| Literature DB >> 34743494 |
Jeong Woo Park1, Chul Young Choi1.
Abstract
PURPOSE: To compare the light transmittance property of seven currently used intraocular lens (IOL) models by spectrophotometer data.Entities:
Keywords: Blue-light filtering intraocular lens; Spectrophotometer; Violet-light filtering intraocular lens
Mesh:
Year: 2021 PMID: 34743494 PMCID: PMC8849998 DOI: 10.3341/kjo.2021.0157
Source DB: PubMed Journal: Korean J Ophthalmol ISSN: 1011-8942
Fig. 1Gross photograph of various intraocular lenses and transparent glass plates used in this experiment. (A) AT LISA 809 M, (B) ZCB00, (C) ZFR00V, (D) TNFT00, (E) SN60WF, and (F) XY-1.
Fig. 2Transmission spectra of various intraocular lenses (IOLs). Data of yellow-tinted IOLs (blue-light filtering IOLs) tends to gradually increase between 400 and 500 nm while transmission in ultraviolet-light filtering IOLs drastically increases at certain wavelengths of less than 400 nm. The violet-light filtering IOL shows a graph similar to that of ultraviolet-light filtering IOLs, but the inflection point is at around 435 nm.
Spectrophotometer data of various IOLs at different wavelengths
| Variable | 375 nm | 400 nm | 425 nm | 450 nm | 475 nm | 500 nm |
|---|---|---|---|---|---|---|
| Hoya XC1-SP | 0.00 ± 0.00 | 7.20 ± 0.01 | 88.46 ± 0.05 | 91.41 ± 0.06 | 91.75 ± 0.06 | 91.89 ± 0.06 |
| Zeiss LISA 809M | 40.8 ± 0.02 | 80.24 ± 0.05 | 83.63 ±0.04 | 84.65 ± 0.05 | 86.43 ± 0.06 | 87.74 ± 0.04 |
| Tecnis ZCB00 | 0.10 ± 0.00 | 73.40 ± 0.03 | 91.31 ± 0.01 | 92.45 ± 0.02 | 92.43 ± 0.02 | 92.39 ± 0.03 |
| Synergy ZFR00V | 0.00 ± 0.00 | 0.00 ± 0.00 | 6.23 ± 0.02 | 86.05 ± 0.02 | 91.51 ± 0.02 | 91.85 ± 0.02 |
| Hoya XY1 | 0.00 ± 0.00 | 0.93 ± 0.01 | 18.44 ± 0.01 | 45.65 ± 0.02 | 73.95 ± 0.02 | 88.81 ± 0.03 |
| Alcon SN60WF | 0.00 ± 0.00 | 4.64 ± 0.02 | 28.67 ± 0.11 | 44.68 ± 0.10 | 66.29 ± 0.04 | 81.28 ± 0.03 |
| Alcon PanOptix | 0.00 ± 0.00 | 3.74 ± 0.02 | 27.70 ± 0.13 | 44.33 ± 0.10 | 67.53 ± 0.02 | 83.86 ± 0.04 |
Values are presented as mean ± standard deviation. Ultraviolet-light filtering IOLs show a plateau after 425 nm while blue-light filtering IOLs show a gradual increase of transmission between 400 and 475 nm. The violet-light filtering IOL (ZFR00V) shows a rapid increase of transmission between 425 and 450 nm while absolutely blocking shorter wavelengths.
IOL = intraocular lens.