Literature DB >> 9721391

Comparative evaluation of the sporicidal activity of new low-temperature sterilization technologies: ethylene oxide, 2 plasma sterilization systems, and liquid peracetic acid.

W A Rutala1, M F Gergen, D J Weber.   

Abstract

OBJECTIVE: This study was undertaken to evaluate the efficacy of 4 new low-temperature sterilization technologies: ethylene oxide with hydrochlorofluorocarbons, a liquid peracetic acid immersion system (Steris System 1 Processor), and 2 plasma sterilization processes that use vaporized hydrogen peroxide (Sterrad 100 and the Sterrad 100S). The Sterrad 100S system potentially improves sterilizer efficacy by using 2 cycles of a diffusion stage and a plasma stage per sterilization cycle.
METHODS: Flat stainless steel carriers were inoculated with approximately 10(6) Bacillus stearothermophilus spores. These carriers were aseptically placed in the middle of 40 cm long stainless steel lumens (hollow tubes). Two types of lumen were used:(1) a lumen test unit with a removable 5 cm center piece (1.2 cm diameter) of stainless steel sealed to the narrower steel tubing by hard rubber septums and (2) a straight lumen. Three different diameters of the lumen test unit (1, 2, and 3 mm) and a single diameter of the straight lumen (3 mm) were studied. At least 40 replicates were performed for each type of lumen and sterilization method. After inoculation, the test unit was evaluated in 1 of the low-temperature sterilization technologies. After sterilization, the carriers were cultured in trypticase soy broth for 14 days at 55 degrees C and assessed for growth of B stearothermophilus spores.
RESULTS: Our results demonstrated that ethylene oxide with hydrochlorofluorocarbons, the Sterrad 100s, and the Sterrad 100S half cycle were highly effective in killing approximately 10(6) B stearothermophilus spores present in the center of narrow-lumen stainless steel tubes. As the lumen diameter decreased with the lumen test unit, the Sterrad 100 demonstrated reduced ability to kill B stearothermophilus spores present on the carrier. At the smallest diameter tested (1 mm), the Sterrad 100 system failed 74% of the time. The Steris System 1 was not effective in completely eliminating the 10(6) inoculum under test conditions.
CONCLUSION: The Sterrad 100S was significantly superior to the Sterrad 100 system and equivalent to ethylene oxide with hydrochlorofluorocarbons. Introduction of this new Sterrad 100S system should improve the margin of safety and reduce processing costs by its use of a shorter cycle time. The Steris System 1 is limited by diffusion of the chemical sterilant into the interior of the lumen test unit.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9721391     DOI: 10.1016/s0196-6553(98)70034-3

Source DB:  PubMed          Journal:  Am J Infect Control        ISSN: 0196-6553            Impact factor:   2.918


  14 in total

1.  Examination of Material Compatibilities with Ionized and Vaporized Hydrogen Peroxide Decontamination.

Authors:  Tohru Kimura; Hiroyuki Yahata; Yoshimichi Uchiyama
Journal:  J Am Assoc Lab Anim Sci       Date:  2020-09-17       Impact factor: 1.232

2.  Sterilization of Lung Matrices by Supercritical Carbon Dioxide.

Authors:  Jenna L Balestrini; Angela Liu; Ashley L Gard; Janet Huie; Kelly M S Blatt; Jonas Schwan; Liping Zhao; Tom J Broekelmann; Robert P Mecham; Elise C Wilcox; Laura E Niklason
Journal:  Tissue Eng Part C Methods       Date:  2016-01-29       Impact factor: 3.056

3.  Effects of Disinfection and Steam Sterilization on the Mechanical Properties of 3D SLA- and DLP-Printed Surgical Guides for Orthodontic Implant Placement.

Authors:  Silvia Izabella Pop; Mircea Dudescu; Sorin Gheorghe Mihali; Mariana Păcurar; Dana Cristina Bratu
Journal:  Polymers (Basel)       Date:  2022-05-21       Impact factor: 4.967

Review 4.  New disinfection and sterilization methods.

Authors:  W A Rutala; D J Weber
Journal:  Emerg Infect Dis       Date:  2001 Mar-Apr       Impact factor: 6.883

5.  The differential effects of heat-shocking on the viability of spores from Bacillus anthracis, Bacillus subtilis, and Clostridium sporogenes after treatment with peracetic acid- and glutaraldehyde-based disinfectants.

Authors:  Jordon K March; Michael D Pratt; Chinn-Woan Lowe; Marissa N Cohen; Benjamin A Satterfield; Bruce Schaalje; Kim L O'Neill; Richard A Robison
Journal:  Microbiologyopen       Date:  2015-07-17       Impact factor: 3.139

Review 6.  Reprocessing and reuse of urological armamentarium: How correct are we!

Authors:  Krutik Vipulbhai Raval; Rajeev Chaudhari; Shahil Rameshbhai Khant; Omkar Joglekar; Dipen Patel
Journal:  Urol Ann       Date:  2017 Apr-Jun

7.  Is plasma sterilization the modality of choice of sterilization today for endourological procedures such as ureterorenoscopy and retrograde intrarenal surgery? A single-center retrospective evaluation of 198 patients.

Authors:  Kandarp Priyakant Parikh; Ravi Jineshkumar Jain; Aditya K Parikh
Journal:  Urol Ann       Date:  2020-04-14

8.  Integrating Quality Tools and Methods to Analyze and Improve a Hospital Sterilization Process.

Authors:  Amira Kammoun; Wafik Hachicha; Awad M Aljuaid
Journal:  Healthcare (Basel)       Date:  2021-05-07

9.  Influence of reactive oxygen species on the enzyme stability and activity in the presence of ionic liquids.

Authors:  Pankaj Attri; Eun Ha Choi
Journal:  PLoS One       Date:  2013-09-16       Impact factor: 3.240

Review 10.  Review of Disinfection and Sterilization - Back to the Basics.

Authors:  Jin Hong Yoo
Journal:  Infect Chemother       Date:  2018-06
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.