| Literature DB >> 34765103 |
Huijing Hu1, Wai Leung Ambrose Lo2, Xiaoyun Wang3, Le Li1, Ping Zhou4.
Abstract
The objective of this study was to evaluate the interrater and intrarater reliability of electrical impedance myography (EIM) using handheld sensors of different sizes. Electrical impedance myography of the biceps brachii muscle of twenty healthy individuals was performed by two raters using both large and small sensors. The procedures were also repeated 5 to 8 days after the first recording session. The repeatability of the resistance, reactance, and phase angle at two different current frequencies (50 and 100 kHz) was assessed by the intraclass correlation coefficient (ICC). The ICCs of the large sensor were higher than those of the small sensor for both the intrarater and interrater reliabilities. High-frequency current tended to improve the ICC for the small sensor. These results indicate reasonable repeatability of the handheld electrode arrays for EIM measurements. The findings suggest that electrode array should be selected appropriately according to the size of the tested muscle.Entities:
Mesh:
Year: 2021 PMID: 34765103 PMCID: PMC8577940 DOI: 10.1155/2021/7296322
Source DB: PubMed Journal: J Healthc Eng ISSN: 2040-2295 Impact factor: 2.682
Figure 1Configuration of electrodes and dimensions of large (a) and small (b) sensors.
Between-day longitudinal biceps EIM measurements.
| Large sensor | Small sensor | ||||||
|---|---|---|---|---|---|---|---|
| 1st measurement | 2nd measurement |
| 1st measurement | 2nd measurement |
| ||
| 50 kHz | Phase | 20.93(3.52) | 21.38 (3.25) | −0.45 | 12.23 (4.31) | 11.75 (3.58) | 0.48 |
| Reactance | 9.83 (1.08) | 10.13 (1.27) | −0.31 | 13.31 (3.10) | 12.61 (2.72) | 0.70 | |
| Resistance | 26.59 (6.91) | 26.70 (7.16) | −0.11 | 64.54 (16.60) | 63.18 (16.40) | 1.36 | |
|
| |||||||
| 100 kHz | Phase | 22.13 (3.54) | 22.56 (3.51) | −0.43 | 15.17 (3.82) | 14.64 (3.27) | 0.53 |
| Reactance | 8.46 (1.11) | 8.63 (1.31) | −0.17 | 14.97 (3.16) | 14.19 (3.08) | 0.77 | |
| Resistance | 21.63 (6.38) | 21.56 (6.52) | 0.07 | 57.11(15.54) | 56.27 (15.53) | 0.85 | |
Note: mean (standard deviation); d: mean difference between first and second measurements.
Within-day longitudinal biceps EIM measurements.
| Large sensor | Small sensor | ||||||
|---|---|---|---|---|---|---|---|
| 1st measurement | 2nd measurement |
| 1st measurement | 2nd measurement |
| ||
| 50 kHz | Phase | 21.54 (3.34) | 21.38 (3.25) | 0.15 | 12.11 (4.07) | 11.75 (3.58) | 0.36 |
| Reactance | 9.85 (1.16) | 10.13 (1.27) | −0.32 | 12.73 (3.51) | 12.61 (2.72) | 0.12 | |
| Resistance | 25.32 (5.37) | 26.70 (7.16) | −1.38 | 61.67 (7.16) | 63.18 (16.39) | −1.38 | |
|
| |||||||
| 100 kHz | Phase | 22.73 (3.80) | 22.56 (3.51) | 0.18 | 14.98 (3.56) | 14.64 (3.27) | 0.34 |
| Reactance | 8.36 (1.01) | 8.63 (1.31) | −0.27 | 14.14 (3.11) | 14.19 (3.08) | −0.05 | |
| Resistance | 20.49 (5.04) | 21.56 (3.52) | −1.06 | 54.59 (14.19) | 56.27 (15.53) | −1.68 | |
Note: mean (standard deviation); d: mean difference between first and second measurements.
Reliability indices for within-day interrater reliability of both sensors.
| Within-day interrater | Large sensor | Small sensor | |||||||
|---|---|---|---|---|---|---|---|---|---|
| ICC | SEM | SRD | 95% LOA | ICC | SEM | SRD | 95% LOA | ||
| 50 kHz | Phase | 0.97 (0.92–0.99) | 0.19 | 1.22 | −6.38–4.17 | 0.92 (0.81–0.97) | 0.59 | 2.13 | −10.44–6.20 |
| Reactance | 0.90 (0.73–0.96) | 0.21 | 1.27 | −4.62–3.29 | 0.44 (−0.41–0.78) | 2.62 | 4.49 | −15.97–8.96 | |
| Resistance | 0.99 (0.98–1.00) | 0.12 | 0.96 | −7.23–4.59 | 0.95 (0.88–0.98) | 1.53 | 3.42 | −30.09–16.03 | |
|
| |||||||||
| 100 kHz | Phase | 0.98 (0.94–0.99) | 0.13 | 0.99 | −5.80–3.88 | 0.92 (0.81–0.97) | 0.51 | 1.98 | −9.41–5.69 |
| Reactance | 0.95 (0.88–0.98) | 0.11 | 0.94 | −4.00–2.98 | 0.67 (0.19–0.87) | 1.76 | 3.68 | −14.31–8.13 | |
| Resistance | 0.99 (0.98–1.00) | 0.11 | 0.92 | −6.34–4.15 | 0.97 (0.93–0.99) | 0.82 | 2.51 | −22.12–12.04 | |
Reliability indices for between-days intrarater reliability of both sensors.
| Between-days intrarater | Large sensor | Small sensor | |||||||
|---|---|---|---|---|---|---|---|---|---|
| ICC | SEM | SRD | 95% LOA | ICC | SEM | SRD | 95% LOA | ||
| 50 kHz | Phase | 0.98 (0.95–0.99) | 0.13 | 0.99 | −5.66–3.81 | 0.84 (0.59–0.94) | 1.87 | 3.79 | −13.59–7.78 |
| Reactance | 0.90 (0.73–0.96) | 0.22 | 1.30 | −4.73–3.35 | 0.63 (0.02–0.85) | 3.30 | 5.04 | −15.28–8.62 | |
| Resistance | 0.95 (0.85–0.98) | 0.54 | 2.03 | −11.73–6.85 | 0.95 (0.88–0.98) | 2.38 | 4.27 | −29.09–15.52 | |
|
| |||||||||
| 100 kHz | Phase | 0.98 (0.96–0.99) | 0.13 | 1.00 | −5.85–3.90 | 0.91 (0.76–0.96) | 0.99 | 2.76 | −10.10–6.30 |
| Reactance | 0.90 (0.74–0.96) | 0.22 | 1.31 | −4.76–3.36 | 0.84 (0.60–0.94) | 1.47 | 3.36 | −11.24–6.60 | |
| Resistance | 0.96 (0.88–0.98) | 0.43 | 1.81 | −10.49–6.22 | 0.96 (0.89–0.98) | 1.98 | 3.90 | −25.98–13.97 | |
Figure 2ICC plots of longitudinal bicep measurements in healthy subjects using large sensors (open triangles) and small sensors (filled circles) taken by two different (inter-) raters. (a) Plots of 50 kHz data. (b) Plots of 100 kHz data.
Figure 3ICC plots of longitudinal bicep measurements in healthy subjects using large sensors (open triangles) and small sensors (filled circles) taken at baseline and 5–8 days later by the same (intra-) rater. (a) Plots of 50 kHz data. (b) Plots of 100 kHz data.