Rafiq M Lubken1,2, Arthur M de Jong3,2, Menno W J Prins1,3,2,4. 1. Department of Biomedical Engineering, Eindhoven University of Technology, Eindhoven 5612 AZ, the Netherlands. 2. Institute for Complex Molecular Systems (ICMS), Eindhoven University of Technology, Eindhoven 5612 AZ, the Netherlands. 3. Department of Applied Physics, Eindhoven University of Technology, Eindhoven 5612 AZ, the Netherlands. 4. Helia Biomonitoring, Eindhoven 5612 AZ, the Netherlands.
Abstract
Sensors for monitoring biomolecular dynamics in biological systems and biotechnological processes in real time, need to accurately and precisely reconstruct concentration-time profiles. This requirement becomes challenging when transport processes and biochemical kinetics are important, as is typically the case for biomarkers at low concentrations. Here, we present a comprehensive methodology to study the concentration-time profiles generated by affinity-based sensors that continuously interact with a biological system of interest. Simulations are performed for sensors with diffusion-based sampling (e.g., a sensor patch on the skin) and advection-based sampling (e.g., a sensor connected to a catheter). The simulations clarify how transport processes and molecular binding kinetics result in concentration gradients and time delays in the sensor system. Using these simulations, measured and true concentration-time profiles of insulin were compared as a function of sensor design parameters. The results lead to guidelines on how biomolecular monitoring sensors can be designed for optimal bioanalytical performance in terms of concentration and time properties.
Sensors for monitoring biomolecular dynamics in biological systems and biotechnological processes in real time, need to accurately and precisely reconstruct concentration-time profiles. This requirement becomes challenging when transport processes and biochemical kinetics are important, as is typically the case for biomarkers at low concentrations. Here, we present a comprehensive methodology to study the concentration-time profiles generated by affinity-based sensors that continuously interact with a biological system of interest. Simulations are performed for sensors with diffusion-based sampling (e.g., a sensor patch on the skin) and advection-based sampling (e.g., a sensor connected to a catheter). The simulations clarify how transport processes and molecular binding kinetics result in concentration gradients and time delays in the sensor system. Using these simulations, measured and true concentration-time profiles of insulin were compared as a function of sensor design parameters. The results lead to guidelines on how biomolecular monitoring sensors can be designed for optimal bioanalytical performance in terms of concentration and time properties.
Biological
systems and biotechnological
processes exhibit time dependencies that are imposed by dynamic changes
of constituting biomolecules, such as nutrients, hormones, proteins,
and nucleic acids. To study dynamic processes in real time, monitoring
sensors that can reveal biomolecular concentration–time profiles
are needed, to support fundamental research,[1−7] patient monitoring,[8−14] and closed-loop control applications.[15−22] Such monitoring sensors should be able to reconstruct concentration–time
profiles accurately and precisely, in both concentration and time,
and the sensors should be suitable for measuring a wide variety of
molecular markers.The developments in biomolecular monitoring
have mainly focused
on measuring high-concentration metabolites, such as glucose and lactate.[8,12,13] Due to their small size and high
concentrations, the transport and detection of these biomolecules
is fast. However, in the case of biomolecular markers at lower concentrations,
fewer molecules are available and transport limitations become important.[22] Furthermore, biochemical reactions are slow
at low concentrations,[23] generating time
delays in the sensors and time-related errors in the concentration
results.To understand and predict how real-time monitoring
of biomolecules
is limited by dynamic processes, we present a comprehensive methodology
for studying affinity-based sensors that continuously interact with
a time-dependent system of interest. Here, concentration changes,
which are present in a system of interest, propagate into a monitoring
sensor by diffusion-based sampling or advection-based sampling. We
focus on sensing by biochemical affinity between binder molecules
and analyte molecules since this is a very generic molecular mechanism
for achieving specific and sensitive measurements. Frequency-dependent
simulations are presented to clarify how concentration gradients and
time delays are caused by mass transport processes and molecular binding
kinetics. The results lead to relationships between sensor design
parameters and measurable concentration change rates, time delays,
and concentration errors. This will help researchers to design biomolecular
sensors for optimal bioanalytical performance in terms of concentration
and time properties.
Biomolecular Monitoring with Continuous Analyte
Exchange
The conceptual layout of the monitoring arrangement
is sketched
in Figure . Figure a shows continuous
analyte exchange between a biological or biotechnological system of
interest and a measurement chamber. The system of interest exhibits
dynamic changes of analyte concentration where the sensing aim is
to achieve minimal differences between the true concentration–time
profile Ca,0(t) and the
measured concentration–time profile Ca,0m(t). The basic modeling approach is to study analyte concentrations
that vary with a sinusoidal time dependence around a mean concentration
valuewith C(t) being the oscillating concentration–time profile, C the mean concentration, ΔC the
top-to-top amplitude of concentration change, f the
oscillation frequency, and ϕ the phase. In the analysis, the
concentration change ΔC is a small perturbation
on the mean value C (a few percent). The advantage
of studying sinusoidal functions is that concentration–time
profiles of arbitrary shape can be reconstructed by frequency decomposition,
as will be discussed later in this paper. Concentration symbols with
subscript “a” refer to analyte concentrations: the analyte
concentration–time profile in the system of interest is denoted
by Ca,0(t), at the sensor
surface by Ca(t), and
the measured analyte concentration–time profile by Ca,0m(t). From the simulations in this paper, it will
become apparent that the response of the monitoring system resembles
a low-pass filter: at low frequencies, the measured and true concentration–time
profiles are close to each other; however, at frequencies higher than
a cutoff frequency fc , the measured
concentration–time profile deviates from the true concentration–time
profile, visible in the concentration change ΔC and in the lag time Δt that corresponds to
the phase lag ϕ.
Figure 1
Conceptual layout of a biomolecular monitoring system
with continuous
analyte exchange. (a) Biomolecular monitoring system with continuous
analyte exchange between a system of interest and a measurement chamber,
where the system of interest exhibits a dynamic concentration–time
profile Ca,0(t) (gray
line), which results in a measured concentration–time profile Ca,0m(t) (orange line). Ideally, the measured concentration–time
profile closely resembles the true concentration–time profile
(dashed line vs solid line). The monitoring system
can be mimicked by a low-pass filter with a cutoff frequency fc. The system of interest supplies an oscillating
concentration–time profile Ca,0(t) with concentration change ΔCa,0, which leads to a measured concentration Ca,0m(t) with concentration change ΔCa,0m. A comparison
of the true and measured concentration–time profiles (dashed
line vs solid line) gives the system response in
terms of the concentration change ratios and lag time Δt. (b) Geometry of the measurement chamber with height H, width W, and length L. The signal of the sensor is generated by an affinity reaction at
the sensor surface, where analyte molecules (orange) associate with
and dissociate from binder molecules (brown), of which the reaction
rates are described by the association rate constant kon, the dissociation rate constant koff, the binder density Γb, the concentration–time
profile Ca(t) at the
sensor surface, and the analyte–binder complex density γab. Two modes of continuous analyte exchange are studied: analyte
exchange by diffusion (top) and by advection (bottom). In the measurement
chamber, mass transport by diffusion occurs in both x- and y-direction, caused by a concentration gradient
(orange gradient) that results in a net molecular flux Ja, which scales with the diffusion coefficient D. Mass transport by advection occurs in the x-direction only, caused by a flow with mean flow velocity vm and flow rate Q. (c) Examples
of how the sensor performance can differ for different sensor design
parameter sets.
Conceptual layout of a biomolecular monitoring system
with continuous
analyte exchange. (a) Biomolecular monitoring system with continuous
analyte exchange between a system of interest and a measurement chamber,
where the system of interest exhibits a dynamic concentration–time
profile Ca,0(t) (gray
line), which results in a measured concentration–time profile Ca,0m(t) (orange line). Ideally, the measured concentration–time
profile closely resembles the true concentration–time profile
(dashed line vs solid line). The monitoring system
can be mimicked by a low-pass filter with a cutoff frequency fc. The system of interest supplies an oscillating
concentration–time profile Ca,0(t) with concentration change ΔCa,0, which leads to a measured concentration Ca,0m(t) with concentration change ΔCa,0m. A comparison
of the true and measured concentration–time profiles (dashed
line vs solid line) gives the system response in
terms of the concentration change ratios and lag time Δt. (b) Geometry of the measurement chamber with height H, width W, and length L. The signal of the sensor is generated by an affinity reaction at
the sensor surface, where analyte molecules (orange) associate with
and dissociate from binder molecules (brown), of which the reaction
rates are described by the association rate constant kon, the dissociation rate constant koff, the binder density Γb, the concentration–time
profile Ca(t) at the
sensor surface, and the analyte–binder complex density γab. Two modes of continuous analyte exchange are studied: analyte
exchange by diffusion (top) and by advection (bottom). In the measurement
chamber, mass transport by diffusion occurs in both x- and y-direction, caused by a concentration gradient
(orange gradient) that results in a net molecular flux Ja, which scales with the diffusion coefficient D. Mass transport by advection occurs in the x-direction only, caused by a flow with mean flow velocity vm and flow rate Q. (c) Examples
of how the sensor performance can differ for different sensor design
parameter sets.The measurement chamber is assumed
to be rectangular with height H, width W, and length L (see Figure b).
The bottom surface of the measurement chamber is a sensor surface
with affinity binder molecules (brown), where association and dissociation
occur of the analyte molecules (orange). The association and dissociation
rates depend on the association rate constant kon, the dissociation rate constant koff, the binder surface density Γb, the analyte concentration–time
profile Ca(t) at the
sensor surface, and the surface density of analyte–binder complexes
γab. The binding of analyte molecules to binder molecules
on the sensor surface causes γab to change as a function
of time, resulting in a time-dependent signal, which relates to the
oscillating analyte concentration Ca,0(t) in the system of interest (see Supplementary Note 1).We study two modes of continuous
analyte exchange, namely, analyte
exchange by diffusion only (top sketch) and analyte exchange by advection
as well as diffusion (bottom sketch). Diffusion-based sampling applies
to a sensor that is worn on the skin or that is fully embedded in
a bioreactor, for example.[8−10] Advection-based sampling applies
to a sensor that is connected to a patient via a catheter or that
is connected to a bioreactor via a sampling line.[19−21] In the case
of diffusion-based sampling, a net molecular flux Ja is caused by a concentration difference (orange gradient),
facilitating mass transport between the system of interest and the
measurement chamber. In the case of advection-based sampling, a laminar
flow with flow rate Q facilitates mass transport
between the system of interest and the measurement chamber. In the
simulations, it is assumed that diffusion occurs in both the longitudinal
(x-direction) and lateral directions (y-direction) and scales with the diffusion coefficient D. In the case of advective exchange, the diffusive transport is superposed
onto the advective transport caused by a flow, of which the transport
scales with the mean flow velocity vm and
thus the flow rate Q. In this paper, different design
parameters will be studied, which lead to different sensor performances,
as exemplified in Figure c.Biomolecular monitoring applications differ widely
in the analyte
molecules that need to be measured, their concentrations, and their
concentration change rates. Figure sketches an overview of analyte concentrations in
blood (in M) and typical concentration change rates (CCRs, in M h–1) for biomedical monitoring applications such as diabetes
(glucose and insulin),[12,13] organ failure (e.g., creatinine),[24,25] and inflammation (e.g., CRP, PCT, cytokines).[1−3,26,27] The CCRs were calculated by estimating characteristic
concentration changes ΔCa,0 and
typical fluctuation times tfluc (see Supplementary Note 2). For example, blood glucose
concentrations vary between 4 and 8 mM in healthy persons, while for
diabetic patients, the glucose level can increase to 10–15
mM and higher within a period of tfluc ∼ 30 min. This results in a typical maximum CCR of about
20 mM h–1. At the low end of the concentration scale,
cytokine biomarker interleukin-6 (IL-6) is indicated. Physiological
IL-6 concentrations are below 0.5 pM, while for patients with acute
inflammatory stress, e.g., due to sepsis or due to
cytokine release syndrome, the IL-6 concentration can increase to
10–100 pM and higher within a period of a few hours (tfluc ∼ 2 h). This results in a typical
maximum CCR of about 30 pM h–1.
Figure 2
Typical concentration
change rates (CCRs) and mean analyte concentrations Ca,0 for various analyte molecules in blood plasma.
CCRs were calculated by estimating a characteristic concentration
change ΔCa,0 and a corresponding
characteristic fluctuation time tfluc (see Supplementary Note 2) based on reported concentration–time
profiles in blood plasma. Abbreviations: IL-6 (interleukin-6), PCT
(procalcitonin), and CRP (C-reactive protein). The black arrow indicates
the standard parameter value for the mean analyte concentration Ca,0 as listed in Table .
Typical concentration
change rates (CCRs) and mean analyte concentrations Ca,0 for various analyte molecules in blood plasma.
CCRs were calculated by estimating a characteristic concentration
change ΔCa,0 and a corresponding
characteristic fluctuation time tfluc (see Supplementary Note 2) based on reported concentration–time
profiles in blood plasma. Abbreviations: IL-6 (interleukin-6), PCT
(procalcitonin), and CRP (C-reactive protein). The black arrow indicates
the standard parameter value for the mean analyte concentration Ca,0 as listed in Table .
Table 1
Standard Parameter Values Used in
the Finite-Element Simulationsa
parameter
value
description
input
H
100 μm
measurement chamber height
L
1 cm
measurement chamber length
W
2 mm
measurement chamber width
D
10–10 m2 s–1
diffusion coefficient of the analyte molecule
Q
120 μL min–1
flow rate
koff
10–2 s–1
dissociation rate constant
kon
106 M–1 s–1
association
rate constant
Ca,0
10 nM
mean analyte concentration in the system of interest
derived
λ = L/H
100
aspect ratio of measurement chamber
τD = H2/D
100 s
characteristic
diffusion time
τA = HLW/Q
1 s
characteristic advection time
τR = (konCa,0 + koff)−1
50 s
characteristic
reaction time
Kd = koff/kon
10 nM
equilibrium dissociation constant
ΔCa,0/Ca,0
0.05 (5%)
relative concentration change
2
Damköhler number
100
longitudinal Péclet number
Details about the simulations are
described in Supplementary Note 1.
In this paper, the dynamic response of sensors with different designs
is characterized by two parameters: first, the lag time Δt of the sensor signal with respect to the input concentration
(see Figure ), and
second, the rate sensitivity, i.e., the minimum CCR that can be measured with an error of 10% (see Supplementary Note 5). We refer to this minimum
CCR as the limit of quantification of CCR (LoCCR). In the next sections,
we study how design parameters influence the lag time and rate sensitivity
using standard parameter values as listed in Table . The sensor signal and its time characteristics are quantified
by finite-element simulations to investigate the consequences of mass
transport and reactions at the sensor surface. The rate sensitivity
is quantified by calculating the stochastic variabilities in the number
of analyte–binder complexes, for concentration–time
profiles with varying concentration levels and CCRs.Details about the simulations are
described in Supplementary Note 1.
Experimental Section
Finite-Element
Analysis
Finite-element simulations
were performed by solving diffusion, advection and reaction equations
simultaneously using COMSOL (COMSOL Multiphysics 5.5) and MATLAB (MATLAB
R2019a, COMSOL Multiphysics LiveLink for MATLAB) (see Supplementary Note 1). The LoCCR was reported
at a distance L/2 in the measurement chamber (Figure b), where the signal
was collected over a signal collection area As = 1 mm2 with a binder molecule density Γb = 10–9 mol m–2 (see Supplementary Note 5).
Frequency Analysis
The amplitude and the phase lag
of the concentration at the sensor surface (Figures a and 4a,b), analyte–binder
complex density (Figure b), and the measured concentration (Figures c and 4c) were calculated
using the Fourier transform of its concentration/density profile.
The calculated values were compared to the amplitude and the phase
(i.e., ϕ = 0) of the input profile. The cutoff
frequency was determined at the frequency where the observed amplitude
was 50% of the input amplitude. The LoCCR was determined according
to Supplementary Notes 2 and 5.
Figure 3
Response of
a biomolecular monitoring system with continuous analyte
exchange by diffusion-based sampling. (a) Frequency response when
only diffusion is considered. Top graph: concentration change ΔCa at the sensor surface (normalized to the concentration
change ΔCa,0 in the system of interest)
as a function of the frequency f (normalized to the
diffusion time τD). The diffusion-induced cutoff
frequency f cD (horizontal dotted black line) is f cDτD ≅ 0.65 (vertical dotted black line).
Bottom graph: lag time Δt (normalized to the
diffusion time τD) as a function of f (normalized to the diffusion time τD). For large f, Δt scales as Δt ∝ 1/√f (dashed black line, see Supplementary Note 3). The inset shows the phase
lag Δϕ as a function of f. The sketch
above the graphs visualizes a measurement chamber with a concentration
flux Ja caused by a concentration gradient
(orange gradient). (b) Frequency response when only the surface reaction
is considered. Top graph: analyte–binder complex density change
Δγab (normalized to the expected analyte–binder
complex density change Δγabexp, see Supplementary Note 4) as a function of f (normalized to
the reaction time τR). The reaction-induced cutoff
frequency f cR is f cRτD ≅ 0.27 (vertical dotted black line). Bottom graph: lag time
Δt (normalized to the reaction time τR) as a function of f (normalized to the reaction
time τR). For large f, Δt scales according to Δt ∝
1/f (dashed black line, see Supplementary Note 3). The inset shows the phase lag Δϕ as a
function of f, where Δϕ reaches a maximum
negative value (see Supplementary Note 3). The sketch above the graphs visualizes a measurement chamber with
an oscillating concentration Ca(t) at the sensor surface and a resulting oscillating analyte–binder
complex density γab. (c) Cutoff frequency fc as a function of measurement chamber height H (top) and mean analyte concentration Ca,0 in the system of interest (bottom). Top graph: for
small H, fc is reaction-limited,
where fc = f cR ∼ 1/τR, while for large H, fc is diffusion-limited with fc = f cD ∼ 1/τD. The inset shows fc, normalized to the reaction time τR, as a function of the Damköhler number Da, with fc = f cD = α1/τD and α1 ≅ 0.65
(dashed black line, cf. panel a). Bottom graph: for low Ca,0, fc is reaction-limited
and fc = f cR ∼ 1/τR, while for high Ca,0, fc is diffusion-limited with fc = f cD ∼ 1/τD. The
inset shows fc, normalized to the diffusion
time τD, as a function of Da with fc = f cR = α2/τD and α2 ≅ 0.27 (dashed black line,
cf. panel b). Note that using standard parameter values as listed
in Table , the full
range of Da cannot be reached by only changing Ca,0 because τR becomes dissociation
rate-limited when Ca,0 ≪ Kd (see Table ); therefore, koff was
varied instead. The black arrows indicate standard parameter values
as listed in Table .
Figure 4
Response of a biomolecular monitoring system
with continuous analyte
exchange by advection-based sampling. (a) Frequency response when
only diffusion and advection are considered, for an advection-dominated
sensor geometry with PeL = 100 (see Table ). Top graph: concentration
change ΔCa at the sensor surface,
normalized to the concentration change ΔCa,0 in the system of interest, as a function of the frequency f (normalized to the advection time τA),
measured at the sensor surface at distance L/2 from
the inlet (see also the sketch in panel b). The diffusion-induced
cutoff frequency f D is taken from Figure a, and the advection-induced
cutoff frequency f cA is found to be f cAτA ≅ 0.39 (vertical dotted black line) and roughly equals f cA = 100·f cD. Bottom graph: lag time Δt, normalized to τA, as a function of the
frequency f, normalized to τA. For
large f, Δt scales according
to (black dashed line). (b) Cutoff frequency
as a function of flow rate Q when only diffusion
and advection are taken into account, measuring in the middle in the
bulk of the measurement chamber at height H/2 (dark
brown line) and at the sensor surface of the measurement chamber (orange
line) both at distance L/2 from the inlet. For increasing Q, measuring in the bulk results in an advection-limited
cutoff frequency (dashed black lines). The inset shows the same data
with the observed cutoff frequency fc ,
normalized to the advection time τA, as a function
of the longitudinal Péclet number PeL. For small PeL, fc for both bulk and surface measurements are comparable with fc = f cD. For increasing PeL, fc increases due to a higher
flow rate, until the system becomes advection-limited. Measuring at
the sensor surface results in a weaker dependency on τA than 1/τA. (c) Cutoff frequency fc as a function of chamber height H with
a fixed chamber length L (top) and mean analyte concentration Ca,0 in the system of interest (bottom) for a
sensor when diffusion, advection, and reaction are taken into account.
Top graph: for small H, fc is reaction-limited where fc = f cR ∼ 1/τ, while for
large H, fc is diffusion-limited
with fc = f cD ∼ 1/τD. The inset shows the same data with the cutoff frequency fc, normalized to the reaction time τR, as a function of the Damköhler number. Bottom graph:
for H = 100 μm (see Table ), the cutoff frequency is reaction-limited
(top graph). Therefore, for low Ca,0, fc is dissociation rate-limited and fc = f c ∼ koff, while for high Ca,0, fc is association rate-limited with fc = f c ∼ konCa,0. The inset shows the
same data with the cutoff frequency fc, normalized to the diffusion time τD, as a function
of the Damköhler number. fc becomes
diffusion-limited at Da ≫ 1 and reaches a
plateau level larger than fcτD = 1 (cf. Figure c) since τD = H2/D, while the actual distance over which molecules
decreases for increasing Q. Note that using the standard
parameter values in Table , the full range of Da cannot be reached
by only changing Ca,0 because τR becomes dissociation rate-limited when Ca,0 ≪ Kd (see Table ); therefore, koff was varied instead. The black arrows indicate
standard parameter values as listed in Table .
Response of
a biomolecular monitoring system with continuous analyte
exchange by diffusion-based sampling. (a) Frequency response when
only diffusion is considered. Top graph: concentration change ΔCa at the sensor surface (normalized to the concentration
change ΔCa,0 in the system of interest)
as a function of the frequency f (normalized to the
diffusion time τD). The diffusion-induced cutoff
frequency f cD (horizontal dotted black line) is f cDτD ≅ 0.65 (vertical dotted black line).
Bottom graph: lag time Δt (normalized to the
diffusion time τD) as a function of f (normalized to the diffusion time τD). For large f, Δt scales as Δt ∝ 1/√f (dashed black line, see Supplementary Note 3). The inset shows the phase
lag Δϕ as a function of f. The sketch
above the graphs visualizes a measurement chamber with a concentration
flux Ja caused by a concentration gradient
(orange gradient). (b) Frequency response when only the surface reaction
is considered. Top graph: analyte–binder complex density change
Δγab (normalized to the expected analyte–binder
complex density change Δγabexp, see Supplementary Note 4) as a function of f (normalized to
the reaction time τR). The reaction-induced cutoff
frequency f cR is f cRτD ≅ 0.27 (vertical dotted black line). Bottom graph: lag time
Δt (normalized to the reaction time τR) as a function of f (normalized to the reaction
time τR). For large f, Δt scales according to Δt ∝
1/f (dashed black line, see Supplementary Note 3). The inset shows the phase lag Δϕ as a
function of f, where Δϕ reaches a maximum
negative value (see Supplementary Note 3). The sketch above the graphs visualizes a measurement chamber with
an oscillating concentration Ca(t) at the sensor surface and a resulting oscillating analyte–binder
complex density γab. (c) Cutoff frequency fc as a function of measurement chamber height H (top) and mean analyte concentration Ca,0 in the system of interest (bottom). Top graph: for
small H, fc is reaction-limited,
where fc = f cR ∼ 1/τR, while for large H, fc is diffusion-limited with fc = f cD ∼ 1/τD. The inset shows fc, normalized to the reaction time τR, as a function of the Damköhler number Da, with fc = f cD = α1/τD and α1 ≅ 0.65
(dashed black line, cf. panel a). Bottom graph: for low Ca,0, fc is reaction-limited
and fc = f cR ∼ 1/τR, while for high Ca,0, fc is diffusion-limited with fc = f cD ∼ 1/τD. The
inset shows fc, normalized to the diffusion
time τD, as a function of Da with fc = f cR = α2/τD and α2 ≅ 0.27 (dashed black line,
cf. panel b). Note that using standard parameter values as listed
in Table , the full
range of Da cannot be reached by only changing Ca,0 because τR becomes dissociation
rate-limited when Ca,0 ≪ Kd (see Table ); therefore, koff was
varied instead. The black arrows indicate standard parameter values
as listed in Table .Response of a biomolecular monitoring system
with continuous analyte
exchange by advection-based sampling. (a) Frequency response when
only diffusion and advection are considered, for an advection-dominated
sensor geometry with PeL = 100 (see Table ). Top graph: concentration
change ΔCa at the sensor surface,
normalized to the concentration change ΔCa,0 in the system of interest, as a function of the frequency f (normalized to the advection time τA),
measured at the sensor surface at distance L/2 from
the inlet (see also the sketch in panel b). The diffusion-induced
cutoff frequency f D is taken from Figure a, and the advection-induced
cutoff frequency f cA is found to be f cAτA ≅ 0.39 (vertical dotted black line) and roughly equals f cA = 100·f cD. Bottom graph: lag time Δt, normalized to τA, as a function of the
frequency f, normalized to τA. For
large f, Δt scales according
to (black dashed line). (b) Cutoff frequency
as a function of flow rate Q when only diffusion
and advection are taken into account, measuring in the middle in the
bulk of the measurement chamber at height H/2 (dark
brown line) and at the sensor surface of the measurement chamber (orange
line) both at distance L/2 from the inlet. For increasing Q, measuring in the bulk results in an advection-limited
cutoff frequency (dashed black lines). The inset shows the same data
with the observed cutoff frequency fc ,
normalized to the advection time τA, as a function
of the longitudinal Péclet number PeL. For small PeL, fc for both bulk and surface measurements are comparable with fc = f cD. For increasing PeL, fc increases due to a higher
flow rate, until the system becomes advection-limited. Measuring at
the sensor surface results in a weaker dependency on τA than 1/τA. (c) Cutoff frequency fc as a function of chamber height H with
a fixed chamber length L (top) and mean analyte concentration Ca,0 in the system of interest (bottom) for a
sensor when diffusion, advection, and reaction are taken into account.
Top graph: for small H, fc is reaction-limited where fc = f cR ∼ 1/τ, while for
large H, fc is diffusion-limited
with fc = f cD ∼ 1/τD. The inset shows the same data with the cutoff frequency fc, normalized to the reaction time τR, as a function of the Damköhler number. Bottom graph:
for H = 100 μm (see Table ), the cutoff frequency is reaction-limited
(top graph). Therefore, for low Ca,0, fc is dissociation rate-limited and fc = f c ∼ koff, while for high Ca,0, fc is association rate-limited with fc = f c ∼ konCa,0. The inset shows the
same data with the cutoff frequency fc, normalized to the diffusion time τD, as a function
of the Damköhler number. fc becomes
diffusion-limited at Da ≫ 1 and reaches a
plateau level larger than fcτD = 1 (cf. Figure c) since τD = H2/D, while the actual distance over which molecules
decreases for increasing Q. Note that using the standard
parameter values in Table , the full range of Da cannot be reached
by only changing Ca,0 because τR becomes dissociation rate-limited when Ca,0 ≪ Kd (see Table ); therefore, koff was varied instead. The black arrows indicate
standard parameter values as listed in Table .
Results and Discussion
Response of a Monitoring System with Diffusion-Based
Sampling
First, we consider the case where the transport
of analyte molecules
between a system of interest and a sensor measurement chamber is governed
by diffusion only. Figure shows how the analyte concentration at the sensor surface
and the analyte–binder complex density respond to an oscillating
concentration Ca,0(t)
in the system of interest, with concentration change ΔCa,0 for various oscillation frequencies (see Supplementary Note 2). Figure a shows how diffusive mass transport influences
the concentration profile Ca(t) at the sensor surface, by quantifying the concentration change
ΔCa at the sensor surface (top,
orange line, normalized to ΔCa,0), and the lag time Δt (bottom, orange line,
normalized to the diffusion time τD), given as a
function of f (normalized to the diffusion time τD). In the top graph, for small f, the concentration
change ratio ΔCa/ΔCa,0 is close to unity, indicating that the concentration
change at the sensor surface is approximately equal to the concentration
change in the system of interest. Since the oscillation time 1/f is larger than τD, the analyte molecules
are evenly distributed throughout the measurement chamber, i.e., there is no concentration gradient. For large f, ΔCa/ΔCa,0 decreases for increasing f, which means that the concentration change at the sensor surface
is smaller than the concentration change in the system of interest.
Since 1/f is now smaller than τD, a concentration gradient is present in the measurement chamber
in the direction of H (see top sketch). This gradient
results in dispersion of analyte molecules, which effectively reduces
ΔCa. A characteristic parameter
to describe this decrease in ΔCa/ΔCa,0 is the cutoff frequency fc , which is the frequency at which ΔCa/ΔCa,0 =
0.5 (horizontal dotted black line). In this case, the diffusion-induced
cutoff frequency f cD is equal to f cDτD ≅ 0.65 (vertical dotted black line). The bottom graph shows
that for f smaller than f cD, the observed
lag time Δt is independent of f, since within a period of 1/f, analyte molecules
can be transported throughout the measurement chamber by diffusion.
This results in a homogeneous analyte concentration in the measurement
chamber where Δt is only determined by diffusion
(Δt ∼ τD). For f larger than f cD, a concentration gradient is
present in the measurement chamber in the direction of H (see top sketch). Now Δt decreases according
to Δt ∝ 1/√f (dashed black line, see Supplementary Note 3), concomitant with a reduction in ΔCa (top graph). The inset shows the phase lag Δϕ
as a function of f. For increasing f, the absolute phase lag increases (Δϕ becomes more negative)
due to the time needed for the transport of analyte molecules from
the top of the measurement chamber to the sensor surface. For a large f, the concentration at the sensor surface can lag multiple
cycles (Δϕ > 2π) with respect to the concentration
in the system of interest (not shown here).Figure b shows how the association
and dissociation of analyte molecules to binder molecules influence
the measured signal. Mass transport effects are neglected and the
concentration profile Ca(t) at the sensor
surface oscillates with a frequency f. The top graph
shows the change in analyte–binder complex density Δγab, normalized to the expected analyte–binder complex
density change Δγabexp based on the concentration profile Ca(t) at the sensor surface
(see Supplementary Note 4). The bottom
graph shows the lag time Δt as a function of
the frequency f (normalized to the reaction time
τR). For small f, Δγab/Δγabexp is close to unity, indicating that the affinity
reaction reaches equilibrium since the oscillation time 1/f is larger than the reaction time τR (see Table ). For a large f, Δγab/Δγabexp decreases,
indicating that fewer analyte molecules bind to binder molecules on
the sensor surface than expected based on Ca(t) under equilibrium conditions. This results in
a reaction-induced cutoff frequency f cR, at f cRτR ≅ 0.27 (vertical dotted black line). For f smaller than f cR, Δt is largely
independent of f. Now equilibrium is reached, causing
the lag time to be determined by the time to equilibrium, i.e., Δt is reaction-limited (Δt ∼ τR). For f larger
than f cR, Δt depends on f as Δt ∝ 1/f (dashed black line, see Supplementary Note 3). The inset shows the phase lag Δϕ as a function of f. For increasing f, the absolute phase
lag increases (Δϕ becomes more negative) since fewer analyte–binder
complexes are formed within a time 1/f. For large f, the phase lag reaches a minimum value of Δϕ
= −π/2 (horizontal black dotted line) with respect to
γabexp since the reaction rates are directly related to the analyte concentration Ca at the sensor surface and therefore the phase
lag cannot be more negative (see Supplementary Note 3).Figure c shows
the cutoff frequency fc as a function
of the measurement chamber height H (top) and the
mean analyte concentration Ca,0 in the
system of interest (bottom, normalized to the equilibrium dissociation
constant Kd) when both diffusion and reaction
processes are considered. Standard values for the chamber height H and mean concentration Ca,0 are indicated by the black arrows (see Table ). For small H, the diffusion
time τD is short since analyte molecules only need
to travel a short distance from the top of the measurement chamber
to the sensor surface. This causes the observed cutoff frequency fc to be reaction-limited where fc = f cR ∼ 1/τR. For
large H, analyte molecules need to travel a long
distance, which causes fc to be diffusion-limited
where fc = f cD ∼ 1/τD. For small Ca,0, the reaction
is slow since the reaction time τR is determined
by the dissociation rate (see Table ), causing fc to be reaction-limited.
For large Ca,0, τR is
short since the reaction time τR is determined by
the association rate, causing fc to be
diffusion-limited. The insets show fc (normalized
to the reaction time τR) as a function of the Damköhler
number Da (see Table ). Da is a dimensionless parameter
describing the relative contribution of reaction and diffusion to
the observed time scale (for Da ≫ 1, diffusion
is slow relative to reaction; for Da ≪ 1,
reaction is slow relative to diffusion). For a high Da, the cutoff frequency is diffusion-limited, while for a low Da, the cutoff frequency is reaction-limited.
Response of
a Monitoring System with Advection-Based Sampling
Figure shows how
dynamic concentration changes generate signals in a monitoring sensor
based on advective sampling, i.e., sampling dominated
by flow. Figure a
visualizes how diffusion and advection jointly influence the concentration
profile Ca(t) at the
sensor surface. The concentration change ΔCa at the sensor surface (top, orange line, normalized
to concentration change ΔCa,0 in
the system of interest) and the lag time Δt (bottom, orange line, normalized to the advection time τA) are given as a function of the oscillation frequency f (normalized to τA) of the analyte concentration Ca,0 in the system of interest. Here, a longitudinal
Péclet number PeL = τD/τA = 100 was assumed (see Table ), where PeL describes the relative contribution of diffusion and
advection to the transport process (for PeL ≫ 1, diffusion is slow relative to advection; for PeL ≪ 1, advection is slow relative to
diffusion). In the top graph, for small f, ΔCa/ΔCa,0 equals
unity, indicating that the concentration is evenly distributed throughout
the measurement chamber. For a large f, ΔCa/ΔCa,0 decreases,
which indicates a concentration gradient in the measurement chamber
perpendicular to the velocity profile. This results in an advection-induced
cutoff frequency f cA (horizontal dotted black line), which
can be found at f cAτA ≅ 0.39 (vertical
dotted black line). Note that the advection-induced cutoff frequency
roughly equals f cA = 100·f cD since PeL = 100, where f cD is the cutoff
frequency for a monitoring system with diffusion-based sampling (cf.Figure ). In the bottom graph, for a small f, the observed
lag time Δt of the concentration at the sensor
surface compared to the concentration in the system of interest is
largely independent of f since the analyte concentration
is homogeneous in the measurement chamber, causing the lag time to
be advection-limited (Δt ∼ τA). For f larger than f cA, a concentration
gradient is present in the measurement chamber, causing a loss in
ΔCa due to dispersion of the analyte
molecules. Here, Δt depends less on the frequency f (, dashed black line) compared to Figure a,b since the observed
time lag is caused by both advection and diffusion, where the contribution
of advection is independent of f (see Supplementary Note 3). Furthermore, diffusion
occurs on a length scale smaller than H, reducing
its contribution to the frequency dependency. The inset shows the
same data visualized as the phase lag Δϕ as a function
of f. For increasing f, the absolute
phase lag increases (Δϕ becomes more negative) due to
the time needed to transport analyte molecules from the bulk of the
measurement chamber to the sensor surface. For large f, Ca(t) can lag for
multiple cycles with respect to Ca,0(t) (not shown here), though ΔCa decreases sharply due to dispersion for f > f cA (see top graph).Figure b visualizes cutoff frequency fc as a function of flow rate Q measured
at two positions, namely, in the bulk of the measurement chamber (position
1, dark brown line) and at the surface of the measurement chamber
(position 2, orange line). The black arrow indicates the standard
parameter value for Q as listed in Table . The inset shows the same data
with fc (normalized to τA) as a function of PeL. For a small Q, there is a concentration gradient in the longitudinal
direction since the distance over which molecules need to diffuse
to the sensor surface is smaller compared to the situation in Figure . This results in
an advection-limiting process at PeL <
1 with a constant fc. For increasing Q, the observed fc becomes different
when measuring in the bulk or at the sensor surface. For measuring
in the bulk, an increased Q results in a change of
the shape of the concentration gradient, namely, perpendicular to
the velocity profile (top sketch, dashed black profile) instead of
in the longitudinal direction. The flow rate where fc becomes advection-limited depends on the measurement
chamber geometry: for a small L, fc is advection-limited at small flow rates since the distance
over which molecules need to be transported is small. For measuring
at the sensor surface, the stationary layer becomes smaller for increasing Q effectively decreasing the distance over which the molecules
need to diffuse to the sensor surface. This effect results in fc scaling with the advection time less than
1/τA (see the main graph, dashed black lines) and
that the normalized cutoff frequency fc decreases (see inset).Figure c visualizes
the cutoff frequency fc as a function
of the flow rate Q (top) and the mean analyte concentration Ca,0 in the system of interest (bottom), where
diffusion, advection, and reaction are included. Standard values for
the flow rate Q and mean concentration Ca,0 are indicated by the black arrows (see Table ). In the top graph, for a low Q, the cutoff frequency is advection-limited, where fc = f cA ∼ 1/τA. For a large Q, the cutoff frequency is limited
by a combination of diffusion and reaction. The inset shows the same
data with fc (normalized to the advection
time τA) as a function of PeL. In the bottom graph, for a low Ca,0, the reaction is slow, which causes the observed cutoff frequency fc to be dissociation rate-limited, where fc = f cR ∼ koff. For a large Ca,0, the reaction
is association rate-limited, where fc = f cR ∼ konCa,0. The inset shows the same data with fc (normalized to the diffusion time τD) as a function of the Damköhler number Da. For a small Da, the cutoff frequency is reaction-limited
(dashed black line). For a large Da, the cutoff frequency
becomes diffusion-limited.
Continuous Biomolecular Monitoring for Arbitrary
Concentration
Profiles
The measured concentration profile of a monitoring
sensor should resemble as closely as possible the true concentration
profile of the analyte. While Figures and 4 discussed the effects
of diffusion, advection, and reaction on the cutoff frequency and
lag time, the question remains how these processes influence an actual
concentration profile and the differences between the measured and
the true concentration profiles. Here, we study an insulin concentration
profile with standard parameter values listed in Table as an example (see Supplementary Note 6). In this paper, the rate
sensitivity is quantified as the limit of quantification of CCR (LoCCR), i.e., the smallest CCR that can be measured with an error
of 10% (see Supplementary Note 5). The
LoCCR is calculated assuming a sensor with noise that is dominated
by Poisson statistics, with a signal collection area As = 1 mm2, and a binder density Γb = 10–9 mol m–2. Poisson
noise represents the fundamental limit of the precision that can be
achieved in a biosensor due to stochastic fluctuations in the number
of detected analyte molecules.[28,29]Figure shows the collective influence
of diffusion, advection, and reaction on LoCCR and the measured concentration
profile, for different measurement chamber heights H (in the case of diffusion-based analyte exchange) and for different
flow rates Q (in the case of advection-based analyte
exchange). Figure a shows LoCCR as a function of frequency f, for
diffusion-based sampling (top) and advection-based sampling (bottom).
The top graph shows results for measurement chamber heights H = 200 μm (dark brown) and H = 800
μm (orange). For a low f, LoCCR equals the
value found for Poisson noise only (dashed black line, see also Supplementary Note 5), which is due to the fact
that the sensor reaches equilibrium and no dispersion occurs (cf. Figure ). For increasing f, the results depend on the measurement
chamber height because a small H gives a larger cutoff
frequency (see Figure c). Here, both lines deviate from the Poisson limit since equilibrium
is not reached within a time equal to 1/f, resulting
in fewer analyte–binder complexes. The inset shows the same
data with the minimum concentration change ΔCa,0 that can be quantified with an error of less than
10% as a function of frequency f. The bottom graph
shows advection-based sampling with flow rates Q =
10 μL/min (dark brown) and Q = 0.1 μL/min
(orange).[30−32] Here, the lines deviate from the Poisson limit at
higher frequencies compared to diffusion-based sampling since the
cutoff frequency is higher in advection-based sampling compared to
diffusion-based sampling (see Figures c and 4c). The inset shows the
same data with the minimum concentration change ΔCa,0 that can be quantified with an error of less than
10% as a function of frequency f.
Figure 5
Measuring concentration-time profiles using two modes
of continuous
analyte exchange. (a) Limit of quantification of CCR (LoCCR) as a
function of frequency f in a monitoring system with
continuous analyte exchange by diffusion-based sampling (top) and
by advection-based sampling (bottom). The insets show the same data
with the concentration change ΔCa,0 as a function of f. For a low f, the precision of the CCR is limited by Poisson noise (dashed black
line). For increasing f, the lines start to deviate
since the frequencies become higher than the corresponding cutoff
frequencies. (b) Concentration-time profiles for a measurement chamber
height H = 200 μm (brown line) and H = 800 μm (orange line), and the true analyte concentration
(black dotted line), for diffusion-based analyte exchange. The bottom
graphs show the frequency spectrum with the CCR component as a function
of frequency. For a small H, the concentration-time
profile closely resembles the true concentration-time profiles. However,
for a large H, the similarity is only visible at
low frequencies; at high frequencies, the measured CCR is close to
0, indicating that sinusoidal components with these frequencies are
not present in the measured signal. (c) Concentration-time profiles
for a measurement chamber with flow rate Q = 10 μL/min
(brown line) and Q = 0.1 μL/min (orange line),
and the true analyte concentration (black dotted line, behind the
brown line), for advection-based analyte exchange. The bottom shows
the frequency spectrum. For both flow rates, the concentration-time
profile closely resembles the true concentration-time profile. (d,
e) LoCCR as a function of f and the corresponding
concentration-time profiles for diffusion-based analyte exchange (d)
and advection-based analyte exchange (e), where koff = 10–3 s–1 (compared
to koff = 10–2 s–1 in panels (a) and (b), see Table ). Due to higher cutoff frequencies, the
similarity of the measured concentration-time profiles is less compared
to panels (b) and (c).
Measuring concentration-time profiles using two modes
of continuous
analyte exchange. (a) Limit of quantification of CCR (LoCCR) as a
function of frequency f in a monitoring system with
continuous analyte exchange by diffusion-based sampling (top) and
by advection-based sampling (bottom). The insets show the same data
with the concentration change ΔCa,0 as a function of f. For a low f, the precision of the CCR is limited by Poisson noise (dashed black
line). For increasing f, the lines start to deviate
since the frequencies become higher than the corresponding cutoff
frequencies. (b) Concentration-time profiles for a measurement chamber
height H = 200 μm (brown line) and H = 800 μm (orange line), and the true analyte concentration
(black dotted line), for diffusion-based analyte exchange. The bottom
graphs show the frequency spectrum with the CCR component as a function
of frequency. For a small H, the concentration-time
profile closely resembles the true concentration-time profiles. However,
for a large H, the similarity is only visible at
low frequencies; at high frequencies, the measured CCR is close to
0, indicating that sinusoidal components with these frequencies are
not present in the measured signal. (c) Concentration-time profiles
for a measurement chamber with flow rate Q = 10 μL/min
(brown line) and Q = 0.1 μL/min (orange line),
and the true analyte concentration (black dotted line, behind the
brown line), for advection-based analyte exchange. The bottom shows
the frequency spectrum. For both flow rates, the concentration-time
profile closely resembles the true concentration-time profile. (d,
e) LoCCR as a function of f and the corresponding
concentration-time profiles for diffusion-based analyte exchange (d)
and advection-based analyte exchange (e), where koff = 10–3 s–1 (compared
to koff = 10–2 s–1 in panels (a) and (b), see Table ). Due to higher cutoff frequencies, the
similarity of the measured concentration-time profiles is less compared
to panels (b) and (c).Figure b shows
a typical insulin profile (dotted black line) and corresponding measured
insulin profiles, using diffusion-based analyte exchange and the standard
parameter values listed in Table . The bottom graphs show the frequency spectrum of
the true and measured insulin profiles plotted as CCR components (see Supplementary Note 2). For H = 200 μm (brown line), the measured concentration profile
is almost identical to the true concentration profile since the cutoff
frequency fc = 9 × 10–4 Hz (see Figure c)
is higher than the frequencies present in the true insulin profile
(see bottom graphs, left). For H = 800 μm (orange
line), the true concentration profile cannot be accurately reconstructed,
only the general up-and-down trend at a 6-hour interval, since the
cutoff frequency fc = 1 × 10–4 Hz (see Figure c) is close to the frequencies in the insulin profile
(see bottom graphs, right). Also, the average lag time Δt of the measured signal is smaller for H = 200 μm than for H = 800 μm since
a smaller distance requires less time for diffusion.Figure c shows
the results for advection-based analyte exchange, for flow rates Q = 10 μL/min (dark brown) and Q =
0.1 μL/min (orange). In both cases, the measured insulin profile
is similar to the true insulin profile since the cutoff frequencies
are fc = 7 × 10–3 Hz and fc = 8 × 10–4 Hz, respectively (see Figure c, top). The strong similarities are also visible in the frequency
spectrum (bottom panel).Figure d,e investigates
the limits of dynamic monitoring when affinity binders with very high
affinity are used, i.e., binders with a very low
dissociation rate constant (koff = 10–3 s–1), which is relevant to study
when low-concentration analytes need to be measured. Figure d shows simulation results
for diffusion-based sampling. The data show that the lower dissociation
rate constant causes a lower cutoff frequency, a longer lag time,
and a higher LoCCR (see Supplementary Note 5). The reversibility of the sensor is worse, particularly for a large
measurement chamber height (H = 800 μm) because
the large volume of the measurement chamber contains many analyte
molecules. Figure e shows results for advection-based sampling. The flow rate increases
the rate of exchange of the large volume above the sensor surface.
A flow rate as low as 0.1 μL/min already significantly improves
the dynamic performance of the sensor. A flow rate of 10 μL/min
gives small differences between the measured and real concentrations
with a lag time that is very close to 1/koff = 1/(10–3 s–1) = 17 min.
Conclusions
To measure in real time the dynamic changes of biomolecular concentrations
in biological systems or biotechnological processes, monitoring sensors
are required that reveal reliable concentration–time profiles.
We have studied the influence of sensor design parameters on the differences
between the true and the measured concentration–time profile,
focusing on the lag time of the sensor signal with respect to the
input concentration and on the rate sensitivity. To quantify a rate
sensitivity, we introduced the concept of concentration change rate
(CCR), which is expressed in the units molar per second. The CCR that
needs to be resolved differs strongly between different biomolecular
monitoring applications, due to their respective concentration changes
and fluctuation times. The limit of the measurable concentration change
rate was evaluated as the limit of CCR (LoCCR), i.e., the lowest CCR that can be quantified with a precision of 10%.In this work, we have presented a comprehensive methodology to
study the properties and limitations of dynamic measurements using
affinity-based sensors, as these represent a very generic and broad
class of bioanalytical measurement techniques. Analyte exchange was
considered between the system of interest and the sensor by diffusive
as well as advective sampling. Finite-element simulations were used
to describe the spatial and temporal dependency of analyte concentration
within the measurement chamber. Sinusoidal concentration–time
profiles were studied as well as arbitrary concentration–time
profiles by frequency decomposition. Using this approach, the effects
of mass transport and biochemical kinetics on the speed of concentration
change, time delays, and concentration errors in the sensing system
were studied.The study of sensor performance was exemplified
for insulin monitoring.
The results show that diffusion-based sampling performs equal to advection-based
sampling in reconstructing the concentration–time profile for
small heights of the measurement chamber (<200 μm). However,
for larger heights, diffusion-based sampling causes an increased lag
time and decreased CCR sensitivity. A monitoring system with advection-based
sampling performs similarly with respect to the CCR sensitivity for
flow rates down to ∼0.1 μL min–1, while
the lag time is larger for low flow rates.For low concentrations
of biomolecules, fewer molecules are available
for the detection and therefore continuous monitoring sensors with
single-molecule resolution are suitable because these sensors can
have Poisson-limited noise levels and therefore a high detection sensitivity.
In the case of binder molecules with a high affinity (koff = 10–3 s–1), the
analytical performance deteriorates for diffusion-based sampling,
but not for advection-based sampling with flow rates of 10 μL
min–1 and higher, allowing the measurement of all
CCR components present in an insulin concentration–time profile.The results and learnings presented in this paper can assist researchers
to identify the most important processes influencing the performance
of continuous monitoring sensors. As a next step, it will be interesting
to compare the simulation results to experimental data, for example,
on how concentration–time profiles with different frequency
components affect the observed signals. Insights into the individual
and combined influence of analyte diffusivity, analyte concentration,
binder affinity, sampling method, measurement chamber geometry, and
flow speed on the observed lag time and rate sensitivity of the measured
concentration–time profile will help researchers to develop
monitoring systems with desirable sensor characteristics for a diverse
range of biomarkers and applications.
Authors: J Heikenfeld; A Jajack; J Rogers; P Gutruf; L Tian; T Pan; R Li; M Khine; J Kim; J Wang; J Kim Journal: Lab Chip Date: 2018-01-16 Impact factor: 6.799
Authors: Jason Heikenfeld; Andrew Jajack; Benjamin Feldman; Steve W Granger; Supriya Gaitonde; Gavi Begtrup; Benjamin A Katchman Journal: Nat Biotechnol Date: 2019-02-25 Impact factor: 54.908
Authors: Yu Shrike Zhang; Julio Aleman; Su Ryon Shin; Tugba Kilic; Duckjin Kim; Seyed Ali Mousavi Shaegh; Solange Massa; Reza Riahi; Sukyoung Chae; Ning Hu; Huseyin Avci; Weijia Zhang; Antonia Silvestri; Amir Sanati Nezhad; Ahmad Manbohi; Fabio De Ferrari; Alessandro Polini; Giovanni Calzone; Noor Shaikh; Parissa Alerasool; Erica Budina; Jian Kang; Nupura Bhise; João Ribas; Adel Pourmand; Aleksander Skardal; Thomas Shupe; Colin E Bishop; Mehmet Remzi Dokmeci; Anthony Atala; Ali Khademhosseini Journal: Proc Natl Acad Sci U S A Date: 2017-03-06 Impact factor: 11.205
Authors: Dhanuka P Wasalathanthri; Matthew S Rehmann; Yuanli Song; Yan Gu; Luo Mi; Chun Shao; Letha Chemmalil; Jongchan Lee; Sanchayita Ghose; Michael C Borys; Julia Ding; Zheng Jian Li Journal: Biotechnol Bioeng Date: 2020-07-01 Impact factor: 4.530
Authors: Sally A N Gowers; Michelle L Rogers; Marsilea A Booth; Chi L Leong; Isabelle C Samper; Tonghathai Phairatana; Sharon L Jewell; Clemens Pahl; Anthony J Strong; Martyn G Boutelle Journal: Lab Chip Date: 2019-07-10 Impact factor: 6.799
Authors: Adrian M Nightingale; Chi Leng Leong; Rachel A Burnish; Sammer-Ul Hassan; Yu Zhang; Geraldine F Clough; Martyn G Boutelle; David Voegeli; Xize Niu Journal: Nat Commun Date: 2019-06-21 Impact factor: 14.919