| Literature DB >> 34739886 |
François Lellouche1, Carole-Anne Lavoie-Bérard2, Emilie Rousseau3, Pierre-Alexandre Bouchard3, Jean-Claude Lefebvre2, Richard Branson4, Laurent Brochard5.
Abstract
Entities:
Mesh:
Year: 2021 PMID: 34739886 PMCID: PMC8563017 DOI: 10.1016/S2213-2600(21)00404-5
Source DB: PubMed Journal: Lancet Respir Med ISSN: 2213-2600 Impact factor: 30.700
FigureAbsolute humidity of devices according to manufacturers’ values and psychrometric measurements
Medline, Sunmed, Inter-Therm, and MR 850 devices were involved in endotracheal tube occlusions in patients with COVID-19; HygrobacS was not. A humidity of 28 mg H2O/L (red line) was considered to be a safe limit, with psychrometric and gravimetric measurements performed in independent and specialised laboratories.9, 15 Blue lines represent the recommended minimum humidity for heat and moisture exchangers (ISO 9360) and heated humidifiers (ISO 8185). ISO=International Organization for Standardization. *Measurements were performed according to ISO standards, with tidal volumes of 500 mL (HygrobacS) or 250 mL (Inter-Therm), or unreported conditions (Medline and Sunmed); data were collected from the manufacturers’ websites or measurements of hygrometry with heated humidifiers and heat and moisture exchangers in our laboratory in June, 2021.