Chloé Guitart1, Yves-Alain Robert2, Nasim Lotfinejad1, Simon Fourquier3, Yves Martin1,2, Daniela Pires1, Julien Sauser1, René Beuchat3, Didier Pittet4.
Abstract
BACKGROUND: We developed SmartRub® powered by iQati®, an electronic device composed of a wristband and an alcohol-based handrub pocket-sized dispenser that measures and provides feedback on the duration of hand friction and the volume poured during each hand hygiene action. We aimed to assess the accuracy of SmartRub®.Entities:
Keywords: Electronic monitoring device; Hand hygiene; Hospital acquired infections; Sensitivity; Specificity
Mesh:
Substances:
Year: 2021 PMID: 34742337 PMCID: PMC8572444 DOI: 10.1186/s13756-021-01026-2
Source DB: PubMed Journal: Antimicrob Resist Infect Control ISSN: 2047-2994 Impact factor: 4.887
Fig. 1Three hardware components of SmartRub® device: the station, the bracelet and the cylinder
Fig. 2Pre-planned path of five true actions and five false actions
Phase 1 (laboratory conditions) predetermined true and false actions of the bottle and the wristband
| Bottle actions | Wristband actions |
|---|---|
| Pouring a volume < 1.5 mL | Slow HH action during [0; 10 s]; [10–15 s] or [15; ∞[ |
| Pouring a volume [ 1.5–2.5 mL] | Fast HH action during [0; 10 s]; [10–15 s] or [15; ∞[ |
| Pouring a volume > 2.5 mL | Usual HH action during [0; 10 s]; [10–15 s] or [15; ∞[ |
| Displacing horizontally the bottle with cap closed | Hands into the pockets |
| Displacing horizontally the bottle with cap open | Moving the bottle with the cap closed |
| Bottle falling down the floor with cap closed | Cleaning the table |
| Bottle falling down the floor with cap open | Start writing |
Shaking the bottle with cap closed Shaking the bottle with cap open | (all actions were performed after an action of the bottle) |
Numbers of actions performed in each study phase
| True positive (n) | False negative ( n) | True negative (n) | False positive (n) | |
|---|---|---|---|---|
| Gel | 214 | 22 | 238 | 0 |
| Rinse | 252 | 2 | 232 | 23 |
| Total | 466 | 24 | 470 | 23 |
| Gel | 188 | 16 | 204 | 0 |
| Rinse | 196 | 8 | 200 | 4 |
| Total | 384 | 24 | 404 | 4 |
| Feedback off | 223 | 20 | 201 | 43 |
| Feedback on | 235 | 6 | 197 | 42 |
| Total | 458 | 26 | 398 | 85 |
| Total | 241 | 8 | 231 | 5 |
Phase 1, controlled laboratory condition; Phase 2, pre-planned clinical path; Phase 3, clinical care condition; N, number of actions
Sensitivity, specificity, PPV and NPV of SmartRub® device in phase I (part 1 and 2), II and III
| Sensitivity % (95% CI) | Specificity % (95% CI) | PPV % (95% CI) | NPV % (95% CI) | |
|---|---|---|---|---|
| Gel | 90.7 (86.2–94.1) | 100.0 (98.5–100.0) | 99.4 (96.5–99.8) | 91.0 (87.2–93.6) |
| Rinse | 99.2 (97.2–99.9) | 91.0 (86.8–94.2) | 91.7 (88.2–94.2) | 99.1 (96.7–99.8) |
| Gel | 92.2 (87.6–95.5) | 100.0 (98.2–100.0) | 99.3 (96.0–99.8) | 92.2 (88.1–94.7) |
| Rinse | 96.1 (92.4–98.3) | 98.0 (95.1–99.5) | 98.0 (94.9–99.2) | 96.2 (92.7–98.0) |
| Gel | 91.8 (87.6–94.9) | 83.5 (78.3–88.0) | 84.8 (80.7–88.1) | 91.0 (86.9–93.9) |
| Rinse | 97.5 (94.7–99.1) | 81.2 (75.7–86.0) | 83.9 (80.0–87.1) | 97.0 (93.7–98.6) |
| Feedback off | 91.8 (87.6–94.9) | 82.4 (77.0–86.9) | 83.9 (79.8–87.3) | 90.9 (86.8–93.9) |
| Feedback on | 97.5 (94.7–99.1) | 82.4 (77.0–87.0) | 84.7 (80.8–88.0) | 97.1 (93.7–98.7) |
| Total | 96.8 (93.8–98.6) | 98.3 (95.7–99.5) | 98.3 (95.6–99.3) | 96.8 (93.9–98.4) |
95% CI, confident interval 95%; PPV, positive predictive value; NPV, negative predictive value; Phase 1, controlled laboratory condition; Part 1, bottle only, Part 2, bottle and wristband combined; Phase 2, pre-planned clinical path; Phase 3, clinical care condition