| Literature DB >> 29273904 |
Alexandra Malachová1, Milena Stránská2, Marta Václavíková1, Christopher T Elliott3, Connor Black3, Julie Meneely3, Jana Hajšlová2, Chibundu N Ezekiel4, Rainer Schuhmacher1, Rudolf Krska5.
Abstract
Liquid chromatography (LC) coupled with mass spectrometry (MS) is widely used for the determination of mycotoxins in cereals and cereal-based products. In addition to the regulated mycotoxins, for which official control is required, LC-MS is often used for the screening of a large range of mycotoxins and/or for the identification and characterization of novel metabolites. This review provides insight into the LC-MS methods used for the determination of co-occurring mycotoxins with special emphasis on multiple-analyte applications. The first part of the review is focused on targeted LC-MS approaches using cleanup methods such as solid-phase extraction and immunoaffinity chromatography, as well as on methods based on minimum cleanup (quick, easy, cheap, effective, rugged, and safe; QuEChERS) and dilute and shoot. The second part of the review deals with the untargeted determination of mycotoxins by LC coupled with high-resolution MS, which includes also metabolomics techniques to study the fate of mycotoxins in plants.Entities:
Keywords: Fungal secondary metabolites; Liquid chromatography–high-resolution mass spectrometry; Liquid chromatography–tandem mass spectrometry; Metabolomics; Validation
Mesh:
Substances:
Year: 2017 PMID: 29273904 PMCID: PMC5775372 DOI: 10.1007/s00216-017-0750-7
Source DB: PubMed Journal: Anal Bioanal Chem ISSN: 1618-2642 Impact factor: 4.142
Fig. 1Chemical structures of the most important mycotoxins belonging to the group of Aspergillus toxins (aflatoxin B1, alfatoxin G1, ochratoxin A, patulin), Fusarium toxins (deoxynivalenol, T-2 toxin, zearalenone, fumonisin B1, beauvericin, enniatin A), Alternaria toxins (alternariol), ergot alkaloids (ergotamine), and Penicillium toxins (patulin, ochratoxin A)
Overview of performance characteristics of an analytical method defined in Commission Decision 2002/657/EC
| Performance characteristic | Definition |
|---|---|
| Accuracy | The closeness of agreement between a test result and the accepted reference value. It is determined by determining trueness and precision |
| Detection capability (CCβ) | The smallest content of the substance that may be detected, identified, and/or quantified in a sample with an error probability of |
| Decision limit (CCα) | The limit at and above which it can be concluded with an error probability of |
| Precision | The closeness of agreement between independent test results obtained under stipulated (predetermined) conditions. The measure of precision is usually expressed in terms of imprecision and computed as the standard deviation of the test results. Less precision is determined by larger standard deviation |
| Recovery | The percentage of the true concentration of a substance recovered during the analytical procedure. It is determined during validation if no certified reference material is available |
| Repeatability | The precision under repeatability conditions. Repeatability conditions means conditions where independent test results are obtained with the same method on identical test items in the same laboratory by the same operator using the same equipment |
| Reproducibility | The precision under reproducibility conditions. Reproducibility conditions means conditions where test results are obtained with the same method on identical test items in different laboratories with different operators using different equipment. Participation in ring trials is needed |
| Ruggedness | The susceptibility of an analytical method to changes in experimental conditions that can be expressed as a list of the sample materials, analytes, storage conditions, and environmental and/or sample preparation conditions under which the method can be applied as presented or with specified minor modifications. For all experimental conditions that could in practice be subject to fluctuation, any variations that could affect the analytical result should be indicated. |
| Specificity | The ability of a method to distinguish between the analyte being measured and other substances. This characteristic is predominantly a function of the measuring technique described, but can differ according to the class of the compound and the matrix |
| Trueness | The closeness of agreement between the average value obtained from a large series of test results and an accepted reference value. Trueness is usually expressed as bias |
Fig. 2An example of the z score compilation obtained by the multimycotoxin liquid chromatography–tandem mass spectrometry method in proficiency testing organized by the Bureau Interprofessionnel des Études Analytique (BIPEA). Green lines borders of acceptable range of z scores, red lines borders of questionable range of z scores, area outside red lines unacceptable values
Fig. 3An overview of use of liquid chromatography (LC)–mass spectrometry (MS) instruments in studies focused on mycotoxin analysis published between 2012 and 2016. LC–MS/MS covers studies using instruments equipped with triple quadrupole, quadrupole–linear ion trap, and ion trap mass analyzers; LC–high resolution MS (HRMS)/MS covers studies using instruments equipped with quadrupole–time of flight or quadrupole–orbital ion trap mass analyzers
Detailed description of the setup of some liquid chromatography (LC)–mass spectrometry (MS) methods for mycotoxin determination
| Reference | Extraction | Cleanup | Analytes | Matrix | LC–MS instrument | LC conditions | MS conditions |
|---|---|---|---|---|---|---|---|
| [ | CH3CN–H2O (84:16, v/v) | MycoSep 226 AflaZON+, MycoSep 227 (both Romer Labs) | NIV, DON, FUS-X, 3ADON, 15ADON, DAS, HT2, T2, ZEN | Maize | QTRAP MS/MS instrument (Sciex) coupled to 1100 series LC system (Agilent Technologies) | Aquasil RP-18 column (100 mm × 4.6 mm, 3 μm) + C18 guard column; 25 °C, flow rate 1000 μL/min, injection volume 25 μL; eluent A H2O–CH3OH (80:20, v/v), eluent B H2O–CH3OH (10:90, v/v), both containing 5 mM NH4CH3COO-; gradient 0.5 min 0% eluent B, linear gradient to 100% eluent B to 4.5 min, 100% eluent B to 7 min, 7.1 min 0% eluent B, reequilibration 3 min, total run 10 min | APCI± MRM, monitoring of 2 transitions (1 quantifier and 1 qualifier), dwell time 100 ms, polarity switching (2 periods) |
| [ | CH3CN–H2O (84:16, v/v) | MycoSep 226 AflaZON+ | AFs, NIV, DON, 3ADON, 15ADON, FUS-X, HT2, T2, ZEN, OTA, STER, CIT, verruculogen | Various foods and feed | Quattro Ultima QqQ instrument (Micromass) coupled to Acquity UHPLC system (Waters) | UPLC BEH C18 (100 mm × 2.1 mm, 1.7 μm); 35 °C, flow rate 300 μL/min, injection volume 5 μL; eluent A ESI+ 10 mM NH4CH3COO-, ESI− 0.1% 0.1% (v/v) aqueous NH3, eluent B CH3OH; gradient initially 20% eluent B, linear increase to 5.5 to 85% eluent B, 100% eluent B within 0.3 min, reequilibration for 2 min at 20% eluent B, total run 10 min | ESI+, ESI−, MRM, 2 chromatographic runs, monitoring of 2 transitions (1 quantifier and 1 qualifier) |
| [ | 2-step extraction: (1) PBS; (2) 70% CH3OH | AOZFDT2 (VICAM) | AFs, OTA, FBs, DON, ZEA, T2, HT2 | Maize | QTRAP MS/MS (Sciex) instrument coupled to 1100 micro LC system (Agilent Technologies) | Gemini C18 column (150 mm × 2 mm, 5 μm) + Gemini C18 guard column (4 mm × 2 mm, 5 μm); 40 °C, flow rate 200 μL/min, injection volume 20 μL; eluent A H2O, eluent B CH3OH, both containing 0.5% CH3COOH and 1 mM NH4CH3COO-; gradient 3 min at 20% eluent B, jump to 40% eluent B, linear increase to 63% eluent B within 35 min, 63% eluent B for 11 min, reequilibration at 20% eluent B for 10 min, total run 59 min | ESI+, ESI−, dMRM, monitoring of 2 transitions (1 quantifier and 1 qualifier), time window of 1 MRM 0.8 min, cycle time 0.55 s |
| [ | CH3CN–H2O–CH3COOH (79.5:20:0.5, v/v/v). Evaporation and redissolution in PBS before IAC | Myco6in1+ (VICAM) | AFs, OTA, FBs, DON, ZEN, T2, HT2 | Barley, maize breakfast cereals, peanuts | QTRAP 4500 instrument (Sciex) coupled to a Prominence UFLC XR chromatography system (Shimadzu) | Acquity UPLC HSS T3 end-capped C18 column (100 mm × 2.1 mm, 1.7 μm), 40 °C, flow rate 400 μL/min, injection volume 10 uL; eluents A H2O, eluent B CH3OH, both containing 5 mM NH4CH3COO-; gradient 5% eluent B increased to 50% eluent B in 1 min, linear increase to 100% eluent B within 6 min, 100% eluent B to 8 min, at 8.1 min initial conditions 5% eluent B, reequilibration at 5% eluent B for 2 min, total run 10 min | ESI± MRM, 2 periods, monitoring of 2 transitions (1 quantifier and 1 qualifier), dwell time 50 ms, polarity switching (2 periods) |
| [ | 2-step extraction: (1) H2O; (2) CH3OH. Evaporation and redissolution in PBS before IAC | Myco6in1+ (VICAM) | AFs, OTA, FBs, DON, ZEN, T2, HT2, NIV | Maize, durum wheat, corn flakes, maize crackers | QTRAP MS/MS instrument (Sciex) coupled to 1100 micro LC system (Agilent technologies) | Gemini C18 column (150 mm × 2 mm, 5 μm) + Gemini C18 guard column (4 mm × 2 mm, 5 μm); 40 °C, flow rate 200 μL/min, injection volume 20 μL; eluent A H2O, eluent B CH3OH, both containing 0.5% CH3COOH and 1 mM NH4CH3COO-; gradient 3 min at 20% eluent B, jump to 40% eluent B, linear increase to 63% eluent B within 35 min, 63% eluent B for 11 min, reequilibration at 20% eluent B for 10 min, total run 59 min | ESI± MRM, monitoring of 2 transitions (1 quantifier and 1 qualifier), dwell time 100 ms, polarity switching (2 periods) |
| [ | NaCl + H2O–CH3OH (30:70, v/v). Dilution with PBS before IAC | OCHRAPREP + DZT MS-PREP, AOF MS-PREP + DZT MS-PREP, AFLAOCHRA PREP + DZT MS-PREP (R-Biopharm) | OTA + DON, ZEN, T2, HT2; AFs, FBs, OTA + DON, ZEN, T2, HT2; AFs, OTA + DON, ZEN, T2, HT2 | Wholemeal bread, maize and maize-based products including infant foods, oat-based muesli | Acquity TQD tandem QqQ MS instrument (Waters) | Gemini C18 column (150 mm × 2 mm, 5 μm), 40 °C, flow rate 300 μL/min, injection volume 20 μL; eluent A H2O–CH3OH (95:5, v/v), eluent B H2O–CH3OH (98:3, v/v), both containing 0.5% HCOOH and 1 mM NH4HCOO-; gradient 20% eluent B for 0.1 min, to 10 min linear increase to 90% eluent B, 90% eluent B to 15 min, reequilibration at 20% eluent B, total run 20 min | ESI+ MRM, monitoring of 2 transitions (1 quantifier and 1 qualifier), 6 acquisition periods, dwell times from 0.1 to 0.27 s |
| [ | QuEChERS | AFs, FBs, NIV, DON, 3ADON, 15ADON, FUS-X, HT2, T2, ZEN, OTA, DAS, NEO | Rice, corn, wheat, rye, oat, barley, infant cereals, soya, corn gluten | QTrap 4000 instrument (Sciex) coupled to 1100 series LC system (Agilent Technologies) | Zorbax Bonus-RP column (150 mm × 2.1 mm, 3.5 μm) + Zorbax RB C8 guard column (12.5 mm x 2.1 mm, 3.5 μm), flow rate 250 μL/min, injection volume 40 μL; eluent A 0.15% (v/v) HCOOH + 10 mM NH4HCOO-, eluent B 0.05% HCOOH (v/v) in CH3OH; gradient: 0% eluent B at 1 min, linear increase to 100% eluent B until 15 min, 100% eluent B for 5 min, reequilibration at 0% eluent B for 5 min, total run 25 min | ESI± MRM, monitoring of 2 transitions (1 quantifier and 1 qualifier), 3 acquisition periods | |
| [ | QuEChERS | AFs, FBs, DON, HT2, T2, ZEN, OTA | Wheat, maize, rice | Micromass Quattro Micro QqQ coupled to Alliance 2695 system (Waters) | Atlantis RP C18 column (150 mm × 2.1 mm, 5 μm), 30 0°C, flow rate 300 μL/min, injection volume 20 μL; eluent A H2O–CH3OH (90:10, v/v), eluent B H2O–CH3OH (10:90, v/v), both containing 5 mM NH4CH3COO-; gradient 20% eluent B for 0.1 min, until 10 min linear increase to 90% eluent B, 90% eluent B to 15 min, reequilibration at 20% eluent B, total run 20 min | ESI± MRM, monitoring of 2 transitions (1 quantifier and 1 qualifier), 3 acquisition periods | |
| [ | QuEChERS, dilute and shoot | 38 mycotoxins and 288 pesticides | Apple baby food, wheat flour, paprika, black pepper, sunflower seed | QTRAP 5500 instrument (Sciex) coupled to Acquity UHPLC system (Waters) | Acquity UPLC HSS T3 end-capped C18 column (100 mm × 2.1 mm, 1.8 μm), 40 °C, flow rate 350–700 μL/min, injection volume 3 μL; ESI+ eluent A: H2O, eluent B: CH3OH, both containing 0.2% HCOOH + 5 mM NH4HCOO-; ESI− eluent A H2O, eluent B CH3OH, both containing 0.2% HCOOH + 5 mM NH4HCOO-; gradient: 10% eluent B with flow rate 350 μL/min increased to 50% eluent B in 1 min, linear increase to 100% eluent B within 10 min and simultaneous increase of flow rate to 550 μL/min, flow rate 0.7 μL/min at 100% eluent B, reequilibration for 2.5 min at 10% eluent B at 450 μL/min, total run 15.5 min | ESI+, ESI−, dMRM, time window for 1 MRM 0.8 min, cycle time 0.55 s | |
| [ | Dilute and shoot | No | 39 mycotoxins | Wheat, maize | QTRAP 4000 instrument (Sciex) coupled to 1100 series LC system (Agilent Technologies) | Gemini C18 column (150 mm × 2 mm, 5 μm) + Gemini C18 guard column (4 mm × 2 mm, 5 μm); 40 °C, flow rate 1000 μL/min, injection volume 5 μL; eluent A CH3OH–H2O–CH3COOH (10:89:1, v/v/v), eluent B CH3OH–H2O–CH3COOH (97:2:1, v/v/v), both containing 5 mM NH4CH3COO-; gradient 2 min at 100% eluent A, linear increase to 100% eluent B within 12 min, held at 100% eluent B for 3 min, reequilibration at 100% eluent A for 4 min, total run 19 min | ESI+, ESI−, dMRM, dwell time 100 ms, pause time 5 ms |
| [ | Dilute and shoot | No | 295 analytes | Apple puree, hazelnut, maize, green pepper | QTRAP 5500 instrument (Sciex) coupled to 1290 series LC system (Agilent Technologies) | Gemini C18 column (150 mm × 2 mm, 5 μm) + Gemini C18 guard column (4 mm × 2 mm, 5 μm); 40 °C, flow rate 1000 μL/min, injection volume 5 μL; eluent A CH3OH–H2O–CH3COOH (10:89:1, v/v/v), eluent B CH3OH–H2O–CH3COOH (97:2:1, v/v/v), both containing 5 mM NH4CH3COO-; gradient: 2 min at 100% eluent A, linear increase to 50% eluent B within 3 min, linear increase zo 100% eluent B within 9 min, hold at 100% eluent B for 4 min, reequilibration at 100% eluent A for 2.5 min, total run 20.5 min | ESI+, ESI−, dMRM, MRM window ±27 s for positive mode, ±42 s for negative mode, scan time 1 s |
| [ | Raw extract, SIDA | No | AFs, FBs, DON, HT2, T2, OTA, ZEN | Maize, cereal-based products | 6490 triple-quadrupole instrument coupled to 1290 series UHPLC system (both Agilent Technologies) | Zorbax RRHL Eclipse Plus C18 column (100 mm × 2.1 mm, 1.8 μm); 30 °C, flow rate 350 μL/min, injection volume 3 μL; eluent A H2O–HCOOH (99.9:0.1, v/v), eluent B CH3OH–HCOOH (99.9:0.1, v/v) both containing 5 mM NH4HCOO-; gradient: 0.5 min at 30% eluent B, linear increase to 100% eluent B in 7.5 min, hold at 100% eluent B for 1.5 min, at 9.6 min back to 30% eluent B, reequilibration at 30% eluent B for 2 min, total run 11.5 min | ESI±, dMRM, monitoring of 2 transitions (1 quantifier and 1 qualifier) |
| [ | CH3CN–H2O (84:16, v/v), SIDA | Bond Elut Mycotoxin SPE cartridges (Agilent Technologies) | NIV, DON, FUS-X, DON-3-Glc, 3ADON, 15ADON, HT2, T2, ENNs, BEA, ZEN | Barley, malt, oat, wheat, maize | QTRAP 4000 instrument (Sciex) coupled to LC-20A Prominence system series LC system (Shimadzu) | Hydrosphere RP-C18 column (100 mm × 3 mm, 3 μm) + C18 guard column; 40 °C, flow rate 200 μL/min, injection volume 10 μL; eluent A H2O–HCOOH (99.9:0.1, v/v), eluent B CH3OH–HCOOH (99.9:0.1, v/v); gradient ESI− 2 min at 10% eluent B, linear increase to 99% eluent B in 6 min, hold at 99% eluent B for 7.5 min, for 2 min back to 10% eluent B, reequilibration at 10% eluent B for 9.5 min, total run 25 min; ESI+ 2 min at 10% eluent B, linear increase to 87% eluent B in 6 min, hold at 87% eluent B for 7 min, increase to 100% eluent B in 5 min, hold at 100% eluent B for 3.5 min, for 2 min back to 10% eluent B, reequilibration at 10% eluent B for 9.5 min, total run 34.5 min | ESI−, ESI+, dMRM, 2 single chromatographic runs, monitoring of 2 transitions (1 quantifier and 1 qualifier) |
| [ | CH3CN–H2O (84:16, v/v), evaporation, reconstitution in CH3OH and H2O | SPE (Oasis HLB columns) | AFs, OTA, DON, ZEN, T2, HT2 | Wheat flour, barley flour, crisp bread | Accela HPLC system, Exactive HRMS instrument (Thermo Fisher Scientific); 1100 micro-LC system (Agilent Technologies), QTRAP instrument (Applied Biosystems) | Kinetex C18 column (100 mm × 2.1 mm, 2.6 μm); 40 °C, flow rate 200 μL/min, injection volume 20 μL; eluent A H2O, eluent B CH3OH, both containing 0.5% CH3COOH and 1 mM NH4CH3COO-; gradient 10% eluent B start, until 4 min linear increase to 40% eluent B, 60% eluent B in 27 min, keep for 5 min, reequilibration at 10% eluent B for 7 min, total run 20 min | HESI-II (heated-electrospray, ESI+, HCD fragmentation (in-source fragmentation) |
| [ | QuEChERS (2 g sample, 10 mL 0.1% HCOOH in H2O, 3 min shaking, 10 mL CH3CN, 3 min shaking, 4 g MgSO4, 1 g NaCl, shaking) | No additional cleanup | 3ADON, 15ADON, DON, DON-3-Glc, FUS-X, NIV, HT2, T2, DAS, NEO, AFs, OTA, FBs, STER, ZEN, penitrem A, BEA, | barley | Accela HPLC system, Exactive HRMS instrument (Thermo Fisher Scientific) | Acquity UPLC HSS T3 column (100 mm × 2.1 mm, 1.8 μm); 40 °C, flow rate 300 μL/min, injection volume 5 μL; eluent A H2O with 5 mM NH4HCOO- and 0.1% HCOOH, eluent B CH3OH; gradient: start with 5% eluent B, increase to 50% eluent B in 6 min, increase to 95% eluent B within 4 min, keep until 15 min of the run, reequilibration at 5% eluent B for 3 min | HESI-II, ESI+/ESI− |
| [ | QuEChERS (2 g sample, 10 mL 0.1% HCOOH in H2O, 3 min shaking, 10 mL CH3CN, 3 min shaking, 4 g MgSO4, 1 g NaCl, 0.5 trisodium citrate dihydrate, shaking) | No additional cleanup | 3ADON,15ADON, DON, DON-3-Glc, FUS-X, NIV, HT2, T2, DAS, NEO, AFs, OTA, FBs, STER, ZEN, mycophenolic acid, MON, BEA, | malting barley | Acquity UHPLC system (Waters), Q-Exactive system (Thermo Fisher Scientific) | Atlantis T3 column (100 mm × 2.1 mm, 3 μm); 30oC, flow rate 300 μL/min, injection volume 2 μL; eluent A CH3CN–H2O–CH3COOH (95:4.9:0.1, v/v/v), eluent B H2O–CH3COOH (99.9:0.1, v/v), both containing 5 mM NH4CH3COO-; gradient 5% eluent A start for 1 min, increase to 15% eluent A in 14 min, increase to 100% eluent A in next 15 min, kept at 100% eluent A for 3 min, reequilibration for 4.4 min | HESI-II (positive, negative) |
15ADON 15-acetyldeoxynivalenol, 3ADON 3-acetyldeoxynivalenol, AFs aflatoxins, APCI atmospheric pressure chemical ionization, BEA beauvericin, CIT citrinin, DON-3-Glc deoxynivalenol 3-glucoside, DAS diacetoxyscirpenol, dMRM dynamic multiple-reaction monitoring, DON deoxynivalenol, ENNs enniatins, ESI electrospray ionization, FBs fumonisins B, FUS-X fusarenon X, HCD high-energy collisional dissociation, HPLC high-performance liquid chromatography, HRMS high-resolution mass specrometry, HT2 HT-2 toxin, IAC immunoaffinity chromatography, MON moniliformin, MRM multiple-reaction monitoring, NEO neosolaniol, NIV nivalenol, OTA typo in ocratoxin A, PBS phosphate-buffered saline, QqQ triple quadrupol, QuEChERS quick, easy, cheap, effective, rugged, and safe, SIDA stable isotope dilution assay, SPE solid-phase extraction, STER sterigmatocystin, T2 T-2 toxin, UFLC ultrafast liquid chromatography, UHPLC ultrahigh-performance liquid chromatography, UPLC ultraperformance liquid chromatography, ZEA zearalanol, ZEN zearalenone
Comparison, advantages and drawbacks of mass spectrometry (MS) instrumentation
| MS technique | Analyzer | Pros | Cons | |
|---|---|---|---|---|
| LRMS(/MS) | QqQ, IT, QLIT | High sensitivity and selectivity in MRM mode | Number of simultaneously detected analytes in MRM mode is limited | |
| HRMS(/MS) | TOF, QTOF | High sensitivity and selectivity in full MS mode | Narrower dynamic range compared with LC–MS(MS/MS) instruments | Lower mass resolving power and mass accuracy compared with Orbitrap systems |
| Orbitrap, Q–Orbitrap | Acquisition speed limited at high mass resolving power settings | |||
HRMS high resolution mass spectrometry, LC liquid chromatography, LRMS low-resolution mass spectrometry, MRM multiple-reaction monitoring, QLIT quadrupole–linear ion trap, Q–Orbitrap quadrupole–Orbitrap, QqQ triple quadrupole, TOF time of flight
Fig. 4Metabolism of deoxynivalenol (DON) in wheat. Sample treated with 1:1 native DON and 13C-labeled DON (tracer). Deoxynivalenol 3-glucoside (DON-Glc) was found as a major metabolite. Mass spectrometry (MS) scan of DON and 13C-labeled DON (A). Extracted ion chromatogram of DON (B). Extracted ion chromatogram of DON-Glc (C). MS scan of DON-3-Glc and 13C-labeled DON-Glc (D). Extracted ion chromatogram of 13C-labeled DON (E). Extracted ion chromatogram of 13C-labeled DON-Glc (F)