| Literature DB >> 34838535 |
L Gard1, M A Fliss2, F Bosma2, D Ter Veen2, H G M Niesters2.
Abstract
BACKGROUND: Throughout the SARS-CoV-2 pandemic, a rapid identification of the virus was essential to quickly recognize positive cases and limit further spread by applying appropriate infection prevention. Many diagnostic laboratories use a multiplex Real-Time PCR assay, as they are not only highly sensitive but also specific. Currently, there are several assays and platforms in the market available which target different SARS-CoV-2 genes. The aim of this study was to validate and verify the GeneFinder™ COVID-19 PLUS RealAmp kit on the ELITe InGenius® instrument and compare to the national reference method.Entities:
Keywords: ELITe InGenius® instrument; GeneFinder™ COVID-19 PLUS RealAmp kit; SARS-CoV-2
Mesh:
Year: 2021 PMID: 34838535 PMCID: PMC8611827 DOI: 10.1016/j.jviromet.2021.114378
Source DB: PubMed Journal: J Virol Methods ISSN: 0166-0934 Impact factor: 2.014
Detection limits using the SARS-CoV-2 Analytical Qnostic panel 01. The limit of detection for the SARS-CoV-2 in-house qPCR and the N-gene using the GeneFinder was 50 dC/mL. The RdRp-gene and E-gene using the GeneFinder was 500 dC/mL.
| SARS-CoV-2 Qnostic analytical panel | In-house SARS-CoV-2 | GeneFinderTM COVID-19 Plus RealAmp kit | ||||
|---|---|---|---|---|---|---|
| Panel number | dC/mL | Log10 dC/mL | E- gene (Ct value) | RdRP- gene (Ct value) | N- gene (Ct value) | E-gene (Ct value) |
| SCVA2AQP01-S01 | 1000000 | 6,00 | 21.87 | 24.22 | 24.17 | 22.27 |
| SCVA2AQP01-S02 | 100000 | 5,00 | 24.66 | 27.09 | 27.11 | 25.46 |
| SCVA2AQP01-S03 | 10000 | 4,00 | 27.87 | 30.07 | 29.74 | 28.77 |
| SCVA2AQP01-S04 | 5000 | 3.7 | 28.52 | 31.11 | 30.62 | 29.73 |
| SCVA2AQP01-S05 | 1000 | 3,00 | 30.57 | 34.39 | 32.8 | 32.88 |
| SCVA2AQP01-S06 | 500 | 2.7 | 31.89 | 43.97 | 34.08 | 35.45 |
| SCVA2AQP01-S07 | 100 | 2,00 | 34.26 | Not detected | 36.4 | Not detected |
| SCVA2AQP01-S08 | 50 | 1.7 | 35.01 | Not detected | 40.14 | Not detected |
| SCVA2AQP01-S09 | Negative | – | Not detected | Not detected | Not detected | Not detected |
Detection limits using the QCMD SARS-CoV-2 EQA 2020. The SARS-CoV-2 in-house qPCR can detect a viral load of up to log 2.48 dC/mL. The GeneFinder ™ COVID-19 Plus RealAmp kit can detect a viral load of up to log 2.48 dC/mL only with the N-gene.
| QCMD SARS-CoV-2 EQA 2020 | In-house SARS-CoV-2 | GeneFinderTM COVID-19 Plus RealAmp kit | ||||
|---|---|---|---|---|---|---|
| Panel number | Log10 dC/mL | Sample content | E- gene (Ct value) | RdRP- gene (Ct value) | N- gene (Ct value) | E- gene (Ct value) |
| SCV2_101S-01 | 3.16 | SARS-CoV-2 | 30 | negative | 33,98 | 36,9 |
| SCV2_101S-02 | 3.93 | hCoV-229E | negative | Negative | negative | Negative |
| SCV2_101S-03 | 2.48 | SARS-CoV-2 | 33 | Negative | 35,77 | Negative |
| SCV2_101S-04 | 4.29 | SARS-CoV-2 | 27 | 29,51 | 29,56 | 28,49 |
| SCV2_101S-05 | 3.27 | SARS-CoV-2 | 30 | 42,04 | 32,64 | 33 |
| SCV2_101S-06 | 2.48 | SARS-CoV-2 | 33 | Negative | 36,49 | Negative |
| SCV2_101S-07 | 3.27 | SARS-CoV-2 | 30 | 35,73 | 32,83 | 32,9 |
| SCV2_101S-08 | 3.16 | SARS-CoV-2 | 32 | 42,01 | 33,62 | 34,46 |
| SCV2_101S-09 | Neg | Coronavirus negative | negative | Negative | negative | Negative |
| SCV2_101S-10 | 4.0 | hCoV-OC43 | negative | Negative | negative | Negative |
Intra- and inter- assay variation of the GeneFinder ™ COVID-19 Plus RealAmp kit using the ELITe InGenius platform. The intra- assay variation of a sample with a high viral load (Ct around 20) was lower compared to a sample with a low viral load (Ct around 31). In contrast, the inter assay variation was comparable for the high and low viral load samples.
| Gene | Sample type | Intra-assay | Inter-assay |
|---|---|---|---|
| Average Ct ± SD (%CV) | Average Ct ± SD (%CV) | ||
| E-gene | high positive | 19.6 ± 0.48 (2.45) | 18.97 ± 1.12 (5.92) |
| low positive | 30.93 ± 1.44 (4.66) | 31.32 ± 0.98 (3.11) | |
| RdRp-gene | high positive | 22.17 ± 0.29 (1.31) | 22.44 ± 0.89 (3.95) |
| low positive | 34.55 ± 1.22 (3.54) | 34.15 ± 1.57 (4.58) | |
| N-gene | high positive | 20.25 ± 0.15 (0.72) | 20.20 ± 1.02 (5.07) |
| low positive | 31.09 ± 1.63 (5.23) | 31.87 ± 0.82 (2.59) |
Fig. 1Overview of the nasopharyngeal samples tested with the in-house SARS-CoV-2 PCR and the GeneFinder ™ COVID-19 Plus RealAmp kit on the ELITe InGenius platform. Overall, twelve samples in the Ct range of 20-25 (number 1-12), twenty-four samples in the Ct range of 26–30 (number 13–36), twenty samples in the Ct range of 31-34 (number 37-56) and ten samples in the Ct range of 35–39 (57–66) were included.
Fig. 2Overview of the sputum samples tested with the in-house SARS-CoV-2 qPCR and the GeneFinder ™ COVID-19 Plus RealAmp kit on the ELITe InGenius platform. Overall, three samples in the Ct range of 20-25 (number 1–3), twelve samples in the Ct range of 26–30 (number 4–15), eleven samples in the Ct range of 31-34 (number 16–25) and four samples in the Ct range of 35-39 (26–30) were included.