Hiroshi Hamasaki1, Takashi Shirasaka1, Yasuhiro Ushijima2, Hiroshi Akamine1, Yukihisa Takayama2, Yuichiro Kubo2, Keisuke Ishimatsu2, Akihiro Nishie3, Toyoyuki Kato1.
Abstract
OBJECTIVES: In CT during hepatic arteriography (CTHA), the addition of a noise power spectrum (NPS) model to conventional hybrid iterative reconstruction (HIR) may improve spatial resolution and reduce image noise. This study aims at assessing the image quality provided by HIR with a NPS model at CTHA.Entities:
Keywords: CT arteriography; hepatocellular carcinoma; hybrid iterative reconstruction; image quality; noise power spectrum model
Year: 2021 PMID: 34611577 PMCID: PMC8447979 DOI: 10.5334/jbsr.2444
Source DB: PubMed Journal: J Belg Soc Radiol ISSN: 2514-8281 Impact factor: 1.894
Quantitative image analysis.
|
| |||
|---|---|---|---|
| FBP† | AIDR‡ | EAIDR§ | |
|
| |||
| CT attenuation (HU) | |||
|
| |||
| Abdominal aorta | 69.0 ± 15.7 | 69.2 ± 15.5 | 69.5 ± 15.5 |
|
| |||
| Tumor | 305.0 ± 109.8 | 303.1 ± 109.2 | 305.3 ± 110.3 |
|
| |||
| Feeding artery | 493.8 ± 107.3 | 448.3 ± 104.5 | 514.3 ± 121.4** |
|
| |||
| Image noise (HU) | 28.5 ± 4.8 | 15.5 ± 1.8* | 15.5 ± 2.2* |
|
| |||
| SNR¶ | |||
|
| |||
| Tumor | 10.8 ± 4.9 | 18.2 ± 5.3* | 18.5 ± 5.5* |
|
| |||
| Feeding artery | 16.7 ± 4.2 | 27.4 ± 6.3* | 34.1 ± 7.9*, ** |
|
| |||
Data are presented as the mean±standard deviation. † FBP: filtered back projection, ‡ AIDR: adaptive iterative dose reduction 3D, § eAIDR: AIDR enhanced, ¶ SNR: the signal-to-noise ratio.
* indicate statistically significant difference as compared to FBP.
** indicate statistically significant difference as compared to AIDR.
Subjective image quality scores.
|
| ||||
|---|---|---|---|---|
| FBP† | EAIDR‡ | |||
|
|
| |||
| MEDIAN (MIN, MAX) | VS AIDR§
| MEDIAN (MIN, MAX) | VS AIDR§
| |
|
| ||||
| Tumor | 0 (–1, 1) | 0.45 | 0 (–1, 1) | 0.45 |
|
| ||||
| Feeding artery | 0 (0, 1) | 0.18 | 1 (0, 2) | 0.02* |
|
| ||||
| Overall image quality | –1 (–2, 1) | 0.01* | 1 (–1, 1) | 0.01* |
|
| ||||
Data are mean subjective image quality ratings regarding the visibility of the tumor, the visibility of its associated feeding arteries, and the overall image quality in comparison with baseline AIDR image rating at 0. †FBP: filtered back projection, ‡eAIDR: adaptive iterative dose reduction 3D enhanced, §AIDR: adaptive iterative dose reduction 3D.
* indicate statistically significant difference (p < 0.05).
Spatial frequency at 10% MTF† at different contrast levels.
|
| |||
|---|---|---|---|
| CONTRAST LEVEL | FBP‡ | AIDR§ | EAIDR¶ |
|
| |||
| High | 0.65 | 0.63 | 0.80 |
|
| |||
| Medium | 0.66 | 0.62 | 0.79 |
|
| |||
| Low | 0.64 | 0.63 | 0.75 |
|
| |||
| cycles/mm. | |||
Data are means. †MTF: modulation transfer function, ‡FBP: filtered back projection, §AIDR: adaptive iterative dose reduction 3D, ¶eAIDR: AIDR enhanced.