| Literature DB >> 34853955 |
Flora H P van Leeuwen1, Beatrice Lena2, Jaco J M Zwanenburg3, Lize F D van Vulpen4, Lambertus W Bartels2, Kathelijn Fischer4, Frank J Nap3,5, Pim A de Jong3, Clemens Bos6, Wouter Foppen3.
Abstract
BACKGROUND: Intra-articular blood causes irreversible joint damage, whilst clinical differentiation between haemorrhagic joint effusion and other effusions can be challenging. An accurate non-invasive method for the detection of joint bleeds is lacking. The aims of this phantom study were to investigate whether magnetic resonance imaging (MRI) T1 and T2 mapping allows for differentiation between simple and haemorrhagic joint effusion and to determine the lowest blood concentration that can be detected.Entities:
Keywords: Haemarthrosis; Image interpretation (computer-assisted); Magnetic resonance imaging; Phantoms (imaging); Synovial fluid
Mesh:
Year: 2021 PMID: 34853955 PMCID: PMC8636530 DOI: 10.1186/s41747-021-00251-z
Source DB: PubMed Journal: Eur Radiol Exp ISSN: 2509-9280
Fig. 1a Nuclear magnetic resonance tubes with different ratios of synovial fluid and blood. Blood percentage from left to right: 0%, 2.5%, 5%, 10%, 25%, 50%, 75%, and 100%. b Phantom setup: a cylindrical water-filled holder with six samples, horizontally placed in a 16-channel knee coil using a 3-T MRI system. The phantom was equipped with a fiberoptic thermometer (the asterisk in all images) and a heating system based on heat exchange between water in the heating system (black arrows) and the water or oil volume in the phantom. c Scan image from the T1 mapping sequence. For each tube, a region of interest was manually placed within the tube region (example in yellow). d Scan image from the T2 mapping sequence
MRI protocols for T1 mapping and T2 mapping with turbo spin-echo and sensitivity encoding acceleration, at 1.5, 3, and 7 T
| Field strength | T1 mapping | T2 mapping* | ||||
|---|---|---|---|---|---|---|
| 1.5 T | 3 T | 7 T | 1.5 T | 3 T | 7 T | |
| Number of slices | 11 | 11 | 11 | 5 | 5 | 5 |
| Number of inversions | 11 | 11 | 11 | 32 | 32 | 32 |
| Voxel size (mm) | 0.8 | 0.8 | 0.8 | 0.8 | 0.8 | 0.8 |
| Field of view (mm) | 128 × 128 | 128 × 128 | 128 × 128 | 128 × 128 | 128 × 128 | 128 × 128 |
| Slice thickness (mm) | 5 | 5 | 5 | 5 | 5 | 5 |
| Inversion time spacing (ms) | 315 | 315 | 907 | – | – | – |
| Repetition time (ms) | 10,000 | 10,000 | 10,000 | 3000 | 3000 | 7800 |
| Echo time (ms) | 7.8 | 7.8 | 4.33 | 20–950 | 20–950 | 30–960 |
| Flip angle | 15° | 15° | 15° | 90° | 90° | 90° |
| EPI frequency encoding bandwidth (Hz) | 614.6 | 614.6 | 1179 | – | – | – |
| Readout bandwidth (Hz/pixel) | 77.1 | 77.1 | 144.2 | 227 | 227 | 208 |
| EPI factor | 5 | 5 | 5 | – | – | – |
| Parallel reduction factor in-plane | 2 | 2 | 2 | 2 | 2 | 2 |
| Scan duration (s) | 220 | 220 | 220 | 252 | 168 | 632 |
EPI Echo-planar imaging, *with turbo-spin echo and sensitivity encoding acceleration
Fig. 2Mean T1 and T2 value estimates for the different blood concentrations, at 3 field strengths. For 7 T, some of the estimates were not available, because artefacts or relaxation times were too short to be measured
Experimentally estimated T1 and T2 values for synovial fluid, blood, LoB, LoD, and blood detection threshold at 1.5, 3, and 7 T
| Field strength | Relaxation time synovial fluid (ms) | Relaxation time blood (ms) | LoB (ms) | LoD (ms) | Blood detection threshold | |
|---|---|---|---|---|---|---|
| T1 | 2641 ± 60 | 1258 ± 33 | 2541 | 2526 | ≥ 10% | |
| 2719 ± 72 | 1310 ± 33 | 2600 | 2560 | ≥ 25% | ||
| 2355 ± 145 | 1272 ± 160 | 2117 | 2040 | ≥ 50% | ||
| T2 | 845 ± 56 | 93 ± 11 | 753 | 298 | ≥ 50% | |
| 592 ± 13 | 28 ± 20 | 579 | 546 | ≥ 5% | ||
| 281 ± 5 | Not measurable | 272 | 218 | = 25% |
Results are reported as mean values, and the standard deviations as data were normally distributed (Kolmogorov-Smirnov test, p > 0.05)
Blood blood concentration 100%, Blood detection threshold blood concentration corresponding to the LoD, LoB limit of blank, LoD limit of detection, synovial fluid blood concentration 0%
Fig. 3Effect of different acceleration techniques on T2 estimates at 1.5 T. T2 exhibited an inverse dependence on the blood concentration with all acceleration techniques. The three accelerated T2 mapping methods showed good agreement with the reference scan (echo spacing 30 ms), except from the 5% blood concentration measurement using the EPI acceleration technique. EPI, echo-planar imaging; TSE, turbo spin-echo; TSE+SENSE, turbo spin-echo with sensitivity encoding
Fig. 4Bland-Altman plots for interobserver agreement regarding T1 and T2 measurements at 1.5 and 3 T. Horizontal dashed lines indicate the upper and lower limit of agreement from the mean by the two observers (ULoA/LLoA). For T1 measurements, the mean was − 28.99 ms, with a limit of agreement (LoA) equal to 143 ms. For T2 measurements, the mean was − 1.164 ms with LoA equal to 148 ms
T1 and T2 relaxation times of blood and synovial fluid at different field strengths: a comparison of the experimental values with literature values
| Field strength | Synovial fluid | Blood | |||||
|---|---|---|---|---|---|---|---|
| Experimental values | Literature values | Author | Experimental values | Literature values | Author | ||
| 2641 ± 60 | 2850 ± 280 | Gold [ | 1258 ± 33 | 1440 ± 120 | Stanisz [ | ||
| 1480 ± 61 | Zhang [ | ||||||
| 2719 ± 72 | 3620 ± 320 | Gold [ | 1310 ± 33 | 1932 ± 85 | Stanisz [ | ||
| 2564 ± 270 | Jordan [ | 1649 ± 68 | Zhang [ | ||||
| 1584 ± 5 | Lu [ | ||||||
| 2355 ± 145 | 4813 ± 700 | Jordan [ | 1272 ± 160 | 2212 ± 53 | Dobre [ | ||
| 2087 ± 131 | Zhang [ | ||||||
| 845 ± 56 | 1210 ± 140 | Gold [ | 93 ± 11 | 290 ± 30 | Stanisz [ | ||
| 592 ± 13 | 767 ± 50 | Gold [ | 28 ± 20 | 275 ± 50 | Stanisz [ | ||
| 652 ± 113 | Jordan [ | 60 ± 6 | Krishnamurty [ | ||||
| 281 ± 5 | 324 ± 60 | Jordan [ | Not measurable* | 21 ± 4 | Krishnamurty [ | ||
Experimental values Values measured in this study in region of interests within the synovial fluid (0% blood) and 100% blood tubes reported as means and standard deviations, Literature values Mean values found in other studies with corresponding standard deviations, *T2 times too short to be measured