DIRECT PEDICLE SCREW INSERTION PULLOUT STRENGTH.

Rômulo Pedroza Pinheiro1, Ariane Zamarioli1, Thibault Chandanson2, Keri George3, Antonio Carlos Shimano1, Helton Luiz Aparecido Defino1.   

Abstract

OBJECTIVE: Study the in vitro pullout strength of SpineGuard/Zavation Dynamic Surgical Guidance Z-Direct Screw (DSG Screw), a screw pedicle designed to be inserted using a direct insertion technique.
METHODS: DSG Screws of 5.5 mm and 6.5 mm were introduced into polyurethane blocks with a density of 10 PCF (0,16 g/cm3). According to the experimental group, screws were inserted without pilot hole, with pilot without tapping, undertapping and line-to-line tapping. Screw pullout tests were performed using a universal test machine after screw insertion into polyurethane blocks.
RESULTS: Screws inserted directly into the polyurethane blocks without pilot hole and tapping showed a statistically higher pullout strength. Insertion of the screw without tapping or with undertapping increases the pullout screw strength compared to line-to-line tapping.
CONCLUSION: DSG Screw showed the highest pullout strength after its insertion without pilot hole and tapping. Level of Evidence V, Expert Opinion.

Entities:  

Keywords:  In Vitro Techniques; Pedicle Screws; Spinal Fusion; Tensile Strength

Year:  2021        PMID: 34566479      PMCID: PMC8443012          DOI: 10.1590/1413-785220212904241648

Source DB:  PubMed          Journal:  Acta Ortop Bras        ISSN: 1413-7852            Impact factor:   0.513


INTRODUCTION

The pedicle of lumbar and thoracic spine has been extensively used as implant anchorage in the spinal surgery. The biomechanical advantages of pedicle screw-based system and the clinical usefulness is supported by the reports of high rate of fusion, deformity correction and clinical outcomes. The use of pedicle screw is related to two topics that still are a challenge in the field of spinal surgery: accuracy of pedicle screw and exposure of surgeon to radiation. ( To improve accuracy and reduce radiation exposure, a Dynamic Surgical Guidance - DSG, called PediGuard® probe was developed. This probe has the ability to identify different tissues by measuring electrical conductivity. ( )-( This device produces a sound, in which changes in pitch and cadence indicates a change in tissues around the tip of PediGuard® probe. A mid-range pitch and cadence audio signal is produced as the probe is in the cancellous bone. A low cadence pitch and cadence audio signal is performed as the probe approaches the pedicle cortical wall and it is the first indication of a potential pedicle breach. ( )-( The ability of Dynamic Surgical Guidance-DSG (PediGuard ®probe) to improve pedicle screw accuracy and to reduce radiation exposure has been shown in vitro using human cadaver specimens as well as in clinical trials. ( A further development of Dynamic Surgical Guidance technique was the combination of Dynamical Surgical Guidance technology and a pedicle screw in just one device to develop a “A Dynamic Surgical Guidance Screw” (DSG Screw). The DSG Screw is a pedicle screw system with a breach anticipation sensor located at the tip of the screw. The device provides a real-time surgical guidance and the ability to insert directly the screw into the pedicle without drilling a pilot hole neither tapping. ( The screw can be introduced directly into the pedicle and redirected during insertion according to the pitch and cadence of the audio signal. Besides, the DSG Screw insertion into the pedicle without drilling it do not require fluoroscopy for guidance, reducing intra-operative radiation and the operating time. This study experimentally evaluate the pullout strength of DSG screw using the direct screw insertion technique. We tested the hypothesis that smart screw has higher pullout strength after its insertion directly in the block without pilot hole and tapping.

MATERIALS AND METHODS

One hundred and five polyurethane blocks of 8 cm height, 5 cm width and 5 cm length, with a density of 10 PCF (0.16 g/cm3) (National Ltda.) were used as test bodies to introduce screws and to perform the mechanical pullout tests. SpineGuard/Zavation Dynamic Surgical Guidance Z-Direct Screw (DSG Screw of 5.5 mm and 6.5 mm outer diameter and 40 mm length) were inserted into the blocks according to the experimental group (Figure 1). The experimental groups were formed according to the use of pilot hole and tap diameter (undertapping and line to line). Thread taps 4.5 mm, 5.5 mm and 6.5 mm were used. The 4.5 mm tap was used as undertap for 5.5 mm screws. The 5.5 mm tap was used as undertap for 6.5 mm. All taps have a 2.9 mm pitch and a double lead design.
Figure 1

Photo of SpineGuard/Zavation Dynamic Surgical Guidance Z-Direct screw of and (A) 6.5 mm and (B) 5.5 mm outer diameter.

The screws were inserted into the blocks according to the experimental group. Each experimental group was formed by ten polyurethane blocks. For the 5.5 mm screws there were four experimental groups, and the screws were inserted: 1 - directly into the polyurethane block (without previous pilot hole and tapping), 2 - with a 2mm pilot hole without tapping, 3 - undertapping (2 mm pilot hole and 4.5mm tap), 4 - line to line (2 mm pilot hole and 5.5 tap). For the 6.5 mm screws there were five experimental groups: 1 - directly into the polyurethane block (without previous pilot hole and tapping), 2 - with a 2 mm pilot hole without tapping, 3 - undertapping (2 mm pilot hole and 4.5 mm tap), 4 - undertapping (2 mm pilot hole and 5.5 mm tap), 5 - line to line (2 mm pilot hole and 6.5 tap). After screw insertion, pullout strength was evaluated using universal test machine (EMIC-DL10000, São José dos Pinhais, PR, Brazil). A rod was attached to the head of the screw and pullout force was applied vertically. This force was applied at a speed of 2.0 mm/min until the screw was pulled out of the polyurethane block.

Statistical methods

Continuous variables were expressed as the means and standard deviations (SD). The results from pullout forces were subjected to statistical analysis of normality using the Kolmogorov-Smirnov test, in order to determine the behavior of the data. The results obtained in the four groups were compared using three-way analysis of variance, followed by Tukey’s post hoc test. Statistically significant differences were noted when p < 0.05. Statistical analyses were determined using Prism v8.4.3 Graphs were generated using Prism v8.4.3 (GraphPad, San Diego, CA).

RESULTS

Pilot hole and tapping have been previously reported to influence the screw pullout strength. To evaluate the influence of pilot hole preparation and techniques, screws without pilot hole, without tapping, with undertapping and line to line tapping were inserted. The results of the 5.5 mm and 6.5 mm screws pullout strength according to the experimental groups are illustrated in the Figures 2 and 3.
Figure 2

Mean maximal pullout strength of 5.5 mm SpineGuard /Zavation screw inserted into polyurethane blocks. The asterisks (*) indicate statistical difference.

Figure 3

Mean maximal pullout strength of 6.5 mm SpineGuard/Zavation screw inserted into polyurethane blocks. The asterisks (*) indicate statistical difference.

The mean pullout strength for 5.5 mm and 6.5 mm screws inserted directly into the blocks without pilot hole or tapping were statistically higher, when compared to the other experimental groups (p < 0.05). As of note, we detected a 21% increase in the pullout strength of the 5.5 mm DSG screw during its direct insertion, when compared to the insertion with the use of pilot hole only. This increase was even higher when the direct insertion was compared to the other experimental groups; 33% increase versus 4.5 mm tapping and 65% versus 5.5 mm tapping (Figure 2). With regards to the 6.5 mm DSG screw, we detected a 15% increase in the pullout strength during its direct insertion, when compared to the insertion with the use of pilot hole only; 27% increase versus 4.5 mm tapping; 40% versus 5.5 mm tapping; and 49% versus 6.5 mm tapping (Figure 3). An increase of screw pullout strength was observed from the experimental groups using line to line tapping to the experimental group, in which the screws were inserted without pilot hole.

DISCUSSION

Our in vitro findings support the hypothesis that SpineGuard/Zavation screw has higher pullout strength after its insertion and tapping directly into the block without pilot hole. Higher pullout strength of DSG screw was recorded after its direct insertion into the polyurethane blocks compared with insertion with pilot hole, undertapping or line-to-line tapping. Since the initial report on the use of pedicle screw for spine fixation, there has been a permanent improvement of this modality of spinal fixation, that is widely used to treat fractures, degenerative disease, tumor, deformities and spinal stability. ( )-( Pedicle screws continue to be studied to improve its locking mechanism in fixation system components, biomechanical screw performance and screw accuracy. ( The main biomechanical requirements of pedicle screw are resistance to cantilever loads (loads oriented perpendicular to the long axis of screw as bending strength) and pullout resistance. The bending strength depends on the material and it is proportional to screw core diameter. ( )-( The strength increases exponentially, it is proportional to the cube of screw core diameter, therefore the largest screw diameter allowed by the bony local anatomy should be used to minimized the likelihood of screw failure. ( )-( Pullout resistance of the pedicle screw is influenced by bone mineral density, ( screw geometry and insertion technique employed by the surgeon. Changes in screw design and optimization of pilot hole has been explored to improve the anchorage of the pedicle screws, as modifications of bone mineral density are not possible to be made acutely. ( In bone with compromised BMD, augmentation of the screw with polymethylmethacrylate, calcium phosphate or hydroxyapatite, or modification to the screw diameter, length, thread design, expandable screws, fenestrated screw or change in screw trajectory has been attempted to increase the purchase of the implants. ( The usual and current surgical technique for pedicle screw insertion is a sequence of pilot hole followed by tapping and screw insertion. ( ),( The screw insertion is generally preceded by a pre-tapping using a smaller diameter than screw. ( If the screw is inserted in untapped pilot hole, rates of misalignment may increase. ( The holding power or pullout strength is influenced by the amount of bone inside the screw threads. ( The perforation and tapping of the pilot hole result in additional trauma, increase operative time and decrease the screw pullout strength. ( To eliminate these drawbacks, changes of screw designs were performed to eliminate drilling of the pilot hole and tapping. ( Self-drilling and self-tapping screws were developed and used in spinal surgery for cervical plate fixation, ( whose screw path is not so critical as pedicle screw. Pedicle screw should be inserted inside the pedicle wall that is surrounded by neural and other anatomical structures that can be injured. ( ),( Pedicles screws are typically placed using pilot holes and the trajectory of the pilot holes can be verified by pedicle sounding. Although pilot holes and tapping do not ensure that the screw will follow the pilot hole trajectory, for tapped pilot holes the risk of screw malposition is lower. ( Pullout strength is proportional to the volume of bone inside the screw thread, ( bone implant-contact with increased area and density of bone inside and outside the screw thread is promoted by self-drilling screw as the DSG screw. However, screw pullout strength depends also on changes induced in bone by insertion trauma, reaction of bone to implant and resorption and remodeling as a result of healing. ( Silva et al. reported increased implant-bone contact in vivo after screws inserted in pilot holes smaller than the screw internal diameter. ( The highest pull-out strength of the direct screw insertion technique can be explained by a higher amount of bone squeezed at the bone-implant interface contact. Furthermore, in vivo studies may be required to confirm these findings. The limitation of the study related to the used experimental model should be considered. Pullout strength test may not be commonly seen in a clinical setting, but its simplicity and reproducibility allow it to be considered as the most efficient method to compare screw anchorage within the bone. ( ),( ),( Axial pullout test is easy to perform, reproducible and is accepted as a good predictor of the mechanical performance of the screw. Yet, pedicles screws are subjected to a complex mechanically demanding situation represented by an association of twisting, bending and pullout force ),( . Most of the time, pedicle screws fail by cyclic loading rather one-time pullout. Screw pullout strength does not represent the only mechanism of screw failure, but it still reflects the magnitude of screw anchorage purchase. (

Clinical applications

The DSG screw combines the characteristics that have been desired for pedicle screws, combining great resistance to pullout and improved positioning accuracy. However, only after the use of DSG screw in clinical settings and evaluation of the outcomes, the true benefits of DSG screw could be confirmed. Preliminarily, the results of the initial experimental evaluation showed advantages of the DSG screw. This component and its direct screw insertion technique provides better pullout strength. In addition, The DSG screw is a time saving approach compared to the traditional pedicle screw placement because screw insertion can be performed without pilot hole and tapping. The accuracy of the screw positioning is not compromised as all along the insertion as the DSG technology is providing guidance in real time to ensure a safe trajectory within the pedicle. And finally, the DSG screw, guided by the bipolar sensor on the tip of the screw, could also reduce intraoperative radiation. Preliminary reports of clinical use of this component was shown to be very successful (although not published yet).

CONCLUSION

The DSG screw and its direct screw insertion technique shows higher pullout strength in experimental in vitro study and it also has the advantage to improve accuracy of pedicle screw insertion with less radiation exposure. The DSG screw has the potential to change the way pedicle screw is inserted, for a faster and more accurate technique with less radiation. However, only after clinical use and evaluation of its cost benefit, its real advantage will be considered.
  16 in total

1.  The holding power of orthopedic screws in vivo.

Authors:  J Schatzker; R Sanderson; J P Murnaghan
Journal:  Clin Orthop Relat Res       Date:  1975-05       Impact factor: 4.176

2.  Electrical conductivity measurement: a new technique to detect iatrogenic initial pedicle perforation.

Authors:  Ciaran Bolger; Michael O Kelleher; Linda McEvoy; M Brayda-Bruno; A Kaelin; J-Y Lazennec; J-C Le Huec; C Logroscino; P Mata; P Moreta; G Saillant; R Zeller
Journal:  Eur Spine J       Date:  2007-06-30       Impact factor: 3.134

3.  The effect of repetitive pilot-hole use on the insertion torque and pullout strength of vertebral system screws.

Authors:  Helton L A Defino; Rodrigo César Rosa; Patrícia Silva; Antonio Carlos Shimano; José Batista Volpon; Francisco José Albuquerque de Paula; Philipp Schleicher; Klaus Schnake; Frank Kandziora
Journal:  Spine (Phila Pa 1976)       Date:  2009-04-20       Impact factor: 3.468

4.  Alignment of pedicle screws with pilot holes: can tapping improve screw trajectory in thoracic spines?

Authors:  Serkan Erkan; Brian Hsu; Chunhui Wu; Amir A Mehbod; John Perl; Ensor E Transfeldt
Journal:  Eur Spine J       Date:  2009-06-14       Impact factor: 3.134

5.  The biomechanics of screws.

Authors:  R L Uhl
Journal:  Orthop Rev       Date:  1989-12

6.  Effect of pilot hole on biomechanical and in vivo pedicle screw-bone interface.

Authors:  Patrícia Silva; Rodrigo César Rosa; Antonio Carlos Shimano; Helton L A Defino
Journal:  Eur Spine J       Date:  2013-05-08       Impact factor: 3.134

7.  Pullout strength for cannulated pedicle screws with bone cement augmentation in severely osteoporotic bone: influences of radial hole and pilot hole tapping.

Authors:  Lih-Huei Chen; Ching-Lung Tai; Po-Liang Lai; De-Mei Lee; Tsung-Tin Tsai; Tsai-Sheng Fu; Chi-Chien Niu; Wen-Jer Chen
Journal:  Clin Biomech (Bristol, Avon)       Date:  2009-05-29       Impact factor: 2.063

8.  Biomechanical effects of polyaxial pedicle screw fixation on the lumbosacral segments with an anterior interbody cage support.

Authors:  Shih-Hao Chen; Ruey Mo Lin; Hsiang-Ho Chen; Kai-Jow Tsai
Journal:  BMC Musculoskelet Disord       Date:  2007-03-10       Impact factor: 2.362

Review 9.  Designs and techniques that improve the pullout strength of pedicle screws in osteoporotic vertebrae: current status.

Authors:  Thomas M Shea; Jake Laun; Sabrina A Gonzalez-Blohm; James J Doulgeris; William E Lee; Kamran Aghayev; Frank D Vrionis
Journal:  Biomed Res Int       Date:  2014-03-03       Impact factor: 3.411

10.  BIOMECHANICAL EVALUATION OF THE INFLUENCE OF CERVICAL SCREWS TAPPING AND DESIGN.

Authors:  Patricia Silva; Rodrigo César Rosa; Antonio Carlos Shimano; Francisco José Albuquerque de Paula; José Batista Volpon; Helton Luiz Aparecido Defino
Journal:  Rev Bras Ortop       Date:  2015-12-08
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.