Literature DB >> 15954488

A comparative study of ethylene oxide gas, hydrogen peroxide gas plasma, and low-temperature steam formaldehyde sterilization.

Keiji Kanemitsu1, Takayuki Imasaka, Shiho Ishikawa, Hiroyuki Kunishima, Hideo Harigae, Kumi Ueno, Hiromu Takemura, Yoshihiro Hirayama, Mitsuo Kaku.   

Abstract

OBJECTIVE: To compare the efficacies of ethylene oxide gas (EOG), hydrogen peroxide gas plasma (PLASMA), and low-temperature steam formaldehyde (LTSF) sterilization methods.
METHODS: The efficacies of EOG, PLASMA, and LTSF sterilization were tested using metal and plastic plates, common medical instruments, and three process challenge devices with narrow lumens. All items were contaminated with Bacillus stearothermophilus spores or used a standard biological indicator.
RESULTS: EOG and LTSF demonstrated effective killing of B. stearothermophilus spores, with or without serum, on plates, on instruments, and in process challenge devices. PLASMA failed to adequately sterilize materials on multiple trials in several experiments, including two of three plates, two of three instruments, and all process challenge devices.
CONCLUSIONS: Our results suggest that PLASMA sterilization may be unsuccessful under certain conditions, particularly when used for items with complex shapes and narrow lumens. Alternatively, LTSF sterilization demonstrates excellent efficacy and is comparable to EOG sterilization. LTSF could potentially act as a substitute if EOG becomes unavailable due to environmental concerns.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15954488     DOI: 10.1086/502572

Source DB:  PubMed          Journal:  Infect Control Hosp Epidemiol        ISSN: 0899-823X            Impact factor:   3.254


  4 in total

1.  Validation of a Clostridium endospore viability assay and analysis of Greenland ices and Atacama Desert soils.

Authors:  Wan-Wan Yang; Adrian Ponce
Journal:  Appl Environ Microbiol       Date:  2011-02-04       Impact factor: 4.792

2.  Applications of a rapid endospore viability assay for monitoring UV inactivation and characterizing arctic ice cores.

Authors:  Hannah S Shafaat; Adrian Ponce
Journal:  Appl Environ Microbiol       Date:  2006-10       Impact factor: 4.792

3.  Eradication of Intracanal Enterococcus Faecalis Biofilm by Passive Ultrasonic Irrigation and RinsEndo System.

Authors:  Ivana Toljan; Ivona Bago; Ivica Anić
Journal:  Acta Stomatol Croat       Date:  2016-03

4.  Monitoring the Effective Sterilization of Low-Temperature Hydrogen Peroxide Gas Plasma Sterilizers in 58 Hospitals - 22 PLADs, China, June 2015-December 2019.

Authors:  Jiaqi Wang; Baoying Zhang; Huihui Sun; Jian Zhang; Hongyang Duan; Haiqun Ban; Jin Shen; Liubo Zhang
Journal:  China CDC Wkly       Date:  2021-07-16
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.