OBJECTIVE: The study objectives were to evaluate the inter-rater and intra-rater reliability; and criterion and construct validity of the 3D printed Toronto Rehabilitation Institute -Hand Function Test (3D TRI-HFT) in individuals with spinal cord injury (SCI).
DESIGN: Observational study.
SETTING: Inpatient Rehabilitation Hospital. PARTICIPANTS: Four individuals with subacute and three individuals with chronic SCI.
INTERVENTIONS: Reliability and validity of the 3D TRI-HFT was assessed within two interventional studies. OUTCOME MEASURES: Participants performed the 3D TRI-HFT, Graded Redefined Assessment of Strength, Sensibility and Prehension (GRASSP), Functional Independence Measure (FIM) and the Spinal Cord Independence Measure (SCIM) at baseline, after 20 and 40 sessions of therapy and at six month follow-up from baseline. 3D TRI-HFT assessments were graded at the time of performance and re-graded from the video recordings for purpose of reliability testing. Validity testing was done by comparing the scores on 3D TRI-HFT with the scores on the GRASSP, and the FIM and SCIM self care sub-scores.
RESULTS: The 3D TRI-HFT had high intra-rater and inter-rater reliability in sub-acute and chronic SCI with ICC values exceeding 0.99. Moderate to strong correlations were found between 3D TRI-HFT object manipulation scores and the FIM and SCIM self care sub-scores, with r values in the range of 0.7-0.8. Strong correlations were found between the various components of GRASSP and the 3D TRI-HFT, with r values exceeding 0.9.
CONCLUSIONS: The 3D TRI-HFT is a reliable and valid measure to assess unilateral hand gross motor function in individuals with SCI.

Entities:  

Keywords:  Function; Hand function assessment; Reliability; Spinal cord injury; TRI-HFT; Toronto Rehabilitation Institute–Hand Function Test; Validity

Mesh:

Year:  2021        PMID: 34779739      PMCID: PMC8604468          DOI: 10.1080/10790268.2021.1961055

Source DB:  PubMed          Journal:  J Spinal Cord Med        ISSN: 1079-0268            Impact factor:   1.985


Introduction

Amongst the spinal cord injury (SCI) population, 59.9% have tetraplegia and their highest priority is to regain upper extremity function.[1,2] Rehabilitation is the mainstay of treatment for retraining function. To scientifically evaluate novel interventions, researchers need access to reliable, valid, easy to use and universally accessible clinical outcome assessment tools that can be administered within reasonable time frames. Although there are many tests of hand function, few reliable and valid assessments have been developed for SCI.[3-9] Hand function tests such as the Action Research Arm Test and the Jebsen Hand Function test are not recommended for use in the SCI population as they are not developed for the patterns of grasp impairments following SCI. The Grasp and Release Test although developed for SCI, specifically to assess function following use of upper limb orthoses, FES, and reconstructive procedures is a time based measure and does not take into account proximal arm function.[10] Thus, there is a need to develop assessments based on the International Classification of Functioning, Disability and Health (ICF) framework and these assessments need to go beyond evaluation of pathophysiology or impairment.[11] The first working group of the International Campaign for Cures of SCI Paralysis aptly recognized that “no SCI therapy will be considered effective for the treatment of patients unless it improves the ability of patients to function in their daily routines or activities”.[12] In a review done by Jones et al.,[13] the authors identified two outcome measures that are SCI specific upper extremity measures in the activity domain. These measures are the Graded Redefined Assessment of Strength, Sensibility and Prehension (GRASSP) and the Capabilities of Upper Extremity Function- Test (CUE-T). The GRASSP is a quantitative impairment measure designed for use in acute and chronic cervical SCI and was developed to address the void in the availability of SCI measures that detect functional capacity of the hand.[14,15] Various sub-components of the GRASSP assess function under the body structure and function domain, and the activity domain of the ICF. The Quantitative Prehension component of the GRASSP is a timed test that does not consider proximal function or quality of movement. The entire test takes approximately 1–1.5 h to administer on bilateral upper extremities and occasionally may have to be completed over two days based on patient tolerance. It is custom manufactured and needs to be purchased for ∼ CAD $2,500. There is a 2-day training module to become certified to perform the test. So, although the psychometric properties of the test have been established in SCI the above barriers reduce its uptake by clinicians and researchers. The CUE-T is intended to evaluate functional limitations as described by Marino RJ[16] and is validated in SCI. It consists of 19 items, 17 of which are unilateral and test basic upper limb actions and grasp patterns. It evaluates upper extremity actions such as reaching, lifting, pulling, and pushing in addition to various grasp patterns. Most items are scored based on time or number of repetitions in 30 s. The test takes approximately 1 h to administer and unlike the 3D TRI-HFT does not integrate the use of the arm and hand during task execution. The test intends to assess the action, not the task.[16] Amongst the other widely used measures are the Functional Independence Measure (FIM);[17] and the Spinal Cord Independence Measure (SCIM), both of which have been validated in the SCI population.[18] Both these scales measure function in the activity domain. However, both scales are administered as questionnaires and hence are subjective in nature. Despite the subjective nature of the SCIM-III measure, the subjective results on the self-care subcategory are found to have a high level of agreement with the scores obtained when this part of the test is objectively administered.[19] In light of the above, we developed the Toronto Rehabilitation Institute-Hand Function Test (TRI-HFT) in 2008.[20] The TRI-HFT is a test to measure unilateral gross motor function of the upper extremity and has been validated in individuals with sub-acute SCI.[20] It consists of 2 parts and is designed to assess both proximal and distal upper extremity function. The first part of the test assesses the individuals’ ability to manipulate the following objects: (1) mug, (2) book, (3) soda can, (4) isosceles triangular sponge, (5) wireless home telephone, (6) paper sheet, (7) Ziploc bag filled with five golf balls, (8) dice, (9) credit card, (10) pencil and (11) nine rectangular blocks in sets of 3 × 100 g, 3 × 200 g and 3 × 300 g. Each of the three blocks in each weight category have surfaces with different levels of friction. The second part measures the ability to withstand eccentric forces (Bar test), the strength of lateral pinch in Newtons using a dynamometer (instrumented credit card test) and the strength of palmar grasp in Nm using a dynamometer (instrumented cylinder). The scoring system for the first part of the test considers ability to reach, grasp and manipulate and is done on a scale of 0–7, where greater scores indicate better performance. Further details on the original test objects, administration and scoring are described in Kapadia et al.[20] While the TRI-HFT has potential for high clinical utility (i.e. low cost, minimal training required to administer), a limitation to its use has been accessibility. To make it more easily accessible we replicated the test in a 3D printed format and demonstrated that the 3D printed objects were within 10% error margin of the original test objects. For details of the 3D printed version please refer to Kapadia et al.[21] The objective of the current study was to test the psychometric properties of the 3D printed Toronto Rehabilitation Institute-Hand Function Test (3D TRI-HFT) in individuals with sub-acute and chronic SCI. Specifically, we assessed the inter-rater reliability, intra-rater reliability and construct (convergent) and criterion validity of the 3D TRI-HFT within two interventional studies.

Methods

The current study was designed to evaluate the psychometric properties of the 3D TRI-HFT in sub-acute and chronic SCI populations. This study was a sub-study conducted within two single arm interventional studies that aimed at assessing the feasibility and efficacy of EEG-Triggered Functional Electrical Stimulation Therapy for Upper Limb Rehabilitation.[22] Inclusion criteria for study 1 were: (a) traumatic SCI classified as American Spinal Injury Association Impairment Scale (AIS) B-D, (b) SCI less than six months prior to baseline assessment, (c) neurological level of injury between C4 to C7 and (d) expected length of stay of at least 80 days at the time of study initiation. The exclusion criteria were (a) history of seizure disorder not effectively managed by seizure medications, (b) an existing electrical stimulation device (e.g. Implantable Cardioverter Defibrillator, Pacemaker, Spinal Stimulation), (c) rash or open wound at electrode site, (d) denervation of the targeted muscles, (e) poorly controlled autonomic dysreflexia, (f) botulinum toxin injection into affected upper extremity within three months prior to the study start, and (g) currently enrolled in another upper limb study. Inclusion criteria for study 2 were: (a) traumatic SCI classified as AIS B-D and (b) SCI at least 24 months prior to the baseline assessment. The exclusion criteria were same as the sub-acute study. Participants in both studies underwent a battery of assessments at baseline, after 20 sessions of therapy (sessions were carried out 3–5 times a week), at discharge (after 40 sessions of therapy) and at 6-month follow-up. Assessment conducted at each time point for each participant for the right and the left upper extremity was treated as an independent data point. These assessments included: (a) FIM,[17] (b) SCIM,[23] (c) 3D TRI-HFT,[21] and (d) GRASSP,[14] and were completed in one session for both studies. The 3D TRI-HFT was video recorded for assessing inter and intra-rater reliability. For construct (convergent) validity, scores on the 3D TRI-HFT were correlated with scores on the self-care sub-scale of the FIM and SCIM. For the assessment of criterion validity, scores on the 3D TRI-HFT were correlated with scores on the different components of the GRASSP, i.e. strength, sensation, qualitative and quantitative prehension. The strong psychometric properties of the GRASSP have resulted in its use in clinical trials to evaluate primary endpoints related to hand function,[24-27] and so we choose to use this test for validity testing. The FIM and SCIM are validated measures in SCI population.[14,17,23] For testing reliability and validity, scores of the left and right upper extremities as well as scores at each time point were treated as separate independent data points. Assessors: Two assessors with extensive experience in SCI research were involved in the psychometric testing of the 3D TRI-HFT. Specifically, assessor AB is a researcher working in the field of SCI research for 20 years and assessor CD is a clinician researcher working in the field of SCI for 12 years. All in-person assessments for both the studies were carried out by assessor AB and all video recorded assessments were rated by assessor CD. Both assessors have eight years on and off experience with the use of original version of the TRI-HFT however they did not receive any additional training for the use of the 3D printed version. Statistical Analysis were independently performed for the sub-acute and the chronic SCI participants as below: The inter-rater reliability of the first component of the test, i.e. object manipulation component which includes the ten objects and the set of nine rectangular blocks, was assessed. For inter-rater reliability, the de-identified video recording from both studies were reviewed by CD and each task was scored. The scores were than compared to the scores from the original testing performed by AB. Intraclass correlation coefficient (ICC) was calculated using SPSS version 16 (SPSS, Inc., USA). For the purpose of intra-rater reliability, AB and CD reviewed the participant videos from both the studies, one month apart and re-rated the participant performance on individual tasks at both time points. ICC was calculated using SPSS version 16 (SPSS, Inc., USA). The construct (convergent) validity of the 3D TRI-HFT was evaluated by computing Spearman’s correlation coefficient with the FIM self-care sub-score and the SCIM self-care sub-score. The criterion validity of the 3D TRI-HFT was established by computing Spearman’s correlation coefficient’s with the strength, sensation, qualitative prehension and quantitative prehension scores of the GRASSP. An r value of 0.0–0.4 was considered as weak correlation, 0.4–0.7 was considered as moderate correlation, and anything above 0.7 was considered as strong correlation.[28] For all statistical procedures, α was set at <0.01.

Results

For the sub-acute study, 4 of the 5 participant’s data were assessed at 3 time points in the study. The 5th participant dropped out of the study after 12 sessions. The remaining 4 participants completed all the outcome assessments at all time points except one participant who missed one assessment. Inter-rater reliability was determined using 22 datasets (n = (4 participants x3 time points)-1missed appointment X 2 (left and right UE)). For intra-rater reliability AB and CD re-assessed the participant videos and scored them again (n = 44). For assessing construct validity and criterion validity, scores on right and left hand as well as at each time point were treated as independent data points (n = 22). For the chronic study, 3 participants were recruited and assessed at 4 time points except one participant who missed one assessment. Inter-rater reliability was determined using 22 datasets and intra-rater reliability was determined using 44 datasets (both AB and CD reassessed the videos). For assessing construct validity and criterion validity scores on right and left hand as well as at each time point were treated as independent data points (n = 22). Participant demographics, FIM and SCIM self-care sub-scores for both the studies are described in Table 1 whereas raw scores on the 3D TRI-HFT and GRASSP are presented in Tables 2 and 3.
Table 1

Participant demographics.

Participant IDSexAgeLevel of injuryAISFIM-Self CareSCIM-Self Care
BaselineMidpointDFBaselineMidpointDF
Sub-acute group            
AAOFM66C4C6N18N0N9N
AAOGF72C4C162326N51013N
AAOHM37C4B666N000N
AAOIM58C4D153235N21616N
AAOJF26C5B88NN22NN
Chronic group            
AAOAM31C4B66660N00
AAOBM53C4C7910101222
AAOCF31C4C777N111N

Note: M, male; F, Female; AIS, American Spinal Injury Association Impairment Scale; FIM, Functional Independence Measure; SCIM, Spinal Cord Independence Measure; D, Discharge; F, Follow-up; N, Not available.

Table 2

3D Toronto Rehabilitation Institute – Hand Function Test scores.

  Sub-acute groupChronic group
Participant ID AAOFAAOGAAOHAAOIAAOJAAOAAAOBAAOC
  LRLRLRLRLRLRLRLR
TRI-HFT                 
Object manipulationB102856240100651616001050300
MNN682610185070NN001058300
D60676840102460702021001058420
FNNNNNNNNNN001054NN
Rectangular blocksB1738501809052181800938220
MNN63229225063NN00954220
D545463289185463181800954400
FNNNNNNNNNN00954NN
Instrumented cylinderB002000012.500000800
MNN5100019NN000800
D415900062500000620
FNNNNNNNNNN00010NN
Credit cardB0010000110000212200
MNN4000638NN0002500
D414100002450000074220
FNNNNNNNNNN00024NN
BarB001500003900000000
MNN30000060NN0002500
D2048470003560000001700
FNNNNNNNNNN00038NN

Note: L, left hand; R, right hand; B, baseline; M, midpoint; D, discharge; F, six month follow up; N, not available.

Table 3

Graded redefined assessment of strength, sensibility and prehension scores.

  Sub-acute groupChronic group
Participant ID AAOFAAOGAAOHAAOIAAOJAAOAAAOBAAOC
  LRLRLRLRLRLRLRLR
GRASSP                 
StrengthB111828174513421310522039235
MNN43289103149NN532240266
D30404531121238501512532240328
FNNNNNNNNNN532241NN
SensitivityB71021234487870094318
MNN2224531211NN00133815
D11824241172116770062921
FNNNNNNNNNN00125NN
Prehension abilityB14620009000041020
MNN11500912NN0041220
D6912600912110031290
FNNNNNNNNNN00312NN
Prehension performanceB0116300017200031260
MNN226001127NN0031550
D1121238021526400041480
FNNNNNNNNNN00313NN

Note: L, left hand; R, right hand; B, baseline; M, midpoint; D, discharge; F, six month follow up; N, not available.

Participant demographics. Note: M, male; F, Female; AIS, American Spinal Injury Association Impairment Scale; FIM, Functional Independence Measure; SCIM, Spinal Cord Independence Measure; D, Discharge; F, Follow-up; N, Not available. 3D Toronto Rehabilitation Institute – Hand Function Test scores. Note: L, left hand; R, right hand; B, baseline; M, midpoint; D, discharge; F, six month follow up; N, not available. Graded redefined assessment of strength, sensibility and prehension scores. Note: L, left hand; R, right hand; B, baseline; M, midpoint; D, discharge; F, six month follow up; N, not available.

Inter-rater reliability

For both the sub-acute and chronic SCI study, we found a statistically significant strong correlation between the two assessors for the 10-object manipulation component of the 3D TRI-HFT with an ICC = 0.994 (95% CI: 0.985–0.998; P < 0.000) (Fig. 1(a)) and ICC = 0.990 (95% CI:0.976–0.996; P < 0.000) respectively. The inter-rater reliability for the 9 rectangular blocks could not be assessed for both studies as the assessor had difficulty identifying the blocks based on their weight and texture from the video recorded assessments.
Figure 1

(a) Inter-rater Reliability of the 10-object manipulation component of the 3D TRI-HFT in sub-acute SCI. (b) Intra-rater Reliability of the 10-object manipulation component of the 3D TRI-HFT in sub-acute SCI.

(a) Inter-rater Reliability of the 10-object manipulation component of the 3D TRI-HFT in sub-acute SCI. (b) Intra-rater Reliability of the 10-object manipulation component of the 3D TRI-HFT in sub-acute SCI.

Intra-rater reliability

We found statistically significant strong correlation between the first and second rating on the 10-object manipulation component of the TRI-HFT within each assessor. ICC values for the sub-acute and chronic study are ICC = 0.995 (95% CI: 0.992–0.998; P < 0.000) (Fig. 1(b)) and ICC = 0.999 (95% CI: 0.999–1.00; P < 0.000) respectively.

Construct and criterion validity

Sub-acute SCI Study

The results showed a moderately strong correlation between the FIM selfcare sub-score and the 10-object manipulation component of the 3D TRI HFT with an r = 0.770 (95% CI: 0.567–0.890, P = 0.01) and the rectangular blocks with an r = 0.760 (95% CI: 0.535–0.887, P = 0.01). Similarly, we found a moderately strong correlation between the SCIM self-care sub-score and the 10-object manipulation component with an r = 0.769 (95% CI: 0.547–0.894, P = 0.01) and the rectangular blocks with an r = 0.758, (95% CI: 0.533–0.875, P = 0.01). The 10-object manipulation component showed very strong correlation with the Strength (r = 0.946; 95% CI: 0.830–0.986 P = 0.01), Qualitative Prehension (r = 0.951; 95% CI: 0.856–0.982 P = 0.01) and Quantitative Prehension (r = 0.974; 95% CI: 0.0.920–0.991 P = 0.01) components of the GRASSP. However, it had a moderate correlation with the Sensation component of the GRASSP (r = 0.523; 95% CI: 0.126–0.740 P = 0.05). Similarly, the rectangular blocks showed a very strong correlation with the Strength (r = 0.918; 95% CI: 0.770–0.968, P = 0.01), Qualitative Prehension (r = 0.941; 95% CI: 0.854–0.978 P = 0.01) and Quantitative Prehension (r = 0.938; 95% CI: 0.830–0.976 P = 0.01) components of the GRASSP, but had a moderate correlation with the Sensation component (r = 0.512; 95% CI: 0.115–0.749 P = 0.05). The instrumented cylinder, instrumented credit card and the bar also showed strong to very strong correlations with all components of the GRASSP except sensation with which it showed only weak correlations. For correlations with individual test components please refer to Table 4.
Table 4

Concurrent and criterion validity of the 3D TRI-HFT.

Outcome Measure3D TRI-HFT Object Manipulation3D TRI-HFT Rectangular Blocks3D TRI-HFT Instrumented Cylinder3D TRI-HFT Instrumented Credit Card3D TRI-HFT Bar
SUB-ACUTE STUDY
FIM-SC0.770**0.760**0.669**0.705**0.636**
SCIM-SC0.769**0.758**0.656**0.700**0.625**
GRASSP
Strength0.946**0.918**0.846**0.774**0.837**
Sensation0.5230.5120.3460.2350.280
Qualitative Prehension0.951**0.941**0.843**0.800**0.841**
Quantitative Prehension0.974**0.938**0.889**0.804**0.885**
CHRONIC STUDY
FIM-SC0.756**0.759**0.4810.608**0.527
SCIM-SC0.725**0.726**0.4640.578**0.540
GRASSP
Strength0.936**0.938**0.737**0.731**0.601**
Sensation0.2740.2840.0530.123−0.035
Qualitative Prehension0.965**0.965**0.779**0.851**0.635**
Quantitative Prehension0.997**0.995**0.775**0.807**0.633**

Note: FIM-SC, Functional Independence Measure Selfcare sub-score; SCIM-SC, Spinal Cord Independence Measure Selfcare sub-score; GRASSP, Graded Redefined Assessment of Strength, Sensibility and Prehension; 3D TRI-HFT, 3D printed Toronto Rehabilitation Institute-Hand Function Test. *Correlation is significant at the 0.05 level (2 tailed); ** Correlation is significant at the 0.01 level (2 tailed).

Concurrent and criterion validity of the 3D TRI-HFT. Note: FIM-SC, Functional Independence Measure Selfcare sub-score; SCIM-SC, Spinal Cord Independence Measure Selfcare sub-score; GRASSP, Graded Redefined Assessment of Strength, Sensibility and Prehension; 3D TRI-HFT, 3D printed Toronto Rehabilitation Institute-Hand Function Test. *Correlation is significant at the 0.05 level (2 tailed); ** Correlation is significant at the 0.01 level (2 tailed).

Chronic SCI study

The results showed a moderately strong correlation between the FIM selfcare sub-score and the 10-object manipulation component of the 3D TRI HFT with an r = 0.756 (95% CI: 0.497–0.898, P = 0.01) and the rectangular blocks with an r = 0.759 (95% CI: 0.530–0.895, P = 0.01). Similarly, we found a moderately strong correlation between the SCIM self-care sub-score and the 10-object manipulation component with an r = 0.725 (95% CI: 0.419–0.888, P = 0.01) and the rectangular blocks with an r = 0.726 (95% CI: 0.448–0.890, P = 0.01). The 10-object manipulation component showed very strong correlation with the Strength (r = 0.936; 95% CI: 0.813–0.980 P = 0.01) (Fig. 2(b)), Qualitative Prehension (r = 0.965; 95% CI: 0.821–0.999 P = 0.01) and Quantitative Prehension (r = 0.997; 95% CI: 0.985–1.0 P = 0.01) components of the GRASSP but had a weak correlation with the Sensation component of the GRASSP (r = 0.274; 95% CI: −0.245–0.709 P = 0.217).
Figure 2

(a) Criterion Validity of the 3D TRI-HFT 10-object manipulation component with the GRASSP Quantitative Prehension component in sub-acute SCI. (b) Criterion Validity of the 3D TRI-HFT 10-object manipulation component with the GRASSP Quantitative Prehension component in chronic SCI.

(a) Criterion Validity of the 3D TRI-HFT 10-object manipulation component with the GRASSP Quantitative Prehension component in sub-acute SCI. (b) Criterion Validity of the 3D TRI-HFT 10-object manipulation component with the GRASSP Quantitative Prehension component in chronic SCI. Similarly, the rectangular blocks also showed a very strong correlation with the Strength (r = 0.938; 95% CI: 0.813–0.980 P = 0.01), Qualitative Prehension (r = 0.965; 95% CI: 0.849–0.999 P = 0.01) and Quantitative Prehension (r = 0.995; 95% CI: 0.974–1.0 P = 0.01) components of the GRASSP but had a statistically non-significant weak correlation with the Sensation component (r = 0.284). The instrumented cylinder, instrumented credit card and the bar also showed moderately strong to strong correlations with all components of the GRASSP except sensation with which it showed only weak correlations. For correlations with individual test components please refer to Table 4.

Discussion

In this study, we assessed the psychometric properties of the 3D TRI-HFT within a small number of individuals with sub-acute and chronic SCI. Despite the small sample size, to the best of our knowledge this is the first 3D printed tool to be validated for use in SCI. The 3D printed test can be accessed from anywhere and hence it practically eliminates variability related to test manufacturing, which may often affect the test scoring, accuracy and ultimately reproducibility of test results.[21] Whilst the 10-object manipulation component of the 3D TRI-HFT showed high inter- and intra-rater reliability, the rectangular blocks could not be assessed as it was difficult to identify the blocks based on weight and texture on the videos. The 9 blocks are now numbered for easy identification. The moderately strong correlations of 3D TRI-HFT with the FIM and SCIM selfcare sub-scores is an expected finding given that FIM and SCIM also measure function at an activity level. In the study by Kalsi-Ryan et al., which looked at the psychometric properties of the GRASSP, the authors showed similar correlations between GRASSP and the FIM and SCIM selfcare sub-scores.[29] The FIM and SCIM have documented floor and ceiling effects,[18,30] and are unable to detect change at either end of the functional spectrum. The first part of the 3D TRI-HFT showed strong correlations with the strength, qualitative and quantitative components of the GRASSP, in both sub-acute and chronic SCI studies. This finding affirms that it produces the same information about a person’s upper extremity function as the Qualitative and Quantitative components of the GRASSP. The first part of the 3D TRI-HFT showed moderate correlations with sensation component of the GRASSP in sub-acute SCI population. This finding is not surprising as literature has shown poor correlations between sensory impairment and motor function. These findings are supported by Kalsi-Ryan et al., as well where the authors showed moderate correlations between International Standard for Neurological Classification of Spinal Cord Injury-Upper Extremity Motor Scores (ISNCSCI-UEMS) and sensation.[5] Advances in medical science has led to a significant improvement in life expectancy in individuals with SCI. It is upon the rehabilitation community to develop tools that can help improve quality of life and independence of this population. Research aimed at developing new rehabilitation therapies require reliable and valid outcome assessment tools that take into account practical aspects such as time needed to administer the test and personnel and financial resources.[31] Amongst the current validated upper extremity activity measures in SCI the main limitations are: (1) self-fabrication of test objects, which often compromises standardization of the test; (2) expensive (average cost ∼$2,500); (3) takes up to or more than 1 h to administer; and (4) requires extensive training on part of the assessor to administer the test correctly, which means that if the assessment is not used for a longer duration than the reliability of the test might be questionable. The 3D TRI-HFT is one of the first upper extremity measures that can assess reaching, grasping and manipulation and addresses each of the above concerns. With respect to standardization, the test is open source (files will be made available on www.kite-uhn.com) and hence every time the test objects will be manufactured in the same fashion. With respect to cost effectiveness, 3D printing technology is becoming popular, accessible and cost effective, we estimate that today the objects of the 3D TRI-HFT can be printed for ∼CAD $500 and this cost should decrease over time. The test takes approximately 15 mins to administer on one upper extremity (Table 5). With respect to training needed to administer the test, the 3D TRI-HFT requires minimal training on part of the assessor,[20] and this is in the form of self- training by reading a two page manual (available in the TRI-HFT original test publication).[20]
Table 5

Time taken (in minutes) to complete the 3D TRI-HFT* for B/L upper extremities for all study participants.

Participant IDAAOFAAOGAAOHAAOIAAOJAAOAAAOBAAOC
Time point        
Baseline14:4311:382:0611:085:253:1318:316:21
Mid-pointN13:5312:1014:33N0:4616:5513:51
Discharge19:0711:4111:5213:3313:250:0419:0012:27
Follow-upNNNNN0:4018:08N

*Tasks that were video recorded and timed include 10 object manipulation, 9 rectangular blocks and the bar test for B/L upper extremities.

Time taken (in minutes) to complete the 3D TRI-HFT* for B/L upper extremities for all study participants. *Tasks that were video recorded and timed include 10 object manipulation, 9 rectangular blocks and the bar test for B/L upper extremities. An added benefit of the 3D TRI-HFT is that since the tasks are routinely used in everyday life the results of the test are meaningful to the patients and their care providers too. By successfully completing a component of the 3D TRI-HFT the patient is shown that they can perform a particular object grasping and manipulation task. The other advantage, specifically in research settings, is that since it is universally accessible, if used uniformly across clinical trials, it will enable researchers across the globe to compare interventions that affect upper limb function and share the data.

Conclusion

In summary, the 3D TRI-HFT is an open-source test that is simple, reliable, valid and takes up to 15 min to administer. It assesses function at an activity level. In the context of interventional trials, it may be recommended to be used as an adjunct measurement to strength and sensibility measurements. It can be easily incorporated in SCI rehabilitation clinics worldwide with minimal expense to the user and since the objects are everyday objects they can double-up as therapy tools where appropriate. Click here for additional data file.
  28 in total

1.  SCIM--spinal cord independence measure: a new disability scale for patients with spinal cord lesions.

Authors:  A Catz; M Itzkovich; E Agranov; H Ring; A Tamir
Journal:  Spinal Cord       Date:  1997-12       Impact factor: 2.772

2.  Measuring patient-reported outcomes after discharge from inpatient rehabilitation settings.

Authors:  Wendy J Coster; Stephen M Haley; Alan M Jette
Journal:  J Rehabil Med       Date:  2006-07       Impact factor: 2.912

3.  The Graded Redefined Assessment of Strength Sensibility and Prehension: reliability and validity.

Authors:  Sukhvinder Kalsi-Ryan; Dorcas Beaton; Armin Curt; Susan Duff; Milos R Popovic; Claudia Rudhe; Michael G Fehlings; Mary C Verrier
Journal:  J Neurotrauma       Date:  2011-08-12       Impact factor: 5.269

4.  An objective and standardized test of hand function.

Authors:  R H Jebsen; N Taylor; R B Trieschmann; M J Trotter; L A Howard
Journal:  Arch Phys Med Rehabil       Date:  1969-06       Impact factor: 3.966

5.  A validation of the functional independence measurement and its performance among rehabilitation inpatients.

Authors:  T A Dodds; D P Martin; W C Stolov; R A Deyo
Journal:  Arch Phys Med Rehabil       Date:  1993-05       Impact factor: 3.966

6.  Targeting recovery: priorities of the spinal cord-injured population.

Authors:  Kim D Anderson
Journal:  J Neurotrauma       Date:  2004-10       Impact factor: 5.269

7.  Outcome of the upper limb in cervical spinal cord injury: Profiles of recovery and insights for clinical studies.

Authors:  Sukhvinder Kalsi-Ryan; Dorcas Beaton; Armin Curt; Milos R Popovic; Mary C Verrier; Michael G Fehlings
Journal:  J Spinal Cord Med       Date:  2014-09       Impact factor: 1.985

Review 8.  Guidelines for the conduct of clinical trials for spinal cord injury as developed by the ICCP panel: spontaneous recovery after spinal cord injury and statistical power needed for therapeutic clinical trials.

Authors:  J W Fawcett; A Curt; J D Steeves; W P Coleman; M H Tuszynski; D Lammertse; P F Bartlett; A R Blight; V Dietz; J Ditunno; B H Dobkin; L A Havton; P H Ellaway; M G Fehlings; A Privat; R Grossman; J D Guest; N Kleitman; M Nakamura; M Gaviria; D Short
Journal:  Spinal Cord       Date:  2006-12-19       Impact factor: 2.772

9.  Brain-computer interface-triggered functional electrical stimulation therapy for rehabilitation of reaching and grasping after spinal cord injury: a feasibility study.

Authors:  Lazar I Jovanovic; Naaz Kapadia; Vera Zivanovic; Hope Jervis Rademeyer; Mohammad Alavinia; Colleen McGillivray; Sukhvinder Kalsi-Ryan; Milos R Popovic; Cesar Marquez-Chin
Journal:  Spinal Cord Ser Cases       Date:  2021-03-19

10.  3-Dimensional printing in rehabilitation: feasibility of printing an upper extremity gross motor function assessment tool.

Authors:  Naaz Kapadia; Mathew Myers; Kristin Musselman; Rosalie H Wang; Aaron Yurkewich; Milos R Popovic
Journal:  Biomed Eng Online       Date:  2021-01-05       Impact factor: 2.819

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.