| Literature DB >> 34740876 |
Dee Wang1, William M Macharia2, Roseline Ochieng2, Dorothy Chomba2, Yifat S Hadida3, Roman Karasik3, Dustin Dunsmuir4, Jesse Coleman5, Guohai Zhou6, Amy Sarah Ginsburg7, J Mark Ansermino4.
Abstract
BACKGROUND: Globally, 2.5 million neonates died in 2018, accounting for 46% of under-5 deaths. Multiparameter continuous physiological monitoring (MCPM) of neonates allows for early detection and treatment of life-threatening health problems. However, neonatal monitoring technology is largely unavailable in low-resource settings.Entities:
Keywords: intensive care units; neonatal; neonatology; technology
Mesh:
Year: 2021 PMID: 34740876 PMCID: PMC9125375 DOI: 10.1136/archdischild-2021-322344
Source DB: PubMed Journal: Arch Dis Child ISSN: 0003-9888 Impact factor: 4.920
Eligibility criteria and study definitions
| Eligibility criteria | |
| Inclusion criteria |
Corrected age of <28 days Caregivers were willing and able to provide informed consent and were available for follow-up for the duration of the study |
| Exclusion criteria |
Receiving continuous positive airway pressure or mechanical ventilation Skin abnormalities in the nasopharynx and/or oropharynx Contraindication to skin sensor application Known arrhythmia Congenital abnormality requiring major surgical intervention Any medical or psychosocial condition or circumstance that would interfere with study conduct or for which study participation could put the neonate’s health at risk |
|
| |
| Epoch | A 60 s period of time |
| Breath | One cycle of neonate-initiated inhalation and exhalation |
| Breath start | End of a waveform trough (low point) where the carbon dioxide level starts to ascend |
| Respiratory rate (RR) manual counting protocol | A breath was counted if the waveform peak reached either 15 mmHg or the average peak of the epoch, AND the waveform trough dipped below the average trough of the epoch plus 10 mmHg Each plot was counted by two independent readers and averaged If the difference in the counts was >5, a third independent reader counted the plot If the third count was within 5 breaths of either previous count, the average of the two closest counts was used |
| RR epoch exclusion criteria | RR epoch excluded if (i) the difference between the epoch count and median RR was >10, (ii) either value was <15, (iii) the capnogram contained a digital artefact or (iv) if there was lack of inter-reader manual count agreement |
| RR median calculation | For each breath in an epoch, an instantaneous RR could be calculated by the breath duration, and the median of all these instantaneous RR values in an epoch was then calculated |
| Heart rate (HR) median calculation | Heart beats were identified by the timing of the Masimo Rad-97 reference device plethysmograph quality index (PO-SQI) which occur at the peak of each heart beat. For each heart beat in an epoch, an instantaneous HR could be calculated by the time between the previous heart beat peak and current heart beat peak, and the median of all these instantaneous HR values in an epoch was then calculated |
| Adequate signal quality | EarlySense device: A signal quality score ≥0.7 for at least 50% of the epoch |
Figure 1(A) Overview of the set-up showing the Masimo Rad-97 device with touchscreen interface (1), pulse oximeter probe (2), Nomoline for capnography (3), EarlySense processing unit (4), EarlySense under-mattress sensor (5). (B) Close-up of the EarlySense sensor under a mattress. The EarlySense sensor is connected to the processing unit that processes, stores data and sends results wirelessly to the remote display unit where the data are presented.
Results from Bland-Altman analysis for EarlySense investigational device versus Masimo Rad-97 reference device heart rate (HR) and respiratory rate (RR) measurements
| Open label | Closed-label round 1 | Closed-label round 2 | Closed-label round 3 | |
|
| ||||
| Mean Masimo Rad-97 HR | 132.3 | 128.2 | 139.1 | 130.2 |
| Normalised bias (95% CI) | −0.5% | −1.2% | −0.4% | −0.6% |
| Normalised spread between 95% LOA (upper and lower 95% LOA) | 19.7% | 15.3% | 20.1% | 6.2% |
| Normalised RMSD | 5.0% | 4.1% | 5.1% | 1.7% |
|
| ||||
| Mean Masimo Rad-97 RR | 48.0 | 51.7 | 54.2 | 48.5 |
| Normalised bias (95% CI) | 3.0% | 0.3% | 2.5% | 3.3% |
| Normalised spread between 95% LOA (upper and lower 95% LOA) | 45.3% | 37.6% | 24.8% | 27.3% |
| Normalised RMSD | 11.9% | 9.6% | 6.8% | 7.7% |
|
| ||||
| Mean Masimo Rad-97 RR | 50.3 | 53.5 | 56.4 | 50.7 |
| Normalised bias (95% CI) | −1.9% | −3.1% | −1.5% | −1.2% |
| Normalised spread between 95% LOA (upper and lower 95% LOA) | 41.6% | 35.9% | 19.2% | 19.8% |
| Normalised RMSD | 10.8% | 9.7% | 5.1% | 5.2% |
LOA, limits of agreement; RMSD, root-mean-square deviation.
Figure 2Bland-Altman plots for heart rate (HR). (A) Open-label round. (B) Closed-label round one. (C) Closed-label round two. (D) Closed-label round three. Colours indicate which participant neonate is associated with the measurement pair.
Figure 3Clarke error grids for closed-label round three measurements. (A) Comparison of heart rate (HR) measurements. (B) Comparison of EarlySense respiratory rate (RR) to Masimo Rad-97 RR manual count. Each dot represents a data pair, with the colour intensity proportional to density of data pairs. Region A (in green) contains data pairs that are within 20% of the Masimo Rad-97 device value. Region B (in yellow) contains data pairs not within 20% that would not lead to unnecessary treatment. Regions C, D and E are in red. C includes data pairs leading to unnecessary treatment. D includes data pairs with a failure in detecting low or high HR/RR events and E includes data pairs where low and high HR/RR events are confused.
Figure 4Bland-Altman plots for manual counts of respiratory rate (RR). (A) Open-label round. (B) Closed-label round one. (C) Closed-label round two. (D) Closed-label round three. Colours indicate which participant neonate is associated with the measurement pair.