BACKGROUND: Magnetic resonance-guided focused ultrasound surgery is an incisionless energy-based thermal method that is used for ablating tumors in the veterinary clinic. AIMS AND OBJECTIVES: In this article we describe a prototype of a veterinary system compatible with magnetic resonance imaging intended for small-to-medium-sized companion animals that was developed and tested in vivo in adult rabbits. METHODS: Real-time monitoring of the ablation during the experiment was possible with MR thermometry. Experiments involved thermal monitoring of sonications applied in the thigh of the rabbits. A 38-mm diameter transducer operating at 2.6 MHz was used with a 60-mm-focal length. The robotic system employed 3 linear axes and one angular axis. For this study, only X and Y axis were enabled. Due to the target size limitations, motion in Z and Θ was not needed. The functionality of the positioning device was evaluated by means of MR thermometry, demonstrating sufficient heating and accurate motion in both axes of operation. RESULTS: The postmortem findings confirm the ability of the system to induce thermal ablations in vivo in the absence of adverse effects. CONCLUSIONS: The device is a reliable and affordable solution for companion animal hospitals, offering and additional tool for the veterinary oncology society. Copyright:
BACKGROUND: Magnetic resonance-guided focused ultrasound surgery is an incisionless energy-based thermal method that is used for ablating tumors in the veterinary clinic. AIMS AND OBJECTIVES: In this article we describe a prototype of a veterinary system compatible with magnetic resonance imaging intended for small-to-medium-sized companion animals that was developed and tested in vivo in adult rabbits. METHODS: Real-time monitoring of the ablation during the experiment was possible with MR thermometry. Experiments involved thermal monitoring of sonications applied in the thigh of the rabbits. A 38-mm diameter transducer operating at 2.6 MHz was used with a 60-mm-focal length. The robotic system employed 3 linear axes and one angular axis. For this study, only X and Y axis were enabled. Due to the target size limitations, motion in Z and Θ was not needed. The functionality of the positioning device was evaluated by means of MR thermometry, demonstrating sufficient heating and accurate motion in both axes of operation. RESULTS: The postmortem findings confirm the ability of the system to induce thermal ablations in vivo in the absence of adverse effects. CONCLUSIONS: The device is a reliable and affordable solution for companion animal hospitals, offering and additional tool for the veterinary oncology society. Copyright:
It is widely accepted among the international veterinary society that animal oncology care is significantly falling behind human oncology. While targeted therapies, immunotherapies, and sophisticated interventional techniques are routinely available for humans, a fraction of animal oncology cases have access to the now traditional treatments such as chemotherapy and radiation therapy, while referral culture is not fully developed.[1] It is also an undisputable fact that veterinary medicine adopts therapies, long after they are validated and established in human medicine, such as immunotherapies[2] and small molecule inhibitors.[3] Research in the veterinary field has the potential to change the dynamics of this imbalanced relationship, which will benefit both animals and humans. Due to a largely common genetic, molecular and histological background, studies in dogs have facilitated the translation of cancer trials to humans.[4] Notably, these studies have successfully answered research questions, where mice and human trials failed.[5] These advancements have been materialized partially due to the increased availability of affordable genomic and molecular characterization; bringing precision cancer medicine a step closer to veterinary clinics.[6] The benefits of these studies not only promote the veterinary oncology field but also benefit humankind, making companion animals valuable partners for translational science.[7] It is worth noting that taking the example of human clinical trials and running multicenter, double-blinded trials to compare existing and novel therapies would enhance the credibility of thereof. However, the logistics and economics of veterinary medicine currently preclude this practice in large scale. Nevertheless, carefully designed studies should be adopted to comply with the ethical parameters of studies in vivo.[8]Magnetic resonance-guided focused ultrasound surgery (MRgFUS) is an incisionless energy-based thermal method to ablate tumors. Up to date, its use has been reported for the treatment of prostate,[910] bone metastases,[11] pancreas,[12] liver,[13] and uterine fibroids.[14] It is among the modalities currently investigated for tumor ablation in veterinary medicine.[15] Currently, only a handful of veterinary hospitals are utilizing MRgFUs for veterinary medicine applications. The scarcity of MRgFUs veterinary centers can be attributed to a number of reasons such as the lack of affordable devices for treatment, the small number of dedicated animal magnetic resonance imaging (MRI) imaging centers where MRI-guidance is an absolute requirement, the limited evidence-based literature in the field, and the lack of interest by MRgFUs vendors to invest and expand in the field of veterinary medicine.Following prostate treatment advancements in human oncology, analogous companion animal studies have been conducted to investigate the potential of the technology for prostate therapies.[16171819202122] A study focusing on naturally occurring solid tumors in canines such as in urinary bladder, oral cavity, hip, mammary gland, orbits, and abdominal cavity, demonstrated a significant decrease in tumor size in some and a noticeable clinical improvement (cessation of bleeding) in subjects with hemorrhagic tumors.[23] In another study, MRgFUS was successful in ablating significant volumes of hepatocellular adenomas in four canines,.[24] Interestingly, the technique is also finding application as a part of novel therapy in combination with advanced therapeutics, such as an anticancer micelle.[25] It is of paramount importance that MRgFUS is becoming a drug delivery component, where it can be utilized as a remote switch for targeted delivery of molecules. These advancements will reveal the full potential of the technique in the stage of advanced oncology therapeutics.In ophthalmology, the hypotensive effect of high-intensity focused ultrasound (HIFU) in managing intraocular pressure in dogs with primary glaucoma was demonstrated successfully without observing intra-operative complications.[26] The technique has also been tested with promising results in interventional cardiology strategies such as for managing ventricular arrhythmia,[2728] cardiac ablation,[2930] and atrial fibrillation treatment.[3132] The effectiveness of HIFU in treating drug-resistant hypertension via bilateral renal nerves ablation was investigated in canine models.[33] The study reported an effective renal sympathetic denervation confirmed by histology that resulted in an improvement in renal function. Another study initially utilized a model of hypersplenism and was tested in secondary cases of splenomegaly.[34] Apart from medical and surgical pathologies, the technique has been proposed as a noninvasive neutering method for male contraception.[3536]In our work, a novel MRgFUS system for veterinary applications was presented. The proposed system is nonmagnetic and thus fully compatible with any commercial 1.5T MRI. The robotic system is highly customizable offering up to 4 degrees of freedom (X, Y, Z, and Θ axes) depending on the intended use of the device. A custom-made platform is designed to controls the positioning device and the sonication parameters of the ultrasound system. The robotic device is compact (44 cm in length, 14.7 cm in width, and 6.15 cm in height) and ergonomic. The device is carrying a single element transducer, which translates into a cost-effective solution for the veterinary market. In this in vivo feasibility study with rabbits, the following aspects of the device were investigated: (1) The coupling efficiency to allow safe and efficient ablation, (2) The positioning device's movement accuracy in standard operating conditions, (3) Animal placement and immobilization for the procedure, (4) The transducer efficiency to create both discrete and overlapping lesions.
MATERIALS AND METHODS
Mechanical design of the positioning device
The positioning device includes four computer-controlled axes (X, Y, Z and Θ). Figure 1 shows the CAD drawings of the robotic system. The moving plate of each axis is coupled to a threaded plastic screw which was attached directly to the shaft of a piezoelectric motor (USR 30-S3, Shinsei Kogyo Corp., Tokyo, Japan). The rotation of the shaft converts the angular motion to the linear translation of the plate. An encoder plastic strip is attached on the axis frame and scanned by an encoder module (US Digital Corporation, Vancouver, WA 98684, USA) provides in real-time displacement feedback to the control platform. The encoder module EM1-0-500-I (US Digital Corporation, Vancouver, WA 98684, USA) was used for all three linear axes. For the theta axis, which is coupled directly on the motor's shaft, an EM1-2-2500-I (US Digital Corporation, Vancouver, WA 98684, USA) encoder was used. The encoder output is connected to the counter input of a data acquisition board USB 6251 (National instruments, Austin, Texas, USA). Images of the developed robotic system are shown in Figure 2.
Figure 1
(a) External CAD drawing of the robotic system. (b) Internal CAD drawing of the robotic system
Figure 2
(a) Photo of the developed robotic system. (b) Photo of the rabbit placement for procedure on the robotic system
(a) External CAD drawing of the robotic system. (b) Internal CAD drawing of the robotic system(a) Photo of the developed robotic system. (b) Photo of the rabbit placement for procedure on the robotic systemCompared to our previous animal positioning device,[37] the main difference is the size of the motors, which is significantly smaller in the new design (almost 50% smaller), allowing the development of a more compact positioning device. The transducer is immersed in a separate watertight compartment of the device, which is filled with degassed water.The transducer's active element consisted of a P762-type piezoceramic crystal (Piezohannas Tech. Co., Wuhan, China). Epoxy capable of withstanding temperatures up to 100°C without losing its integrity was used as a backing material of the transducer element. The transducer's impedance was matched to 50 Ω. To reduce the transmission of high-frequency harmonics, a custom made low-pass filter was used (10 MHz cut-off frequency).The positioning device was modeled using CAD software (Microstation V8, Bentley Systems, Inc., Beijing, China). The positioning device's parts were fabricated using fusion deposition modelling technology (FDM400, Stratasys, Eden Pairie, Minnesota, USA) that employed acrylonitrile butadiene styrene filament.
Software
The positioning device is controlled via an in-house developed user-friendly program developed in C # (Visual Studio 2010 Express, Microsoft Corporation, Redmond, USA). The software holds the following functionalities: (a) Communication with MRI, (b) 2 axes movement either manually or automatically by specifying the algorithm,[38] the step and the number of steps, (c) MR thermometry, and (d) Ultrasound control (frequency, power, sonication time, and cooling time).
Electronic system
The motor drivers were hosted in a metallic enclosure placed outside the MRI room. The enclosure includes a DC supply (24 V, 2 A), which drives the Shinsei motors. Wires connect the motor drivers and a data acquisition interface card (USB 6251, NI) via a connecting block. The USB 6251 interface card includes the timing and digital I/O modules. The motors are driven in clockwise rotation when the ground and clockwise terminals of the motor drivers are connected to the same potential or in the anti-clockwise rotation when the ground and anti-clockwise terminals of the motor drivers are connected to the same potential.
Magnetic resonance-guided-focused ultrasound surgery system
The proposed MRgFUS system is modular system composed of the positioning device described above, a signal generator (HP 33220A, Agilent technologies, Englewood, CO, USA), an RF amplifier (Model: BT00250-AlphaS-CW, Tomco Technologies, Stepney, Australia) and a spherical transducer made from piezoelectric ceramic (Ferroperm). The transducer operates at 2.578 MHz, has focal length of 60 mm and diameter of 38 mm.
In vivo study
The experimental protocol was approved by the Animal Welfare Committee of the Veterinary Services, Ministry of Agriculture, Republic of Cyprus (CY/EXP/PR.L6/2019). Ten female locally bred rabbits provided by an approved vendor were included in the study. A qualified veterinary surgeon was assigned to continuously monitor the vital signs of the animals throughout the experiment. The rabbits underwent injectable anesthesia with the combined intramuscular administration of 0.5 mg/kg medetomidine (Medeson, Invesa SA, Barcelona, Spain) and 0.15 mg/kg Ketamine (Narketan-10, Vetoquinol LTD, Towcester, UK). The thighs of the rabbit were depilated (VEET, Reckitt Benkiser, Slough, UK), and the rabbit was placed on the platform with its thigh immersed in the degassed water above the transducer. Upon completion of the experimental procedure, the animals were humanely euthanized with intracardial injection of an agent containing embutramide, mebezonium, and tetracaine (T61, MSD Animal Health, Madison, USA). The tissues of the animals were dissected postmortem for macroscopic observation and measurements.
X-ray imaging
X-rays were taken prior treatment with an X-ray system (IMS001, Shenzhen Browiner Tech Co., Ltd, Shenzhen, China). The animal was set on left recumbency above a CR (Computed Radiography) cassette. On taking the X-ray, the latent image of the cassette was digitized with a CR reader (Vita Flex, Carestream Health, Inc., NY, USA). The purpose of the X-ray taken was to estimate the available muscle area for ablation with dedicated software (Carestream image suite, Carestream Health, Inc., NY, USA).
Magnetic resonance imaging
For the real-time assessment of tissue ablation in situ, the full procedure was run on the anesthetized rabbit while being tested in a 1.5 T MR system (Signa Excite 1.5 T, by General Electric, Fairfield, CT, USA). High-resolution MRI was performed to visualize the animal/transducer setup. Thus, a T2-weighted fast spin-echo sequence was used with the following parameters: Repetition time (TR) was 2500 ms, echo time (TE) was 60 ms, slice thickness was 3 mm (gap 0.3 mm), matrix 256 × 256, field of view (FOV) was 16 cm, the number of excitations (NEX) was 3, while echo train length (ETL) was 8.
Magnetic resonance thermometry
Temperature maps were reconstructed in quasi real time after data processing. Image data were acquired in the coronal plane using echo-planar imaging (EPI) single-shot sequences for monitoring sonications of 23 W acoustic power. The main scanning parameters used to obtain these images were: TE: 24 ms, TR: 80 ms, NEX: 8, ETL: 1, FOV: 20 cm, slice thickness: 3 mm, while flip angle was 25°.Temperature increment during thermal ablation was estimated using the proton resonance frequency shift equation as described in detail.[39] The equation relates to the measured MR signal's phase shift with temperature elevation (ΔT). This relationship is given by:where φ(T0) and φ(T) are the phases at a starting and final temperature T0 and T, respectively, γ is the gyromagnetic ratio in MHz/T, α is the PRF change coefficient in ppm/°C, B0 is the magnetic field strength in T and TE is the echo time.
RESULTS
As shown in Figure 3, X-ray imaging of the rabbit thigh, and demarcation of the muscle tissue, supported calculation of area available for thermal ablation. The thigh area ranged from 21 cm2 to 30 cm2 with an average of 25.57 cm2 and standard deviation of 3.16 cm2.
Figure 3
Radiography of a rabbit thigh and calculation of the muscle area reveals the tissue available for ablation
Radiography of a rabbit thigh and calculation of the muscle area reveals the tissue available for ablationHigh-intensity sonications were able to produce 5-mm diamater discrete lesions, as shown in Figure 4a. These lesions are shown in a plane perpendicular to the ultrasound beam. Settings: Acoustic Power was set at 15.64 W, focal depth at 10 mm and sonication time was 30 s. The lesion length varied from 12 mm to 15 mm with an average of 13.5 mm and a standard deviation of 1.5 mm. Figure 4b shows the same discrete lesions after dissection. These lesions are shown in a plane parallel to the ultrasound beam. Settings: Acoustic Power was set at 15.64 W, focal depth at10 mm and sonication time was 30 s. The lesion length varied from 12 mm to 15 mm with an average of 13.5 mm and standard deviation of 1.5 mm.
Figure 4
(a) Macroscopic appearance of discrete lesions inflicted by ablation with the system demonstrating: (a) The lesion size in a plane perpendicular to the transducer. (b) The lesion after dissection in a plane parallel to the transducer (f = 2.578 MHz, Acoustic Power = 15.64 W, focal depth = 10 mm, sonication time = 30 s) Red arrows indicate lesions
(a) Macroscopic appearance of discrete lesions inflicted by ablation with the system demonstrating: (a) The lesion size in a plane perpendicular to the transducer. (b) The lesion after dissection in a plane parallel to the transducer (f = 2.578 MHz, Acoustic Power = 15.64 W, focal depth = 10 mm, sonication time = 30 s) Red arrows indicate lesionsThe designed pathway of the grid pattern is shown in Figure 5. Each numbered point corresponds to an ablation site, 5 mm apart. Sonication time was 30 s and cooling time was 60 s.
Figure 5
Pathway of the grid pattern that was performed on the rabbit's thigh. A 3 × 3 grid pattern with 5 mm spacing and a cooling time of 140 s. Between ablations was followed. Ablation settings: f = 2.578 MHz, Acoustic Power = 27 W, focal depth = 10 mm, sonication time = 60 s)
Pathway of the grid pattern that was performed on the rabbit's thigh. A 3 × 3 grid pattern with 5 mm spacing and a cooling time of 140 s. Between ablations was followed. Ablation settings: f = 2.578 MHz, Acoustic Power = 27 W, focal depth = 10 mm, sonication time = 60 s)Figure 6a shows overlapping lesions created with a series of high intensity sonications. These lesions are shown in a plane perpendicular to the ultrasound beam. Settings: Acoustic Power was set at 29.9 W, focal depth was 10 mm, sonication time at 30 s. A 3 × 3-point grid with 60 s cooling time between successive ablations was followed. Figure 6b shows the same overlapping lesions after dissection. These lesions are shown in a plane parallel to the ultrasound beam. Settings: Acoustic Power was set at 29.9 W, focal depth was 10 mm, sonication time at 30 s. A 3 × 3-point grid with 60 s. Cooling time between ablations was followed. The overall diameter of the overlapping lesions was 20 mm, while the lesion length was 14 mm.
Figure 6
(a) Macroscopic appearance of overlapping lesions inflicted by ablation with the FUSROBOT system demonstrating the lesion size in a plane perpendicular to the transducer. (b) The lesions after dissection in a plane parallel to the transducer. Ablation settings: f = 2.578 MHz, acoustic power = 29.9 W, focal depth = 20 mm, sonication time = 30 s. A 3 × 3 grid pattern with 5 mm spacing and a cooling time of 140 s. Between ablations was followed. Red arrow indicates lesions
(a) Macroscopic appearance of overlapping lesions inflicted by ablation with the FUSROBOT system demonstrating the lesion size in a plane perpendicular to the transducer. (b) The lesions after dissection in a plane parallel to the transducer. Ablation settings: f = 2.578 MHz, acoustic power = 29.9 W, focal depth = 20 mm, sonication time = 30 s. A 3 × 3 grid pattern with 5 mm spacing and a cooling time of 140 s. Between ablations was followed. Red arrow indicates lesionsTemperature maps reconstructed after data processing and which were obtained in the coronal plane are demonstrated in Figure 7a. Settings: Acoustic power was set at 27 W, focal depth was 20 mm sonication time of 10 s, 20 s, 30 s, 40 s, 50 s, and 60 s using single-shot EPI gradient echosequence. MR parameters: NEX: 8, TR: 80 ms, TE: 24 ms, temporal resolution of the sequence was 1.64 s/image. The temperature change recorded at 60 s was 17.06°C. Figure 7b shows the temperature maps reconstructed after data processing and which were obtained in the sagittal plane. Settings: Acoustic power was set at 27 W, focal depth was 20 mm sonication time of 10 s, 20 s, 30 s, 40 s, 50 s, and 60 s. Within the rabbit using EPI single-shot sequence. MR parameters: NEX: 8, TR: 80 ms, TE: 24 ms, was temporal resolution of the sequence was 1.64 s/image. The temperature change recorded at 60 s was 39.59°C.
Figure 7
(a) Temperature maps (coronal plane) recorded using single shot echo-planar imaging sequence for sonication of 10 s, 20 s, 30 s, 40 s, 50 s and 60 s. MR parameters: Number of excitations = 8, TR = 80 ms, TE = 24 m. Yellow circle indicates the rabbit thigh. Red arrow indicates lesion. (b) Temperature maps (sagittal plane) recorded using single shot echo-planar imaging sequence and the transducer. Acquisition time of each image = 1.64. Ablation settings: f = 2.578 MHz, Acoustic Power = 27 W, focal depth = 20 mm) for sonication of 10 s, 20 s, 30 s, 40 s, 50 s and 60 s. s. Yellow circle indicates the rabbit thigh. Red arrow indicates lesion
(a) Temperature maps (coronal plane) recorded using single shot echo-planar imaging sequence for sonication of 10 s, 20 s, 30 s, 40 s, 50 s and 60 s. MR parameters: Number of excitations = 8, TR = 80 ms, TE = 24 m. Yellow circle indicates the rabbit thigh. Red arrow indicates lesion. (b) Temperature maps (sagittal plane) recorded using single shot echo-planar imaging sequence and the transducer. Acquisition time of each image = 1.64. Ablation settings: f = 2.578 MHz, Acoustic Power = 27 W, focal depth = 20 mm) for sonication of 10 s, 20 s, 30 s, 40 s, 50 s and 60 s. s. Yellow circle indicates the rabbit thigh. Red arrow indicates lesionCoronal MR images with T2-weighted fast spin-echo sequence with the following parameters: TR: 2500 ms, TE: 60 ms, slice thickness: 3 mm (gap 0.3 mm), matrix 256 × 256, FOV: 16 cm, NEX: 3, and ETL: 8. Settings: Acoustic Power was set at 27 W, focal depth was 20 mm, sonication time at 60 s. A 3 × 3-point grid with 140 s cooling time between successive ablations was followed. The lesions appear in a plane perpendicular to the beam. The dimensions of the ablated area were 20 × 25 mm, while the lesion length was 24 mm [Figure 8].
Figure 8
Coronal magnetic resonance images showing lesions with T2W-weighted fast spin echo sequence was used with the following parameters: TR = 2500 ms, TE = 60 ms, slice thickness = 3 mm (gap 0.3 mm), matrix = 256 × 256, field of view = 16 cm, number of excitations = 3, echo train length = 8. Ablation settings: Frequency = 2.578 Mhz, acoustic power = 27 W, focal depth = 20 mm, sonication time = 60 s. A 3 × 3 grid pattern with 5 mm spacing and a cooling time of 140 s. Between ablations was followed. The lesions appear in a plane perpendicular to the beam
Coronal magnetic resonance images showing lesions with T2W-weighted fast spin echo sequence was used with the following parameters: TR = 2500 ms, TE = 60 ms, slice thickness = 3 mm (gap 0.3 mm), matrix = 256 × 256, field of view = 16 cm, number of excitations = 3, echo train length = 8. Ablation settings: Frequency = 2.578 Mhz, acoustic power = 27 W, focal depth = 20 mm, sonication time = 60 s. A 3 × 3 grid pattern with 5 mm spacing and a cooling time of 140 s. Between ablations was followed. The lesions appear in a plane perpendicular to the beamA T2-weighted fast relaxation fast spin-echo (T2W-FRFSE) coronal image with fat suppression was acquired after the grid pattern to observe the ablated area [Figure 9a]. The dimensions of the ablated area were 2.17 cm × 1.84 cm. Settings: Acoustic power was set at 27 W, focal depth was 20 mm, sonication time at 60 s. A 3 × 3-point grid with 140 s. Cooling time between successive ablations was followed. Figure 9b shows a T2W-FRFSE axial image with fat suppression was acquired after the grid pattern to observe the ablated area. Settings: Acoustic Power was set at 27 W, focal depth was 20 mm, sonication time at 60 s. A 3 × 3-point grid with 140 s. Cooling time between successive ablations was followed. The dimensions of the ablated area are 26.8 mm × 29.8 mm.
Figure 9
(a) Coronal T2-weighted fast relaxation fast spin-echo image with fat suppression that was obtained after the completion of the ablation plan. (b) Axial T2-weighted fast relaxation fast spin-echo image with fat suppression that was obtained after the completion of the ablation plan
(a) Coronal T2-weighted fast relaxation fast spin-echo image with fat suppression that was obtained after the completion of the ablation plan. (b) Axial T2-weighted fast relaxation fast spin-echo image with fat suppression that was obtained after the completion of the ablation planFigure 10a demonstrates the corresponding tissue damage induced by ablation under MR imaging, as described in Figure 8. Settings: Acoustic Power was 27 W, focal depth was 20 mm, sonication time was 60 s. with 140 s. The cooling time between ablations. The lesions appear in a plane perpendicular to the beam. The dimensions of the ablated area were 20 mm × 25 mm. Figure 10b demonstrates the same lesions after dissection. These lesions are shown in a plane parallel to the ultrasound beam. The lesion length was 24 mm.
Figure 10
(a) Macroscopic appearance of discrete lesions after dissection inflicted by ablation with the system in a plane perpendicular to the transducer. (b) Dissected tissue of the ablated area that was created. Ablation settings: f=2.578 MHz, acoustic power of 27 W focal depth 20 mm, sonication time was 60 s. A 3×3 grid pattern with 5 mm spacing and a cooling time of 140 s. between ablations was followed. The lesions appear in a plane parallel to the transducer. Red square indicated area of lesions; Red arrow indicates lesion
(a) Macroscopic appearance of discrete lesions after dissection inflicted by ablation with the system in a plane perpendicular to the transducer. (b) Dissected tissue of the ablated area that was created. Ablation settings: f=2.578 MHz, acoustic power of 27 W focal depth 20 mm, sonication time was 60 s. A 3×3 grid pattern with 5 mm spacing and a cooling time of 140 s. between ablations was followed. The lesions appear in a plane parallel to the transducer. Red square indicated area of lesions; Red arrow indicates lesion
DISCUSSION
The proposed device is capable of producing both discrete and overlapping lesions in the healthy muscle tissue of rabbits. The software and the positioning apparatus are moving the transducer with accuracy to the location selected by the user. The depth of the lesion corresponds with the focal depth of the transducer element. The accuracy of the device to ablate tissue is confirmed both by real-time MR imaging and postmortem dissection. A reliable MRgFUs system, compatible with multiple MR systems, will offer a solution for incisionless intervention for tumor ablation in veterinary hospitals. Its utilization can be of therapeutic or palliative intent. Tumors of superficial tissues such as the integumentary system, skeletal structures, mammary and testicular glands can be ablated with precision with minimal animal discomfort in the clinic. Such tumors can be ablated under ultrasound guidance from skilled surgeons without the need of MRI guidance. Tumors of the alimentary, reproductive, lymphatic, circulatory system require MRI guidance; however, general anesthesia for such a procedure does not exceed the time of surgery and does not carry the risks associated with surgery complications (i.e., hemorrhage, infection, and thrombosis).Contrary to human medicine, the treatment cost is a major factor that drives decision-making in Veterinary medicine. It is, therefore, important that a newly launched technology will be affordable to become available in companion animal hospitals. The proposed system has been designed with the intention to be affordable for most referral veterinary hospitals with an existing MRI system in place. The investment required is reduced not only due to its compatibility with multiple MRI systems but also because it carries a single-element transducer. More specifically, the system requires a laptop computer to carry the software, a bed modification to host the robotic system, while no upgrade is required for the MRI system. In a future study, we will assess the efficacy of the system to ablate naturally occurring tumors of companion animals. The disease recurrence and patient survival of treated patients need to be compared to those achieved with chemotherapy and radiation. In addition, the system can be tested not only as a stand-alone option but also in combination with other modalities (chemotherapy, targeted therapies, and immunotherapies). The technology is also expected to play a major role in drug delivery applications for cancer. Overall, MRgFUs is a promising technology that will be influencing cancer therapeutics for the years to come.
CONCLUSIONS
This study demonstrated the capability of the proposed system of thermally ablating in a controllable and reproducible manner. Tissue ablation remained within the targeted zone, sparing normal surrounding tissue in the absence of adverse effects. In accordance with the designed focal depth of the device, no major cutaneous tissue lesion was observed during the procedure. MRgFUs is designed to target tumors avoiding bone structures in the path of the beam. In addition, during the procedure, there was neither evidence of animal distress nor reaction, suggesting that under the proper anesthetic regime, the welfare of the animal is not compromised.The system used in this experiment has shown to be capable of producing ablations in healthy tissue in vivo. Under MR guidance, the system can ablate tissues in depth with precision and efficiency. MRI enabled near real-time temperature monitoring of the treatment.