WHAT IS ALREADY KNOWN ON THIS TOPIC?: Hydrogen peroxide sterilizeation is widely used for luminal devices. However, the low penetrability of the sterilant is of major concern. WHAT IS ADDED BY THIS REPORT?: This report investigated the effective sterilization of low-temperature hydrogen peroxide gas plasma sterilizers and compared the applicability of different biological monitoring methods based on medical luminal devices. WHAT ARE THE IMPLICATIONS FOR PUBLIC HEALTH PRACTICE?: It is recommended to use a biological process challenge device for monitoring the sterilization of luminal devices with low-temperature hydrogen peroxide gas plasma sterilizers. Copyright and License information: Editorial Office of CCDCW, Chinese Center for Disease Control and Prevention 2021.Entities:
Keywords: low-temperature hydrogen peroxide gas plasma sterilizers; process challenge device; sterilization monitoring
Year: 2021 PMID: 34594949 PMCID: PMC8393050 DOI: 10.46234/ccdcw2021.152
Source DB: PubMed Journal: China CDC Wkly ISSN: 2096-7071
Biological monitoring of low-temperature hydrogen peroxide sterilizers in 58 sentinel hospitals across 22 PLADs from June 2015 to December 2019.
|
|
|
|
|
|
| Note: Biological PCD, the number of sterilizers using biological PCD in daily monitoring for luminal devices. Common BI, the number of sterilizers using common biological indicators in daily monitoring for luminal devices.
| ||||
| Secondary | CSSD | 3 | 11 | 14 |
| Operating room | 3 | 8 | 11 | |
| Tertiary | CSSD | 23 | 35 | 58 |
| Endoscopic Center | 1 | 2 | 3 | |
| Operating room | 8 | 20 | 28 | |
| Total | − | 38 | 76 | 114 |
Comparison of 338 times biological PCD monitoring and common biological monitoring of low-temperature hydrogen peroxide sterilizers
|
|
|
| ||
|
|
| |||
| Note: Biological PCD, the number of sterilizers using biological PCD in daily monitoring for luminal devices. Common BI, the number of sterilizers using common biological indicators in daily monitoring for luminal devices.
| ||||
| Qualified | 323 | 12 | 335 | |
| Unqualified | 0 | 3 | 3 | |
| Total | 323 | 15 | 338 | |