Quality Control of Magnetic Resonance Elastography Using Percent Measurable Liver Volume Estimation.

David H Ballard1, Daniel R Ludwig1, Tyler J Fraum1, Amber Salter2, Vamsi R Narra1, Anup S Shetty1.   

Abstract

BACKGROUND: Although studies have described factors associated with failed magnetic resonance elastography (MRE), little is known about what factors influence usable elastography data.
PURPOSE: To identify factors that have a negative impact on percent measurable liver volume (pMLV), defined as the proportion of usable liver elastography data relative to the volume of imaged liver in patients undergoing MRE. STUDY TYPE: Retrospective.
SUBJECTS: A total of 264 patients (n = 132 males, n = 132 females; mean age = 57 years) with suspected or known chronic liver disease underwent MRE paired with a liver protocol MRI. FIELD STRENGTH/SEQUENCE: MRE was performed on a single 1.5 T scanner using a two-dimensional gradient-recalled echo phase-contrast sequence with a passive acoustic driver overlying the right hemiliver. ASSESSMENT: Stiffness maps (usable data at 95% confidence) and liver contours on magnitude images of the MRE acquisition were manually traced and used to assess mean stiffness and pMLV. Hepatic fat fraction and R2 * values were also calculated. The distance from the acoustic wave generator on the skin surface to the liver edge was measured. Two radiologists performed the MR analyses with 50 overlapping cases for inter-reader analysis. STATISTICAL TESTS: Linear regression was performed to identify factors significantly associated with pMLV. Intraclass correlation was performed for inter-reader reliability.
RESULTS: pMLV was 31% ± 20% (range 0%-86%). Complete MRE failure (i.e. pMLV = 0%) occurred in 10 patients (4%). Multivariate linear regression identified higher hepatic fat fraction, R2 *, BMI, and driver-to-liver surface distance; male sex; and lower mean liver stiffness was significantly independently associated with lower pMLV. Intraclass correlation for pMLV was 0.96, suggestive of excellent reliability. DATA
CONCLUSION: Higher fat fraction, R2 *, BMI, driver-to-liver surface distance, male sex, and lower mean liver stiffness were associated with lower pMLV. Optimization of image acquisition parameters and driver placement may improve MRE quality, and pMLV likely serves as a diagnostic utility quality control metric. LEVEL OF EVIDENCE: 3 TECHNICAL EFFICACY STAGE: 2.
© 2021 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  elastography; hepatic steatosis; liver fibrosis; magnetic resonance elastography; proton density fat fraction

Mesh:

Year:  2021        PMID: 34704644      PMCID: PMC9043030          DOI: 10.1002/jmri.27976

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   5.119


  28 in total

1.  Technical Failure of MR Elastography Examinations of the Liver: Experience from a Large Single-Center Study.

Authors:  Mathilde Wagner; Idoia Corcuera-Solano; Grace Lo; Steven Esses; Joseph Liao; Cecilia Besa; Nelson Chen; Ginu Abraham; Maggie Fung; James S Babb; Richard L Ehman; Bachir Taouli
Journal:  Radiology       Date:  2017-01-03       Impact factor: 11.105

2.  Assessment of advanced hepatic MR elastography methods for susceptibility artifact suppression in clinical patients.

Authors:  Jin Wang; Kevin J Glaser; Tianhui Zhang; Qungang Shan; Bingjun He; Jun Chen; Meng Yin; Bogdan Dzyubak; Jennifer L Kugel; Scott A Kruse; Roger C Grimm; Sudhakar K Venkatesh; Richard L Ehman
Journal:  J Magn Reson Imaging       Date:  2017-08-11       Impact factor: 4.813

3.  Can MR elastography be used to measure liver stiffness in patients with iron overload?

Authors:  Suraj D Serai; Andrew T Trout
Journal:  Abdom Radiol (NY)       Date:  2019-01

4.  Comparison of technical failure of MR elastography for measuring liver stiffness between gradient-recalled echo and spin-echo echo-planar imaging: A systematic review and meta-analysis.

Authors:  Dong Wook Kim; So Yeon Kim; Hee Mang Yoon; Kyung Won Kim; Jae Ho Byun
Journal:  J Magn Reson Imaging       Date:  2019-08-27       Impact factor: 4.813

5.  Epidemiology of Hepatic Steatosis at a Tertiary Care Center: An MRI-based Analysis.

Authors:  Tyler J Fraum; Daniel R Ludwig; Scott Kilian; William A Curtis; Thomas K Pilgram; Claude B Sirlin; Kathryn J Fowler
Journal:  Acad Radiol       Date:  2017-12-02       Impact factor: 3.173

Review 6.  Diagnostic performance of magnetic resonance elastography in staging liver fibrosis: a systematic review and meta-analysis of individual participant data.

Authors:  Siddharth Singh; Sudhakar K Venkatesh; Zhen Wang; Frank H Miller; Utaroh Motosugi; Russell N Low; Tarek Hassanein; Patrick Asbach; Edmund M Godfrey; Meng Yin; Jun Chen; Andrew P Keaveny; Mellena Bridges; Anneloes Bohte; Mohammad Hassan Murad; David J Lomas; Jayant A Talwalkar; Richard L Ehman
Journal:  Clin Gastroenterol Hepatol       Date:  2014-11-20       Impact factor: 11.382

7.  Diagnostic Performance of MR Elastography for Liver Fibrosis in Children and Young Adults with a Spectrum of Liver Diseases.

Authors:  Andrew T Trout; Rachel M Sheridan; Suraj D Serai; Stavra A Xanthakos; Weizhe Su; Bin Zhang; Daniel B Wallihan
Journal:  Radiology       Date:  2018-02-22       Impact factor: 11.105

8.  Liver MR Elastography at 3 T: Agreement Across Pulse Sequences and Effect of Liver R2* on Image Quality.

Authors:  Ely R Felker; Kang-Sun Choi; Kyung Sung; Holden H Wu; Steven S Raman; Bradley D Bolster; Stephan Kannengiesser; Kari Sorge; David S K Lu
Journal:  AJR Am J Roentgenol       Date:  2018-07-11       Impact factor: 3.959

9.  re: Comparison of Technical Failure of MR Elastography for Measuring Liver Stiffness Between Gradient-Recalled Echo and Spin-Echo Echo-Planar Imaging: A Systematic Review and Meta-Analysis.

Authors:  Richard L Ehman
Journal:  J Magn Reson Imaging       Date:  2019-10-06       Impact factor: 4.813

10.  Stereological Analysis of Liver Biopsy Histology Sections as a Reference Standard for Validating Non-Invasive Liver Fat Fraction Measurements by MRI.

Authors:  Tim G St Pierre; Michael J House; Sander J Bangma; Wenjie Pang; Andrew Bathgate; Eng K Gan; Oyekoya T Ayonrinde; Prithi S Bhathal; Andrew Clouston; John K Olynyk; Leon A Adams
Journal:  PLoS One       Date:  2016-08-08       Impact factor: 3.240

View more
  1 in total

1.  Free-breathing radial magnetic resonance elastography of the liver in children at 3 T: a pilot study.

Authors:  Sevgi Gokce Kafali; Tess Armstrong; Shu-Fu Shih; Grace J Kim; Joseph L Holtrop; Robert S Venick; Shahnaz Ghahremani; Bradley D Bolster; Claudia M Hillenbrand; Kara L Calkins; Holden H Wu
Journal:  Pediatr Radiol       Date:  2022-04-02
  1 in total