔS = p1/(p2 + Cb,suppl), where p1 scales the change in fractional occupancy to signal
change and p2 = Γb/H. For H = 200 μm, it was found that
Γb = (7 ± 4)·10–10 mol
m–2, which is comparable to the previously found
value for Γb and the standard parameter value, as
listed in Table .
The insets show the same data on linear-logarithmic scales. The errors
reported in the figure (smaller than the symbol size) and the caption
are fitting errors based on a 68% confidence interval.
Experimental study of a limited-volume
assay with varying binder
concentrations using BPM. (a) Sketch of the measurement chamber in
a BPM measurement (see Note S7) without
(top) and with (bottom) supplemented binders with concentration Cb,suppl. For simplicity, the particles of the
BPM sensor are not shown in the sketch. (b) Experimentally observed
time-to-equilibrium τ (left) and normalized signal change ΔS (right) as a function of supplemented binder concentration Cb,suppl in a BPM measurement. The model assay
is based on DNA–DNA interactions, with ssDNA analytes at a
concentration of 200 pM (see Note S7).
Left: the dashed line shows the fitted curve τ = p1/(p2 + Cb,suppl) + p3, where p1 = 1/kon (kon is assumed to be equal for all binders), p2 = Γb/H,
and p3 is the delay contributed by diffusion
(see τD, black line, see also Figure b) and experimental steps (see the Experimental Section). Assuming H = 200 μm (see Table ), the fit gives Γb=(3 ± 1)·10–10 mol m–2, which is comparable to
the standard parameter value, as listed in Table . The fitted association rate constant is kon=(1.5 ± 0.4)·105 M–1 s–1, which is in the range of values
reported in the literature for comparable DNA–DNA hybridization
reactions.[33,34] Right: In the depletion condition
(finit < feq,IV), the fractional occupancy scales according to f ∝ 1/Cb,tot = Ca,0/(Cb,0 + Cb,suppl). The dashed line shows the fitted curve ΔS = p1/(p2 + Cb,suppl), where p1 scales the change in fractional occupancy to signal
change and p2 = Γb/H. For H = 200 μm, it was found that
Γb = (7 ± 4)·10–10 mol
m–2, which is comparable to the previously found
value for Γb and the standard parameter value, as
listed in Table .
The insets show the same data on linear-logarithmic scales. The errors
reported in the figure (smaller than the symbol size) and the caption
are fitting errors based on a 68% confidence interval.
Analytical Performance of the Limited-Volume Assay
Figure shows how the analytical performance of the limited-volume
assay depends on the sensor design. The results are based on numerical
simulations with parameters, as listed in Table . The analyte exchange is assumed to be instantaneous
and the incubation includes mass transport by diffusion and reaction
kinetics within the measurement chamber only. All panels show curves
for different values of the initial fractional occupancy finit of the binder molecules.Analytical performance
of the limited-volume assay, derived from
simulations of a single measurement cycle. (a) Fractional occupancy
at the end of the incubation fend as a
function of analyte concentration Ca,0 for different initial fractional occupancies finit. The right y-axis indicates the number
of surface-bound analyte molecules at the end of the cycle γabend. (b) Absolute
change in fractional occupancy Δf as a function
of Ca,0 for various finit. The right y-axis indicates Δγab. A positive Δf and Δγab indicate depletion; negative values indicate repletion.
Note that the lines for finit = 10–3 and lower are overlapping. The inset shows the same
data on a linear–linear scale. (c) Coefficient of variation
CVC with which the analyte concentration Ca,0 can be determined as a function of analyte concentration Ca,0 for various initial fractional occupancies finit. CVC scales as for low finit and high Ca,0; CVC scales
as 1/Ca,0 for high finit and low Ca,0. The horizontal
dashed line indicates the 10% CV level that defines the LoQ (limit
of quantification). (d) CVC as a function of measurement
chamber height H (top) and binder density Γb (bottom) for various initial fractional occupancies finit and Ca,0 =
0.1 pM. The arrows on the x-axes indicate the standard
parameter values for H and Γb, as
listed in Table .
(e) CVC as a function of the observed time-to-equilibrium
τ when varying the measurement chamber height H (left) or binder density Γb (right) for various
initial fractional occupancies finit and Ca,0 = 0.1 pM. The sketches above the graphs
visualize a measurement chamber with an increasing height or a decreasing
binder density. The arrows on the x-axes indicate
the obtained time-to-equilibrium using the standard parameter values
for H and Γb, as listed in Table .Figure a shows
the fractional occupancy of binders by analyte molecules at the end
of the incubation (fend) as a function
of the input analyte concentration Ca,0. For finit = 0 (dashed black line), fend scales linearly with the analyte concentration,
which makes the sensor suitable for analyte quantification. For larger
values of finit, the curves start with
a rather flat segment, from which one might erroneously conclude that
under those conditions, low analyte concentrations cannot be determined.
Interestingly, the limited-volume assay has a linear dependence on
concentration by focusing not on the absolute value of fend but rather on the change in fractional occupancy Δf (see Note S2)This equation shows that Δf depends linearly
on Ca,0, independent of the value of finit. This fact is also illustrated by the simulation
results in Figure b. The response scales linearly with concentration Ca,0 and is downshifted for increasing values of finit, in agreement with eq (note that the steep increase in the curves
relates to the logarithmic x-axis). Positive values
of Δf relate to depletion behavior and negative
values to repletion. The curves cross the x-axis
(Δf = 0) when finit corresponds to the equilibrium condition, that is, when there is
no net association or dissociation during incubation because finit is equal to the equilibrium fractional
occupancy of the infinite-volume case: . For example, the curve for finit = 10–3 crosses Δf = 0 at Ca,0 = finitKd = 0.1 pM, as is highlighted
in the inset of Figure b.Figure c
shows
the precision of the concentration output of the sensor, that is,
the precision with which the analyte concentration in an unknown sample
can be determined for a signal collection area of 1 mm2. The precision is calculated based on Poisson noise, which gives
the fundamental limit of the precision that is achievable with a molecular
biosensor due to stochastic fluctuations in the number of analyte
molecules (see Note S3 and S8).[35] To calculate the precision, a sensor with initial
fractional occupancy finit is provided
with a sample with analyte concentration Ca,0, resulting in a Δf with variability σΔ, which via the slope of the calibration
curve, given in Figure b, leads to a variability σC in the concentration
output of the sensor (see Note S3). The
precision is indicated as the concentration-based coefficient of variation
CVC = σC/μC, with σC being the variability and μC being the mean
of the concentration output. Figure c shows how the concentration precision depends on
the analyte concentration and the initial fractional occupancy finit. For finit =
0 (dashed line), the CVC scales as , in agreement with number fluctuations
in a Poisson process (see Note S3). For
higher finit, a stronger dependency is
observed (CVC ∝ 1/Ca,0), caused by the smaller relative change in the fractional occupancy
(see Note S3). The graph indicates the
10% precision level that is used to define the limit of quantification
(LoQ) of the sensor (see Figure c, dashed horizontal line). The results show that analyte
concentrations in the sub-picomolar range can be measured with a precision
better than 10%, even for high initial fractional occupancies.Figure d shows
the precision of the concentration output of the sensor as a function
of two design parameters, namely, the measurement chamber height H (top panel) and the binder density Γb (bottom panel), at an analyte concentration Ca,0 = 0.1 pM, for an initial fractional occupancy finit between 0 and 0.01. The arrows indicate the height
and density, as listed in Table . For an increasing H, a decrease
in CVC is observed, caused by an increase in the number
of analyte molecules present in the measurement chamber. The CVC is the smallest for finit = 0
and increases for increasing finit since
the absolute change in fractional occupancy decreases. The CVC decreases for increasing Γb caused by an
increase in the number of analyte molecules captured from solution.
The CVC reaches a plateau for finit = 0 due to a limited number of analyte molecules in the measurement
chamber. For larger finit, the absolute
change in fractional occupancy decreases and causes a less precise
concentration determination; this effect is in particular visible
at high Γb where the absolute number of analyte–binder
complexes increases due to finit.The trade-off between precision and time-to-equilibrium is illustrated
in Figure e, for sensors
with different heights of the measurement chamber (left) and different
binder densities (right). The arrows indicate the time-to-equilibrium
that results from the height and density, as listed in Table . The left panel shows that
an increase in H gives, on the one hand, a slower
sensor response (due to a larger diffusion distance) but, on the other
hand, a lower CVC due to a larger number of analyte molecules
present in the measurement chamber. At low H, the
CVC strongly depends on finit due to the low number of analyte molecules in the solution. The
right panel shows again that the CVC decreases for a slower
sensor response, now controlled by decreasing the binder density Γb. At high Γb, the time-to-equilibrium is
diffusion-limited (resulting in τ = 130 s). At low Γb, the time-to-equilibrium is reaction-limited with τ
= 1/koff = 104 (see Figure b). At high Γb, the CVC increases for increasing finit due to the larger amount of analyte molecules on
the sensor surface. At low Γb, the CVC strongly increases due to the small absolute change in the number
of analyte−binder complexes during the incubation phase.
Conclusions
We have presented a sensing methodology suitable
for monitoring
low-concentration biomolecules with high precision, with small time
delays and short time intervals, over an endless time span. The sensing
methodology is based on a limited-volume assay, using high-affinity
binders, a fully reversible detection principle, and time-controlled
analyte exchange. We studied by simulations how the kinetics of the
sensor depend on mass transport and on the surface reaction in the
measurement chamber, and how time-controlled analyte exchange determines
the system response and enables precise measurements of analyte concentration.
Experimental results show the ability to control the sensor response
time by tuning the total binder concentration in the measurement chamber.
Finally, simulations show that the sensing principle allows picomolar
and sub-picomolar concentrations to be monitored with a high precision
over long time spans.Approaches described in the literature
for measuring low-concentration
biomolecules have focused primarily on assays in which every concentration
determination involves consumption of reagents.[17−20,36,37] When numbers of assays become high, due
to frequent measurements over long time spans, then reagent-consuming
approaches are complex and costly. The sensing methodology described
in this paper is based on a fully reversible assay principle, without
consuming reagents with each newly recorded concentration data point,
enabling measurements with high frequency over an endless time span.
The described assay principle can be implemented on several sensing
platforms, for example, based on optical, electrical, or acoustical
transduction methods, where especially sensing platforms with single-molecule
resolution seem suitable since these allow digital measurements with
very high precision, limited only by the number of observed molecular
interactions. The sensing method is suited for the monitoring of a
wide variety of analytes, including small molecules, proteins, and
viral particles (see Note S9). Furthermore,
the sensing methodology can be combined with various sampling methods,
including remote advection-based sampling through a sampling line
or a catheter and proximal diffusion-based sampling methods for on-body
and in-body monitoring devices. The presented sensing principle warrants
further experimental studies, for example, to investigate trade-offs
between time characteristics and precision, for various transduction
methods, sampling methods, measurement chamber geometries, binder
types, and for various complex biological matrices. Due to its sensing
performance and generalizability, we believe that the limited-volume
assay with time-controlled analyte exchange will enable research on
time dependencies of low-concentration biomolecules and novel applications
in the fields of dynamic biological systems, patient monitoring, and
biotechnological process control.
Authors: Yu Shrike Zhang; Julio Aleman; Su Ryon Shin; Tugba Kilic; Duckjin Kim; Seyed Ali Mousavi Shaegh; Solange Massa; Reza Riahi; Sukyoung Chae; Ning Hu; Huseyin Avci; Weijia Zhang; Antonia Silvestri; Amir Sanati Nezhad; Ahmad Manbohi; Fabio De Ferrari; Alessandro Polini; Giovanni Calzone; Noor Shaikh; Parissa Alerasool; Erica Budina; Jian Kang; Nupura Bhise; João Ribas; Adel Pourmand; Aleksander Skardal; Thomas Shupe; Colin E Bishop; Mehmet Remzi Dokmeci; Anthony Atala; Ali Khademhosseini Journal: Proc Natl Acad Sci U S A Date: 2017-03-06 Impact factor: 11.205
Authors: Krzysztof Bielec; Krzysztof Sozanski; Marco Seynen; Zofia Dziekan; Pieter Rein Ten Wolde; Robert Holyst Journal: Phys Chem Chem Phys Date: 2019-05-14 Impact factor: 3.676
Authors: John Todd; Bob Freese; Ann Lu; Douglas Held; Jennifer Morey; Richard Livingston; Philippe Goix Journal: Clin Chem Date: 2007-09-21 Impact factor: 8.327
Authors: Leonardo B Pinheiro; Victoria A Coleman; Christopher M Hindson; Jan Herrmann; Benjamin J Hindson; Somanath Bhat; Kerry R Emslie Journal: Anal Chem Date: 2011-12-21 Impact factor: 6.986
Authors: Alissa D Buskermolen; Yu-Ting Lin; Laura van Smeden; Rik B van Haaften; Junhong Yan; Khulan Sergelen; Arthur M de Jong; Menno W J Prins Journal: Nat Commun Date: 2022-10-13 Impact factor: 17.694