Tim Gruene1, Enrico Mugnaioli2. 1. University of Vienna, Faculty of Chemistry, Department of Inorganic Chemistry, AT-1090 Vienna, Austria. 2. Center for Nanotechnology Innovation@NEST, Istituto Italiano di Tecnologia, Piazza S. Silvestro 12, IT-56127 Pisa, Italy.
Abstract
In the past few years, many exciting papers reported results based on crystal structure determination by electron diffraction. The aim of this review is to provide general and practical information to structural chemists interested in stepping into this emerging field. We discuss technical characteristics of electron microscopes for research units that would like to acquire their own instrumentation, as well as those practical aspects that appear different between X-ray and electron crystallography. We also include a discussion about applications where electron crystallography provides information that is different, and possibly complementary, with respect to what is available from X-ray crystallography.
In the past few years, many exciting papers reported results based on crystal structure determination by electron diffraction. The aim of this review is to provide general and practical information to structural chemists interested in stepping into this emerging field. We discuss technical characteristics of electron microscopes for research units that would like to acquire their own instrumentation, as well as those practical aspects that appear different between X-ray and electron crystallography. We also include a discussion about applications where electron crystallography provides information that is different, and possibly complementary, with respect to what is available from X-ray crystallography.
Electron crystallography is gaining acceptance as a method of structure
elucidation in chemistry, structural chemistry, and materials science,
complementary to X-ray crystallography. In 2018, two publications
about structure determination of organic compounds with electron diffraction[1,2] triggered a lot of attention. Matthew Warren asked “Why didn’t
we think to do this earlier? [...]”[3] With a better account for the developments in 3D electron diffraction
(3D ED), one would better phrase this as “Why didn’t
we realize this earlier?”, because chemical structure determination
with electron diffraction has a much longer history. The first paper[1] is in fact the main publication of the research
project “Applicability of 3D electron diffraction in the pharmaceutical
industry” (A3EDPI). The motivation for this project proposal
was the realization that the majority of crystallographers, and chemists
who make use of crystallography, were even unaware of the existence
of the 3D ED method. The project proposal aimed at changing this lack
of awareness.Several recent reviews light the field from various
angles. Nannenga[4] and Clabbers and Xu[5] cover the sample preparation, data processing,
and refinement process,
with an emphasis on organic and macromolecular compounds. Gemmi et
al.[6] have provided an overview of the historical
development, the various kinds of data collection techniques, and
examples from a broad range of applications. Huang et al.[7] describe a long history of electron diffraction
in the field of metal organic and covalent organic frameworks (MOFs
and COFs).The name “3D ED” was suggested by Gemmi
et al.[6] as an umbrella term for a number
of techniques
that collect reflection intensities in the 3-dimensional reciprocal
space by an electron diffraction experiment. Several very similar
techniques have been developed in the past 15 years. The very original
term is automated diffraction tomography (ADT[8]); the most popular term is probably microcrystal electron diffraction
(microED[9]). Other terms include rotation
electron diffraction (RED[10]), continuous
rotation electron diffraction (cRED[11]),
electron diffraction tomography (EDT[12]),
integrated electron diffraction tomography (IEDT[13]), and precession electron diffraction tomography (PEDT[14]).The crystal size is the most striking
difference between electron
and X-ray diffraction, especially for X-ray crystallographers who
are new to ED. Due to the strong interaction between electrons and
matter, crystals can be as small as a few tens of nanometers and up
to about 1 μm thick. Between 1 and 5 μm, depending on
the elemental composition of the crystals, electrons with an energy
of 200 keV get absorbed, so that ED closes the gap to X-ray crystallography,
where 5 μm is about the lower end of what can be measured with
single crystal X-ray diffraction. This enables the analysis of minute
single crystals that compose powder samples. Individual single crystal
structures can be determined from powder blends. This resolves the
ambiguity that X-ray powder diffraction can suffer from. The importance
of this was recently recognized in the extensive study of 50 organic
samples by Bruhn et al.[15] and readily realized
during the study of a new lithium-rich zincolithosilicate.[16] The strong interaction of electrons with matter
leads to a significant deviation of the Bragg intensities from the
kinematic theory of diffraction. Ignoring this effect, i.e. applying
a kinematic refinement, yields already fairly good results.[15] More details can be revealed with dynamical
refinement that takes the dynamical theory of diffraction into account.
Although this slows down the refinement process, dynamical refinement
makes ED very sensitive for the determination of chirality and the
location of hydrogen.[17,18] We refer the reader to refs (19−21) for a detailed discussion and to ref (22) for an overview of dynamical
refinement and its possibilities in ED.The present review focuses
on instrumental aspects for chemists
who wish to include ED as an analytical tool in crystallography. It
complements the perspective of Gruene et al.,[22] which addresses the experimental aspects of sample preparation and
the data analysis process, and it complements the review of Gemmi
et al.,[6] which covers the broad field of
applications of ED, explains the different types of data collection
strategies, and summarizes the historical development of 3D ED. This
paper also adds some of the most recent results since Gemmi et al.[6]
Access to Electron Diffraction
Any lab that carries out X-ray diffraction has, in principle, the
expertise to carry out electron diffraction too. The costs for an
instrument suitable for electron diffraction are of the same order
of magnitude as a modern X-ray diffractometer. The skill required
to operate a transmission electron microscope for electron diffraction
is comparable to the skill required to operate an X-ray diffractometer.
The same software for all steps of data analysis can be used, although
programs specifically dedicated to electron diffraction are available.[23] The main difference is the possibility of taking
dynamical scattering into account during structure refinement. Dynamic
refinement results in a higher level of detail for the interpretation
of the structure, e.g. a better observation of hydrogen atoms.[17] It can also be used to determine the chirality
of chiral molecules.[18]Table is a compilation
of laboratories where electron diffraction is carried out. It has
been compiled from the structures deposited at the Cambridge Crystallographic
Database, CCDC, by searching for the keyword “electron diffraction”,[24,25] and from recent publications. Table is likely bound to be incomplete. The fact that the
number of laboratories still fits into a table illustrates that 3D
ED has not unfolded its potential, yet.
Table 1
Collection of Institutes
That Recently
Carried out and Published Structures Determined from 3D Electron Diffraction
Institute
PI
refs
Czech Academy of Sciences (CZ)
L. Palatinus
P. Brazda
G. Steciuk
(26, 27)
Univ. of California, Los Angeles (US)
T. Gonen
(2, 28)
H. Nelson
(29, 30)
J. A. Rodriguez
(2, 31)
Univ. of Stockholm (SE)
X.
Zou
(32, 33)
H. Xu
(5, 34)
Univ. of Vienna
(AT)
T. Gruene
(1, 35)
Nanoimaging Services
(US)
J. Bruhn
(15, 36)
Univ. of Tokyo
(JP)
E. Nakamura
(37)
Paul Scherrer Inst. (CH)
E. Müller Gubler
(1)
E. Poghosyan
J. P. Abrahams
(38, 39)
RIKEN SPring-8
Centre (JP)
K. Yonekura
(40, 41)
S. Maki-Yonekura
Univ. of
Hamburg (DE)
R. Bücker
(42)
Center for Nanotechnology Innovation, Pisa (IT)
M.
Gemmi
(43, 44)
E. Mugnaioli
Electron Bio-Imaging Centre (eBIC) (UK)
P. Zhang
(45, 46)
D. G. Waterman
(15)
Johannes Gutenberg Univ. Mainz (DE)
U. Kolb
(47, 48)
CNRS ENSICAEN (FR)
P. Boullay
(17, 49)
Univ. of Antwerpen
(BE)
J. Hadermann
(50)
Univ. Grenoble Alpes and CNRS (FR)
S. Kodjikian
(51, 52)
H. Klein
Univ. of Manchester (UK)
A. S. Eggemann
(53)
Arizona State Univ. (US)
B. Nannenga
(54)
Univ. Lille (FR)
D. Jacob
(21, 55)
Instrumentation
X-ray crystal structures
are determined with X-ray diffractometers.
Electron diffractometers, in an analogue sense, are currently under
development. Vainshtein presented their horizontal electron diffraction
chamber in his book on electron diffraction in 1964,[56] but it appears that a commercial solution has never existed.
There are now two commercial electron diffractometers under development.
Both ELDICO-Scientific and Rigaku have presented a first model of
their instruments. The control software of both instruments is close
to what X-ray crystallographers are familiar with from their X-ray
home sources.[57,58] Rigaku presented their electron
diffractometer XtaLAB Synergy-ED in spring 2021. It is based on a
JEOL JEM2100Plus and fully controlled through the Rigaku CrysAlis
Pro software, which also operates their X-ray diffractometers.[58] The instrument by ELDICO is an entirely new
design, much more akin to an X-ray diffractometer. The Swiss Innovation
Park Basel Area also expects a dedicated electron diffractometer during
summer 2021 (priv. commun. Dr. G. Santiso-Quĩnones, ELDICO
Scientific).Up to now, transmission electron microscopes (TEMs)
have been used
instead of electron diffractometers. Two manufacturers of TEMs, Thermofisher
and JEOL, offer “microED” packages as software upgrades
for their instruments. This mainly addresses the need in crystallography
to rotate the crystal about an axis with a constant angular velocity.
Hitachi is another TEM manufacturer, although we are not certain whether
extensions for 3D electron diffraction are available. It should be
noted that several freely available software solutions such as Instamatic/Insteadmatic,[59,60] parallelEM,[41,61] fast-ADT,[62] and a collection of scripts based on serialEM[63] are independent of such “microED”
packages. SerialEM is a generic software interface, which works on
a broad range of instruments.[64]Unlike
X-ray diffractometers, TEMs usually have a single axis goniometer.
This axis is typically called the α-axis. Depending on the make,
the sample-holder can have a second, β-axis for positioning
the sample at an angle nonperpendicular to the incident beam. However,
in our experience, the β-axis is not eucentric, and crystal
alignment becomes rather difficult.
Choice
of Electron Source: LaB6 or FEG
TEMs can be equipped with two types
of electron source: thermo-ionic cathodes, such as a Tungsten (W) hairpin or a pointed lanthanum hexabromide (LaB6) crystal, or field emission guns (FEGs).[65] TEMs with a thermo-ionic electron source are
more commonly found and can readily be used for 3D ED, as long as
the rotation range of the goniometer is reasonably wide.The
energy bandwidth is about 0.2–0.7 eV for a FEG and 1–2
eV for a LaB6 source. At 200 keV, the
energy spread of the LaB6 source corresponds to the energy spread of a perfect
silicon monochromator
used at X-ray synchrotron beamlines.[66] Thus,
for electron diffraction the advantages of a FEG source are not as
striking as for imaging, and a TEM equipped with an LaB6 source can deliver comparable results at a significantly
smaller cost and easier maintenance. A modern, high-end TEM with an LaB6 source with energy varying from 80 to 200
keV is available for less than 800 k€. This is of the same
order of magnitude as a modern X-ray diffractometer (400–500
k€). Refurbished second-hand, good quality TEMs are available
for about 200–300 k€. Moreover, an LaB6 electron source is much less sensitive to damage from
impacting positive ions and requires a vacuum P ≤
10–6 mbar. FEGs require at least P ≤ 10–8–10–10 mbar.[65]The divergence of the electron beam of
a LaB6 source is about the same order
of magnitude as the X-ray
beam of a third-generation synchrotron and at least 1 order of magnitude
less than the X-ray beam of an in-house X-ray diffractometer. This
is sufficient for most crystallographic applications, and the lower
divergence (“greater parallelity”) of FEGs brings no
extra benefit.On the other hand, the coherence length for FEGs
is much larger
than for LaB6. The electron beam for a
FEG allows probing the sample with a quasi-parallel beam of a few
tens of nanometers, which results advantageous for the study of nanoparticles,
multiphase nanodevices, precipitates casted on crystalline matrices,
and materials characterized by nanotwinning or aggregated nanodomains.
Moreover, the higher brilliance of an FEG can improve the performance
of serial and roto-serial diffraction acquisitions (see section ).Independent
of the electron source, all TEMs can be equipped with
an energy filter. Energy filters significantly reduce the background
scatter,[67] especially at low voltage. However,
background subtraction is well advanced in data reduction programs,[68] so that an energy filter produces visually appealing
diffraction patterns, but the experimental proof of better data quality
is not available to date. A recent theoretical study suggests that
an energy filter may actually enhance the relative contribution of
dynamical scattering.[69] However, hitherto
no experimental data in support of this effect have been produced.In addition to the above technical aspects, high-end TEMs are available
with an “autoloader”, a cryogenic sample manipulation
robot, which may be convenient for high-throughput screening of cryo-protected
samples. In our experience, though, the time-limiting step is crystal
search rather than changing grids, and we would not consider an autoloader
a “must-have” accessory.
Choice
of Energy, 120 keV vs 200 keV vs 300
keV
There are instruments available with a maximum acceleration
voltage of 120 kV, 200 kV, or 300 kV. The choice is mainly a matter
of costs and of accessories, that may be available for a high-end
TEM at 300 kV equipped with a FEG but not a midrange 120 kV or 200
kV TEM with a LaB6 electron source. In
addition to financial considerations, the penetration depth of 300
keV electrons is slightly greater than at 200 keV so that thicker
samples can be studied. The difference in cross section, however,
is rather small and can be compensated with preparing smaller crystals.[70,71] The upper limit for organic crystals is about 1 μm at 200
keV. As the thickness increases, the diffraction pattern becomes fuzzy,
before the electrons are fully absorbed from a few micrometer thick
crystal. Like with X-rays, absorption increases with heavier weight
elements but remains within the same order of magnitude.The
wavelength of the diffraction energy is related to the acceleration
voltage of the electron beam: the energy of the electron beam equals
the elementary charge (1 e) times the voltage. The
wavelength λ results from the de Broglie equation (where c = 299, 792 km/s is the speed of light, h = 6.626070 × 10–34 J s is the Planck constant, m0 = 510.999 keV/c2 is the rest mass of the electron, E is the total
energy E + m0c2, and E is the potential energy, i.e. e*voltage,
measured in keV).The common energies 120, 200, and 300
keV correspond to the wavelengths
0.03349 Å, 0.02508 Å, and 0.01969 Å, respectively.
Note that the 2θ ranges for a typical data set with a resolution
range between 0.8 and 25 Å for these three wavelengths are 2.4°–0.08°,
1.8°–0.06°, and 1.4°–0.05°, respectively.
For Mo Kα X-rays, λ = 0.7107 Å, the range is 52.7°–1.6°.
This requires a much larger detector distance. This distance is set
only virtually by the electromagnetic lens system of the TEM. Higher
energies result in a higher penetration width and reduced dynamical
scattering.[40] The effect, however, is not
dramatic, and there have been excellent results at 120 keV.[17,43,44] A 300 kV TEM can also be operated
at any lower energy and thus covers a 120 kV and 200 kV TEM, likewise
for a 200 keV TEM. The different setting requires a realignment of
the optics and is rarely done. Lower energies result in less spread
of the electrons in the detector surface and better focused intensity
spots, and indeed, hybrid pixel detectors are nearly ideal at about
120 keV.[72−74]The energy of the electron beam also slightly
affects the radiation
damage during sample search and data collection. The effect is only
small and sample dependent and does not provide a clear-cut case for
an optimal energy. At energies below 80 keV, so-called knock-on effects
cannot take place for carbon. The limit is 240 keV for silicon. Knock-on
refers to the kicking out of the entire atom. However, even at higher
energies, knock-on effects only play a small role for radiation damage.
The major contribution to radiation damage comes from radiolysis,
even if the specific radiolysis damage (deposited energy per elastic
scattering) is orders of magnitude lower for electrons than for X-rays.[75] The impacting electrons trigger the release
of secondary electrons from the compound, which are the main cause
of damage. The physics of radiation damage by both knock-on and radiolysis
are detailed in various textbooks.[65,76,77] Radiolysis by electrons is similar to X-rays and
can be reduced by cooling the sample to cryogenic temperatures.[78−80] With X-ray diffraction, radiation damage is a major concern for
macromolecular compounds and supramolecular compounds but much less
so for the typical organic small molecule sample and certainly not
for inorganic compounds. The minute crystal volume makes radiation
damage a concern for a much larger class of materials in electron
diffraction—many electron microscopists consider zeolites as
radiation sensitive. A good training in macromolecular crystallography
helps to appreciate a fast working mode and helps to understand the
fact that looking at the crystal means damaging the crystal. Hence,
wherever possible, instrument alignment is carried out with the crystal
just off the field of view and only moved in position at the beginning
of data collection.
Detectors for ED
ED patterns are
traditionally recorded by charge-coupled device (CCD) cameras. Due
to the limited dynamical range and the sensitivity, a beam stop is
normally required not to damage such detectors after intensive ED
data collections. The read-out time and the read-out noise of CCD
cameras may also hinder full data acquisition from very beam-sensitive
materials (like pharmaceuticals or macromolecules) before significant
beam damage is introduced. Complementary metal–oxide–semiconductor
(CMOS) detectors have the significant advantages of better sensitivity,
lower background, and faster read-out compared to CCD cameras. Most
CMOS detectors are not specifically designed for ED experiments and
may suffer a certain deterioration after extensive diffraction data
experiments. Still, many CMOS detectors are available with full integration
into the instrument software control, and many groups collect their
ED data with CMOS detectors. Companies that offer such CMOS detectors
include TVIPS, Thermofisher, Gatan, and Direct Electron.[40,54,81−85]On the other hand, hybrid pixel detectors (HPDs)
have been dominating the field of X-ray crystallography since their
introduction about two decades ago.[86] Nowadays,
they are abundant at synchrotron beamlines for crystallography, both
for powder and for single crystal diffraction, and they also become
more and more popular with in-house X-ray diffractometers. Their excellent
suitability for the detection of electrons was investigated soon after.[72] Both with X-rays and with electrons, HPDs stand
out with their high dynamic range of typically 20 bit or more, zero
read-out time, zero read-out noise, and high image rate above 1 kHz.
In addition, with electron radiation, HPDs are radiation hard and
do not require a beamstop: even long time exposure to the direct beam
with electrons at 200 keV and below creates no damage, and they even
stand at 300 keV, when the exposure time and beam intensity are not
extreme.[73] The absence of a beam stop facilitates
data processing, as the direct beam position can be read directly
from the diffraction images. HPDs are excellent also for imaging,
and crystals are visible at very low beam intensity, comparable to
STEM imaging,[65] with the advantage of a
live, jitter-free view during sample search.[87] Several companies offer hybrid pixel detectors for TEMs: ASI, DECTRIS,
Quantum Detectors, Rigaku, and X-Spectrum.[88−92] In addition to these commercially available products,
the JUNGFRAU detector, developed at the PSI Switzerland, was recently
used to discriminate silicon from aluminum in ED data from aluminosilicates.
The JUNGFRAU detector is a charge integrating detector designed for
studies at free electron lasers. It can be operated with a 1 kHz or
2 kHz frame rate. Each pixel switches its gain automatically when
a certain charge threshold is reached. This way it covers a very large
dynamic range of 120 MeV per pixel and frame. For data processing,
50 or 100 frames, say, can be summed to get an effective frame rate
of 100 Hz. This still yields very fine-sliced frames and greatly reduces
the data volume.[35] Readers interested in
a JUNGFRAU detector can contact the PSD Detector group at PSI (https://www.psi.ch/en/detectors).In terms of data quality, all of the mentioned hybrid pixel
detectors
are properly calibrated and have a proper gain setting. Improper gain
correction leads to pixels with negative pixel counts. While the use
of negative Bragg intensities from profile fitting do contain valuable
information for the diffraction experiment,[93] raw pixels with negative intensities do not make sense. They require
a work-around solution or scripts to scan for the proper pedestal
setting.[15,81,94,95]The total number of pixels is usually smaller
for hybrid pixel
detectors than for imaging detectors, e.g. 512 × 512 pixels for
the Dectris QUADRO or ASI Timepix2 versus 4k × 4k for a Gatan
OneView or the Thermofisher Ceta-D detector. However, 512 × 512
pixels is sufficient for high-resolution small molecule diffraction.
For protein diffraction studies, a low- and high-resolution scan can
be combined, as should be good practice anyway.[96] In our pilot study with the JUNGFRAU detector,[35] the detector area was limited to 340 pixels
across due to an aperture of a film box in the aged TEM, yet it collected
complete data across the entire resolution range of 12.3–0.68
Å of zeolite A.[35]
Setting up for Data Collection
The principles of data collection
are rather similar between X-ray
diffraction and 3D ED. Two significant differences are the fact that
the sample chamber of the TEM is under vacuum and that centering of
the crystal is carried out at a much smaller scale. Both aspects are
covered in the following. An additional aspect is the difference between
nanobeam electron diffraction (NBD, also called nanoelectron diffraction,
NED) and selected area electron diffraction (SAED). These two modes,
their advantages, and disadvantages are explained very comprehensibly
in Lanza et al.,[43] especially section 3.1.
Working under Vacuum
Electrons are
absorbed by air. The flight tube of a TEM is under vacuum. Usually,
the pressure is below 10–6 mbar. Many types of crystals
will deteriorate under these conditions and need to be protected.
Crystals with solvent in their channels are particularly sensitive
to vacuum. Solvent is always present in macromolecular compounds and
is very common in supramolecular compounds. Organic and inorganic
porous materials can also host guest molecules in their channels,
which may be critical for the stability of the framework structure
or may be the very topic of interest. In the early stages of cryo-EM
development, molecules were typically coated by sugar, while more
recently cryo-plunging emerged as the technique of choice for the
study of vacuum-sensitive materials.[97] Both
techniques immobilize the sample and preserve it from vacuum deterioration.
In principle, one can also use ionic liquids with very low vapor pressure,
which do not evaporate under high vacuum. Ionic liquids are interesting,
because they are conductive and may have a positive effect with respect
to radiation damage. They also enhance the visibility of the crystals,
because their electro-optical density differs more strongly from organic
compounds than aqueous solutions differ from organic compounds. Another
option are liquid-cell holders, where the crystals can be measured
under local ambient conditions up to 1 bar.[50] A recent review goes into greater depth about these different possibilities
of sample preparation.[98] Finally, the crystals
can be immersed into a solidifying material and the crystal can be
cut out with a focused ion beam (FIB milling).[45] FIB milling can even improve the diffraction quality of
the crystal. In any case, one needs to ensure that the crystals are
embedded in a film thin enough to avoid absorption of the electron
beam. Sample preparation is often a trial-and-error approach, and
the success-rate improves with experience.[5]
Centering of the Crystal
In X-ray
diffraction, the crystal must be centered at the intersection point
of the rotation axes of the goniometer. Due to imperfections, the
intersection point is actually a little volume. The X-ray beam has
to be aligned to also cross this point of intersection. In electron
diffraction, there is (currently) only one rotation axis, and the
beam can, in principle, be shifted in position and focused at different
heights. As a terminology, the word “eucentric” height
is used in microscopy (eu-: representing Greek ευ-, combining form of ευζ good,
used in neuter form ευ as adverb = well).
However, when working with a submicrometer sized beam and a submicrometer
sized crystal, the requirements to the stability of the goniometer
are much higher than for a goniometer used in X-ray diffractometers.
Furthermore, exposure of the crystal to the beam damages it, and centering
should be done with the crystal translated along the rotation axis
just outside of the beam. In general, the goniometer can be balanced
to one particular sample holder. Ideally, a crystal can be aligned
such that it stays even within a very small beam for the entire rotation
range supported by the goniometer. In most instruments, this range
is ≈ ±70° from the horizontal of the grid plane.Figure shows a room-temperature
holder and a cryo-tomography holder. The room-temperature holder is
balanced with the goniometer at the TEM in Vienna. A 500 nm crystal
stays inside a 750 nm beam diameter throughout 80°, the maximum
rotation range of this instrument. Using a different holder disturbs
the balance. This can reduce the total available rotation range. The
same needs to be taken into consideration for cryo-holders when the
liquid nitrogen evaporates with time. One solution to unstable goniometers
is to track the crystal. This can be done before data collection with
a “dummy rotation” in imaging mode,[62] or during data collection, by switching the instrument
to imaging mode in regular intervals and discarding the respective
frames from data processing.[59] However,
switching between imaging and diffraction mode may suffer from hysteresis
effects of the electromagnetic lenses and is therefore not suitable
for all instruments. A third option, which was actually developed
long before continuous rotation data collection was introduced to
ED by Nannenga et al.[9] and consists of
the combination of still images with beam precession at each position,[99−101] cf. section .
Figure 1
Holder must be balanced to the goniometer for stable rotation.
(A) Room-temperature holder used in Vienna. (B) Typical TEM sample
grid, 3.05 diameter, covered with crystalline powder. (C) Opened clip
of the holder with grid already placed. (D) Sample crystal with a
beam diameter of 750 nm. The crystal is 125 nm wide; the visible length
is 500 nm (courtesy J. T. C. Wennmacher). (E) Example of a cryo-transfer
holder inside the cryo-transfer station (Fischione Model 2550, with
kind permission). (F) The well-aligned crystal (D) diffracts throughout
80° rotation (the maximum possible with the CM200 in Vienna;
courtesy J. T. C. Wennmacher).
Holder must be balanced to the goniometer for stable rotation.
(A) Room-temperature holder used in Vienna. (B) Typical TEM sample
grid, 3.05 diameter, covered with crystalline powder. (C) Opened clip
of the holder with grid already placed. (D) Sample crystal with a
beam diameter of 750 nm. The crystal is 125 nm wide; the visible length
is 500 nm (courtesy J. T. C. Wennmacher). (E) Example of a cryo-transfer
holder inside the cryo-transfer station (Fischione Model 2550, with
kind permission). (F) The well-aligned crystal (D) diffracts throughout
80° rotation (the maximum possible with the CM200 in Vienna;
courtesy J. T. C. Wennmacher).Once the sample grid has been centered near the crystal, the instrument
can be set from imaging mode to diffraction mode. Electron microscopes
feature a simple push button for this. With a perfect instrument,
the diffraction pattern is focused on the detector plane, resulting
in optimally focused Bragg spots. When the direct beam is visible,
as is possible with HPDs, the pattern can be focused manually with
the focus button. This process takes only a few seconds. For detectors
that get damaged from exposure to the strong direct beam, this can
be done at low beam intensity, before placing the beam stop. Once
the crystal is centered and the beam is focused, data collection is
started as with X-ray diffraction; that is, the crystal is set to
rotate, and the detector starts to collect data.
Data Collection with Precession vs Rotation
Data collection
with the rotation method consists of the rotation
of the crystal on a single axis. Ideally, the crystal volume is evenly
illuminated throughout the rotation. The rotation method as described
by Arndt and Wonacott[102] results in a contiguous section of reciprocal space. Later, when detectors
became available with negligible readout-time, contiguous data collection
turned into continuous, i.e. shutterless data collection.[103]3D ED data collection uses a beam diameter
of usually less than 1.5 μm. Some instruments permit a beam
diameter down to 30 nm. This is instrument specific and affected by
the diameter of the condenser lens aperture. In electron diffraction,
the rotation range is usually limited to a maximum of about 150°.
At angles greater than ±75° off the horizontal, the sample
holder blocks the beam. This rotation range sets high requirements
to the mechanical reliability of the goniometer. The crystal must
be centered to within half the beam diameter to ensure it does not
leave the beam during exposure. When the crystal is very small, and
therefore the beam diameter is very small, or when the sample holder
is not very well balanced with respect to the instrument goniometer,
the crystal is likely to move out of the beam during rotation. When
the holder itself is not calibrated well to the goniometer, it may
not be possible at all to correct for the drift out of the beam. As
a compromise, the beam diameter can be increased, at the cost of increased
background noise.The first effective approach for 3D electron
diffraction was a
combination of a stepwise rotation of the crystal with a small precession
of the beam at each step.[104] The crystal
was then visualized and recentered after each tilt step, at the expense
of an increase in electron dose. At the time the precession method
was developed, electron diffraction data were always collected from
crystals oriented along a crystallographic direction. Beam precession
was used to reduce the dynamical scattering, which is enhanced in
such conditions and makes the intensities deviate more strongly from
the kinematic approximation.[99−101] On the other hand, when coupled
with 3D electron diffraction, beam precession is mostly used for smoothing
the excitation error that affects intensities when the Ewald sphere
does not cross the center of the reflection.[105]Both 3D data collection strategies, either precession-assisted
stepwise or by continuous rotation, reduce the effects of dynamical
scattering and make the observed intensities better match with the
kinematic approximation. This facilitates structure solution ab initio, for example by direct methods.[106] Also model refinement is then possible with programs that
X-ray crystallographers are familiar with, such as CRYSTALS, OLEX2,
or SHELXL.[107−109]
Serial Electron Diffraction
Electrons
interact strongly with matter, and time-resolution with TEMs reaches
femtoseconds.[110,111] Both properties make a single
TEM as powerful as an X-ray free electron laser. A milestone for crystallography
in this context was the structure determination of two proteins with
serial electron diffraction (serialED).[42] ED structures for small organic and inorganic compounds reach the
same resolution limits as X-ray structures (0.55 Å in Figure F). This is very
different for macromolecular structures: all ED structures of macromolecules
published so far were about a factor of 2 worse in resolution compared
with X-ray diffraction of the respective samples. Bücker et
al.[42] have been the first to demonstrate
that ED of macromolecules can reach a similar resolution limit as
macromolecular X-ray crystallography.Serial crystallography
for structure determination appears to be of little interest for small
molecules. First, when crystals, or even crystalline powders, are
available, they will be suitable for structure determination by 3D
ED. Second, diffraction images of still crystals show much less Bragg
peaks than for macromolecular compounds. This makes indexing and thus
merging of the data extremely difficult. Time-resolution in the femtosecond
regime of serial crystallography, however, would be interesting to
study reactions, e.g. by stopped-flow experiments carried out in liquid-cell
sample holders. The indexing problem could be solved with roto-serial
crystallography.[112−114] Roto-serial crystallography combines the
diffraction with a very intense beam with a small, say 1°–5°
rotation of the crystal. This is possible with TEMs, because unlike
with samples for free electron lasers, the crystals are visible in
the TEM and can thus be located, and the roto-shot can be prepared.
Roto-serial ED appears promising especially for beam-sensitive phase
mixtures of small-molecules, when combined with reliable clustering
analysis. The software solution SerialRED might act as a prototype
for this technique.[115] The instrumental
preconditions are already met by current TEMs.
Types
of Grids and Sample Preparation
Crystals for electron diffraction
are deposited on TEM grids. TEM
grids are thin metal grids with a diameter of typically 3.05 mm. The
metal gives mechanical stability to the grid and acts as conductor
for captured electrons. The grids are commonly characterized by their
mesh. A small mesh number, such as 150 or 200, results in large squares
and thus increases the maximum available rotation range. Grids can
be covered with a variety of thin layers, which support the sample
and are electron-transparent. With less than 10 nm thickness, amorphous
carbon is the thinnest material and therefore the preferable one for
general electron diffraction applications. The stability can be increased
with a layer of Formvar, at the cost of greater background noise.There are special sample holders, with electron-transparent sample
supports, allowing for a larger rotation range, even 360°.[116] These special thin tips, however, are not produced
commercially and have a large failure rate in production. Most data
are therefore collected with normal TEM grids, coated with a nanometer-layer
of continuous or lacy carbon. In case of low-symmetry space groups,
complete data may not be collected from a single crystal. Either the
crystal morphology favors various orientations of the crystals on
the flat support layer or three-dimensional support grids can be used
in order to achieve 100% complete data by merging data from several
crystals. In order to create a three-dimensional support grid, the
continuous carbon layer can be caused to coil by gently stroking it
with a brush. Crystals stick with the coiled carbon layer, and therefore
crystals in various orientations are conveniently found. This increases
the chance to obtain high data completeness from only very few crystals,
even in low-symmetry space groups.[117] Moreover,
continuous carbon has no visible features in the TEM imaging view.
This makes orientation very difficult, and the alignment of the microscope
for diffraction just off the crystal is impossible. Even coiled continuous
carbon may not have features suitable for crystal centering. Holey
and lacy carbon are much better suited for the purpose of crystal
alignment and for orientation. At high magnification, even a single
square of the grid is a huge area to scan. Therefore, orientation
by the irregular patterns of lacy carbon is a great aid.During
data collection, the crystallographer is at risk of bias
toward well diffracting crystals. This is particularly important,
when mixtures of various phases or mixtures of different compounds
are investigated. In such cases, conclusions about the sample can
be supported with X-ray powder diffraction. A valuable approach in
this context is the development of automated data collection, possibly
combined with automated clustering of the sample, based on unit cell
parameters.[61,115] This technique can supplement
X-ray powder diffraction and possibly find compounds in mixtures even
if present in small trace amounts.
Applications
Chemists who apply X-ray crystallography for their research most
likely would like to know whether 3D ED is a suitable technique for
the compounds of their research. Generally speaking, whenever X-ray
diffraction is part of the analytical toolbox, electron diffraction
will also be useful. Gemmi et al.[6] listed
a number of applications where 3D ED was able to solve crystallographic
problems that could not be addressed by X-ray. More recently, several
studies exploited the fact that 3D ED can determine the single crystal
structure when crystal size is the limiting factor for single crystal
X-ray diffraction.[16,30,32,118] A most impressive demonstration for organic
compounds was published recently by Bruhn et al.[15] 3D ED has been used to overcome local minima in crystal
structure prediction.[119] In our experience,
sub-micrometer crystals of chemical compounds diffract to similar
resolution with electron diffraction, as crystals tens of micrometers
thick diffract with X-ray diffraction (Figure D and F). A crystal structure provides 3D
coordinates of the atoms, that compose the compound, and their atomic
displacement parameters, which indicate their thermal vibrations.
When data quality is good, not only atom positions but also their
element type can be determined.The 3D coordinates from a crystal
structure also provide information
about distances of bonded and nonbonded contacts, about crystal packing,
and about the space group. The space group has consequences, e.g.
for electronic or optical properties. In materials science, the crystal
structure helps with understanding crystal defects and describing
the polymorphism of a material.[44] In organic
chemistry, a crystal structure provides qualitative proof for a synthesis
pathway. In pharmacology, diffraction data helps to understand the
kinematic and thermodynamic stability of a formulation.[120] X-ray diffraction is also an established technique
to determine the chirality.[121]The
large number of publications and reviews on 3D ED illustrate
to what extend it complements X-ray crystallography in the above listed
aspects, especially for those cases where single crystals cannot be
grown to a size sufficiently large for X-ray diffraction.[4,6,7,43,122−124] In particular, this
complementarity also includes the determination of chirality with
3D ED.[18,125] The following sections reflect our opinion,
for which scientific applications of 3D ED will not only complement
but actually extend X-ray crystallography.A natural product is a chemical compound
or substance produced
by a living organism—that is, found in nature. [...] Within
the field of organic chemistry, the definition of natural products
is usually restricted to organic compounds isolated from natural sources
that are produced by the pathways of primary or secondary metabolism.
Within the field of medicinal chemistry, the definition is often further
restricted to secondary metabolites.[126]
Natural
Products
The interest in “natural products”
lies in the interest
to understand, improve, and exploit the diverse machinery available
in nature. In many cases, their functionality involves modified standard
amino-acids and circular peptides. The fields of interest include
the health-section, with the research for antibacterial, antiviral,
and anticancer drugs.[127−129] They also include literally green sources
of energy through the construction of metallochemical compounds that
mimic the process of photosynthesis or the improvement of chemical
reactions.[130] Typically, natural products
are only available in very small milligram amounts.[30] This places a restriction on the possibilities of structure
elucidation and makes electron diffraction a very attractive alternative
to structure determination with X-rays or NMR. It is hard to imagine
how little amounts are necessary for electron crystallography. A grain,
just about visible with the bare eye, can contain hundreds or thousands
of single crystals. This explains the interest in 3D ED for structure
determination for cases where the amount of material is very limited,
e.g. due to difficulties in purification.That said, one has
to bear in mind that sample preparation for 3D ED takes some experience.
The main difficulty is the preparation of a TEM grid with the right
density of particles. Too many particles result in agglomeration of
crystals and make it difficult to find single crystals. This is certainly
true for natural products that can often be quite sticky. Too few
particles make it very cumbersome to find any crystals at all. With
some experience in grid preparation, testing less than 5–10
different grids should lead to success.[5]It is convenient to allow for a certain size distribution
of particles.
Small crystals, that are suitable for data collection, are often close
to larger clusters on the sample grid. These large clusters can be
found when scanning the TEM grid at low magnification. The hit score
can be improved with a TEM equipped with a STEM unit.[65] The dose required for STEM imaging is much lower than for
a real image at the same resolution, which reduces radiation damage
in the case of very sensitive samples.[8,87,122] In addition to the reduced dose, the STEM imaging
takes place in the diffraction mode of the TEM, so that no switching
of the lens system is required before data collection.
Charges: Oxidation States, Noninnocent Ligands
In terms
of structure determination, electron diffraction is quite
analogous to X-ray diffraction. The result is a map, and the chemical
model is an interpretation of the map. The atoms are located at the
map peaks. Model building is guided through the difference map. Physically,
electron diffraction occurs through the interaction of the electron
beam with the local electrostatic potential inside the crystal. This
causes an interesting feature for electron diffraction which is different
from X-ray diffraction: the scattering factors of ions diverge at
low resolution. For cations, the scattering factor diverges to infinity;
for anions, it diverges to minus infinity.[131,132]Figure shows the theoretical scattering factors for a few
selected elements. The scattering factor for the negative O– ion diverges to minus infinity. The
scattering factors of different oxidations states are rather different
and allow for their differentiation. Figure includes Ru, Ru3+, and Ru4+ as an example.
In order to demonstrate that this difference is strong enough to differentiate
between these oxidation states, the scattering curves for Al3+ and Si4+ are
also shown. These two elements were recently differentiated with electron
diffraction in the case of two aluminosilicates.[35] It is worth noting that for aluminosilicates, neither element
is expected to be at their full oxidation state and the actual difference
in their scattering power should be less than the difference between
their theoretical scattering curves. As Figure illustrates, electron diffraction is sensitive
to the oxidation states especially at low-resolution data. Large unit
cells result in more data points at low resolution and may make it
easier to differentiate ionic states.[131] The lowest resolution reflection of albite, one of the samples where
silicon was differentiated from aluminum,[35] is only at d = 6.35 Å, i.e. sin θ/λ
= 0.0787 1/Å. Therefore, the use of electron diffraction to probe
for ionic states not only applies to proteins, but indeed to most
chemical compounds, either organic or inorganic. In fact, the electrostatic
potential in salts can be determined quantitatively with electron
diffraction data.[133,134]
Figure 2
Electron scattering factors for ions and
noncharged elements. The
graph illustrates the divergence of ionic scattering factors to ±∞ for s → 0 Å–1.
Electron scattering factors for ions and
noncharged elements. The
graph illustrates the divergence of ionic scattering factors to ±∞ for s → 0 Å–1.This characteristic of electron
diffraction becomes interesting
in cases where the oxidation states are not purely ionic and where
they are not easily assigned to individual atoms. Such complex situations
arise when both cation and anion can have multiple oxidation states,
e.g. when noninnocent ligands are studied. A “non-innocent
ligand is a ligand in a metal complex, where the oxidation state is
not clear”.[135] Noninnocent ligands
are only one example where 3D ED may become the tool of choice. In
many chemical systems, the electronic state escapes simulation studies,
while charge density studies with X-ray diffraction require such high-resolution
data that may not be available.[136,137]
Twinning
Ideally, diffraction data
are collected from only a single crystal. Twinning refers to a data
set from two or more crystal lattices. This can occur on the domain
level; that is, a seemingly single crystal is composed of unit cells
which are not related to pure translation. It can also occur at the
macroscopic level, when two crystals are attached to one another.[138] Twinning may introduce space group ambiguities
or destabilize the convergence of the refinement procedure. In case
of an arbitrary twin law, the integration program can pick out the
major lattice and ignore the contribution of smaller lattices. In
case of overlapping reflections, however, this leads to corruption
of the observed intensity. Advanced programs can deconvolute several
lattices during data processing. The algorithms for the treatment
of twinned data do not rely on the radiation source and therefore
are also applicable to 3D ED data.[139,140] In this context,
3D ED has a great advantage compared with X-ray diffraction: Since
3D ED can collect data from domains as small as the beam diameter,
i.e. down to 30 nm, it offers the chance of getting single crystal
data from crystals, that macroscopically appear twinned, but microscopically
are single.[44,141,142] At high magnification, individual domains may become apparent, and
data can be collected from single domains which are too small to be
seen with the optical microscope.This has been recently achieved
for the structure of orthocetamol (Figure ). Light microscopy shows strongly intergrown
crystals, which could not be singularly sampled by X-ray diffraction[140] (Figure (a)). Indeed, orthocetamol domains are often smaller than
100 nm and only using electron diffraction it was possible to obtain
structural data from volumes of sample that are mostly single crystal
(Figure (b)). The
structure of orthocetamol is monoclinic, space group C2/c, and is characterized by two cell parameters
with relatively close lengths (a = 10.5612 Å
and b = 10.3856 Å) and by a β-angle not
far from 90°. Orthocetamol tends to form twin domains after a
rotation of 90° around the [001] axis. However, the occurrence
of diffuse scattering makes it difficult to recognize the splitting
of reflections originated by two twin domains (Figure (c)). The overlap of the two lattices results
also in the apparent violation of the extinction rule related with
the c-glide plane, and therefore, the structure was
first assigned to the tetragonal systems. Still, all structural solution
attempts failed, until the actual monoclinic symmetry was recognized
and imposed. The ab initio model was later refined
with SHELXL, and it was found that the secondary twin domain accounts
for 34% of the diffraction intensities (Figure (d)).
Figure 3
(a) Optical microscope image of a typical
orthocetamol polycrystalline
aggregate. (b) Dark-field STEM image of orthocetamol fragments. The
fragment on the bottom right corner has an optimal size for 3D ED
data collection. (c) Reconstructed h0l slice of reciprocal space extracted from a 3D ED data collection.
Extinctions due to C-centering are clearly visible,
while extinctions due to the c-glide plane are partially
violated by reflections coming from a twin domain. (d) Orthocetamol
structure, made of alternating layers of orthocetamol chains oriented
along [110] and [−110].
(a) Optical microscope image of a typical
orthocetamol polycrystalline
aggregate. (b) Dark-field STEM image of orthocetamol fragments. The
fragment on the bottom right corner has an optimal size for 3D ED
data collection. (c) Reconstructed h0l slice of reciprocal space extracted from a 3D ED data collection.
Extinctions due to C-centering are clearly visible,
while extinctions due to the c-glide plane are partially
violated by reflections coming from a twin domain. (d) Orthocetamol
structure, made of alternating layers of orthocetamol chains oriented
along [110] and [−110].
Conclusions and Outlook
Less than two decades
after the first milestone experiments[143] and only three years after its nomination as
a breakthrough technology in Science,[144] 3D ED has become a routine technique for chemical analysis.[6,15] It complements X-ray crystallography where crystal size matters.
It also extends X-ray crystallography for greater sensitivity to chirality,[18] for the detection of light atoms in the presence
of heavy atoms,[17] and 3D ED is more sensitive
to charge, including partial charges. Especially the latter property
has not been fully exploited yet, and we are confident that future
development will open new possibilities for chemists and materials
scientists. For its exciting future to happen, broader acceptance
and application of the technology will be essential. We previously
elaborated that 3D ED is best practised with a crystallographer’s
mind.[22] In this review we explain the required
instrumentation. For general purposes, we recommend a LaB6 TEM equipped with a hybrid pixel detector. The maximum
available energy for a LaB6 TEM is currently
200 keV. With costs comparable to a laboratory X-ray diffractometer,
every X-ray facility is now equipped with the knowledge to complement
their instruments with 3D electron diffraction.
Authors: Maxime Debost; Paul B Klar; Nicolas Barrier; Edwin B Clatworthy; Julien Grand; Fabien Laine; Petr Brázda; Lukas Palatinus; Nikolai Nesterenko; Philippe Boullay; Svetlana Mintova Journal: Angew Chem Int Ed Engl Date: 2020-10-16 Impact factor: 15.336
Authors: James M Parkhurst; Graeme Winter; David G Waterman; Luis Fuentes-Montero; Richard J Gildea; Garib N Murshudov; Gwyndaf Evans Journal: J Appl Crystallogr Date: 2016-10-21 Impact factor: 3.304
Authors: Felix Engelhardt; Christian Maaß; Diego M Andrada; Regine Herbst-Irmer; Dietmar Stalke Journal: Chem Sci Date: 2018-02-08 Impact factor: 9.825
Authors: Joachim M Matz; Benjamin Drepper; Thorsten B Blum; Eric van Genderen; Alana Burrell; Peer Martin; Thomas Stach; Lucy M Collinson; Jan Pieter Abrahams; Kai Matuschewski; Michael J Blackman Journal: Proc Natl Acad Sci U S A Date: 2020-06-29 Impact factor: 11.205
Authors: K Naydenova; G McMullan; M J Peet; Y Lee; P C Edwards; S Chen; E Leahy; S Scotcher; R Henderson; C J Russo Journal: IUCrJ Date: 2019-10-11 Impact factor: 4.769
Authors: Edward T Broadhurst; Hongyi Xu; Max T B Clabbers; Molly Lightowler; Fabio Nudelman; Xiaodong Zou; Simon Parsons Journal: IUCrJ Date: 2020-01-01 Impact factor: 4.769
Authors: Nicholas Pearce; Katherine E A Reynolds; Surajit Kayal; Xue Z Sun; E Stephen Davies; Ferdinando Malagreca; Christian J Schürmann; Sho Ito; Akihito Yamano; Stephen P Argent; Michael W George; Neil R Champness Journal: Nat Commun Date: 2022-01-20 Impact factor: 14.919