| Literature DB >> 34564604 |
Iris Fiby1, Marta Magdalena Sopel1, Herbert Michlmayr2, Gerhard Adam2, Franz Berthiller1.
Abstract
The Fusarium mycotoxin deoxynivalenol (DON) is a common contaminant of cereals and is often co-occurring with its modified forms DON-3-glucoside (D3G), 3-acetyl-DON (3ADON) or 15-acetyl-DON (15ADON). A stable-isotope dilution liquid chromatography-tandem mass spectrometry (LC-MS/MS) based method for their determination in cereals was developed and validated for maize. Therefore, 13C-labelled D3G was enzymatically produced using 13C-DON and [13C6Glc]-sucrose and used as an internal standard (IS) for D3G, while uniformly 13C labelled IS was used for the other mycotoxins. Baseline separation was achieved for the critical peak pair DON/D3G, while 3ADON/15ADON could not be fully baseline separated after testing various reversed phase, fluorinated phase and chiral LC columns. After grinding, weighing and extracting the cereal samples, the raw extract was centrifuged and a mixture of the four 13C-labelled ISs was added directly in a microinsert vial. The subsequent analytical run took 7 min, followed by negative electrospray ionization and selected reaction monitoring on a triple quadrupole MS. Maize was used as a complex cereal model matrix for validation. The use of the IS corrected the occurring matrix effects efficiently from 76 to 98% for D3G, from 86 to 103% for DON, from 68 to 100% for 15ADON and from 63 to 96% for 3ADON.Entities:
Keywords: masked mycotoxins; mass spectrometry; modified mycotoxins; mycotoxins; stable-isotope dilution assay; trichothecenes
Mesh:
Substances:
Year: 2021 PMID: 34564604 PMCID: PMC8470870 DOI: 10.3390/toxins13090600
Source DB: PubMed Journal: Toxins (Basel) ISSN: 2072-6651 Impact factor: 4.546
Figure 1Chemical structures of (a) deoxynivalenol (DON) and its modified forms (b) deoxynivalenol-3-glucoside (D3G), (c) 3-acetyldeoxynivalenol (3ADON) and (d) 15-acetyldeoxynivalenol (15ADON).
Figure 2Enhanced Product Ion (MS/MS) spectra of D3G (left) and 13C-D3G (right) at a collision energy of 30 eV. The spectra derived from the acetate adducts are shown on top, the spectra using the deprotonated precursors on the bottom.
List of analytes with optimized ESI-MS/MS parameters.
| Analyte ID | Q1 Mass (Da) | Q3 Mass (Da) | DP (V) | CE (eV) | CXP (V) |
|---|---|---|---|---|---|
| D3G 1 | 517.1 | 457.1 | −80 | −22 | −7 |
| D3G IS 1 | 538.1 | 478.1 | −80 | −22 | −7 |
| D3G 2 | 517.1 | 59.0 | −80 | −74 | −9 |
| D3G IS 2 | 538.1 | 59.0 | −80 | −74 | −9 |
| D3G 3 | 517.1 | 427.0 | −80 | −32 | −5 |
| D3G IS 3 | 538.1 | 447.0 | −80 | −32 | −5 |
| DON 1 | 355.0 | 59.0 | −70 | −36 | −9 |
| DON IS 1 | 370.0 | 59.0 | −70 | −36 | −9 |
| DON 2 | 355.0 | 295.0 | −70 | −16 | −13 |
| DON IS 2 | 370.0 | 310.0 | −70 | −16 | −13 |
| DON 3 | 355.0 | 265.0 | −70 | −24 | −13 |
| DON IS 3 | 370.0 | 279.0 | −70 | −24 | −13 |
| ADONs 1 | 397.1 | 59.0 | −70 | −34 | −9 |
| ADON IS 1 | 414.1 | 49.0 | −70 | −34 | −9 |
| ADONs 2 | 397.1 | 337.1 | −70 | −12 | −7 |
| ADON IS 2 | 414.1 | 354.1 | −70 | −12 | −7 |
| 3ADON | 397.1 | 307.0 | −70 | −22 | −5 |
| 3ADON IS | 414.1 | 323.0 | −70 | −22 | −5 |
D3G deoxynivalenol-3-glucoside, DON deoxynivalenol, ADONs acetyl-deoxynivalenols, 3ADON 3-acetyl-deoxynivalenol, IS internal standard, DP declustering potential, CE collision energy, CXP cell exit potential.
Evaluated analytical columns for the LC-MS/MS method.
| Column | tR (DON) (min) | tR (D3G) (min) | tR (3ADON) (min) | tR (15ADON) (min) | FWHM (min) | RS (DON/D3G) | RS (3/15ADON) |
|---|---|---|---|---|---|---|---|
| Agilent Zorbax Eclipse Plus | 2.63 | 2.68 | 4.16 | 4.17 | 0.045 | −1.11 | −0.22 |
| Agilent Zorbax Extend-C18 | 2.15 | 2.26 | 3.85 | 3.86 | 0.077 | −1.43 | −0.13 |
|
|
|
|
|
|
|
|
|
| Agilent Zorbax SB C18 | 5.87 | 5.87 | 9.00 | 9.02 | 0.085 | 0.00 | −0.24 |
|
|
|
|
|
|
|
|
|
| Daicel Chiralcel | 3.59 | 3.45 | 11.00 | 9.00 | 3.000 | 0.05 | 0.67 |
| Daicel Chiralpak | 2.29 | 2.06 | 9.29 | 9.08 | 0.143 | 1.60 | 1.47 |
| Phenomenex Kinetex C18 | 4.07 | 4.03 | 5.81 | 5.82 | 0.075 | 0.53 | −0.13 |
| Phenomenex Kinetex F5 | 4.26 | 4.15 | 6.33 | 6.23 | 0.067 | 1.65 | 1.50 |
| Sigma Discovery HS F5 | 6.14 | 5.95 | 9.88 | 9.68 | 0.122 | 1.56 | 1.64 |
| Thermo Hypersil Gold | 1.84 | 2.13 | 3.68 | 3.69 | 0.112 | −2.60 | −0.09 |
| Waters Acquity BEH C18 | 2.98 | 3.21 | 4.37 | 4.40 | 0.040 | −5.75 | −0.75 |
|
|
|
|
|
|
|
|
|
tR retention time, FWHM full width half maximum, Rs resolution. A water–acetonitrile gradient (5% → 60% acetonitrile in 9 min) containing 2 mM ammonium acetate was used for separation. The flow rate was 0.2 or 0.4 mL/min, depending on the column length.
Figure 3LC-MS/MS selected ion monitoring (SRM) chromatogram of the optimized method. Separation was achieved on a Waters Acquity UPLC HSS T3 C18, 1.8 µm, 2.1 × 100 mm column using a water–acetonitrile gradient.
Method performance parameters.
| Analyte | LOQ Solution (μg/L) | LOQ Maize (μg/kg) | RE (%) | SSE (%) | RA (%) | RSDr (%) |
|---|---|---|---|---|---|---|
| D3G (ext.) | <10 | <40 | 94.7 | 75.9 | 92.3 | 3.1 |
| D3G (int.) | 97.5 | 8.3 | ||||
| DON (ext.) | <10 | <40 | 101 | 86.4 | 104 | 5.9 |
| DON (int.) | 103 | 5.3 | ||||
| 15ADON (ext.) | <30 | <120 | 105 | 67.8 | 105 | 9.4 |
| 15ADON (int.) | 100 | 7.8 | ||||
| 3ADON (ext.) | <10 | <40 | 94.4 | 63.2 | 90.8 | 7.0 |
| 3ADON (int.) | 96.2 | 5.7 |
LOQ: limit of quantification; RE: extraction recovery; SSE: signal suppression or enhancement; RA: apparent recovery; RSDr: relative standard deviation under conditions of repeatability; ext. external calibration; int. internal calibration.
Evaluated analytical columns for the LC-MS/MS method.
| Supplier | Brand Name | Dimensions (mm) | Particle Size (µm) |
|---|---|---|---|
| Agilent | ZORBAX RRHD Eclipse Plus C18 | 2.1 × 50 | 1.8 |
| Agilent | ZORBAX RRHT Extend-C18 | 2.1 × 50 | 1.8 |
| Thermo | Hypersil GOLD C18 | 2.1 × 50 | 1.9 |
| Waters | ACQUITY UPLC BEH C18 | 2.1 × 50 | 1.7 |
| Agilent | ZORBAX RRHD StableBond C18 | 2.1 × 100 | 1.8 |
| Agilent | ZORBAX RRHD Eclipse XDB-C18 | 2.1 × 100 | 1.8 |
| Phenomenex | Kinetex C18 | 2.1 × 100 | 2.6 |
| Waters | ACQUITY UPLC HSS T3 (C18) | 2.1 × 100 | 1.8 |
| Agilent | InfinityLab Poroshell 120 EC-C18 | 2.1 × 150 | 2.7 |
| Phenomenex | Kinetex F5 | 2.1 × 100 | 2.6 |
| Sigma-Aldrich | Discovery HS F5 | 2.1 × 100 | 5.0 |
| Daicel | CHIRALPAK AD-3R | 2.1 × 150 | 3.0 |
| Daicel | CHIRALCEL OJ-3R | 2.1 × 150 | 3.0 |